The largest database of trusted experimental protocols
> Physiology > Organism Function > Heat Loss

Heat Loss

Heat Loss refers to the process in which thermal energy is transferred from a warmer object or environment to a cooler one.
This phenomenon is crucial in various fields, including engineering, physics, and biology.
It occurs through the mechanisms of conduction, convection, and radiation, and can have significant implications for energy efficiency, thermal comfort, and overall system performance.
Understanding and quantifying heat loss is essential for optimizing designs, improving insulation, and enhancing the effectiveness of heating and cooling systems.
Researchers in these domains can leverage advanced tools like PubCompare.ai to enhance the reproducibility of their heat loss experiments, locate the best protocols from literature, and optimize their findings for maximum impact.

Most cited protocols related to «Heat Loss»

The whole body phenotype was analyzed at the beginning and end of the study by dual-energy X-ray absorptiometry (DEXA) utilizing a Lunar PIXImus densitometer (Lunar Corp., Madison, WI). DEXA determined the relative proportion of fat tissue mass (FTM) and bone-free lean tissue mass (LTM) as described earlier [34 (link)]. Prior to PIXImus analysis each mouse was anesthetized as described in the previous section. To facilitate recovery after the procedure, mice were injected with yohimbine (0.11 μg/g body weight), as well as with warm saline solution for rehydration. Mice were kept warm during recovery with heat pads to minimize heat loss, and checked every 10 min until recovery was complete. Whole body composition was determined by exposing the mouse (except the head region) to sequential beams of low- and high-energy X-rays with image acquisition of the X-rays impacting a luminescent panel. Bone mass was separated from soft tissue mass by measurement of the ratios of signal attenuation at the different energy levels. Soft tissue mass was further resolved into lean tissue mass (LTM) and fat tissue mass (FTM) for accurate measurement of whole body composition. Calibration was performed using a phantom mouse with known bone mineral density and FTM as described in Refs. [33 (link), 34 (link)]. Data were expressed as gain in FTM and LTM at the end of the dietary study compared to the beginning as described earlier [34 (link)].
Publication 2009
Body Composition Body Weight Bone Density Bone Tissue Diet Dual-Energy X-Ray Absorptiometry Head Heat Loss Human Body Luminescence Measure, Body Mus Phenotype Rehydration Saline Solution Silver Tissue, Adipose Tissues Yohimbine
Chlorophyll fluorescence imaging was performed using a customised FluorCam imaging fluorometer fitted with a white and red LED panel (Additional file 1: Fig. S1). Shutter time and sensitivity of the charge-coupled device (CCD) were adjusted in accordance with sample. The FluorCam is located in a temperature controlled dark room maintained between 20 and 22 °C.
Modern fluorometers commonly use a modulated light source at a known frequency to induce chlorophyll fluorescence—otherwise known as pulse-amplitude modulated (PAM) fluorescence—where the detector is set to measure at the same frequency as the excitation [37 (link)]. This methodology allows measurements to occur when the plant is illuminated. During a typical measurement, the plant is dark adapted (between 20 and 60 min) to allow maximal plastoquinone A (QA) oxidation after which the leaf is exposed to a saturating flash of light that maximally reduces QA, closing all PSII reaction centres. This procedure gives a maximum fluorescence value (Fm) and, in the light, allows the separation of the photochemical (e.g. PSII operating efficiency—Fq′/Fm′) and non-photochemical (e.g. Non-photochemical quenching—NPQ) processes in the leaf under specific photosynthetic photon flux density (PPFD) conditions. The parameter Fq′/Fm′, also termed ɸPSII or quantum yield (QY), is a measure of the proportion of absorbed light utilised by PSII and therefore can also be used, in combination with measurements of leaf absorbance, to calculate linear electron transport rate (ETR). These parameters (Table 1) are key in the identification of differences between different lines, treatments (biotic or abiotic) or genotypes [20 (link)–22 (link), 38 (link), 39 ]. Many instruments are available for assessment of these parameters either as spot measurements or as images. The benefit of imaging chlorophyll fluorescence is the ability to analyse both temporal and spatial variation in PSII efficiency [28 (link)].

