The largest database of trusted experimental protocols

486 tunable absorbance detector

Manufactured by Waters Corporation
Sourced in United States

The 486 Tunable Absorbance Detector is a laboratory instrument designed to measure the absorbance of light in a sample. It operates by passing light through a sample and detecting the amount of light that is absorbed, which can provide information about the chemical composition of the sample. The detector is capable of tuning the wavelength of the light source to optimize detection for specific analytes.

Automatically generated - may contain errors

13 protocols using 486 tunable absorbance detector

1

HPLC Analysis of Fermentation Supernatants

Check if the same lab product or an alternative is used in the 5 most similar protocols
Fermentation supernatants of the feces ( = 6) used in the fermentation study were analyzed for lactate and SCFA contents with a Waters 2690 HPLC system fitted with an Aminex HPX 87 H column (300 mm × 78 mm; Bio-Rad Laboratories) combined with a UV absorbance detector (Waters 486 tunable absorbance detector) set at 210 nm. The sterile bottle containing 15 mL of fermentation supernatants was mixed on a vortex for 1 minute, and 2 mL was sampled and centrifuged at 13,000 rcf for 15 min; 1.5 mL of the supernatants was transferred to a vial, and pH was adjusted between 1 and 3 using 0.1 M HCl. The SCFAs were eluted as described by Murugesan et al. (31 (link)) using SIGMA standard. Fermentation values were corrected for the content of the blanks as well as the inocula.
+ Open protocol
+ Expand
2

Molecular Weight Distribution Analysis

Check if the same lab product or an alternative is used in the 5 most similar protocols
The molecular weight distribution was determined using size-exclusion chromatography (SEC). The system consisted of a controller (Waters 600E, Waters, Milford, MA, USA) connected to a refractive index detector (Waters 2414 Differential Refractometer) and UV detector (Waters 486 Tunable Absorbance Detector) set to 234 nm. The column used was a TSKgel (G4000PWXL, TOSOH Bioscience GmbH, Griescheim, Germany) and the eluent was deionized water, which was pumped using a Waters 600 gradient pump at a flow rate of 0.5 mL/min and degassed using a Waters in-Line degasser. The injection volume was 20 µL, which was performed using an autosampler (Waters 717 plus autosampler). The standards used were polyethylene glycol (400 Da, Merck Schuchardt OHG, Germany) and dextran (2000, 500, 100, 150, 60, 10 and 4 kDa Merck Schuchardt OHG, Germany). The same standards, instrument parameters and column were used for an alkali SEC (100 mM NaOH eluent) with a Shimadzu (Shimadzu Corp., Kyoto, Japan) system (SIL-10AXL autosampler, LC-10AT pump, CTO-10A column oven, RID-10A refractive index detector and SPD-10AV UV-detector).
+ Open protocol
+ Expand
3

HPLC Quantification of Serum Betaine

Check if the same lab product or an alternative is used in the 5 most similar protocols
Serum betaine (trimethylglycine) levels were measured using high-performance liquid chromatography (HPLC) as described42 (link). Serum samples (50 μL) were mixed with an equal volume of 100 mM KH2PO4 in screw-top microcentrifuge tubes, then derivatization solution (900 μL; 12.5 mM 18-crown-6, 50 mM 4-bromophenacyl bromide in acetonitrile) was added. The tubes were vacuum-packed in plastic bags and placed in a water bath at 80 °C for 1 h. After cooling to room temperature, the samples were centrifuged at 1000×g. Supernatants (15 μL) were directly injected into the HPLC comprising a 600 Controller (Waters Corp., Milford, MA, USA), a 486 Tunable Absorbance Detector (Waters Corp.), Supelcosil™ LC-SCX HPLC column, particle size 5 μm, 25 cm × 4.6 mm) (Sigma-Aldrich, St. Louis, MO, US), mobile phase, acetonitrile:ultra-pure-water (9:1) containing 22 mM choline, at a flow rate of 1.5 mL/min.
+ Open protocol
+ Expand
4

