The largest database of trusted experimental protocols

Loctite

Loctite is an AI-driven platform that helps researchers optimize their research protocols and enhance reproducibility.
The platform allows users to locate protocols from literature, pre-prints, and patents, and use AI-driven comparisons to identify the best protocols and products for their needs.
Loctite's cutting-edg technology improves research efficiency and reliability, making it a valuable tool for scientists seeking to improve their research process.

Most cited protocols related to «Loctite»

The basic protocol for each perfusion comprises three sections: calibration/testing of the system, cannulation of the eye and acclimatisation, and a multi-step perfusion regimen.
For each day of experiments, the pressure sensor was calibrated using an automated 8-point calibration with the automated linear actuator, which has a resolution of 1.25 μm per step. Subsequently, the resistance of the flow sensor was measured to ensure that it was blockage free. Prior to each perfusion, the resistance of a glass capillary of known resistance (comparable to a mouse eye) was measured to confirm that the system was functioning properly. The perfusion tubing (downstream of the flow sensor) and needle were then filled with perfusate and the resistance of the needle was measured before cannulating the eye. Note that, as the fluid in the system was water, there was a water-perfusate interface in the perfusion tubing. However, given that the volume of the tubing (≈2ml) was large compared to the total volume perfused in an experiment (≈25μl), neither advection nor diffusion could have altered the perfusate entering the eye. If any of the readings from the tests were unexpected, for example if the system time-response was too long or measured resistances were not correct, leaks/bubbles/blockages were identified and removed before the ocular perfusion commenced.
All animal experiments were done ex vivo in accordance with the Animals (Scientific Procedure) Act with the authority of a UK Home Office project licence (PPL 70/7306). All perfusions were carried out on 10–14 week old male C57BL/6J (B6) mice (Charles River UK, Ltd.), that were euthanised via cervical dislocation. All mice were fed ad libitum and housed in clear cages at 21°C with a 12 hour light-dark cycle (lights on at 7AM).
After enucleation, the eyes were stored in PBS at room temperature to await perfusion. Each eye was then affixed to a platform inside the heated bath with a small amount of cyanoacrylate glue (Loctite, UK). An XYZ micromanipulator (World Precision Instruments, USA) was used to cannulate the eye via the anterior chamber with a 33 gauge needle (Nanofil, World Precision Instruments, USA) under a stereomicroscope. The control perfusate was DBG: Dulbecco’s PBS containing divalent cations, supplemented with 5.5 mM glucose and passed through a 0.2 μm filter. After cannulation, the bath was filled with PBS to fully immerse the eye and the temperature was raised to 35°C. The applied pressure was held at 9 mmHg for a period of 30–45 minutes to allow the eye to acclimatise to the pressure and temperature environment.
After acclimatisation, a nine-step perfusion protocol was carried out, consisting of applied pressures of 4.5, 6, 7.5, 9, 10.5, 12, 15, 18 and 21 mmHg. A sample perfusion tracing is shown in Fig 1b and 1c.
In order to avoid a subjective element in the definition of steady state at each pressure step, a steady state condition was defined and automatically monitored by the perfusion software. A parameter Γ(t) = Q(t)/P(t) was continuously evaluated and /dt was estimated by linear regression over a moving window of 5 minutes. When |/dt| was continuously less than 0.1 nl/min/mmHg/min for one minute, the system was considered to be at steady state. The measured flow and pressure were then averaged over the four previous minutes to yield Qj and Pj respectively, for pressure step j. The red lines in Fig 1b and 1c indicate these averaging periods. This approach ensures that the measured facility changes by less than 0.4 nl/min/mmHg over the averaging period. For the eyes in the present study, the median change in facility over the averaging window was 0.16 nl/min/mmHg.
Publication 2016
Acclimatization Animals ARID1A protein, human Bath Cannulation Capillary Resistance Cations, Divalent Chambers, Anterior Cyanoacrylates Diffusion Eye Glucose Joint Dislocations Light Loctite Males Mus Neck Needles Perfusion Pressure Rivers Treatment Protocols Vision

