The largest database of trusted experimental protocols

5 protocols using glu1 fibrinopeptide b

1

Peptide Separation and Characterization

Check if the same lab product or an alternative is used in the 5 most similar protocols
Peptides were loaded on reversed-phase trap column PST C18, 100 Å, 5 μm, 180 μm × 20 mm (Waters Corporation, UK) with a flow rate of 15 μL/min using loading buffer of 0.1% formic acid and subsequently separated on HSS-T3 C18 1.8 μm, 75 μm × 250 mm analytical column (Waters Corporation, UK) in 90 min linear gradient (A: 0.1% formic acid, B: 100% CH3CN, and 0.1% formic acid) at a flow rate of 300 nl per min.
The nano-LC was coupled with HDMS Synapt G2 mass spectrometer (Waters Corporation, UK). Data were acquired using Masslynx version 4.1 software (Waters Corporation, UK) in positive ion mode. LC-MS data were collected using data independent acquisition (DIA) mode MSE in combination with online ion mobility separation. The trap collision energy of mass spectrometer was ramped from 18 to 40 eV for high-energy scans in MSE mode. The trap and transfer collision energy for high-energy scans in HDMS mode was ramped from 4 to 5 eV and from 27 to 50 eV. For both analyses, the mass range was set to 50–2,000 Da with a scan time set to 0.9 seconds. A reference compound [Glu1]-fibrinopeptide B (Waters Corporation, UK) was infused continuously (500 fmol/μL at flow rate 500 nL per min) and scanned every 1 minute for online mass spectrometer calibration purpose. The samples were run in triplicate.
+ Open protocol
+ Expand
2

UPLC-MS Analysis of N-glycans

Check if the same lab product or an alternative is used in the 5 most similar protocols
Ten microlitres (10 μL) of reconstituted released N-glycans were injected into an ACQUITY H-Class UPLC (Waters Corporation, United States) coupled to a Xevo G2S QTof mass spectrometer (Waters Corporation, United States). Samples were separated using an ACQUITY UPLC Glycan BEH amide column (130 Å, 1.7 μm, 2.1 mm × 150 mm, Waters Corporation, United States) at 60°C and 400 μL/min, with a 40 min gradient from 25 to 49% of 50 mM Ammonium Formate (mobile phase A). 100% ACN was used as mobile phase B. RFMS-labelled glycans were excited at 265 nm and measured at 425 nm with an ACQUITY UPLC FLR detector (Waters Corporation, United States). The MS1 profile scans of m/z 400–2,000 were acquired using the Xevo G2S-QTof in positive mode with an acquisition rate of 1 Hz. The electrospray ionisation capillary voltage was set at 2.75 kV, cone voltage at 15 V, desolvation gas flow at 800 L/h, ion source temperature and desolvation temperature were kept at 120°C and 300°C, respectively. Glu1-fibrinopeptide B (Waters Corporation, United States) was used as the LockSpray compound for real-time mass correction. RapiFluor-MS Dextran Calibration ladder (Waters Corporation, United States) was also injected into LC-MS to calibrate the retention time of sample peaks. The retention times were normalised using the dextran calibration curve to Glucose Units (GU).
+ Open protocol
+ Expand
3

Comprehensive Proteomics Sample Preparation

Check if the same lab product or an alternative is used in the 5 most similar protocols
Ammonium bicarbonate was from Sigma Aldrich (St. Louis, MO). Emplura absolute ethanol and high-purity sodium deoxycholate (99 %) were from VWR International (Radnor, PA). Trypsin/Lys-C Mix (mass spectrometry grade) was from Promega (Madison, WI). Glu-1-Fibrinopeptide B was from Waters (Milford, MA). All other items were purchased from ThermoFisher Scientific (Waltham, MA). Nanopure water was obtained from a Purelab Flex 3 water purification system (Elga, Veolia Environment S. A., Paris, France).
+ Open protocol
+ Expand
4

Nanoflow LC-IMS-MS Protocol for Vesicle Analysis

Check if the same lab product or an alternative is used in the 5 most similar protocols
Purified vesicles were analysed by ultra-performance nanoLC (ACQUITY M Class, Waters) coupled to an IMS mass spectrometer (SYNAPT G2-Si, Waters) fitted with a NanoLockSpray source (Waters). Samples were loaded via a Symmetry C18 5 µm, 180 µm × 20 mm trap column (Waters) and separated through a HSS T3 C18 1.8 µm, 75 µm × 150 mm analytical column (Waters). Samples were separated using a 5 to 95% acetonitrile 0.1% formic acid gradient over 35 min at a flow rate of 300 nL/min. The mass spectrometer was operated in positive ion mode with a capillary voltage of 3.0 kV, cone voltage of 40 V and a source offset of 80 V. Before analysis the instrument was calibrated with NaI and during analysis a LockMass reference, Glu-1-fibrinopeptide B (Waters), was delivered to the NanoLockSpray source. Mass spectra were collected, over 50–2000 m/z, alternating between low (4 eV) and elevated (15–45 eV) collision energies at a scan speed of 1.0 s. Data were processed using the UNIFI software package from Waters (Wilmslow, U.K.).
+ Open protocol
+ Expand
5

Trypsin-Cleaved Peptide Separation by nanoLC-MS

Check if the same lab product or an alternative is used in the 5 most similar protocols
The liquid chromatography (LC) separation of trypsin-cleaved peptides was performed with nanoAcquity UPLC system (Waters Corporation, UK). Peptides were loaded on a reversed-phase trap column PST C18 (Waters Corporation) at a flow rate of 15 ml/min using loading buffer of 0.1% formic acid and subsequently separated on HSS-T3 250 mm analytical column (Waters Corporation) in 30-min linear gradient (A: 0.1% formic acid, B: 100% CH 3 CN and 0.1% formic acid at a flow rate of 300 nl/min). The nano-LC was coupled online through HDMS Synapt G2 mass spectrometer (Waters Corporation) . The data was acquired using Masslynx version 4.1 software (Waters Corporation) in a positive ion mode. LC-MS data were collected using data-independent acquisition mode MSE with online ion mobility separation. Mass range was set to 50-2,000 Da with a scan time set to 0.75 s. A reference compound [Glu1]-Fibrinopeptide B (Waters Corporation) was continuously infused (500 fmol/ml at a flow rate 500 nl/min) and scanned every 30 s for online mass spectrometer calibration purpose.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!