The largest database of trusted experimental protocols
> Physiology > Organ or Tissue Function > Dominance, Ocular

Dominance, Ocular

Dominance, Ocular refers to the tendency of one eye to be preferred over the other in visual processing and perception.
This phenomenon, also known as ocular dominance, can have implications for various visual and oculomotor functions.
The dominant eye may exhibit superiority in tasks such as visual acuity, depth perception, and binocular fusion.
Ocular dominance is thought to be influenced by a combination of genetic, developmental, and environmental factors, and can show variability across individuals.
Understanding the mechanisms and implications of ocular dominance is an area of ongoing research in the field of vision science and ophthalmology.
PubCompare.ai's cutting-edge AI-powered platform can assist researchers in navigating the latest literature on this topic, optimizeing their research process, and enhancing the reproducibility of their findings.

Most cited protocols related to «Dominance, Ocular»

Cats (postnatal days 36–49) were anesthetized with isoflurane (1–2% in surgery, 0.5–1.0% during imaging)22 (link) and paralyzed with vecuronium bromide22 (link). A craniotomy was performed over area 18 of the visual cortex, the dura reflected, and the underlying cortex covered with agarose. Movement of the brain from respiratory and heart beat pulsations were negligible (Supplementary Fig. S8). The cell-permeant calcium indicator Oregon Green 488 Bapta-1 AM (1 mM) was prepared22 (link),30 (link) and co-loaded with 40 μM Alexa Fluor 594 into a glass patch pipette (2.5 μm diameter tip). Under continuous visual guidance, the pipette tip was advanced 200–250 μm below the cortical surface and the indicators were then pressure ejected (5–10 psi). This particular method of loading produces minimal staining of glial cells (see Ref 22 (link)) but it is possible that some of the stained cells in the present study were not neuronal. Fluorescence was monitored with a custom-built microscope (Prairie Technologies) coupled with a Mai Tai XF (Newport Spectra-Physics) mode-locked Ti:sapphire laser (850 or 920 nm). Drifting sine-wave gratings (2 Hz, 50% contrast) were presented on a CRT (100 Hz refresh rate) in a variety of configurations for orientation, direction of motion, spatial frequency, ocularity (left or right eye—for ocular dominance), and eight inter-ocular spatial phase disparities (0°, 45°, 90°, 135°, 180°, 225°, 270°, 315°). For ocular dominance and binocular disparity assays, animals viewed the monoptic and dichoptic visual stimuli through ultra-fast ferroelectric liquid crystal shutters (7 KHz switching time, 1,000:1 extinction contrast ratio, DisplayTech). Each stimulus period (8 s) was preceded by an equal blank period, repeated 3–8 times. Coarse retinotopic positions of monocular receptive fields were determined by using 5° wide flashing bars of light or strips of gratings at ten retinotopic positions. Two-photon images were analyzed inMatlab (Mathworks)—see Full Methods.
Full Methods and any associated references are available in the online version of the paper at www.nature.com/nature.
Publication 2009
ADRB2 protein, human Alexa594 Animals Biological Assay Brain Calcium Cells Cortex, Cerebral Craniotomy Dominance, Ocular Dura Mater Extinction, Psychological Felis catus Fluorescence Isoflurane Light Liquid Crystals Microscopy Movement Neuroglia Neurons Operative Surgical Procedures Oregon green 488 BAPTA-1 Pressure Pulse Rate Respiratory Rate Sapphire Sepharose Short Interspersed Nucleotide Elements Vecuronium Visual Cortex
Cats (postnatal days 36–49) were anesthetized with isoflurane (1–2% in surgery, 0.5–1.0% during imaging)22 (link) and paralyzed with vecuronium bromide22 (link). A craniotomy was performed over area 18 of the visual cortex, the dura reflected, and the underlying cortex covered with agarose. Movement of the brain from respiratory and heart beat pulsations were negligible (Supplementary Fig. S8). The cell-permeant calcium indicator Oregon Green 488 Bapta-1 AM (1 mM) was prepared22 (link),30 (link) and co-loaded with 40 μM Alexa Fluor 594 into a glass patch pipette (2.5 μm diameter tip). Under continuous visual guidance, the pipette tip was advanced 200–250 μm below the cortical surface and the indicators were then pressure ejected (5–10 psi). This particular method of loading produces minimal staining of glial cells (see Ref 22 (link)) but it is possible that some of the stained cells in the present study were not neuronal. Fluorescence was monitored with a custom-built microscope (Prairie Technologies) coupled with a Mai Tai XF (Newport Spectra-Physics) mode-locked Ti:sapphire laser (850 or 920 nm). Drifting sine-wave gratings (2 Hz, 50% contrast) were presented on a CRT (100 Hz refresh rate) in a variety of configurations for orientation, direction of motion, spatial frequency, ocularity (left or right eye—for ocular dominance), and eight inter-ocular spatial phase disparities (0°, 45°, 90°, 135°, 180°, 225°, 270°, 315°). For ocular dominance and binocular disparity assays, animals viewed the monoptic and dichoptic visual stimuli through ultra-fast ferroelectric liquid crystal shutters (7 KHz switching time, 1,000:1 extinction contrast ratio, DisplayTech). Each stimulus period (8 s) was preceded by an equal blank period, repeated 3–8 times. Coarse retinotopic positions of monocular receptive fields were determined by using 5° wide flashing bars of light or strips of gratings at ten retinotopic positions. Two-photon images were analyzed inMatlab (Mathworks)—see Full Methods.
