The largest database of trusted experimental protocols
> Procedures > Laboratory Procedure > Energy Dispersive X Ray Spectroscopy

Energy Dispersive X Ray Spectroscopy

Energy Dispersive X-Ray Spectroscopy (EDXS) is a powerful analytical technique used to identify and quantify the elemental composition of materials.
It utilizes the interaction between X-rays and a sample to generate characteristic X-ray spectra, providing detailed information about the elements present.
EDXS is widely employed in materials science, geology, biology, and other fields to study the chemical makeup of a wide range of specimens.
This non-destructive method offers rapid, high-resolution analysis, making it an invaluable tool for researchers.
Enahnce your EDXS research accuracy with PubCompare.ai, an AI-driven platform that helps you locate the best protocols from literature, preprints, and patents.
Utilize advanced comparison tools to identify the optimal protocols and products for your needs, streamlining your research process and improving results.

Most cited protocols related to «Energy Dispersive X Ray Spectroscopy»

Method A: In the silver reduction, the methodology described previously was followed [17 (link)]. Briefly, approximately 10 g of F. oxysporum biomass was taken in a conical flask containing 100 mL of distilled water. AgNO3 solution (10-3 M) was added to the erlenmeyer flask and the reaction was carried out in the dark. Periodically, aliquots of the reaction solution were removed and the absorptions were measured using a UV-Vis spectrophotometer (Agilent 8453 – diode array).
Method B: Another test was also carried out as following: approximately 10 g of F. oxysporum biomass was taken in a conical flask containing 100 mL of distilled water, kept for 72 h at 28°C and then the aqueous solution components were separated by filtration. To this solution, AgNO3 (10-3 M) was added and kept for several hours at 28°C.
The silver nanoparticles were characterized by scanning electron microscopy (SEM) and energy-dispersive spectroscopy (EDS) at a voltage of 20 kV (Jeol – JSM-6360LV) and previously coated with gold under vacuum.
Full text: Click here
Publication 2005
colloidal silver Energy Dispersive X Ray Spectroscopy Filtration Gold Scanning Electron Microscopy Silver Vacuum
To detect endothelial glycocalyx using electron microscopy [21 (link)], mice were anesthetized and perfused with a solution composed of 2% glutaraldehyde, 2% sucrose, 0.1 M sodium cacodylate buffer (pH 7.3), and 2% lanthanum nitrate through a cannula placed in the left ventricle 48 h after LPS administration [22 (link)]. Before perfusion, an incision was made in the right atrial appendage, and the neck was ligated with a silk suture. In addition, a perfusion pump was used for injection at a steady rate of 1 ml/minute.
Thereafter, the left ventricle, liver, and kidney were harvested and diced. Three or four pieces of approximately 1 mm3 each were immersed in the perfusion solution for 2 h for fixation and then soaked overnight in a solution without glutaraldehyde before being washed in alkaline (0.03 mol/L NaOH) sucrose (2%) solution. The specimens were then dehydrated through a graded ethanol series.
The frozen fracture method was used to prepare samples for examination using scanning electron microscopy (SEM). Each sample was laid on an iron plate chilled with liquid nitrogen, and ethanol was sprinkled onto it. Once the ethanol was frozen, the sample was fractured using a chisel such that it was not touched directly. The samples were then incubated in tert-butyl alcohol at room temperature. After the tert-butyl alcohol had solidified, it was freeze-dried, and the specimens were examined using SEM (S-4500; Hitachi, Tokyo, Japan). In addition, for further elemental analysis of each sample, energy-dispersive X-ray spectroscopy was performed under SEM.
To prepare samples for transmission electron microscopy (TEM), each specimen was embedded in epoxy resin. Ultrathin sections (90 nm) stained with uranyl acetate and lead citrate were then examined using TEM (HT-7700; Hitachi). For usual electron microscopy, 2.5% glutaraldehyde in 0.1 mol/L phosphate buffer (pH 7.4) was used instead of perfusion buffer as described above.
Full text: Click here
Publication 2017
Atrium, Right Auricular Appendage Buffers Cacodylate Cannula Citrates Electron Microscopy Endothelium Energy Dispersive X Ray Spectroscopy Epoxy Resins Ethanol Freezing Glutaral Glycocalyx Infusion Pump Iron Kidney lanthanide nitrate Left Ventricles Liver Mice, House Neck Nitrogen Perfusion Phosphates Scanning Electron Microscopy Silk Sodium Sucrose Sutures tert-Butyl Alcohol Transmission Electron Microscopy uranyl acetate
The variable pressure scanning electron microscopy used in this study (VP-SEM, Hitachi SU3500) is equipped with dual energy dispersive X-ray spectroscopy (dEDS, Bruker XFlash® 6|60) detectors. This instrument has the ability to perform simultaneously multimodal imaging and spatial distribution chemical mapping, a truly powerful analytical approach to study biological surfaces in their native state. The XFlash® 6|60 is particularly suitable for applications with relatively low X-ray yield, as common in the area of nanoanalysis.