Protocol full text hidden due to copyright restrictions

Open the protocol to access the free full text link

Publication 2016
Autopsy Cartilage Condyle Cyanoacrylates Euthanasia Femur Ligaments Loctite Menisci, Lateral Neoplasm Metastasis Operative Surgical Procedures Protease Inhibitors Radius Tendons Tissues

Protocol full text hidden due to copyright restrictions

Open the protocol to access the free full text link

Publication 2016
Acclimatization Chambers, Anterior Cyanoacrylates Eye Glucosephosphates Loctite Mus netarsudil Perfusion Pharmaceutical Preparations physiology Pressure Saline Solution Sulfoxide, Dimethyl

Protocol full text hidden due to copyright restrictions

Open the protocol to access the free full text link

Publication 2011
Animals Asepsis Bicarbonate, Sodium Brain Equus caballus GDNF protein, human Glucose Glutamine HEPES Loctite Nerve Growth Factor paraform Phosphates Rats, Sprague-Dawley Saline Solution Serum Sterility, Reproductive Tissue, Membrane

Protocol full text hidden due to copyright restrictions

Open the protocol to access the free full text link

Publication 2017
Bath Cannula Capillaries Chambers, Anterior Cyanoacrylates Glucose Hydrodynamics Loctite Needles Perfusion Pressure Sterility, Reproductive Syringes Tonometry, Ocular

Most recents protocols related to «Loctite»

