The largest database of trusted experimental protocols

Agilent 5973c network mass selective detector

Manufactured by Agilent Technologies
Sourced in United States

The Agilent 5973c Network Mass Selective Detector is a laboratory analytical instrument designed for the detection and identification of chemical compounds. The device utilizes mass spectrometry technology to analyze the molecular composition of samples.

Automatically generated - may contain errors

7 protocols using agilent 5973c network mass selective detector

1

Quantitative Analysis of Cobalt Protoporphyrin IX

Check if the same lab product or an alternative is used in the 5 most similar protocols
The extracted organic compounds were separated using an Agilent 7890A Series Gas Chromatograph interfaced to an Agilent 5973c Network Mass Selective Detector and an Agilent 7683 Series Injector (Agilent Technologies, Santa Clara, CA, USA). A 5 µL sample was injected with a split ratio of 1:5 by 0.3% standard deviation into an HP-5MS column (30 m × 0.25 mm, 0.25 µm film thickness; Agilent Technologies, USA) using helium as the carrier gas at 1 mL/min. The ion source was maintained at 250 °C. The GC oven was programmed with a temperature gradient starting at 100 °C (for 3 min), which was gradually increased to 300 °C (for 5 min) at 8 °C/min.
MS was carried out in the electron-impact mode at an ionizing potential of 70 eV by selected ion monitoring (SIM). The presence of CoPPIX-specific m/z 621 ion was monitored.
The qualitative and quantitative analyses of CoPPIX were performed using synthetic CoPPIX as a standard (Figure S1). Its concentration was determined based on the standard curve (0.08, 0.8, 1.6, 16, and 32 µM (50, 500, 1000, 10,000, and 20,000 µg/L)). A concentration curve was plotted with the peak areas corresponding to the concentration of CoPPIX tested (Figure S2). The analysis was performed in triplicate. Error bars represented the classical standard deviation for 3 replicates. The statistical significance of the obtained results was tested using the Student’s t-test.
+ Open protocol
+ Expand
2

GC-MS Analysis of Organic Compounds

Check if the same lab product or an alternative is used in the 5 most similar protocols
The separation of organic compounds was performed using an Agilent 7890A Series Gas Chromatograph (GC) interfaced to an Agilent 5973c Network Mass Selective Detector and an Agilent 7683 Series Injector (Agilent Technologies, Santa Clara, CA, USA). A 5 µL sample was injected with split 1:5 (sample; carrier gas) by 0.3% SD to a HP-5MS column (30 m × 0.25 mm I.D., 0.25 µm film thickness, Agilent Technologies, Santa Clara, CA, USA) using He as the carrier gas at 1 mL min−1. The ion source was maintained at 250 °C; the GC oven was programmed with a temperature gradient starting at 100 °C (for 3 min), and this was gradually increased to 300 °C (for 5 min) at 8 °C min−1. Mass spectrometry analysis was carried out in the electron impact mode at an ionizing potential of 70 eV. Mass spectra were recorded from m/z 40 to 800 (0–30 min).
+ Open protocol
+ Expand
3

GC-MS Analysis of Gaseous Compounds

Check if the same lab product or an alternative is used in the 5 most similar protocols
Gaseous compounds were separated using an Agilent 7890A Series Gas Chromatograph interfaced to an Agilent 5973c Network Mass Selective Detector (Agilent Technologies, Santa Clara, CA, USA). A 5 cm3 gas sample was injected into an HP-1MS column (30 m × 0.25 mm I.D., 0.25 µm film thickness; Agilent Technologies, Santa Clara, CA, USA) with a splitless injector using helium as the carrier gas (1 mL/min). The ion source was maintained at a temperature of 250 °C. The GC oven was programmed with a temperature gradient starting at 35 °C (for 5 min) and gradually increasing to 100 °C at 12 °C/min.
MS analysis was carried out in the electron-impact mode at an ionizing potential of 70 eV. The mass spectra of the samples were recorded at the mass-to-charge (m/z) ratio of 0–150 (0–10.5 min).
Selected compounds were identified using Agilent Technologies Enhanced ChemStation (G1701EA ver. E. 02.00.493) and Wiley Registry of Mass Spectral Data (version 3.2; Palisade Corporation, 8th Edition with Structures (Copyright 1988–2000), and John Wiley and Sons, Inc. (Copyright 2000)) using a 3% cutoff threshold.
+ Open protocol
+ Expand
4