Full Methods and any associated references are available in the online version of the paper at www.nature.com/nature.
Publication 2009
ADRB2 protein, human Alexa594 Animals Biological Assay Brain Calcium Cells Cortex, Cerebral Craniotomy Dominance, Ocular Dura Mater Extinction, Psychological Felis catus Fluorescence Isoflurane Light Liquid Crystals Microscopy Movement Neuroglia Neurons Operative Surgical Procedures Oregon green 488 BAPTA-1 Pressure Pulse Rate Respiratory Rate Sapphire Sepharose Short Interspersed Nucleotide Elements Vecuronium Visual Cortex
Experimental protocols were carried out in strict accordance with the University of Rochester Committee on Animal Resources (UCAR) and conformed to the National Institute of Health’s “Guide for the Care and Use of Laboratory Animals, 8th Edition, 2011.” Experiments characterizing baseline microglial function were carried out at ages representing early development (postnatal day (p) 15), adolescence (p28), or early adulthood (p60). Experiments examining visual cortical plasticity were carried out during the visual critical period for ocular dominance plasticity, between p26 and p34. Monocular deprivations were performed between p26 and p30 by removing the right eye lid margins and suturing the lid shut. Examination of retino-geniculate projections in the lateral geniculate nucleus were carried out during the same time period, as reorganization of these projections is complete by this time and the final organization of eye-specific layers can be assessed. Experiments examining microglial infiltration into thalamocortical axon (TCA) clusters were carried out at p7 and p10 to replicate previously published methods (Hoshiko et al. 2012 (link)). Both female and male mice were included in all experiments and all mouse lines were generated on a C57Bl/6 background. C57Bl/6 (Jackson Labs), Cx3cr1-EGFP (Jung et al. 2000 (link)), Cx3cr1-knockout (Taconic Biosciences), and thy1-YFP line H (Feng et al. 2000 (link)) mouse lines were used and bred together as follows: The Cx3cr1-EGFP mouse line was used both to visualize microglia and to achieve manipulation of CX3CR1. For experiments involving in vivo imaging of microglia, because visualization of microglia requires at least one copy of GFP, Cx3cr1-EGFP heterozygous mice (Cx3cr1G/+) were used as controls. It is important to note that some studies have observed gene dosage effects in Cx3cr1-EGFP (Jung et al. 2000 (link); Lee et al. 2010 (link); Rogers et al. 2011 (link)). While this finding comes from a small subset of studies conducted under mostly pathological conditions, it is therefore possible that heterozygous mice might not behave the same as wild-type mice. However, given that these experiments cannot be carried out without a fluorescent label, this question will need to be explored using a different approach in the future. Similarly, to assure similar levels of GFP expression and therefore similar visualization of microglia in Cx3cr1-null mice as in control mice, Cx3cr1-EGFP homozygous mice (Cx3cr1G/G) were crossed to Cx3cr1-knockout mice (Cx3cr1−/−) to generate Cx3cr1-null mice with a single copy of GFP (Cx3cr1G/−). Cx3cr1G/G mice were included in imaging experiments to assay the potential impact of additional GFP expression on visualization and/or GFP toxicity. We did not observe any differences in the dynamics of microglia expressing different levels of GFP, but we cannot rule out the possibility that GFP overexpression alters microglial behavior. For experiments examining in vivo interactions between neurons and microglia, thy1-YFP mice were crossed to generate Cx3cr1G/+/YFP control mice, as well as both Cx3cr1G/−/YFP and Cx3cr1G/G/YFP Cx3cr1-null mice, again to assay the impact of additional GFP expression. For experiments examining in vivo dendritic spine turnover, YFP, Cx3cr1G/+/YFP, and Cx3cr1G/G/YFP mice were used, as microglia were not studied in these experiments.
Publication 2017
Animals Animals, Laboratory Axon Biological Assay Dendritic Spines Dominance, Ocular Eyelids Females Heterozygote Homozygote Lateral Geniculate Body Males Mice, Knockout Mice, Laboratory Microglia Neurons Pathologic Processes Visual Cortex
We wanted to measure the contribution that each eye makes to the fused binocular percept. To do this each eye views a grating of each but opposite spatial phase (−22.5° for one eye and +22.5° for the other eye). If the contribution from each eye is equal then the binocularly fused percept will be of a grating of zero phase. If the contributions are not equal then the perceived phase can be reset to zero by offsetting the contrasts in each eye (see Fig. 5). The interocular contrast ratio that produces equal contribution (i.e. zero phase) is our measure of the ocular dominance.