Full text: Click here
Publication 2020
Biopharmaceuticals Energy Dispersive X Ray Spectroscopy Pressure Radiography Scanning Electron Microscopy
The purification and functionalization of the MWCNT were accomplished as previously described. 36 –38 This activity was carried out in a Microwave Accelerated Reaction System CEM Mars (Matthews, NC, USA) fitted with internal temperature and pressure controls. The 100 mL reaction chamber was lined with Teflon PFA® with an operating range of 0~200 °C and 0~200 psi. The residual metals were removed via microwave induced reaction with 1N HNO3 followed by a reaction with saturated ethylenediamine tetra-acetate in CH3COOH. 39 (link) The purified MWCNT (PD-MWCNT) were added to the reaction chamber together with 40 mL of a mixture of 1:1 70 % nitric acid and 97 % sulfuric acid. With the microwave power set to 95 % of the maximal possible 1600 watt output, the reaction was carried out for 20 min at a temperature of 140 °C. The suspension was filtered through a 10 μm PTFE membrane, washed with deionized water (DI H2O) to a neutral pH and vacuum dried at 70 °C. This sample derivative was designated COOH-MWCNT. A variety of different analytical techniques were used to characterize the MWCNT, including Energy Dispersive X-ray Spectroscopy (EDS, Oxford Instrument, Oxfordshire, UK) to identify the elemental composition of the MWCNT. The Fourier Transform Infrared spectroscopy (FTIR) measurements were carried out in purified KBr pellets using a PerkinElmer Spectrum One instrument (Downers Grove, IL, USA). Thermogravimetric analyses (TGA) were performed using a Pyris 1 TGA (Perkin-Elmer Inc., Covina, CA, USA) from 30 °C to 900 °C under a flow of air at 10 mL/min and a heating rate of 10°C per min. Scanning Electron Microscopy (SEM) using a LEO 1530 VP SEM equipped with an energy-dispersive X-ray analyzer was used to study the morphology of the samples. Zeta-potential measurements of the MWCNT suspensions were performed using a ZetaSizer Nano-ZS Instrument (Malvern Instruments, Worcestershire WR, UK). The relative hydrodynamic radius of MWCNT suspended in H2O was measured using high throughput dynamic light scattering (HT-DLS, Dynapro Plate Reader, Wyatt Technology, Santa Barbara, CA, USA).
Publication 2010
Acetate Energy Dispersive X Ray Spectroscopy Ethylenediamines Hydrodynamics Metals Microwaves Nitric acid Pellets, Drug Polytetrafluoroethylene Pressure Radiography Radius Scanning Electron Microscopy Spectroscopy, Fourier Transform Infrared Sulfuric Acids Teflon Tetragonopterus Tissue, Membrane Vacuum
A sample of copper anode slime provided from a refining factory available as a powdered solid with a mean particles size less than 30 µm was used for this study. The sample composition was determined by diverse analytical methods, but only the results of scanning electron microscopy-energy dispersive spectroscopy (SEM-EDS) and X-ray diffraction (XRD) analyses were selected to be presented in this paper (see Section 3). SEM examinations were performed by using a HITACHI S-4800 device (Hitachi Ltd., Tokyo, Japan) equipped with an EDS elemental analysis microprobe. SEM imaging and chemical microanalysis were performed mostly at an accelerating voltage of 15 kV. The samples were carbon coated prior to analysis. XRD was performed on samples using a Bruker D8 Advance device (Bruker, Karlsruhe, Germany), (Co Kα1 radiation, λ = 1.789 Å), under 35 kV and 45 mA operating conditions. The acquisition of diffraction patterns was performed between 3° and 64° (2θ) range at step scan 0.034° with 3 s per step. The XRD-patterns were analyzed thanks to the DIFFRAC.EVA software (version 5, Bruker) and PDF-2 release 2011 database.
Isothermal experimental tests for the CAS sample treatment were conducted in horizontal set-ups, including a system of static tubular furnaces that were able to reach 1600 °C. To achieve this, a pre-weighted sample of several grams was introduced directly into the furnace preheated at the desired temperature. When the dwell time was reached, the sample was removed from the furnace and cooled down to room temperature. Air was used as a flowing gas and it also performed the oxidation of CAS components. The outlet gases were cooled at room temperature, leading to the condensation of the vapor phase and the recovery of a solid condensate. Initial CAS sample and solid products obtained from CAS thermal treatment were examined by visible microscopy, SEM-EDS and XRD. The advantage of using SEM-EDS analysis here is the ability to gather punctual information about elemental content, phases differentiation, and about the morphological and textural evolution of the thermally treated samples, all this contributes to a better understanding of the involved reaction mechanism and processing steps.
Full text: Click here
Publication 2019
Biological Evolution Carbon Copper Energy Dispersive X Ray Spectroscopy Gases Medical Devices Microscopy Physical Examination Radionuclide Imaging Radiotherapy Scanning Electron Microscopy Specimen Handling X-Ray Diffraction

