The largest database of trusted experimental protocols
> Procedures > Diagnostic Procedure > Skinfold Thickness

Skinfold Thickness

Skinfold Thickness refers to the measurement of the thickness of a fold of skin and the underlying subcutaneuous tissue, typically used to estimate body composition and fat distribution.
This technique involves pinching the skin at specific anatomical sites, such as the biceps, triceps, subscapular, or suprailiac regions, and using a caliper to measure the thickness of the skinfold.
Skinfold thickness is a non-invasive, cost-effective method for assessing body fat percentage and is commonly used in clinical, fitness, and research settings.
However, accurate and reliable measurement requires adherence to standardized protocols and proper technique to minimize variability and ensure reproducibility of results.

Most cited protocols related to «Skinfold Thickness»

At present, simple, accurate methods for measuring percent of body fat and, in particular, body fat in different fat depots are not available. The indirect methods currently in use for estimating total percent of body fat include underwater weighing, an air displacement and density determination using a Bod Pod, a bioelectrical impedance analyzer, and a determination of the isotopically labeled water mass. In the past, determination of the total body radioactive potassium and thus metabolizing tissue mass have been used to estimate lean body mass, and by difference, the fat mass.86 (link)
Anthropometric determination of fat mass directly has been done using skin-fold thickness measured at various sites.87 (link) A dual-energy x-ray absorptiometry (DEXA) scan, which provides a 3-dimensional picture of body organ densities, can be used for estimating total body fat. Its location also can be determined. Single computed tomography (CT) slices of the abdomen and thigh can be used to obtain 2 dimensions of those fat depots from which a 3-dimensional fat area can be reconstructed. This also can be done using magnetic resonance imaging, but magnetic resonance imaging is very expensive. One cannot do serial sections of the body using CT to determine fat mass because of the excess radiation associated with this procedure.
Because of their convenience, bioelectric impedance methods or DEXA scans are the most commonly used to estimate the amount and, with DEXA scans, the location of body fat depots. Estimates of abdominal and thigh fat depots also can be estimated using CT slices.52 (link),72 (link),88 (link)
All of the previously mentioned methods use certain assumptions in the calculation of body fat mass, and all are subject to potential error. Nevertheless, there are more specific methods of determining body fat mass than is the BMI. Important information regarding the location of the stored fat also can be determined with some methods.
It now is generally accepted that a relationship between BMI and mortality risk should be applied only to large populations. It should not be applied to an individual in an unqualified fashion. As indicated previously, there is the issue of being “overweight” versus “over fat.” In addition, a segment of the population is now considered to be “fat” by any criteria but “fit” and not at risk for early mortality.74 (link),75 (link),89 (link)–91 (link)
Publication 2015
Abdomen Bioelectrical Impedance Body Fat Human Body Potassium Radiation Radioactivity Skinfold Thickness Thigh Tissues X-Ray Computed Tomography
START has been granted ethical approval locally from the Research Ethics Board, Hamilton Health Sciences/McMaster Health Sciences (REB#: 10-640) and in India, Institutional Ethics Review Board Reference #: 114/2010). In both countries, pregnant mothers are recruited during their antenatal visits (1st or 2nd trimester) to their primary care practitioner or obstetrician. The study is described by the study personnel to the pregnant mothers and consent for participation is obtained. Information concerning medical and pregnancy history, health status, health behaviors, and socioeconomic status is obtained by questionnaires. Anthropometric measurements (height, weight, skinfold thickness), blood pressure, urine sample, and a fasting blood sample for glucose, insulin, micronutrients (i.e. vitamin B12, RBC folate, plasma homocysteine, methylmalonic acid MMA), lipids and a buffy coat for future DNA extraction will be collected, and processed using a standardized protocol at 24-28 weeks of gestation. Mothers who are not known to have diabetes will undergo a 75 oral glucose tolerance test between 24-28 weeks gestation. The results of an ultrasound performed between 18-24 weeks to assess for congenital anomalies and for precise determination of gestational age will be collected from each pregnant mother. At the time of delivery, details of the delivery, birth outcomes for the mother and baby will be collected, and a cord blood sample for DNA, glucose, insulin, lipids and additional aliquots for future analysis of adiponectin, and leptin will be taken from each baby. The placenta will be weighed, and where possible a biopsy of the placenta will be collected and stored in RNAlater for future analysis of RNA and methylation patterns. In addition, the infant’s anthropometry including birth weight, triceps and sub-scapular skin fold thickness, length, abdominal, head, and arm circumference will be measured by a trained research assistant.