The temporal sequence of the binocular phase combination task and an illustration of fitting data to a binocular combination model. (a) Two trials of the binocular phase combination and two configurations are shown. There were 80 trials in the baseline test, and 30 trials in the post-patching test. The reference line was placed at the right side of the sinusoidal gratings. (b) Perceived phases from the binocular phase combination task during the baseline measurement of one subject were plotted as a function of contrast ratio. We fitted data from each measurement (baseline and post-patch) to a binocular combination model30 (link),31 (link).

Publication 2018
Dominance, Ocular Patch Tests Sinusoidal Beds
Participants viewed a gamma-corrected ViewSonic V3D231 white (x = 0.305; y = 0.315 z = 0.380) 3D LED display (refresh rate = 60 Hz; mean luminance through polarizers = 25.6 cd/m2) with polarized glasses (worn over their corrective lenses if necessary- crosstalk = 1%) from a distance of 72 cm. The display was driven by an iMac 11.1 (MacOSX, Intel Core i7, 2.8 GHz, 8 GB RAM). Stimuli were created using Psychtoolbox20 (link)21 for MATLAB (MathWorks, Natick, MA). On each trial the stimulus was presented for 200 ms, accompanied by a simultaneous auditory signal (100 ms, 880 Hz click). A single-interval, 4-alternative choice design was used. The task was to report the binocular percept using a keyboard. The responses were: single flat edge (no tilt), single right-tilted edge (i.e. right side higher), a single left-tilted edge (i.e. left side higher), or two (tilted) edges. These reflected fusion, suppression of either eye, or diplopia, respectively. Each response initiated the next trial after a 300 ms interstimulus interval.
The whole experiment consisted of two blocks of eight experimental conditions (4 contrast offsets × 2 scales), each containing 144 trials, performed before and after 3 hours of patching of the non-dominant eye. The eye dominance was established by the the Miles test22 . The total number of trials was therefore 2304. Before each block, participants aligned the percepts of the two eyes using a dichoptically presented nonius cross. Every participant was provided with at least 144 training trials to familiarize him or her with the task prior to any data collection.
Publication 2017
Auditory Perception Cross Reactions Dominance, Ocular Eye Eyeglasses Gamma Rays imidazole-4-acetic acid Lens, Crystalline

Most recents protocols related to «Dominance, Ocular»