Sample imaging was performed using a Zeiss Lightsheet 7 microscope operated with Zen Black version 3.1 and equipped with a detection objective (Fluar 2.5x/0.12 M27; WD=8.7mm), two illumination objectives (5x/0.1 foc), two PCO.edge 4.2 cameras, and a Mesoscale Imaging System from Translucence (imaging chamber with glass windows, objective adapter, and magnetic sample holders). The posterior portion of each hemisphere was securely mounted to a magnetic sample holder with Loctite Ultra Gel Control Super Glue and submerged into the imaging chamber filled with ethyl cinnamate. Dual side illumination (with online fusion of left and right light sheet images) was used to simultaneously acquire images of autofluorescence (later used for registering samples to the mouse brain atlas as it provides structural information) and IF labeling with separate cameras at 3.53 μm x and y resolution (0.52x zoom), moving the sample 3.5 μm for each z-step to acquire images spanning each hemisphere. Autofluorescence was captured using 488 nm excitation (505–530 nm band pass emission) at 8% of 30 mW laser power, while IF labeling was captured using 638 nm excitation (660 nm long pass emission) at 20% of 75 mW laser power. Exposure time for both excitation wavelengths was 50 ms. Pivot scanning was used to limit image artifacts with light sheet thickness set to 10.61 μm.
Publication Preprint 2023
Brain ethyl cinnamate Light Loctite Mice, Laboratory Microscopy
In a second step, we embedded the implants in the same position using resin casts, in such a way that 5 mm of the rough surface was exposed. This approach was chosen to simulate horizontal bone resorption of 5 mm (50% of the total implant length), which is 2 mm more than the International Standardization Organization (ISO) 14801:2016 specifications. The epoxy resin was EA 3471 A and B Loctite® (Henkel AG and Company, Düsseldorf, Germany) to simulate bone (Young’s modulus of elasticity ≥ 3 GPa). In Figure 3 can be observed the preparation of the samples.
Publication 2023
Bone Resorption Bones CD3EAP protein, human Epoxy Resins Loctite Resins, Plant
After the cavity filling, the roots of each tooth were removed 3 mm below the enamel–cementum junction with a diamond disc (Disc DS022, Clinique, manufacturer information is not provided), under continuous water cooling.
Then, the pulp tissue was removed using small excavators, and a 17% EDTA solution (Endo-Solution, Cerkamed, Stalowa Wola, Poland) was applied to the pulp chamber for 1 min. Then the pulp chamber was thoroughly rinsed with distilled water.
Each tooth fragment was fixed using a cyanoacrylate adhesive (Loctite, Westlake, OH, USA) to the inner face of the metal cap of a glass container. Through the cap, a 3 mm high syringe needle was inserted into the pulp chamber. The needle was connected to a perfusor coupled to a 100 mL saline solution vial. The vial was fixed at a height of 20 cm to produce a hydrostatic pressure of 20 cm H2O in the pulp chamber [29 ,30 (link)].
Publication 2023
Cyanoacrylates Dental Caries Dental Cementum Dental Enamel Dental Pulp Diamond Edetic Acid Endometriosis Face Hydrostatic Pressure Loctite Metals Needles Pulp Chamber Saline Solution Syringes Tissues Tooth Tooth Root
As a first step, the sensing element ADXL 1005Z was individually calibrated to evaluate the intrinsic responsiveness as a function of deterministic mechanical vibrations. Calibration was carried out by gluing (Loctite®) the sensing element on a hexagonal prism mass and the assembly was screwed on a vertical instrumented vibrating table (PCB Piezotronics, Depew, NY, USA), as shown in Figure 5.
The instrumented vibration table was equipped with a single axis reference transducer (PCB Piezotronics, Depew, the United States, NY model 080A199/482A23), calibrated according to ISO 16063-11:1999 [44 ] against INRiM primary standard and integrated within the stroke of the shaker, allowing to measure the reference acceleration along the vertical axis, aref (parallel to g). The calibration of the individual sensing element was carried out according to ISO 16063-21:2003 [45 ]. The reference acceleration was acquired by an acquisition board NI 4431 (National Instruments, Austin, TX, USA) with a sampling rate of 50 kHz, fed to the PC and processed through LabVIEW® 22 software to provide the RMS reference value in ms−2. Actually, the calibration of the individual sensing element, by using the set-up shown in Figure 5, was affected by some systematic errors induced by the vibrational modes of the hexagonal prism mass, and by the small (but not negligible) damping of the glue. The errors depend on both frequency and amplitude and cannot be experimentally evaluated with this calibration setup; thus, it needs to be considered in the evaluation of the overall uncertainty [46 (link),47 ,48 (link)].
Once the sensitivity of the individual sensing element was known, the calibration of the MDUT, embedded in resin, was carried out. Calibration was performed by screwing the MDUT directly onto the vibrating instrumented table, as shown in Figure 6. The aim of the investigation was to characterize the possible effects on the sensing elements response due to the embedding resin and to quantify the occurring transverse accelerations in the horizontal direction.
Publication 2023
Acceleration austin Cerebrovascular Accident Epistropheus Hypersensitivity Loctite prisma Resins, Plant Response Elements Transducers Vibration
In addition to the transfer of silicon dies, the transferability of electronic components is also of practical relevance (see column (c) in Figure 1 as well as Figure 2). Hence, bulky packaged LEDs of 2.0 mm × 2.0 mm × 1.4 mm width, length and depth, which are usually designed for surface mounting (SMD), were used to investigate the transferability. The LEDs are obtained from CML innovative Technologies (CMD28-21 Series SMT LEDs) and have a red color, a maximum forward voltage of 2.8 V and a current of 20 mA.
For the sample preparation, two contact pads and two electrical connectors were printed on the PVA using the aerosol jet and the GenesInk S-CS31506 again, as described in Section 2.1. Afterwards, the LED was glued onto the printed electrodes using conductive glue LOCTITE ABLESTIK ICP 4015, which is a silver and silicone-based glue. The glue was cured in accordance with the datasheet at 80 °C for 30 min in air. After this, the Parylene deposition and lift-off were performed. In order to prove the successful transfer of the LED, it was tested as to whether it was still glowing after the transfer when applying a forward voltage using a Statron power supply, type 3225.
Publication 2023
Dietary Fiber Electric Conductivity Electricity Loctite parylene Silicon Silicones Silver