Full text: Click here
Publication 2013
Abdomen Adiponectin Biopsy Birth Birth Weight Blood Glucose Blood Pressure Cobalamins Congenital Abnormality Diabetes Mellitus DNA, A-Form Folate Gestational Age Glucose Head Homocysteine Infant Insulin Leptin Lipids Methylation Methylmalonic Acid Micronutrients Mothers Obstetric Delivery Obstetrician Oral Glucose Tolerance Test Placenta Plasma Pregnancy Primary Health Care Scapula Skinfold Thickness Ultrasonography Umbilical Cord Blood Urine
At the in-person visit, trained research assistants measured height to the nearest 0.1 cm using a calibrated stadiometer (Shorr Productions, Olney, Maryland) and weight to the nearest 0.1 kg using a calibrated scale (Tanita model TBF-300A, Tanita Corporation of America, Inc., Arlington Heights, IL). We computed each child’s BMI using the following formula: BMI=weight/height2 (kg/m2). We calculated age-sex-adjusted BMI z-score and BMI percentile using the 2000 Centers for Disease Control and Prevention reference data [20 ].
We measured subscapular (SS) and triceps (TR) skinfold thicknesses to the nearest 0.1 mm using Holtain calipers (Holtain Ltd, Crosswell, Wales) and calculated the sum (SS + TR) and ratio (SS:TR) of the two thicknesses. The correlations of other measures of adiposity with subscapular or triceps thickness individually were very similar to the correlations with sum of the two, so we chose to show results for only sum of skinfolds. We measured hip and waist circumferences to the nearest 0.1 cm using a Hoechstmass measuring tape (Hoechstmass Balzer GmbH, Sulzbach, Germany), and calculated waist-to-hip circumference ratios. We measured middle upper arm circumference using a Ross measuring tape (Ross Products Division, Abbott Laboratories Inc., Columbus, OH).
Research assistants performing the measurements followed standardized techniques [21 ] and participated in biannual in-service training to ensure measurement validity. Inter- and intra-rater measurement errors were within published reference ranges for all of the measurements [22 ]. Experienced field supervisors provided ongoing quality control by observing and correcting measurement technique every 3 months.
We measured bipolar bioelectrical impedance using a Tanita scale model TBF-300A (Tanita Corporation of America, Inc., Arlington Heights, IL) foot-to-foot body composition analyzer. We calculated fat mass and fat-free mass indices for DXA and bioelectrical impedance measurements using the following formula: (mass in kg)/(height in meters)2.
Trained research assistants performed whole body DXA scans on the children (n=875) using a Hologic model Discovery A (Hologic, Bedford, MA) that they checked for quality control daily by scanning a standard synthetic spine to check for machine drift. We used Hologic software QDR version 12.6 for scan analysis. A single trained investigator (CEB) checked all scans for positioning, movement, and artifacts, and defined body regions for analysis. Intrarater reliability on duplicate measurements was high (r=0.99).
Full text: Click here
Publication 2013
Adiposity Arm, Upper Bioelectrical Impedance Body Composition Body Regions Child Foot Movement Radionuclide Imaging Skinfold Thickness Vertebral Column Waist-Hip Ratio Waist Circumference Whole Body Imaging
A single survey team including three nurse observers carried out all measurements during school terms between October 2004 and February 2007; each measured approximately one-third of all children in each ethnic group and visited schools in different parts of London, Leicester and Birmingham in rotation. The observers received initial training in all measurement techniques in accordance with standard recommendations.24 (link) The observers’ measurement performance was reviewed before the study and at regular intervals during the study. For each anthropometric measure, a single measurement was obtained for each child. Height was measured to the last complete millimetre using a portable stadiometer (Chasmors Ltd, London, UK). Weight was measured to the nearest 0.1 kg using an electronic digital scale (Tanita Inc., Tokyo, Japan). Waist circumference was measured at the mid-point between the lower margin of the ribs and the pelvic crest in the mid-axillary line. Skinfold thickness was measured in subscapular, suprailiac, biceps and triceps locations (right sided) using a Holtain skinfold caliper. Sum of all skinfolds was based on the sum of all four measurements of skinfold thickness, sum of trunk skinfolds on the sum of subscapular and suprailiac skinfolds and sum of limb skinfolds on the sum of biceps and triceps skinfolds. Leg to arm bioimpedance (right sided) was measured using the Bodystat 1500 bioimpedance monitor (Bodystat Ltd, Isle of Man, UK). Fat free mass was derived using an equation derived specifically for children of a similar age using DXA validation:25