Most recents protocols related to «Energy Dispersive X Ray Spectroscopy»

Example 1

95 g of manganese (purity: 99.95%; purchased from Taewon Scientific Co., Ltd.) and 5 g of high-purity graphite (purity: 99.5%; purchased from Taewon Scientific Co., Ltd.) were placed in a water-cooled copper crucible of an argon plasma arc melting apparatus (manufactured by Labold AG, Germany, Model: vacuum arc melting furnace Model LK6/45), and melted at 2,000 K under an argon atmosphere. The melt was cooled to room temperature at a cooling rate of 104 K/min to obtain an alloy ingot. The alloy ingot was crushed to a particle size of 1 mm or less by hand grinding. Thereafter, the obtained powders were magnetically separated using a Nd-based magnet to remove impurities repeatedly, and the Mn4C magnetic powders were collected. The collected Mn4C magnetic powders were subjected to X-ray diffraction (XRD) analysis (measurement system: D/MAX-2500 V/PO, Rigaku; measurement condition: Cu—Kα ray) and energy-dispersive X-ray spectroscopy (EDS) using FE-SEM (Field Emission Scanning Electron Microscope, MIRA3 LM).

FIGS. 2(a) and 2 (b) show an X-ray diffraction pattern and an energy-dispersive X-ray spectroscopy graph of the Mn4C magnetic material produced according to Example 1 of the present disclosure, respectively.

As can be seen in FIG. 2(a), the Mn4C magnetic material showed diffraction peaks of (111), (200), (220), (311) and (222) crystal planes at 2θ values of 40°, 48°, 69°, 82° and 88°, respectively, in the XRD analysis. Thus, it can be seen that the XRD patterns of the Mn4C magnetic material produced according to Example 1 are well consistent with the patterns of the cubic perovskite Mn4C. In addition, the Mn4C magnetic material shows several very weak diffraction peaks that can correspond to Mn23C6 and Mn. That is, the diffraction peak intensity at 2θ values of 43° and 44°, which correspond to Mn and Mn23C6 impurities, is as very low as about 2.5% of the diffraction intensity of the peak corresponding to the (111) plane. Through this, it can be seen that the powders obtained in Example 1 have high-purity Mn4C phase. The lattice parameter of the Mn4C is estimated to be about 3.8682 Å.

FIG. 2(b) shows the results of analyzing the atomic ratio of Mn:C in the powder by EDS. The atomic ratio of Mn:C is 80.62:19.38, which is very close to 4:1 within the experimental uncertainties. Thus, it can be seen that the powder is also confirmed to be Mn4C.

The M-T curve of the field aligned Mn4C powder obtained in Example 1 was measured under an applied field of 4 T and at a temperature ranging from 50 K to 400 K. Meanwhile, the M-T curve of the randomly oriented Mn4C powder was measured under an applied field of 1 T. The Curie temperature of Mn4C was measured under 10 mT while decreasing temperature from 930 K at a rate of 20 K/min.