Top products related to «Loctite»

Sourced in Germany
Loctite 401 is a cyanoacrylate adhesive product designed for industrial and commercial applications. It is a fast-acting, single-component adhesive that provides strong bonding between a variety of materials, including metal, plastics, rubber, and more. The product is suitable for applications where a rapid and durable bond is required.
Sourced in United States, Germany, China, Australia, United Kingdom, Belgium, Japan, Canada, India, France
Sylgard 184 is a two-part silicone elastomer system. It is composed of a siloxane polymer and a curing agent. When mixed, the components crosslink to form a flexible, transparent, and durable silicone rubber. The core function of Sylgard 184 is to provide a versatile material for a wide range of applications, including molding, encapsulation, and coating.
Sourced in United States
The PICM03050 is a laboratory equipment product manufactured by Merck Group. It is a piece of apparatus used for performing various scientific experiments and analyses in a controlled environment. The core function of this equipment is to facilitate controlled conditions for conducting research and testing activities. Detailed specifications and intended use are not available.
Sourced in Germany, United States
Loctite 454 is a cyanoacrylate-based adhesive product designed for industrial use. It is a fast-curing, single-component adhesive that forms strong bonds between a variety of materials, including metals, plastics, and rubbers. The product is formulated to provide high-strength, impact-resistant, and temperature-resistant properties.
Sourced in United States, United Kingdom
Metabond is a self-curing acrylic resin formulated for the fabrication of dental prosthetics. It is a two-component system that mixes a polymer powder and a liquid monomer to create a workable material for dental laboratory procedures.
Sourced in Germany, United States
The VT1000A is a vibrating-blade microtome device manufactured by Leica. It is used for the preparation of thin tissue sections for microscopic analysis. The VT1000A provides precise and consistent sectioning of samples, allowing for high-quality specimen preparation.
Sourced in Australia, Germany
Loctite Super Glue is a fast-acting adhesive designed for instant bonding. It is a solvent-based, cyanoacrylate adhesive formulation.
Sourced in Germany, United States, United Kingdom, Italy, India, Spain, China, Poland, Switzerland, Australia, France, Canada, Sweden, Japan, Ireland, Brazil, Chile, Macao, Belgium, Sao Tome and Principe, Czechia, Malaysia, Denmark, Portugal, Argentina, Singapore, Israel, Netherlands, Mexico, Pakistan, Finland
Acetone is a colorless, volatile, and flammable liquid. It is a common solvent used in various industrial and laboratory applications. Acetone has a high solvency power, making it useful for dissolving a wide range of organic compounds.
Sourced in Germany, Austria, United States, Netherlands
6-well plates are a type of cell culture vessel used in laboratories. Each plate has six individual wells to hold and grow cells or perform other experiments. The plates are made of tissue culture-treated polystyrene, providing a suitable surface for cell adhesion and proliferation.

More about "Loctite"

Loctite is an innovative AI-driven platform that empowers researchers to optimize their research protocols and enhance the reproducibility of their studies.
The platform allows users to access a vast repository of protocols from scientific literature, preprints, and patents, and then utilize AI-driven comparisons to identify the most suitable protocols and products for their specific needs.
Loctite's cutting-edge technology, comparable to solutions like Sylgard 184 and PICM03050, revolutionizes the research process by improving efficiency and reliability.
Researchers can use the platform to locate protocols for a variety of applications, including the use of adhesives like Loctite 454, Metabond, and Loctite Super Glue, as well as the handling of solvents like Acetone and the preparation of 6-well plates.
By leveraging the power of artificial intelligence, Loctite's AI-driven comparisons help researchers make informed decisions and select the best protocols and products, ultimately enhancing the overall quality and consistency of their research.
This innovative tool, analogous to VT1000A, is a valuable asset for scientists seeking to streamline their research process and improve the reproducibility of their findings.