SEC Characterization of Protein Size

Check if the same lab product or an alternative is used in the 5 most similar protocols
SEC was performed on a Waters system (Waters Corp., Milford, MA) comprising a 600S controller, 616 pump, 717plus autosampler, 486 Tunable Absorbance Detector, and PeakSimple Software. Ten µL of 4.26 mg/mL protein in 10 mM sodium phosphate, pH 7.4, was loaded onto a 7.8 × 300 mm SRT-SEC1000 column (5 µm particle size, 100 Å pore size; Sepax Technologies, Inc., Newark, DE) fitted with a 7.8 × 50 mm SRT SEC-100 guard column (5 µm particle size, 100 Å pore size; Sepax Technologies, Inc., Newark, DE). The elution was carried out isocratically with 10 mM sodium phosphate buffer, pH 7.5, at 22.5°C at a flow rate of 0.5 mL/min and monitored by UV absorbance at 280 nm. Column standardization was done using a mixture of ovalbumin (65 kDa), myoglobin (17 kDa), oxidized insulin B-chain (3.5 kDa), and vitamin B12 (1.3 kDa). Peak detection and area calculations were done using Peak 3.56 Software.
+ Open protocol
+ Expand
5

Homo PSt Characterization using SEC

Check if the same lab product or an alternative is used in the 5 most similar protocols
The analyses of the extracted homo PSt were performed on a system with an M510 pump, U6K universal injector, 486 tunable absorbance detector (all from Waters Corporation, Milford, MA, USA) and 250 dual refractometer/viscometer detector (Viscotek/Malvern Corporation). The separation was achieved over a set of three 5 µm Styragel columns (HR 2, 3 abnd 5, Waters Corporation) and calibrated with 17 narrow dispersity PSt standards with molecular masses between 0.162 kDa and 956 kDa.
+ Open protocol
+ Expand
6

Quantification of Fucoxanthin in Diatom Cultures

Check if the same lab product or an alternative is used in the 5 most similar protocols
The growth of P. tricornutum cells was determined by measuring the absorbance of the culture at 680 nm using a spectrophotometer (UV-5200, China). Cell numbers were also counted using a hemocytometer under a microscope.
For fucoxanthin quantification, algal cells were pelleted from an 80 mL culture by centrifugation at 12,000× g at 4 °C. Harvested cells were frozen-dried, weighed and then ground into fine powder. One mL anhydrous ethanol was added to each of the algal samples, and the extraction was performed by vigorous vortex for 3 times (30 s each). The extraction was repeated three times. After each extraction, the mixture was centrifuged at 13,000× g at 4 °C for 10 min and the supernatants were combined, filtered through a 0.22-µm pore size membrane (Sangon Biotech, Shanghai, China) and then stored at 4 °C in the dark for further analysis.
Carotenoids in the extract were separated by a 2695 high performance liquid chromatography (HPLC) Separations Module (Waters, Milford, MA, USA). Fucoxanthin was detected by monitoring the absorbance at 445 nm using a 486 Tunable Absorbance Detector (Waters). A YMC Carotenoid HPLC Column (250 mm × 4.6 mm with 5-µm particle size) (YMC, Kyoto, Japan) was used and the elution was performed using a gradient of methanol/water ratio from 90/10 to 100/0 over 30 min, followed by a further 20 min with 100% methanol.
+ Open protocol
+ Expand
7

