The largest database of trusted experimental protocols

11 protocols using as 4150

1

Quantification of m-Cresol Permeation

Check if the same lab product or an alternative is used in the 5 most similar protocols
m-Cresol in the other side cell in the diffusion cell experiment was measured using High-performance liquid chromatography (HPLC, photo diode array detector: MD-4015, column oven: CO-4061, autosampler: AS-4150, pump: PU-4180) (JASCO, Tokyo, Japan) on C18 column (Mightysil, RP-18 GP 150–4.6 mm, 5 μm). The mobile phase was water containing 25% acetonitrile, the flow rate was 1.2 mL/min, and the column oven was set to 60 °C. The m-cresol remaining in the formulations was calculated from the measurement of the amount that permeated the membrane.
+ Open protocol
+ Expand
2

HPLC Analysis of Polyphenols

Check if the same lab product or an alternative is used in the 5 most similar protocols
High-performance liquid chromatography was performed with the Jasco LC Net II, equipped with the AS-4150 autosampler, the PU-4180 pump and the MD-4010 PDA detector. The system was controlled with the JASCO ChromNAV Version 2.01.00 (JASCO International Co., Ltd., Tokyo, Japan). The experiments were performed on a C18 Kinetex column (150 mm × 4.6 mm, 2.6 μm; Phenomenex, Torrance, CA, USA). The mobile phase consists of A (water containing 1% formic acid) to B (methanol containing 1% formic acid). The gradient program was from 95% A to 80% in 10 min, from 80% to 70% in 5 min, from 70% to 50% in 5 min, from 50% to 0% in 5 min and isocratic for 10 min at a flow rate of 1 mL/min. Five µL of the sample was injected in duplicate onto the column kept at 50 °C.
The UV-Vis absorption spectra of the standards, as well as the samples, were recorded in the range of 190 to 600 nm. Polyphenols were detected at 280, 320 and 360 nm (Figure 3) and identified by the comparison of their retention times and UV-Vis spectra to those of pure standards. Quantification was performed by external standard calibration. The amount of polyphenols was expressed as mg/100 g of dw.
+ Open protocol
+ Expand
3

Monitoring Microbial Growth and Metabolites

Check if the same lab product or an alternative is used in the 5 most similar protocols
Cell growth during cultivation was monitored by measuring OD600 on a Genesys 20 spectrophotometer (Thermo Fisher).
The concentration of glucose and organic acids in cultivation broth was analyzed by a Jasco (Tokyo, Japan) LC-4000 high-performance liquid chromatography system equipped with an autosampler (AS-4150), a pump (PU-4180), a column oven (CO-4061), a UV detector (RI-4030), and an RI detector (UV-4075). The mobile phase consisted of 5 mM H2SO4 and was pumped at constant flow rate of 0.8 mL/min. A Rezex ROA-organic acid H + column (Phenomenex, Torrance, CA, USA) maintained at 80 °C was used to separate the metabolites in the cultivation broth. The concentration of each compound was determined by Jasco ChomNAV software (version 2.03.03).
+ Open protocol
+ Expand
4

RP-HPLC Method Validation for Pharmaceutical Analysis

Check if the same lab product or an alternative is used in the 5 most similar protocols
RP-HPLC analysis was performed on a Jasco Extrema (JASCO, Inc., 28600 Mary’s court, Easton, Maryland 21601, USA) connected with an in built autosampler (AS-4150), quaternary pump (PU-4180) with an on-line degasser, column oven (CO-4061), UV/Vis detector (UV-4075) and an interface box LC-NetII/ADC. Chromatographic data was collected and analysed using ChromNAV Ver.2 software. A second HPLC system (Agilent 1100 series) equipped with variable wavelength detector (G1314A VWD), autosampler (G1313A) and isocratic pump system (G1310A). This system automated with a Chemstation (Agilent technologies) software and which was used for intermediate precision (repeatability) testing and data collection. Chromatographic separations were achieved by using SOLAS 100 Å C18 150 mm × 4.6 mm 5 µm column (Glantreo limited, Ireland). Different HPLC columns such as SOLAS ODS C18 150 mm × 4.6 mm 5 µm, SOLAS ODS C18 150 mm × 4.6 mm 3 µm, SOLAS BDS C18 150 mm × 4.6 mm 5 µm, SOLAS BDS C18 150 mm × 4.6 mm 3 µm and EIROSHELL C18 150 mm × 4.6 mm 2.6 µm (Glantreo limited) were used for this method development purpose. Each Chemical and test samples were accurately weighed using an analytical balance (Adventurer Pro AS214, Ohaus corporation, Pine Brook, NJ USA).
+ Open protocol
+ Expand
5

HPLC Analysis of Organic Acids

Check if the same lab product or an alternative is used in the 5 most similar protocols
Quantification of trimesic acid (H3BTC), H4TazBz, AzA and Nic was performed using HPLC using a reversed phase Jasco LC-4000 series system, equipped with a PDA detector MD-4015 and a multisampler AS-4150 controlled by ChromNav software (Jasco Inc., Easton, MD, USA). A Purple ODS reverse-phase column (5 µm, 4.6 × 150 mm2 Análisis Vínicos) was employed. Note here that AzA was analyzed after its derivatization (der-AzA), as previously described [42 (link)]. The mobile phase consisted of a 50:50 solution (v/v) of buffer (0.04 M, pH = 2.5) and methanol (MeOH) for H3BTC and H4TazBz; 20:80 solution (v/v) acetonitrile:buffer (0.5 mM, pH = 4) for der-AzA; and 5:95 solution (v/v) acetonitrile:buffer pH = 4 for Nic. The injection volume was set at 30 µL with a flow rate of 1 mL min−1 and the column temperature fixed at 25 °C (H3BTC and H4TazBz) and 40 °C (der-AzA and Nic). The standard calibration curves showed a good correlation coefficient ≥0.99. The chromatogram of standard solutions showed a retention time (rt) of 3.51 min (identified as H3BTC, λmax at 225 nm, Figure S1), 21.36 min (identified as H4TazBz, λmax at 205 nm, Figure S2), 9.12 min (identified as der-AzA, λmax at 255 nm, Figure S3), and 5.56 min (identified as Nic, λmax at 213 nm, Figure S4).
+ Open protocol
+ Expand
6