Fat mass was obtained by subtracting fat free mass from total body weight.
Pubertal status was assessed in girls using the Tanner breast development scoring system.26
Publication 2010
Axilla Body Weight Breast Child Costal Arch Crista Ampullaris Ethnicity Nurses Pelvis Puberty Skinfold Thickness Waist Circumference Woman
We used data from the National Health and Nutrition Examination Survey (NHANES) from 8 cycles conducted from 1999-2000 through 2013-2014 (17 ). NHANES employs a multi-stage, stratified, cluster sampling design to select a representative sample of the US civilian, non-institutionalized population. The surveys were approved by the ethics review board, and parental permission was obtained for subjects < 18 y of age.
We focus on 2- to 19-year-olds who had weight and height measurements, and after excluding pregnant girls, the resulting sample size was 30,003. Race and ethnicity were self-reported, and subjects were classified as white non-Hispanic, black non-Hispanic, Mexican-American, or other (which includes other Hispanics and multi-racial persons). Weight, height, waist circumference, and mid upper-arm circumference were measured in a standardized fashion (18 ). Triceps skinfold thickness was measured through the 2009-10 surveys, but not in more recent cycles.
DXA scans were acquired in NHANES 1999-2006 for boys and non-pregnant girls who were ≥ 8 y using a Hologic QDR 4500A fan-beam densitometer (Hologic Inc., Bedford MA) (19 , 20 ). The current analyses focus on fat mass (kg) and fat mass index (fat mass/ht2). The 1999-2000 DXA data for girls are not available in the publicly released data and are not included in the current analyses.
We used the NHANES DXA Multiple Imputation Data Files (19 ) in the analyses. These imputations, which were performed by the National Center for Health Statistics, used sequential regression to impute (estimate) missing DXA values for 5 complete datasets from non-missing DXA measurements and characteristics such as sex, race-ethnicity, age, BMI and waist circumference (21 (link)).
Publication 2017
Arm, Upper Boys Ethnicity Hispanics Mexican Americans Parent Sex Characteristics Skinfold Thickness Waist Circumference Woman

Most recents protocols related to «Skinfold Thickness»