Commonly used abbreviations and equations employed when measuring chlorophyll fluorescence

ParameterFormulaDefinition
F, F′, Fsn/aSteady state fluorescence emission from dark- or light-adapted (‘) leaf, respectively. F′ is sometimes referred to as Fs′ when at steady state.
Fm, Fmn/aMaximal chlorophyll fluorescence measured in a dark- or light-adapted state respectively
Fo, Fon/aMinimal chlorophyll fluorescence measured in a dark- or light-adapted state respectively
Fv, Fvn/aVariable chlorophyll fluorescence measured as the difference between dark- or light-adapted Fm/Fm′ and Fo/Fo′.
Fv/Fm(FmFo)/FmMaximum quantum efficiency of PSII.
Fv′/Fm(Fm′–Fo′)/FmMaximum efficiency of PSII in the light.
Fq′/Fm(Fm′–F′)/FmPSII operating efficiency: the quantum efficiency of PSII electron transport in the light. AKA ΦPSII, quantum yield or ΔF/Fm
ETR or JΦPSII (AKA Fq′/Fm′) × PPFDa × (0.5)Linear electron transport rate; where PPFDa is absorbed light (μmol m−2 s−1) and 0.5 is a factor that accounts for the partitioning of energy between PSII and PSI.
NPQ(FmFm′)/FmNon-photochemical quenching: estimates the rate constant for heat loss from PSII.
qL(Fq′/Fv′)/(Fo′/F′)Estimates the fraction of open PSII centers (QA oxidized); considered a more accurate indicator of the PSII redox state than qP
qP(Fm′–F′)/(Fm′–Fo′) AKA Fq′/FvPhotochemical quenching: relates PSII maximum efficiency to operating efficiency. Non-linearly related to proportion of PSII centers that are open. 1–qP has also been used to denote proportion of closed centers

A summary table of the commonly used chlorophyll fluorescence parameters and corresponding equations. For a more comprehensive review please refer to Murchie and Lawson [21 (link)], Baker [20 (link)] and Maxwell and Johnson [22 (link)]

Full text: Click here
Publication 2019
Chlorophyll Electron Transport Fluorescence Genotype Heat Loss Hypersensitivity Light Medical Devices Oxidation-Reduction Photosynthesis Plant Leaves Plants Plastoquinone Pulse Rate
The hand immersion bath had a dimension of 49.5 (l) × 29 (w) × 13 (h) cm, and the foot water bath had a dimension of 46 (l) × 36 (w) × 22.5 (h) cm. To minimize the heat loss or heat gain of the water in the baths, they were insulated with polystyrene. The temperature was controlled within 1°C using a thermostat bath (TLC 15, PM Tamson Instruments, Bleiswijk, The Netherlands).
Finger and toe temperatures were continuously monitored using thermistors (type P-8432, ICBT, Tokyo, Japan) attached to the skin by one layer of Leukoplast tape (BSN medical & GmbH & Co.KG, D-22771, Hamburg, Germany) and connected to a Mobi8 data acquisition system (TMS International BV, Oldenzaal, The Netherlands). The temperature of the fingers and toes was sampled every second. The lowest value over the 30 min immersion interval was defined as the minimum temperature (Tmin). The mean (Tmean) and maximum (Tmax) temperatures were calculated over the 5- to 30-min interval. CIVD reactions were defined as a continuous rise of at least 1°C. To exclude minor fluctuations, we averaged the values over a period 20 s before and 20 s after the measurement for all temperatures. When the rise was <1°C, the response was counted as ‘No CIVD’, when it was 1°C or more it was counted as a CIVD response. The onset time is the time in seconds from start of the immersion until the start of a continuous increase of temperature of at least 1°C. Tpeak is the temperature at the peak of the first CIVD wave. The CIVD analysis was completely automated to exclude human subjectivity.
Pain was assessed every 5 min using a 0–10 visual analog scale (VAS) Numeric Pain Distress scale. Tactile sensitivity at the tip of the index finger was assessed using Semmes–Weinstein monofilaments (Bell-Krotoski and Tomancik 1987 ). The subjects turned the hand under water every 5 min for about 10 s to enable determination of tactile sensitivity.
Publication 2011
Bath Fever Fingers Foot Heat Loss Homo sapiens Hypersensitivity Leukoplasts Neoplasm Metastasis Pain Polystyrenes Skin Submersion Toes Visual Analog Pain Scale