LC-MS Analysis of Protein-Thiol Adducts

Check if the same lab product or an alternative is used in the 5 most similar protocols
All the analyses were performed on an integrated LC-4000 Series JASCO RHPLC system (JASCO, Tokyo, Japan) equipped with a quaternary pump (model No. PU-4180, Tokyo, Japan), a vacuum degasser, an autosampler (model No. AS-4150, Tokyo, Japan), a column oven (model No. CO-4062, Tokyo, Japan), and a fluorescence detector (model No. FP-4020, Tokyo, Japan) operating at the excitation and emission wavelengths of 378 nm and 492 nm, respectively. The detector signal was amplified tenfold. System control and data acquisition processes were performed using the ChromNAV2 software. Spectra Manager ver. 2 was used to analyze the spectra.
The samples (5 μL) were injected onto a 150 × 4.6 mm, 3.6 μm particle size Aeris™ WIDEPORE XB-C18 column (Phenomenex, CA, USA) using an autosampler. The mobile phase consisted of 0.1% TFA in water (A) and 0.1% TFA in ACN (B). All the analyses were performed at room temperature and the flow rate of the mobile phase was 1 mL min−1. Chromatographic separation of HSA and low-molecular-weight thiols was achieved in 35 min with gradient elution as follows: 0–5 min, 5% B; 5–11 min, 5–8% B; 11–18 min, 8–15% B; 18–20 min, 15–30% B; 20–25 min, 30–40% B; 25–26 min, 40–50% B; 26–30 min, 50–70% B; 30–33 min, 70–5% B; 33–35 min, 5% B. Peaks were identified by comparing retention times and fluorescence spectra with corresponding data from the authentic standard.
+ Open protocol
+ Expand
7

Polyphenol Identification and Quantification

Check if the same lab product or an alternative is used in the 5 most similar protocols
Polyphenol identification was performed on a Jasco LC Net II, equipped with the AS-4150 autosampler, the PU-4180 pump and the MD-4010 PDA detector. The method and instrumental conditions were previously optimised (A1-A5) by the authors Lončarić et al. [27 (link)]. Quantification has been performed by external standard calibration and the calibration range for each phenolic standard was 0.1–10 µg/g. The concentration of individuals polyphenols was expressed as µg/g of dry weight (dw).
+ Open protocol
+ Expand
8

HPLC Quantification of Trimesic Acid

Check if the same lab product or an alternative is used in the 5 most similar protocols
Quantification of trimesic acid (H3BTC) was performed using HPLC using a reversed phase Jasco LC-4000 series system, equipped with a PDA detector MD-4015 and a multisampler AS-4150 controlled by ChromNav software v.2 (Jasco Inc., Easton, MD, USA). A Purple ODS reverse-phase column (5 µm, 4.6 × 150 mm2 Análisis Vínicos, Ciudad Real, Spain) was employed. The mobile phase consisted of a 50:50 solution (v/v) of buffer (0.04 M, pH = 2.5) and methanol (MeOH). The injection volume was set at 30 µL with a flow rate of 1 mL·min−1 and the column temperature fixed at 25 °C. The standards used for the calibration curve consisted of trimesic acid solutions in MilliQ water solution with a concentration range from 9.65 to 0.01 µg·mL−1 (correlation coefficient > 0.99). The chromatogram of the standard solution showed a retention time (rt) of 2.70 min (λmax at 225 nm).
+ Open protocol
+ Expand
9

Integrated LC-JASCO System for Fluorescence Detection

Check if the same lab product or an alternative is used in the 5 most similar protocols
Experiments were performed on an integrated LC-4000 Series JASCO RHPLC system (JASCO, Tokyo, Japan) equipped with a quaternary pump (model No. PU-4180, Tokyo, Japan), a vacuum degasser, an autosampler (model No. AS-4150, Tokyo, Japan), a column oven (model No. CO-4062, Tokyo, Japan), and a fluorescence detector (model No. FP-4020, Tokyo, Japan). The detector operated at the excitation and emission wavelengths of 264 nm and 338 nm, respectively. System control and data acquisition processes were performed using the ChromNAV2 software. Spectra Manager ver. 2 was used to analyze the spectra.
+ Open protocol
+ Expand
10

RP-HPLC Analysis of Amino Acids

Check if the same lab product or an alternative is used in the 5 most similar protocols
The total and free amino acids contents were analyzed by RP-HPLC (Reverse Phase-High Performance Liquid Chromatography) using an integrated system from Jasco (Jasco, Japan) equipped with two high-pressure pumps (PU-980), an automatic injector (AS-4150), a fluorescence detector (FP-2020 Plus), and a UV/Vis absorption detector (MD-2015 Plus). The samples (extracts of total and free amino acids) were derivatized online with OPA/3-MPA and FMOC as described by Machado et al. [25 (link)]. Amino acids were identified based on the retention time of the corresponding standards. The quantification of each amino acid was based on the response of the fluorescence signal of each standard, converted into units of concentration through calibration curves obtained for each compound, using the internal standard method. The determination of the total and free amino acid contents was performed in triplicate.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!