Neuro-ophthalmological examinations were performed to exclude an ophthalmological cause for the reading difficulty. These included: Distance and near visual acuity, accommodation, refraction, crowding (the difference of near visual acuity between single optotypes and grouped optotypes—assessed by the Oculus near vision charts, Landolt rings), eye position, strabismus, motility, convergence, stereopsis, saccades, eye dominance and pupil reaction, and morphology of the outer and inner segments of the eye.
We used the Zürcher reading Test (ZLT II) [82 ], for measuring reading performance and speed, because it provides standardized and evaluated reading texts for each class from second to sixth grade. We used sentences for the 4th and 5th grade. In the test, reading time and reading errors were recorded using word lists and texts of varying difficulty. Reading speed in words per minute (wpm) was calculated as follows: Reading time in seconds divided by the sum of correctly read words multiplied by 60.
Publication 2023
Depth Perception Dominance, Ocular Motility, Cell Neoplasm Metastasis Ocular Accommodation Ocular Refraction Physical Examination Strabismus Vision Visual Acuity
A Scanning Laser Ophthalmoscope (SLO 101, Rodenstock Instruments, Ottobrunn, Germany) was used to assess the fixation locations on the stimulus. During the examination, the SLO scans stimuli directly onto the retina, which is visible simultaneously with the stimuli on a video monitor (Fig 2). This shows the absolute position of the center of the fovea relative to the stimulus with a spatial resolution of 0.1 deg. The temporal resolution of the SLO is limited by its 25 Hz sampling rate (50 half-frames/s). The SLO examination was performed on the dominant eye. Eye dominance was determined by the pinhole test [83 (link)]. The video signal could be recorded for further analysis by image processing software.
A total of 58,694 half-frames (eye coordinates) were included in our analysis, which was a very time-consuming procedure. In order to determine the fixation location, we used custom-designed image analysis software that tracks the position of a user-defined landmark, e.g. a retinal vessel branching. This detects retinal movements and allows determining the coordinates of the fovea relative to the image of the stimulus, both of which are visible on the image of the retina (Fig 2). The definition of the eye movement variables was the same as for the IR-ET method.
After separate analysis of the eye movements by both methods, we found good agreement between the data and therefore, we combined them.
Publication 2023
Dominance, Ocular Eye Movements Movement Ophthalmoscopes Radionuclide Imaging Reading Frames Retina Retinal Vessels
Human recordings were taken from two free-viewing tasks of natural images. The Ethics Committee for Human Research of Universidad de Chile approved both protocols.
In Experiment 1, Subject A freely explored natural scenes while EEG and video eye-tracking data were recorded (a female subject who had normal vision, was 28 years old, right-handed, and had right ocular dominance). The subjects explored several image types (plain gray, plain white, plain black, natural scenes on grayscale, and noise) freely. A total of 46 natural scenes from the IAPS database were used [12] . These stimuli images were interleaved with control images. A trial started with a blank screen presented for a period between 900 and 1500 ms, followed by a stimulus image explored for 3950 ms. In the current work, we further analyzed only eye traces when the subject was exploring natural scenes. Details of the methods are on [3] .
In Experiment 2, Subject B freely explored natural images where, contingent on a saccadic eye movement, we introduced a Gabor patch of 1.5–2.5 visual degrees (the subject was a 33-year-old male with normal vision and right ocular and hand dominance). Stimulus presentation was developed in Python 2.7. We further analyzed the eye trace when the subject was exploring natural scenes (n = 80 trials).
The same general setup was used on both experiments. Subjects were sited 70 cm away from the screen. Images were presented on a screen of 1920 × 1080 pixels (32 px/cm), equivalent to 39.38 pixels per visual degree. Images were 1024 × 768 pixels in the center of the screen over a mid-gray background. Screen luminance was gamma-corrected. We recorded binocular eye movements with an eye tracker (Eyelink 1000, SR Research, Ontario, Canada) at a sampling rate of 500 Hz using the pupil area to collect pupillometry. To avoid head movement, the subject rested on a chin rest during recording.