Protocol full text hidden due to copyright restrictions

Open the protocol to access the free full text link

Publication 2011
Animals Calorimetry Heat Loss Obstetric Delivery Telemetry
The objective of the computational study was the investigation of means to improve the performance of PCB-based microPCR in terms of duration, energy consumption, and temperature uniformity for a two-temperature (65 °C and 95 °C) PCR protocol. The geometry of the microPCR used in the computational study is shown in Figure 2. It included a PCB layer (with thickness of 1.68 mm) with an embedded microheater, i.e., a copper line. The thickness, width, and total length of the copper line were 25 μm, 100 μm, and 4.973 m. On top of the PCB layer, there was a PMMA layer (with thickness of 300 μm) with meander-shaped microchannel which was sealed with a polyolefin layer (with thickness of 50 μm). The depth, width, and total volume of the microchannel were 100 μm, 2.5 mm, and 41 μL, respectively. The footprint of the microPCR design was 54.1 × 24.9 mm2. The footprint as well as the stack of materials in the PCB layer were very close to the thin chip used in the experiments (cf. Section 2.1).
The study was based on a modeling framework [19 (link)], including an energy balance in the solid domains and the fluid in the microchannel, which read: ρCpdTdt=(kT)+Q,
where T, k, ρ, and Cp are the temperature, the thermal conductivity, the density, and the heat capacity of the solid or the fluid. Q is the heat generation rate at the microheater. It was zero for all domains except for the microheater.
The Joule heating mechanism and the details of the geometry of the microheaters (operating as resistances) were taken into account. The heat generation rate according to the Joule heating read: Q=JE.
E is the electric field in the microheater and J is the current density, which read: J=σE,
and was calculated by the current conservation equation, i.e.,
J=0.
σ is the electrical conductivity of the microheater, which, for the case of copper, was linear, with the following formula
σ=1ρ0[1+α(TT0)],
where ρ0 is the electrical resistivity at temperature equal to T0, and a is the temperature coefficient of resistivity.
Heat losses by convection and radiation were applied on all external surfaces of the device. The heat transfer coefficient is a function of the surface temperature, the latter coming from a computational study for the heat losses of microfluidic devices [29 ]. A time varying voltage was applied across the microheater in order to achieve the desired thermal cycle, resembling the functionality of a simplified temperature controller. During heating, a constant voltage was applied, during cooling the temperature controller was switched off. Finally, electrical insulation was applied to all other boundaries of the heaters.
The numerical calculations required for the study were performed by the finite element method implemented with the commercial code COMSOL (COMSOL Inc., Stockholm, Sweden).
Full text: Click here
Publication 2020
Convection Copper DNA Chips Electric Conductivity Electricity Heat Loss Medical Devices Microchip Analytical Devices Polymethyl Methacrylate polyolefin Radiation

Most recents protocols related to «Heat Loss»

Example 3

This Example describes a method for modeling a reactor which is configured to allow torrefaction therein. The behavior of the reactor predicted by the model is compared to data generated experimentally from a reactor with similar characteristics to those modeled. In the present Example, a mathematical description of the reactor is developed. This mathematical description may produce reasonable fit to experimental data. It is demonstrated that at the small test-reactor scale, heat loss mechanism through the side wall may affect biomass torrefaction. Furthermore, by interrogating the scaling behaviors of the reactor, it is shown that as the reactor is scaled up, at the same operating condition, the mass yield of the torrefied biomass may improve by 10-20%.