FIGS. 3(a) to 3(c) show the M-T curves of the Mn4C magnetic material, produced according to Example 1 of the present disclosure, under magnetic fields of 4 T, 1 T, and 10 mT, respectively.

FIG. 3 shows magnetization-temperature (M-T) curves indicating the results of measuring the temperature-dependent magnetization intensity of the Mn4C magnetic material, produced in Example 1, using the vibrating sample magnetometer (VSM) mode of Physical Property Measurement System (PPMS®) (Quantum Design Inc.).

According to the Néel theory, the ferrimagnets that contain nonequivalent substructures of magnetic ions may have a number of unusual forms of M-T curves below the Curie temperature, depending on the distribution of magnetic ions between the substructures and on the relative value of the molecular field coefficients. The anomalous M-T curves of Mn4C, as shown in FIG. 3(a), can be explained to some extent by the Néel's P-type ferrimagnetism, which appears when the sublattice with smaller moment is thermally disturbed more easily. For Mn4C with two sublattices of MnI and MnII, as shown in FIG. 1, the MnI sublattice might have smaller moment.

FIG. 3(a) shows the temperature dependence of magnetization of the Mn4C magnetic material produced in Example 1. The magnetization of Mn4C measured at 4.2K is 6.22 Am2/kg (4 T), corresponding to 0.258μB per unit cell. The magnetization of the Mn4C magnetic material varies little at temperatures below 50 K, and is quite different from that of most magnetic materials, which undergo a magnetization deterioration with increasing temperature due to thermal agitation. Furthermore, the magnetization of the Mn4C magnetic material increases linearly with increasing temperature at temperatures above 50 K. The linear fitting of the magnetization of Mn4C at 4 T within the temperature range of 100 K to 400 K can be written as M=0.0072T+5.6788, where M and T are expressed in Am2/kg and K, respectively. Thus, the temperature coefficient of magnetization of Mn4C is estimated to be about ˜2.99*10−4μB/K per unit cell. The mechanisms of the anomalous thermomagnetic behaviors of Mn4C may be related to the magnetization competition of the two ferromagnetic sublattices (MnI and MnII) as shown in FIG. 1.

FIG. 3(b) shows the M-T curves of the Mn4C powders at temperatures within the range of 300 K to 930 K under 1 T. The linear magnetization increment stops at 590 K, above which the magnetization of Mn4C starts to decrease slowly first and then sharply at a temperature of about 860 K. The slow magnetization decrement at temperatures above 590 K is ascribed to the decomposition of Mn4C, which is proved by further heat-treatment of Mn4C as described below.

According to one embodiment of the present disclosure, the saturation magnetization of Mn4C increases linearly with increasing temperature within the range of 50 K to 590 K and remains stable at temperatures below 50 K. The increases in anomalous magnetization of Mn4C with increasing temperature can be considered in terms of the Néel's P-type ferrimagnetism. At temperatures above 590 K, the Mn4C decomposes into Mn23C6 and Mn, which are partially oxidized into the manganosite when exposed to air. The remanent magnetization of Mn4C varies little with temperature. The Curie temperature of Mn4C is about 870 K. The positive temperature coefficient (about 0.0072 Am2/kgK) of magnetization in Mn4C is potentially important in controlling the thermodynamics of magnetization in magnetic materials.

The Curie temperature Te of Mn4C is measured to be about 870 K, as shown in FIG. 3(c). Therefore, the sharp magnetization decrement of Mn4C at temperatures above 860 K is ascribed to both the decomposition of Mn4C and the temperature near the Tc of Mn4C.

FIG. 4 is a graph showing the magnetic hysteresis loops of the Mn4C magnetic material, produced according to Example 1 of the present disclosure, at 4.2 K, 200 K and 400 K. The magnetic hysteresis loops were measured by using the PPMS system (Quantum Design) under a magnetic field of 7 T while the temperature was changed from 4 K to 400 K.

As shown in FIG. 4, the positive temperature coefficient of magnetization was further proved by the magnetic hysteresis loops of Mn4C as shown in FIG. 4. The Mn4C shows a much higher magnetization at 400 K than that at 4.2 K. Moreover, the remanent magnetization of Mn4C varies little with temperature and is Δ3.5 Am2/kg within the temperature range of 4.2 K to 400 K. The constant remanent magnetization of Mn4C within a wide temperature range indicates the high stability of magnetization against thermal agitation. The coercivities of Mn4C at 4.2 K, 200 K, and 400 K were 75 mT, 43 mT, and 33 mT, respectively.