Publication 2023
Chin Dominance, Ocular Ethics Committees, Research Eye Movements Females Gamma Rays Head Movements Homo sapiens Intracisternal A-Particle Elements Males Pupil Python Vision
Repeated optical imaging of intrinsic signals and quantification of ocular dominance were performed as described (24 (link)). Briefly, during recording mice were anesthetized with 0.7% isoflurane in oxygen applied via a homemade nose mask, supplemented with a single intramuscular injection of 20 to 25 µg chlorprothixene. Images were recorded transcranially; the scalp was sutured closed at the end of each session and re-opened at the same location in subsequent sessions. Intrinsic signal images were obtained with a Dalsa 1M30 CCD camera (Dalsa, Waterloo, Canada) with a 135 × 50 mm tandem lens (Nikon Inc.) and red interference filter (610 ± 10 nm). Frames were acquired at a rate of 30 fps, temporally binned by 4 frames, and stored as 512 × 512 pixel images after binning the 1024 × 1024 camera pixels by 2 × 2 pixels spatially. The visual stimulus for recording the binocular zone, presented on a 40 × 30 cm monitor placed 25 cm in front of the mouse, consisted of 2°-wide bars, which were presented between −5° and 15° on the stimulus monitor (0° = center of the monitor aligned to center of the mouse) and moved continuously and periodically upward or downward at a speed of 10°/s. The phase and amplitude of cortical responses at the stimulus frequency were extracted by Fourier analysis as described (52 (link)). Response amplitude was an average of at least 4 measurements. Ocular dominance index (ODI) was computed as described (53 (link)). Briefly, the binocularly responsive region of interest (ROI) was chosen based on the ipsilateral eye response map after smoothing by low-pass filtering using a uniform kernel of 5 × 5 pixels and thresholding at 40% of peak response amplitude. We then computed OD score, (C−I)/(C+I), for each pixel in this ROI, where C and I represent the magnitude of response to contralateral and ipsilateral eye stimulation, followed by calculation of the ODI as the average of OD score for all responsive pixels.
The reliability of the quantitative use of intrinsic signal imaging for assessment of the strength of the responses to the two eyes has repeatedly been confirmed by showing that it provides a measure consistent with single unit recording in at least 6 papers that we have published since 2008 (24 (link)), and by 2-photon calcium imaging of the responses of single neurons (54 (link)).
Publication 2023
Calcium Chlorprothixene Cortex, Cerebral Dominance, Ocular Eye Intramuscular Injection Isoflurane Lens, Crystalline Mus Neurons Nose Oxygen Reading Frames Scalp
A total of 26 young, healthy subjects (12 females and 14 males) with a mean age of 26.0 ± 4.6 years and a mean body mass index (BMI) of 22.6 ± 3.0 kg/m2 were enrolled in this crossover study. The mean refractive error (spherical equivalent) was −1.16 ± 1.77 D. The inclusion criteria were: best-corrected monocular and binocular visual acuity ≥1.0 (in decimal notation), normal stereoacuity near and at a distance (40 arcsecs or lower), no pathology or pharmacological treatment that could affect visual performance, and no contraindication for alcohol consumption—being a social drinker with a score eight or less obtained on the alcohol use disorders identification test (AUDIT) [66 (link),67 (link)].
Finally, sensory ocular dominance was determined using the line anterior to the best visual acuity and alternating a +1.50 D lens in front of each eye under binocular viewing conditions. The dominant sensory eye was the eye with the positive lens reporting the most blurred vision [68 (link)]. Before beginning the experiment, all participants signed an informed consent form in accordance with the Declaration of Helsinki. The study was approved by the Human Research Ethics Committee of the University of Granada.
Publication 2023
Alcohol Use Disorder Dominance, Ocular Ethics Committees, Research Females Forehead Healthy Volunteers Homo sapiens Index, Body Mass Lens, Crystalline Males Pharmacotherapy Refractive Errors Vision Visual Acuity