Full text: Click here
Patent 2024
Heat Loss
A 20 MHz 33220A function/arbitrary waveform generator (Agilent Technologies Inc., Santa Clara, CA, USA) was used to drive the coils. Amplification of this input signal was performed using a 1000 W RF power amplifier, model 1140LA (Electronics & Innovation Ltd., Rochester, NY, USA). The amplified signal runs through the L-matching circuit that enables maximum power transfer from the source to load (coil). Since the coil dissipates heat due to resistive losses, a Masterflex® peristaltic pump circulates water through the instrument’s body to remove excess heat. A simple air-cooled heat exchanger configuration maintains the circulating water at a constant temperature of about 24 °C to avoid the temperature increasing throughout the duration of the experiment. Temperatures of the test samples (e.g. cells, MNPs) and surroundings were monitored with a FOTEMP-PLUS signal conditioner using TS3–10 mm-06 fiber optic temperature sensors (Optocon AG, Dresden, Germany) and stored in a computer using the FOTEMP-ASSISTANT 2.3 configuration and data logging software, also from Optocon AG. The temperature of the samples was measured by passing the fiber optic thermal sensor through a lid with a hole and the surrounding temperature was measured by placing another fiber optic sensor close to the plate. The reusable lid was cleaned with ethanol after each run to avoid contamination. Subsection E in Supplementary Information shows top-view and side-view of the samples placed on the devices (see figure S9 and figure S10), as well as an example of the temperatures reached during MFH (see figures S11 and S12).
Full text: Click here
Publication 2023
Cells Ethanol Heat Loss Medical Devices Peristalsis Water, Body
Participants wore a long-sleeved t-shirt (100% cotton, 160 g), a ventilation jacket (100% cotton, 430 g), undershorts (100% polyester, 80 g), work pants (100% cotton, 315 g), socks (100% cotton, 40 g), and shoes in all three trials. The intrinsic clothing insulation value of the jacket was 0.21 clo and that for the total clothing was 0.53 clo, was determined using a thermal manikin (Thermal manikin, Kyoto Electronics Manufacturing Co., Ltd., Tokyo, Japan). The size of the t-shirt was selected for each participant to adhere tightly to their body surface. To increase the wetted surface area, the long-sleeved type of t-shirt was selected. The ventilation jacket (KU91400, Kuchofuku, Tokyo, Japan) was long sleeved and equipped with two fans (8 cm blade diameter, 194 g) on the lower back (Supplementary Figure S1). The fans used a battery box (4 AA nickel-metal hydride, 140 g) and took in outside air. The airflow rate of the fan was 15.1 L s−1, which was set at the maximum of the apparatus because heat loss increases with ventilation flow rates (Yang et al., 2022 (link)). The jacket was tightened around the waist with an elastic band, and then, the air was exhausted from the cuff and neck openings. The airflow velocity underneath clothing was 4.5, 2.5, and 7.0 m s−1 at the cuff, in front of the neck, and behind the neck, respectively (measured using an anemometer; 6006, Kanomax Japan Inc., Suita, Japan).
After baseline measurements at 25°C (50% RH, <0.3 m/s air velocity), the t-shirt was soaked with 350 ± 5 mL of tap water (37°C) using an electric vaporizer in the WET trial, and water was never added later. This amount of water was chosen to saturate the t-shirt without leaving dry spots or dripping. To ensure the volume of water, the clothed body weight was measured before and immediately after the soaking (±5 g error) because the participants had trouble wearing a pre-determined and pre-soaked t-shirt. At the same time, the fans of the ventilation jacket were turned on in the DRY and WET trials. In the CON trial, the fans remained off. Immediately after each preparation of the clothing, the room temperature was elevated to 37°C (50% RH, <0.3 m/s air velocity) and stabilized within 10 min. The participants remained seated during the Ta transition, then they performed three 20-min bouts of walking exercise (Ex1, Ex2, and Ex3) separated by 10-min breaks (B1, B2, and B3). The walking was conducted at a predefined speed (all participants, 4.5 km h−1) and inclines (young, 6.6% ± 2.4%; older, 3.4% ± 2.1%) for a target heat production of 200 W m-2 on the treadmill (%VO2peak, young 39% ± 9%, older 54% ± 8%, as averaged across trials). During the breaks, drinking water (37°C) was provided ad libitum.
Full text: Click here
Publication 2023