The magnetic properties of Mn4C measured are different from the previous theoretical results. A corner MnI moment of 3.85μB antiparallel to three face-centered MnII moments of 1.23μB in Mn4C was expected at 77 K. The net moment per unit cell was estimated to be 0.16μB. In the above experiment, the net moment in pure Mn4C at 77 K is 0.26μB/unit cell, which is much larger than that expected by Takei et al. It was reported that the total magnetic moment of Mn4C was calculated to be about 1μB, which is almost four times larger than the 0.258μB per unit cell measured at 4.2 K, as shown in FIG. 4.

FIG. 5 is an enlarged view of the temperature-dependent XRD patterns of the Mn4C magnetic material produced according to Example 1 of the present disclosure.

The thermomagnetic behaviors of Mn4C are related to the variation in the lattice parameters of Mn4C with temperature. It is known that the distance of near-neighbor manganese atoms plays an important role in the antiferro- or ferro-magnetic configurations of Mn atoms. Ferromagnetic coupling of Mn atoms is possible only when the Mn—Mn distance is large enough. FIG. 5 shows the diffraction peaks of the (111) and (200) planes of Mn4C at temperatures from 16 K to 300 K. With increasing temperature, both (111) and (200) peaks of Mn4C shifted to a lower degree at temperatures between 50 K and 300 K, indicating an enlarged distance of Mn—Mn atoms in Mn4C. No peak shift is obviously observed for Mn4C at temperatures below 50 K. The distance of nearest-neighbor manganese atoms plays an important role in the antiferro- or ferro-magnetic configurations of Mn atoms and thus has a large effect on the magnetic properties of the compounds.

Thus, it can be seen that the abnormal increase in magnetization of Mn4C with increasing temperature occurs due to the variation in the lattice parameters of Mn4C with temperature.

The powder produced in Example 1 was annealed in vacuum for 1 hour at each of 700 K and 923 K, and then subjected to X-ray spectroscopy, and the results thereof are shown in FIG. 6.

The magnetization reduction of Mn4C at temperatures above 590 K is ascribed to the decomposition of Mn4C, which is proved by the XRD patterns of the powders after annealing Mn4C at elevated temperatures. FIG. 6 shows the structural evolution of Mn4C at elevated temperatures. When Mn4C is annealed at 700 K, a small fraction of Mn4C decomposes into a small amount of Mn23C6 and Mn. The presence of manganosite is ascribed to the spontaneous oxidation of the Mn precipitated from Mn4C when exposed to air after annealing. The fraction of Mn23C6 was enhanced significantly for Mn4C annealed at 923 K, as shown in FIG. 6.

These results prove that the metastable Mn4C decomposes into stable Mn23C6 at temperatures above 590 K. The presence of Mn4C in the powder annealed at 923 K indicates a limited decomposition rate of Mn4C, from which the Tc of Mn4C can be measured. Both Mn23C6 and Mn are weak paramagnets at ambient temperature and elevated temperatures. Therefore, the magnetic transition of the Mn4C magnetic material at 870 K is ascribed to the Curie point of the ferrimagnetic Mn4C.

The Mn4C shows a constant magnetization of 0.258μB per unit cell below 50 K and a linear increment of magnetization with increasing temperature within the range of 50 K to 590 K, above which Mn23C6 precipitates from Mn4C. The anomalous M-T curves of Mn4C can be considered in terms of the Néel's P-type ferrimagnetism.

