The largest database of trusted experimental protocols

H 9000 microscope

Manufactured by Hitachi

The H-9000 is a high-performance microscope designed for laboratory use. It features advanced optics and a user-friendly interface. The H-9000 is capable of providing clear, detailed images for a variety of scientific and research applications.

Automatically generated - may contain errors

3 protocols using h 9000 microscope

1

Multifunctional BN-PCPA Composite Characterization

Check if the same lab product or an alternative is used in the 5 most similar protocols
Thermogravimetric
analysis (TGA) was performed on TA SDT650 thermal analysis mass spectrometry
under nitrogen. The surface composition of BN and BN-PCPA was conducted
via an X-ray photoelectron spectroscopy (XPS, Thermo Electron Corporation)
system by using ESCALAB 250. Scanning electron microscopy (SEM; FEI
Quattro S field-emission microscope) and high-resolution transmission
electron microscopy (HR-TEM; Hitachi H9000 microscope) were utilized
to observed the microsturcture of samples. An Instron 3366 tensile
apparatus was used to measure the stress–strain curves at a
speed of 50 mm/min. The elastic modulus values of samples were determined
by the slopes of stress–strain curves at 10% strain. The dielectric
behavior of composites was determined by using a broadband dielectric
spectrometer (Novocontrol, Alpha-A, GmbH Germany). The thermal conductivity
of the sample with a diameter of 20 mm and a thickness of 1 mm was
characterized via a DRL-III flat thermal conduction instrument (Xiangtan,
China). The following equation was as follows where λ, Q, X, and A were the thermal
conductivity
W/(m·K), heat flux (W), thickness (m), and area (m2) of the samples, respectively. ΔT was the
temperature difference.
+ Open protocol
+ Expand
2

Characterization of CTAB-Stabilized Gold Nanorods

Check if the same lab product or an alternative is used in the 5 most similar protocols
The CTAB-stabilized AuNR (catalogue no. A12-10-750) with axial dimension = 10 nm, long size dimension = 35 nm, and absorbance peak = 750 nm were supplied by Nanopartz TM (Salt Lake City, UT, USA) and stored according to manufacturer's recommendations. Size and morphology of the CTAB-AuNR were assessed by Transmission Electron Microscopy (TEM) either in the stock or in working suspensions using a Hitachi H-9000 microscope operated at 300 kV. For this analysis, a drop of the suspensions under study was placed on a carbon-coated copper (Cu) grid and the solvent was left to evaporate at room temperature. Particle size distribution was measured in TEM images (n = 150 NP) using the ImageJ software (NIH, USA). To determine the crystallographic nature and purity of the tested AuNR, a TEM-energy dispersive X-ray (EDX) analysis was performed. Zeta potential was measured by Laser Doppler electrophoresis using a Zetasizer Nano ZS (Malvern Instruments, Worcestershire, UK). Measurements were performed in triplicate, either in distilled water or in zebrafish water (ZW).
+ Open protocol
+ Expand
3

Synthesis and Characterization of EDTA-Functionalized Magnetic Nanoparticles

Check if the same lab product or an alternative is used in the 5 most similar protocols
A detailed description of NPs@EDTA (patent register PT107608) preparation has been described elsewhere [7] (link) but, essentially, comprised: i) the synthesis of the magnetic core made of Fe 3 O 4 through oxidative hydrolysis of an Iron (II) salt in alkaline medium; ii) surface coating with a shell of amorphous silica (SiO 2 ) in alkaline environment using tetraethyl ortho-silicate as a precursor; and iii) surface chemical functionalization with N-(trimethoxysilylpropyl) ethylenediamine triacetate trisodium salt (EDTA-TMS), as illustrated in Figure S1.
Morphology and size of NPs@EDTA particles were obtained through transmission electron microscopy (TEM) using a Hitachi H-9000 microscope operated at 300 kV. Sample for TEM analysis was prepared by evaporating dilute suspensions of the nanoparticles on a copper grid coated with an amorphous carbon film. The Fourier Transform Infrared (FTIR) spectrum was collected using a spectrometer Mattson 7000 with 256 scans and 4 cm À1 resolution, using a horizontal attenuated total reflectance cell. The surface charge of the NPs was assessed by zeta potential measurements, using a Zetasizer Nanoseries equipment from Malvern Instruments.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!