Body Weight Electricity Exanthema Gossypium Heat Loss Human Body M-200 Metals Neck Nickel Polyesters Respiratory Rate Thermogenesis Vaporizers
All behavioural testing was performed during the light phase of the 12 h light/dark cycle in a sound-controlled room with diffuse overhead lighting. Testing was conducted in the mornings prior to feeding. Additionally, control testing was interspersed throughout the treatment conditions to account for any time-of-day effects and other confounding variables. Fish were randomly assigned to either a control group or one of the three treatment groups. During testing, a three-sided enclosure of white corrugated plastic encircled the testing arena to minimize exposure to external stimuli. Water temperature was maintained between 26 °C and 28 °C. The arena was placed on a pad heated to 35 °C to reduce heat loss between trials. Luminance in all testing arenas was measured at ~ 32 cd/m3 (cal SPOT photometer; Cooke Corp. CA, USA). A Basler GenICam acA1300-60gc Area Scan video camera (Basler Inc., USA) was suspended approximately 1 m above testing arenas to record zebrafish behaviour. Zebrafish movement was tracked using Noldus EthoVision XT ® tracking software (v. 11.0, Noldus, Wageningen, NL) using differencing settings. Quantification of behaviour began immediately after the fish or shoal was placed in the center of the arena. The time the fish was immobile (immobility) was quantified in EthoVision with a 5% threshold49 (link),50 (link). The individual tests, novel tank dive test (n = 80) and light/dark test (n = 81), consisted of ~ 20 fish per group for each 0, 1, 5, and 15 mg/L concentration of chlordiazepoxide. For the shoaling test (n = 250), each shoal consisted of 5 fish. There were 10 trials conducted in each of the shoaling treatment groups, however, in the control group 20 trials were conducted as they were interspersed throughout the testing trials. Data were only excluded from analyses if the tracking software did not acquire data for the total time spent in arena.
Full text: Click here
Publication 2023
Chlordiazepoxide Fishes Heat Loss Light Movement Radionuclide Imaging Sound Zebrafish
Upon exposure to the atmosphere following the hydrogen explosion, CsMP experienced rapid cooling. It was assumed convective heat loss occurred only at the surface, with conductive cooling due to the heat gradient formed between the centre and surface36 (link) dominating within the particle bulk. To further simplify the model, radiative heat loss at the particle surface was taken to be negligible. The radial and temporal heat profile of the particle due to just conductive heat loss was then modelled by solving the one-dimensional, spherically symmetric heat equation ρpcpTt=kpr2rr2Tr, where ρp is density of the particle, cp the heat capacity of the particle, kp the thermal conductivity and r the radial coordinate. The physical particle properties (e.g. density, heat capacity, thermal conductivity) were assumed to be constant throughout the model. The convective heat loss at the particle surface was accounted for by imposing the boundary condition kpTrr=R=-qc, where qc=h(T-Ts) is the convective heat flux. h is the heat transfer coefficient, Ts is the surface temperature of the particle and T the temperature of the surroundings23 (link). The heat transfer coefficient was calculated using the equation h=Nukg2rp, where Nu is the Nusselt number, kg the thermal conductivity of the air surrounding the particle and rp the particle radius. Calculation of the Nusselt number required the particle Reynolds number, Re, determined by Re=2|(vg-vp)|rpρgηg, where vg is the velocity of the surrounding air, vp the particle velocity, ρg the air density and ηg the air viscosity. In previous studies26 (link), the Nusselt number was calculated for a negligible internal temperature gradient by invoking the lumped capacitance approximation. This states that for Biot numbers close to zero (Bi 0 ) , the convective heat transfer to the surrounding gas limits the surface heat flux, and internal conduction is large enough to equilibrate the internal temperature gradient of the pyroclast. This contradicts our key hypothesis that radial variations in viscosity caused the unique internal texture observed in ‘Type B’ CsMP. For this reason the Nusselt number used in this study was calculated using newer data from Moitra et al.37 (link), that is not dependent on the lumped capacitance approximation: Nu=a+bRe1/2Pr1/3 with fitting parameters a=76 and b=1.9 , and Pr is the Prandtl number, taken to be 0.71 for ambient air37 (link).