Full text: Click here
Patent 2024
Alloys Argon Atmosphere Biological Evolution Cells Copper Cuboid Bone Debility Energy Dispersive X Ray Spectroscopy Face Fever fluoromethyl 2,2-difluoro-1-(trifluoromethyl)vinyl ether Graphite Ions Magnetic Fields Manganese perovskite Physical Processes Plasma Powder Radiography Scanning Electron Microscopy Spectrum Analysis Vacuum Vision X-Ray Diffraction
Scanning
electron microscopy
was performed with a Sigma 500 Gemini (Zeiss) at 4 kV using an InLens
detector. Energy-dispersive X-ray spectroscopy analysis was performed
using a Bruker Quantax EDS. X-ray photoelectron spectroscopy (XPS)
measurements were carried out on a custom-build spectrometer (Moses)
equipped with an X-ray source from Al Ka radiation (1486.6 eV) and
a hemispherical electron analyzer in the analysis chambers, under
ultra-high vacuum with the base pressure being lower than 1 ×
10–9 mbar. To minimize any beam damage on the sample,
a large probe area of 1 × 10 mm2 and a low X-ray powder
of 30 W were employed. All measurement was done at room temperature
with a 45° emission angle. The binding energy was calibrated
based on the position of the Au 4f7/2 peak (84.0 eV) and
the Fermi level (0 eV).
Full text: Click here
Publication 2023
Electrons Energy Dispersive X Ray Spectroscopy Pressure Radiation Radiography Vacuum X-Ray Photoelectron Spectroscopy
The morphologies of the prepared samples were studied using scanning electron microscopy (SEM; Hitachi SU 5000 VPSEM) and transmission electron microscopy (TEM; FEI Talos F200X) with a field-emission gun at 200 kV. For TEM studies, the samples were scratched into an ethanol solution for dispersion. The ethanol solution was then dropped onto carbon-coated molybdenum TEM grids and dried under ambient conditions. The compositions of the samples were studied using energy-dispersive X-ray spectroscopy (EDS) system (Bruker Quantax EDS) coupled to the SEM and TEM instruments. X-ray diffraction (XRD) patterns were collected using an X’Pert-Pro MPD diffractometer (PANalytical) with a Cu Kα X-ray source (λ = 1.540598 Å). X-ray photoelectron spectroscopy (XPS) studies were conducted using an SSI S-Probe XPS spectrometer. The carbon peak (284.6 eV) was used as a reference to correct the charging effect. To investigate the bond vibrations during CO2R, in situ Raman spectra (200–3200 cm−1) were collected using a microspectrophotometer (Horiba-LabRAM HR), a long-working-distance water immersion objective (40×; U M Plan Semi Apochromat), a grating with 100 lines mm−1, and a homemade flow cell with a 633 nm laser as the excitation source (excitation energy of 1.96 eV). The objective lens was immersed into electrolytes in the homemade flow cell. After the sample focus, the working distance was ~2 mm from the lens to the electrode surface. All spectroscopic data were baseline corrected.
Full text: Click here
Publication 2023
Carbon Cells Electrolytes Energy Dispersive X Ray Spectroscopy Ethanol Lens, Crystalline LINE-1 Elements Molybdenum Radiography Scanning Electron Microscopy Spectrum Analysis Submersion Transmission Electron Microscopy Vibration X-Ray Diffraction
Scanning electron microscopic (SEM) images were obtained on a SU8030 scanning electron microscope at the voltages of 10/15 kV, while the energy-dispersive X-ray spectroscopy (EDS) elemental maps were recorded at 15 kV.
Full text: Click here
Publication 2023
Energy Dispersive X Ray Spectroscopy Microtubule-Associated Proteins Scanning Electron Microscopy
After BSTS crystal growth, the ampoule was gently broken and the material was cleaved via razor blade and exfoliated to produce a thin layer for elemental composition detection using energy-dispersive X-ray spectroscopy (EDS, HITACHI S-4800 High Resolution Field Emission Scanning Electron Microscopy (SEM)) operating at an acceleration voltage of 20 to 30 kV. Both solid and powder samples were prepared for both crystal plane and powder X-ray diffraction (XRD, Philips PANalytical X’Pert, Cu Kα wavelength). For peak identification and crystal structure Rigaku data analysis software (PDXL—version 2) was used. Part of the crystal was prepared via fracturing for Thermo-gravimetric analysis (TGA, TA Instruments Discovery SDT 650).
Full text: Click here
Publication 2023
Acceleration Crystal Growth Energy Dispersive X Ray Spectroscopy Powder Scanning Electron Microscopy Thermogravimetry X-Ray Diffraction

Top products related to «Energy Dispersive X Ray Spectroscopy»