Determining HER2(scFv)-PE24B Purity

Check if the same lab product or an alternative is used in the 5 most similar protocols
To determine the HER2(scFv)-PE24B purity, the final product was analyzed by HPLC using a Protein-pak 300SW SEC 7.5 × 300 mm column. The column was equilibrated with at least 10 CVs of 1 × PBS buffer, pH 7.4, using a Waters 600 Controller connected to a Waters 486 Tunable Absorbance Detector and a Waters 717 Plus Autosampler from Waters Corporation (Milford, MA, USA). The protein was injected onto the column at a flow rate of 1 mL/min over 25 min. The protein elution peaks were detected at 280 nm and checked by SDS-PAGE. The HER2(scFv)-PE24B purity was also evaluated using a Silver Stain Plus kit. The reaction was completed by adding 5% acetic acid (v/v) for 15 min when the bands became visible.
+ Open protocol
+ Expand
8

Isolation and Purification of Phytochemicals

Check if the same lab product or an alternative is used in the 5 most similar protocols
AG (98%) was supplied by Sigma-Aldrich, USA. The other two phytochemicals (DDAG and NAG) were obtained from in-house standards collection. Solvents (AR grade) used for isolation and purification of the compounds were supplied by Fisher Scientific (UK). Silica gel (70–230 MESH) and 20×20 cm silica gel 60 F254-coated TLC plates were purchased from Merck (Darmstadt, Germany). In addition, HPLC grade solvents including methanol and acetonitrile were provided by Merck (Darmstadt, Germany). The HPLC system was supported by Waters™ and consisted of Waters™ 600 Controller pumps, Waters™ 717plus Autosampler injector with a capacity of 96 samples. LiChrocart® HPLC-Column RP-18 (150×4.6 mm, Merck, Germany) was used as the stationary phase. The isocratic mobile phase was implemented with acetonitrile- water (40∶60 v/v) and 0.1% (v/v) analytical grade phosphoric acid dissolved in ultra-pure water at a flow rate of 1 mL/min [28] . The water used in this research was purified using the MilliporeTM water purification system. Detection was done at 223 nm using Waters™ 486 Tunable Absorbance Detector (photodiode array detector).
+ Open protocol
+ Expand
9

Polymer Standards Analysis by SEC

Check if the same lab product or an alternative is used in the 5 most similar protocols
The analyses of the extracted linear PSt were performed on a liquid chromatography line consisting of M510 pump, U6K universal injector, 486 tunable absorbance detector (Waters Corporation, Milford, MA, USA), and 250 dual refractometer/viscometer detector (Viscotek/Malvern Panalytical Ltd, Malvern, UK). The separation was carried out with a set of three 5 µm Styragel columns (HR 2, 3, and 5, Waters Corporation) and calibrated with 17 narrow dispersity PSt standards with molecular masses between 0.162 kDa and 956 kDa (Polymer Standards Service, Amherst, MA, USA).
+ Open protocol
+ Expand
10

Spectroscopic Analysis of Isolated Compounds

Check if the same lab product or an alternative is used in the 5 most similar protocols
Optical rotation was obtained on a JASCO P-2000 polarimeter in MeOH. The UV spectra were recorded with a PerkinElmer UV-vis spectrophotometer, whereas the IR spectra were obtained using a PerkinElmer FTS FT-IR spectrophotometer. The Bruker Ultrashield 500 MHz NMR spectrometer was used to measure the 1D and 2D NMR spectra of the isolated compounds. Chemical shifts (δ) are expressed in ppm with CDCl3 (δH 7.24 and δC 77.0) using TMS as an internal reference, reference to the solvent signals. A MicroTOF, Bruker Daltonics mass spectrometer was employed to acquire HRESIMS spectra. Column chromatography (CC) was performed on silica gel 60H (5–40 μm), silica gel 100 (63–200 μm) and Sephadex LH-20 (Pharmacia Fine Chemical Co., Ltd., Sweden). Semipreparative HPLC separations were performed on a Waters 626 liquid chromatography system equipped with a Grace C18 column (Econosil C18, 10 μm, 10.0 i.d. × 250 mm) and a Waters 486 Tunable Absorbance detector (Waters, USA). Fractions obtained from CC were monitored by TLC using precoated plates of silica gel 60 F254 (Merck).
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!