Schematic outlining the physical processes at each scale of the model. Particle-scale: the cooling of the isotropic, spherical particle from convective heat transfer to the surrounding environment and conductive cooling within the particle. Bubble-scale: The bubble growth model in which growth is limited by viscosity and halted once the temperature is lower then the glass transition temperature.

Full text: Click here
Publication 2023
Atmosphere Blast Injuries Convection Dietary Fiber Electric Conductivity Heat Loss Hydrogen Physical Processes Radius Surface Radiotherapy Viscosity Vitrification

Top products related to «Heat Loss»

Sourced in United States, Sweden
COMSOL Multiphysics 5.5 is a powerful software package for modeling and simulating physics-based problems. It provides a comprehensive environment for applying numerical methods to a variety of engineering and scientific applications. The software supports multiple physics interfaces, allowing users to couple different physical phenomena within a single simulation.
Sourced in United States, Germany, Italy
The Stereotaxic frame is a laboratory instrument used to immobilize and position the head of a subject, typically an animal, during surgical or experimental procedures. It provides a secure and reproducible method for aligning the subject's head in a three-dimensional coordinate system to enable precise targeting of specific brain regions.
Sourced in United States
The ECM 830 is a single-channel, computer-controlled electrophysiology stimulator designed for tissue culture and in vitro studies. It provides programmable electrical stimulation with adjustable voltage, current, and pulse duration parameters. The device is capable of delivering both monophasic and biphasic stimuli to biological samples.
Sourced in United States
F7252 is a laboratory equipment product. It is designed to perform specific functions within a laboratory setting. The core function of F7252 is to facilitate certain laboratory processes, but a detailed description cannot be provided while maintaining an unbiased and factual approach.
Sourced in Germany, United States
The Dual-PAM-100 is a laboratory instrument designed for simultaneous measurement of chlorophyll a fluorescence and P700 absorbance changes. It provides reliable data on the photosynthetic performance of plants and algae.
Sourced in United States, Sweden, Germany
COMSOL Multiphysics is a simulation software that allows users to model and analyze multi-physics problems. It provides a unified environment for finite element analysis, solver, and visualization tools to solve a variety of engineering and scientific problems.
Sourced in China
Calcein-AM/PI double staining kit is a laboratory tool used to assess cell viability. The kit contains Calcein-AM and Propidium Iodide (PI), which are fluorescent dyes that enable the simultaneous detection of live and dead cells, respectively, through fluorescence microscopy or flow cytometry.
Sourced in United States
ExpeData is a data acquisition and analysis software for laboratory equipment. It is designed to record and process data from various scientific instruments. The software provides tools for data visualization, analysis, and management.
Sourced in Japan
The EZ Test is a compact and versatile laboratory equipment designed for a wide range of analytical applications. It offers reliable and accurate measurements, providing essential data for research and testing purposes.
Sourced in United States
The Model 687 is a laboratory instrument designed to measure the displacement of a sample in response to an applied load. It features a high-resolution displacement transducer, a programmable load cell, and a user-friendly software interface. The core function of the Model 687 is to provide accurate and reliable measurements of the mechanical properties of materials under controlled loading conditions.

More about "Heat Loss"

Thermal Energy Transfer: Exploring the Dynamics of Heat Loss.
Heat transfer is a fundamental process in various scientific and engineering disciplines, where the flow of thermal energy from a warmer object or environment to a cooler one is a crucial consideration.
This phenomenon, known as heat loss, can occur through conduction, convection, and radiation, and has significant implications for energy efficiency, thermal comfort, and overall system performance.
Conduction, the transfer of heat through direct contact between materials, plays a vital role in heat loss.
Understanding the thermal conductivity of different materials, as measured by tools like the EZ Test and COMSOL Multiphysics 5.5, is essential for optimizing insulation and minimizing heat transfer.
Convection, the movement of heat through the flow of fluids like air or water, is another key mechanism of heat loss, influenced by factors such as air velocity and temperature gradients.
Radiation, the emission of electromagnetic waves, can also contribute to heat loss, particularly in systems with significant temperature differences, like those involving a Stereotaxic frame or Dual-PAM-100 device.
Quantifying and predicting heat loss is crucial for enhancing the performance of heating, ventilation, and air conditioning (HVAC) systems, as well as industrial processes and biological applications.
Researchers in these domains can leverage advanced tools like the ExpeData software and ECM 830 to collect and analyze data on heat transfer, ensuring the reproducibility and reliability of their findings.
By optimizing heat loss through improved insulation, airflow management, and other strategies, practitioners can enhance energy efficiency, reduce operating costs, and improve the overall comfort and well-being of occupants.
In summary, the study of heat loss, facilitated by technologies like COMSOL Multiphysics, Calcein-AM/PI double staining kit, and F7252, is a vital area of research and application, with far-reaching implications across various fields.
Mastering the principles of thermal energy transfer and leveraging the latest tools and techniques can unlock new possibilities for enhancing energy efficiency, thermal comfort, and overall system performance.