Sourced in Japan, United States, China, Germany, United Kingdom, Spain, Canada, Czechia
The S-4800 is a high-resolution scanning electron microscope (SEM) manufactured by Hitachi. It provides a range of imaging and analytical capabilities for various applications. The S-4800 utilizes a field emission electron gun to generate high-quality, high-resolution images of samples.
Sourced in Germany, United States, Japan, United Kingdom, China, France, India, Greece, Switzerland, Italy
The D8 Advance is a versatile X-ray diffractometer (XRD) designed for phase identification, quantitative analysis, and structural characterization of a wide range of materials. It features advanced optics and a high-performance detector to provide accurate and reliable results.
Sourced in Japan, United States, Germany, United Kingdom
The JEM-2100F is a transmission electron microscope (TEM) designed and manufactured by JEOL. It is capable of high-resolution imaging and analytical capabilities. The JEM-2100F is used for a variety of research and industrial applications that require advanced electron microscopy techniques.
Sourced in United States, United Kingdom, Japan, China, Germany, Netherlands, Switzerland, Portugal
The ESCALAB 250Xi is a high-performance X-ray photoelectron spectroscopy (XPS) system designed for surface analysis. It provides precise and reliable data for the characterization of materials at the nanoscale level.
Sourced in Japan, United States, Germany, United Kingdom, China, France
The JEM-2100 is a transmission electron microscope (TEM) manufactured by JEOL. It is designed to provide high-quality imaging and analysis of a wide range of materials at the nanoscale level. The instrument is equipped with a LaB6 electron source and can operate at accelerating voltages up to 200 kV, allowing for the investigation of a variety of samples.
Sourced in Japan, Germany, United States, China
The SU8010 is a field emission scanning electron microscope (FE-SEM) manufactured by Hitachi. It is designed for high-resolution imaging of a wide range of sample types. The SU8010 provides stable and reliable performance, with a high-resolution capability and a wide range of analytical functions.
Sourced in United States, United Kingdom, China, Japan, Germany, Canada, Hong Kong
The K-Alpha is a high-performance X-ray photoelectron spectroscopy (XPS) system designed for surface analysis. It provides detailed information about the chemical composition and bonding of materials at the surface. The K-Alpha system features advanced technology to deliver accurate and reproducible data for a wide range of sample types and applications.
Sourced in Japan, United States, Germany, United Kingdom, China
The SmartLab is a versatile X-ray diffractometer designed for materials analysis. It features automated alignment and data collection capabilities, providing efficient and reliable performance for a wide range of sample types and research applications.
Sourced in Japan, United States, Germany, United Kingdom, Canada
The S-4700 is a field emission scanning electron microscope (FE-SEM) manufactured by Hitachi. It provides high-resolution imaging and analytical capabilities for a wide range of applications. The S-4700 utilizes a field emission electron source to produce a stable, high-brightness electron beam, enabling high-resolution imaging of samples.
Sourced in Japan, United States, Germany, China
The Rigaku Ultima IV is a compact and versatile X-ray diffractometer designed for a wide range of analytical applications. It features a high-performance X-ray generator, advanced optics, and a sophisticated detector system to provide reliable and accurate data. The Ultima IV is capable of performing various types of X-ray diffraction analyses, including phase identification, structural characterization, and quantitative analysis.

More about "Energy Dispersive X Ray Spectroscopy"

Energy Dispersive X-Ray Spectroscopy (EDXS), also known as EDS or EDX, is a powerful analytical technique used to identify and quantify the elemental composition of materials.
This non-destructive method utilizes the interaction between X-rays and a sample to generate characteristic X-ray spectra, providing detailed information about the elements present.
EDXS is widely employed in materials science, geology, biology, and other fields to study the chemical makeup of a wide range of specimens, including those analyzed using instruments like the S-4800, D8 Advance, JEM-2100F, ESCALAB 250Xi, JEM-2100, SU8010, K-Alpha, SmartLab, and S-4700.
The technique offers rapid, high-resolution analysis, making it an invaluable tool for researchers.
To enhance the accuracy of your EDXS research, consider using PubCompare.ai, an AI-driven platform that helps you locate the best protocols from literature, preprints, and patents.
Utilize its advanced comparison tools to identify the optimal protocols and products for your needs, streamlining your research process and improving your results.
Whether you're studying the chemical composition of materials, geological samples, or biological specimens, EDXS can provide you with the detailed insights you need to advance your work.
Enhance your research with the power of EDXS and the tools available at PubCompare.ai.