GC-MS Analysis of Organic Compounds

Check if the same lab product or an alternative is used in the 5 most similar protocols
The separation of organic compounds was performed using an Agilent 7890A Series Gas Chromatograph (GC) interfaced to an Agilent 5973c Network Mass Selective Detector and an Agilent 7683 Series Injector (Agilent Technologies, USA). A 5 μl sample was introduced (split by 0.3% SD) to an HP-5MS column (30 m × 0.25 mm I.D., 0.25 μm film thickness, Agilent Technologies, USA) using He as the carrier gas at 1 ml/min. The ion source was maintained at 250°C; the GC oven was programmed with a temperature gradient starting at 100°C (for 3 min) and this was gradually increased to 300°C (for 5 min) at 8°C/min. A mass spectrometry (MS) analysis was performed in electron-impact mode at an ionizing potential of 70 eV. Mass spectra were recorded from m/z 40 to 800 (0–30 min).
+ Open protocol
+ Expand
5

Organotin Compound Separation by GC-MS

Check if the same lab product or an alternative is used in the 5 most similar protocols
The separation of organotin compounds was performed using an Agilent 7890A Series Gas Chromatograph interfaced to an Agilent 5973c Network Mass Selective Detector and an Agilent 7683 Series Injector (Agilent Technologies, Santa Clara, CA, USA). A 5 μL sample was injected in splitless mode (volume relative standard deviation was 0.3%) to an HP-5MS column (30 m × 0.25 mm I.D., 0.25 μm film thickness, Agilent Technologies, Santa Clara, CA, USA) using helium as the carrier gas at 1 mL/min flow. The ion source was maintained at 230 °C; the GC oven was programmed with a temperature gradient starting at 40 °C for 3 min, then increased with a 10 °C/min rate to 220 °C, held for 5 min, after that increased to 20 °C/min rate to 300 °C, and held for 10 min. MS was carried out in the electron-impact mode at an ionizing potential of 70 eV. Mass spectra were recorded in the range of 40–800 mass-to-charge ratio (m/z). Identification of the selected organic compounds was performed with an Agilent Technologies Enhanced ChemStation (G1701EA ver. E.02.00.493) and The Wiley Registry of Mass Spectral Data (version 3.2, Copyright 1988–2000 by Palisade Corporation with, 8th 213 Edition with Structures, Copyright 2000 by John Wiley and Sons, Inc., Hoboken, NJ, USA) using a 3% cut-off threshold.
+ Open protocol
+ Expand
6

Gas Chromatography-Mass Spectrometry Analysis

Check if the same lab product or an alternative is used in the 5 most similar protocols
The separation of organic compounds was performed using an Agilent 7890A Series Gas Chromatograph interfaced with an Agilent 5973c Network Mass Selective Detector and an Agilent 7683 Series Injector (Agilent Technologies, USA). A 5 µl sample was injected with split 1:5 (sample; carrier gas) by 0.3% SD to an HP-5MS column (30 m × 0.25 mm I.D., 0.25 µm film thickness, Agilent Technologies, USA) using He as the carrier gas at 1 ml min−1. The ion source was maintained at 250 °C; the GC oven was programmed with a temperature gradient starting at 100 °C (for 3 min), and this was gradually increased to 300 °C (for 2 min) at 6 °C min−1. Mass spectrometry analysis was carried out in the electron-impact mode at an ionizing potential of 70 eV. Mass spectra were recorded from m/z 40 to 800 (0–39 min).
+ Open protocol
+ Expand
7

Gas Chromatography-Mass Spectrometry Analysis

Check if the same lab product or an alternative is used in the 5 most similar protocols
The separation of single mass was performed using an Agilent 7890A Series Gas Chromatograph interfaced to an Agilent 5973c Network Mass Selective Detector and an Agilent 7683 Series Injector (Agilent Technologies, Palo Alto, CA, USA). A 5 µL sample was injected with split 1:5 (sample/carrier gas) to an HP-5MS column (30 m × 0.25 mm I.D., 0.25 µm film thickness) using He as the carrier gas at 1 mL min−1. The ion source was maintained at 250 °C; the GC oven was programmed with a temperature gradient starting at 50 °C (for 8 min) and this was gradually increased to 325 °C (for 10 min) at 7 °C min−1. Mass spectrometry analysis was conducted in the electron-impact mode at an ionizing potential of 70 eV. The mass spectra were recorded from the selected ion monitoring (SIM) mode. In the SIM mode, the GC-MS collected signals from the individual ions.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!