Top products related to «Dominance, Ocular»

Sourced in United States, United Kingdom, Germany, Canada, Japan, Sweden, Austria, Morocco, Switzerland, Australia, Belgium, Italy, Netherlands, China, France, Denmark, Norway, Hungary, Malaysia, Israel, Finland, Spain
MATLAB is a high-performance programming language and numerical computing environment used for scientific and engineering calculations, data analysis, and visualization. It provides a comprehensive set of tools for solving complex mathematical and computational problems.
Sourced in United States, United Kingdom, Germany, Morocco, Sweden
MATLAB is a high-performance numerical computing software that provides a powerful programming environment for technical and scientific computing. It is designed to perform matrix and array operations, data analysis, and visualization tasks. MATLAB offers a wide range of built-in functions and toolboxes for various applications, such as signal processing, control systems, and image processing.
Sourced in United Kingdom
The VSG 2/5 is a video signal generator produced by Cambridge Research Systems. It is a multi-channel device capable of generating high-resolution visual stimuli for various experimental and research applications.
Sourced in United States, United Kingdom, Japan, Spain, Italy
SPSS is a software package developed by IBM that provides statistical analysis capabilities. It allows users to perform a wide range of statistical analyses, including data management, descriptive statistics, and advanced analytics. SPSS is designed to help users explore, visualize, and interpret data, though its specific use cases may vary depending on the needs of the user or organization.
Sourced in United States, Japan
SPSS version 26.0 is a statistical software package developed by IBM. It is designed for data management, analysis, and reporting. The software provides a user-friendly interface for working with various data types and performing a range of statistical operations.
Sourced in United States, Japan, United Kingdom, Germany, Belgium, Austria, Italy, Poland, India, Canada, Switzerland, Spain, China, Sweden, Brazil, Australia, Hong Kong
SPSS Statistics is a software package used for interactive or batched statistical analysis. It provides data access and management, analytical reporting, graphics, and modeling capabilities.
Sourced in United States
Transpore surgical tape is a medical-grade tape designed for use in healthcare settings. It is a transparent, microporous film tape that adheres securely to the skin while allowing it to breathe. The tape is designed to provide a secure and comfortable fit for a variety of medical applications.
Sourced in Japan
The KR-800 is a fully automated refraction system designed for efficient and accurate eye examinations. It features an advanced optical system and a user-friendly interface to provide reliable measurements of refractive errors.
Sourced in United Kingdom
The SpectroCAL MKII is a spectroradiometer produced by Cambridge Research Systems. It is a device used for measuring the spectral power distribution of light sources. The SpectroCAL MKII provides accurate and reliable measurements of the spectral characteristics of various light sources.
Sourced in United Kingdom
The CP-400 vision chart is a medical device designed for visual acuity assessment. It features a standardized optotype display for measuring a patient's visual function. The chart provides a consistent and reliable tool for eye care professionals to evaluate and monitor visual acuity.

More about "Dominance, Ocular"

Ocular Dominance, Eye Dominance, Preferred Eye, Dominant Eye, Visual Asymmetry, Binocular Vision, Visual Acuity, Depth Perception, Binocular Fusion, Genetic Factors, Developmental Factors, Environmental Factors, Vision Science, Ophthalmology, Research Optimization, Reproducibility, PubCompare.ai, MATLAB, VSG 2/5, SPSS Statistical Package, SPSS ver. 26.0, SPSS Statistics, Transpore Surgical Tape, KR-800, SpectroCAL MKII Spectroradiometer, CP-400 Vision Chart.
Ocular dominance, also known as eye dominance or preferred eye, refers to the tendency of one eye to be favored over the other in visual processing and perception.
This phenomenon has implications for various visual and oculomotor functions, such as visual acuity, depth perception, and binocular fusion.
The dominant eye may exhibit superior performance in these tasks, though the exact mechanisms behind ocular dominance are not fully understood.
Researchers believe that a combination of genetic, developmental, and environmental factors contribute to the establishment of ocular dominance, which can vary across individuals.
Understanding the nuances of ocular dominance is an active area of research in the fields of vision science and ophthalmology.
Researchers often utilize advanced tools like MATLAB, VSG 2/5, SPSS Statistical Package, and specialized equipment like the SpectroCAL MKII spectroradiometer and CP-400 vision chart to study this topic.
PubCompare.ai's cutting-edge AI-powered platform can assist researchers in navigating the latest literature, optimizing their research process, and enhancing the reproducibility of their findings related to ocular dominance and related visual phenomena.
By leveraging PubCompare.ai's intelligent comparison capabilities, researchers can access the best research protocols from literature, pre-prints, and patents, ultimately driving scientific discovery forward.