Fourteen trained, and recreationally trained cyclists (7 females & 7 males, 74.1 ± 10.5 kg, 32.1 ± 7.6 years of age, 170.0 ± 11.0 cm, 11.3 ± 5.4 mm VL skinfold thickness, and 55.0 ± 9.1 ml·kg·min−1 maximum oxygen uptake) volunteered and provided written informed consent to participate in the study (19 (link)). To obtain sufficient power of β = 0.8 with α = 0.05, an a priori sample size calculation was made in G*Power software (version 3.1.9.7, Kiel, Germany) using previously reported data from other groups that compared SmO2 values within and between sessions during ramp incremental tests and severe intensity efforts (16 (link), 20 (link)–23 (link)). This study was approved by the research ethics committee of The University of British Columbia and was conducted in accordance with principles established in the Declaration of Helsinki, except for registration in a database.
Full text: Click here
Publication 2023
Ethics Committees, Research Females Males Oxygen Skinfold Thickness
Two wearable NIRS sensors (Moxy Monitor, Fortiori Design LLC., Hutchinson, USA) were used during the test. The Moxy monitor employs four wavelengths of near-infrared light (680, 720, 760, and 800 nm), with source detector separation of 12.5 and 25 mm (10 (link)). The sensors were placed on the right and left VL, and the right side was used for analysis and the left was used in case of right sensor failure. In all but one participant, the right sensor was used. The anatomical location on the VL was 1/3 the distance from the proximal pole of the patella to the greater trochanter. Left and right sensors were held in place by the participants' elastic cycling shorts, and both sensors were covered using a light shield supplied by the manufacturer to minimize ultraviolet light interference.
During instrumentation, skinfold thickness was measured and recorded from the right VL with a Harpenden skinfold caliper (Creative Health, Dallas, USA). According to the manufacturer, the Moxy sensor does not require calibration (10 (link), 11 (link)). Prior to each trial, the sensor was charged, and both the emitter and receiving optodes were cleaned.
Full text: Click here
Publication 2023
ARID1A protein, human Infrared Rays Light Patella Skinfold Thickness Spectroscopy, Near-Infrared Trochanters, Greater Ultraviolet Rays
Anthropometric data were assessed, including standing body height, leg length, body mass (BM), and body fat percentage (% fat). All measurements were taken by the same person three times using techniques established by the international biological program [31 ]. Circumferences and skin-fold thickness at different levels of the thigh and the calf, the length of the lower, and the femoral condyles breadth are measured to estimate the muscle volume of the lower limbs [29 (link)]. Muscle volumes were estimated in accordance with Eq. 1 Musclevolume=totallimbvolume-fatvolume+bonevolume
The total limb volume was estimated as the volume of a cylinder, based on its length (L), corresponding to the distance from the trochanter major to the lateral malleolus for the lower limb, and the mean of five limb circumferences lower limbs (for the maximal thigh, mid-thigh, just below the patella, maximal calf and just above the ankle) in accordance with Eq. 2 Totallimbvolume=C2·L/62.8 where ∑C2 is the sum of the squares of the five circumferences of the corresponding limb. Skin folds were assessed using a standard Harpenden calliper (Baty International, Burgess Hill, Sussex, UK). The fat volume was calculated using Eq. 3 C/5·S/2nL where ∑S is the sum of four skinfolds for the lower limb (front of mid-thigh, back of mid-thigh, back of calf and outside of calf) and ''n'' represents the number of skin folds measured.
Bone volume was calculated as π·F·D2·L where D is the femoral intercondylar diameter, F is a geometric factor (0.235 for the lower limb), and L is the limb length as measured above.
Bone volume was calculated as π ∙ (F ∙ D) 2∙ L, where D is the femoral intercondylar diameter, F is a geometric factor (0.235 for the lower limb), and L is the limb length as measured above.
Full text: Click here
Publication 2023
Ankle Biopharmaceuticals Body Fat Body Height Bones Condyle Femur Human Body Lower Extremity Muscle Tissue Patella Skin Skinfold Thickness Thigh Trochanters, Greater
Patients’ weight and height were measured with a calibrated scale and stadiometer, respectively. Subsequently, waist, hip, and mid-arm circumference were assessed using a flexible, nonelastic tape measure. We employed the Harpenden Skinfold Caliper (Baty International, West Sussex, United Kingdom) to determine triceps skinfold thickness and recorded the mean of three repeated measurements.
Body composition analysis was performed with the seca mBCA 525 (seca, Hamburg, Germany), an eight-electrode, phase-sensitive, segmental bioelectrical impedance analysis (BIA) device. Patients were instructed to restrain from eating for 4 hours, from strenuous physical activity for 12 hours, and from alcohol consumption for 24 hours; as well as to empty their bladder prior to the assessment. The measurement was conducted in a supine position using adhesive gel electrodes placed at specified anatomical sites on the dorsal surfaces of hand, wrist, ankle and foot.
Handgrip strength was tested employing the Jamar Plus+ Digital Hand Dynamometer (Patterson Medical, Warrenville, IL, USA). Three measurements were taken with the patients seated, the elbow in 90° flexion, and the wrist in a neutral position using their dominant hand. The maximum value of the three attempts was considered for analysis.
Assessment of blood pressure followed a standardized protocol employing a fully automated device (boso medicus, BOSCH+ SOHN, Jungingen, Germany). The measurement was taken in a seated position after a minimum rest of 5 minutes on the right arm.
Blood testing was performed in all patients. Selection of blood parameters that were determined was based on the guideline’s recommendations and included the following: Complete blood count, electrolytes, creatinine, blood glucose and HbA1c, vitamins A, B1, and B12, folic acid, 25-hydroxyvitamin D, parathormone, albumin, calcium, ferritin as well as zinc, copper and selenium.
Full text: Click here
Publication 2023
3-methoxybutylcyanoacrylate Albumins Ankle Bioelectrical Impedance BLOOD Blood Glucose Blood Pressure Body Composition Body Regions Calcifediol Calcium, Dietary Cardiac Arrest Complete Blood Count Copper Creatinine Elbow Electrolytes Ferritin Fingers Folic Acid Foot Medical Devices Parathyroid Hormone Patients Physicians Selenium Sitting Skinfold Thickness Urinary Bladder Vitamins Wrist Zinc
Anthropometry was completed by a single observer following standardized procedures11
. Stature was measured to the nearest 0.1 cm using a stadiometer (Harpenden model 98.603, Holtain LTD, Crosswell, UK). A portable balance (SECA model 770, Hanover, MD, USA) was used to measure BM to the nearest 0.1 kg. Skinfold thickness was measured at two sites (triceps and subscapular) to the nearest 1 mm using a Lange calliper (Beta Technology Incorporate Cambridge, MD, USA). A non-invasive equation recommended for male adolescents12 (link)
was used to determine body fat expressed as a percentage of BM (%FM). Subsequently, fat mass (FM) and FFM in kg were derived.
Publication 2023
Body Fat Body Height Males Skinfold Thickness

Top products related to «Skinfold Thickness»

Sourced in United Kingdom
The Skinfold Caliper is a device used to measure the thickness of a fold of skin and underlying fat tissue. It provides an estimate of body composition by assessing subcutaneous fat.
Sourced in United Kingdom
Holtain calipers are a type of precision measuring instrument used to accurately measure the dimensions of objects. They consist of two jaws that can be adjusted to fit around an object, and a scale that displays the measured value.
Sourced in United Kingdom
The Skinfold Calliper is a device used to measure the thickness of a fold of skin and underlying tissue. It consists of two flat parallel arms that exert a standardized pressure on the skin, allowing the user to accurately measure the skinfold thickness.
Sourced in Germany, United Kingdom, United States, Japan, Jamaica, Switzerland, Brazil
A stadiometer is a medical device used to measure a person's height. It consists of a vertical scale, typically marked in centimeters or inches, with a horizontal headpiece that can be lowered to rest on top of the person's head, allowing for an accurate measurement of their stature.
Sourced in United Kingdom, Germany, United States
The Stadiometer is a medical device used to measure an individual's height. It consists of a vertical ruler or scale mounted on a stable base, with a sliding horizontal headpiece that is lowered onto the top of the person's head to record their height.
Sourced in United Kingdom
The Tanner/Whitehouse Skinfold Caliper is a device used to measure the thickness of subcutaneous fat layers. It consists of two parallel jaws that gently pinch a fold of skin and the underlying fat, allowing the thickness to be measured accurately.
Sourced in United Kingdom
The Holtain Calliper is a precision measuring instrument used to measure the thickness or diameter of various objects. It consists of two opposing arms that can be adjusted to fit around an object, and a scale that displays the measured value.
Sourced in Germany, United Kingdom, Jamaica
The Seca 416 Infantometer is a medical device designed for measuring the length of infants and small children. It features a stable base and a sliding measuring rod that allows for precise length measurements.
Sourced in Japan, United States
The HEM-780 is a digital blood pressure monitor that measures systolic and diastolic blood pressure as well as heart rate. It is designed for home and clinical use.
Sourced in Germany, Jamaica, United States
The Seca model 881 is a digital column scale designed for medical use. It features a stable, easy-to-read LCD display and can measure weight up to 250 kg with an accuracy of 100 g.

More about "Skinfold Thickness"

Skinfold thickness, also known as subcutaneous fat thickness or skin caliper measurement, is a widely used technique for assessing body composition and fat distribution.
This non-invasive method involves pinching the skin at specific anatomical sites, such as the biceps, triceps, subscapular, or suprailiac regions, and using a specialized caliper device like the Holtain caliper, Seca 416 Infantometer, or Tanner/Whitehouse Skinfold Caliper to measure the thickness of the skinfold.
Skinfold thickness is a cost-effective and convenient way to estimate body fat percentage, which is particularly useful in clinical, fitness, and research settings.
Accurate and reliable measurements require adherence to standardized protocols and proper technique, as outlined in the literature, to minimize variability and ensure reproducibility of results.
The HEM-780 and Seca model 881 are examples of handheld devices that can be used to measure skinfold thickness, providing a quick and easy-to-use solution.
Researchers and healthcare professionals often utilize these tools alongside stadiometers, which measure height, to obtain a comprehensive assessment of body composition.
Unlocking the power of skinfold thickness measurement can be achieved through the use of innovative platforms like PubCompare.ai, which leverages AI-driven comparisons to identify the most reliable and reproducible methods from the available literature, preprints, and patents.
By streamlining the research process and providing access to the best protocols, these solutions can help optimize results and drive more accurate body composition analysis.