The largest database of trusted experimental protocols

13 protocols using fluorocube

1

Trp Fluorescence Anisotropy Decay Analysis

Check if the same lab product or an alternative is used in the 5 most similar protocols
The fluorescence properties of Trp in proteins were probed by exciting at 295 nm and monitoring the emission for 310–450 nm (λem = 338 nm) at 25 °C using Jobin Yvon Fluoromax‐4 spectrofluorometer equipped with a PMT detector. The slit width (2 nm), step size (0.1 nm), and integration time (0.05 s) were maintained for all experiments.
Fluorescence anisotropy decay measurements were performed using TCSPC (Fluorocube, Horiba Jobin Yvon, Kyoto, Japan). For decay measurements, the emission polarizer was set to 0° with respect to excitation polarizer for parallel measurements and at 90° for perpendicular measurements. A 293 nm laser diode was used as an excitation source, and the emission monochromator for tryptophan was fixed at 342 nm, at a slit width of 8 nm. The instrument response function was measured using 2% LUDOX (Sigma‐Aldrich).
The anisotropy decay was calculated using Equation (2) and was fitted to biexponential decay fit to determine the values of rotational correlation times using Equation (3). The values are tabulated in Table S6. rt=It-ItIt+2It
rt=r0A1e-tϕ1+A2e-tϕ2
where I is the vertical emission and I is the horizontal emission.
+ Open protocol
+ Expand
2

Absorption, Emission, and Lifetime Spectroscopy

Check if the same lab product or an alternative is used in the 5 most similar protocols
UV/Vis absorption spectra were recorded on a PerkinElmer (Lambda 750) UV-VIS-NIR spectrophotometer. Fluorescence and phosphorescence emission spectra were recorded on a PerkinElmer (LS-55) spectrofluorophotometer. Fluorescence lifetimes were measured on a HORIBA Scientific time-correlated single-photon counting system (FluoroCube) with the laser light (DeltaDiode, laser diode head, 404 nm, pulse width: 100 ps) as the excitation source. Phosphorescence lifetimes were measured on a JASCO FP-8500. The absolute fluorescence quantum yields were determined by a Hamamatsu Photonics C9920-02 system equipped with an integrating sphere and a red-sensitive multichannel photodetector (PMA-12): excitation wavelength = 300 nm.
+ Open protocol
+ Expand
3

Time-Resolved Fluorescence Spectroscopy

Check if the same lab product or an alternative is used in the 5 most similar protocols
Fluorescence decay lifetimes were measured using a time-correlated single photon counting instrument (Fluorocube, Horiba JobinYvon, Edison, NJ) with a 457 nm pulsed LED (NanoLED-293, Horiba JobinYvon, Edison NJ) light source having a 1 ns pulse width and 1 MHz repetition rate. Decays were recorded at the emission maximum of each PDI with a 32 nm bandwidth.
+ Open protocol
+ Expand
4

Fluorescence Lifetime Analysis of QBA-DNA Interactions

Check if the same lab product or an alternative is used in the 5 most similar protocols
QBAs in 3 × 10-6M concentration were prepared in 0.1M borate buffer, pH 9.45 (alkanolamine form) or in 0.1M acetate buffer, pH 3.95 (iminium form). For measurement of lifetimes of QBA-DNA complexes, ctDNA was added (DNA base pair:drug ratio 1.6:1). Measurements were performed on a Fluorocube (Horiba Jobin Yvon, France) with a 329 nm excitation LED diode (pulse width 1.2 ns) and emission monochromator set to 440 nm (alkanolamine form) or 610 nm (iminium form, 570 nm for chelerythrine). Instrument response function was obtained by measuring a solution of colloidal silica. When necessary, intensity was lowered by neutral filters. For each QBA three measurements were performed and lifetimes were globally fitted using DecayFit 1.3. Quality of a fit was considered by inspecting residuals and reduced χ2 value around 1. SE are reported in Table 1.
+ Open protocol
+ Expand
5

Fluorescence Lifetime Measurement of HSA

Check if the same lab product or an alternative is used in the 5 most similar protocols
Fluorescence decay lifetimes were measured using a time-correlated single photon counting (TCSPC) instrument (Fluorocube, Horiba Scientific, Edison, NJ) with a 294 nm pulsed LED (NanoLED-293, Horiba JobinYvon, Edison NJ) source of ~1 ns pulse width and 1 MHz repetition rate. Decays were recorded at the emission maximum of HSA with a 12 nm bandwidth.
+ Open protocol
+ Expand
6

Spectroscopic Characterization of Fluorescent Proteins

Check if the same lab product or an alternative is used in the 5 most similar protocols
The absorption
spectra were recorded with a spectrophotometer (V-670, Jasco). The
fluorescence spectra were recorded using a photonic multi-channel
analyzer (C10027, Hamamatsu Photonics). The fluorescence decay curves
of EGFP emissions were measured by excitation at 485 nm using a time-correlated
single-photon counting kit (Horiba Fluoro Cube). The fluorescence
autocorrelation curves were measured on a compact FCS system (C9413-01MOD,
Hamamatsu Photonics, Japan) at an excitation of 473 nm using a laser-diode-pumped
solid-state laser. The size of the pinhole was 25 μm, and the
spectral range of the detection wavelengths was 500–900 nm.
For the determination of the concentration of GSH–QDs, the
number of QD particles in a 10 μL solution was measured by using
FCS, and the QD concentration was estimated by using a 20 nM solution
of rhodamine 6G as a reference. Fluorescence quantum yields were measured
using an absolute quantum yield measurement system (C9920, Hamamatsu
Photonics). Excitation wavelengths were set to be 475 nm for annexin
V–EGFP and 488 nm for GSH–QDs.
+ Open protocol
+ Expand
7

Fluorescence Lifetime Spectroscopy of Tubulin

Check if the same lab product or an alternative is used in the 5 most similar protocols
Fluorescence lifetime was recorded using a time-correlated single-photon-counting (TCSPC) instrument (Fluorocube, Horiba Scientific, Edison, NJ). The details of the method can be found elsewhere [21 , 38 (link)] and are only summarized in the following: tubulin fluorescence decay was recorded upon excitation with a pulsed LED at 294 nm (NanoLED-295, 1.2 ns pulsewidth, Horiba Scientific), operated at a repetition rate of 1.0 MHz. The instrument response function (prompt) was recorded after each decay curve using a suspension of glycogen (1 mg/mL) in deionized water. All signals were recorded by accumulating 104 peak counts at a counting rate below the suggested maximum (2 × 104 cts/s) corresponding to 2% of the repetition rate of the source. The decays shown in Figure 4 and 9 are plots of the logarithm of the number single photons counted as a function of the time delay from the trigger of the pulse of the Nano-LED source.
+ Open protocol
+ Expand
8

Characterization of Graphene Quantum Dots

Check if the same lab product or an alternative is used in the 5 most similar protocols
HR-TEM images were taken using a 2100F field emission gun TEM (JEM 2100F, JEOL, USA, 200 kV) for on-GQDs and off-GOQDs samples. XPS spectra were recorded for both samples using VG ESCALAB 220i (Thermo scientific, USA). XPS survey and high-resolution scans were performed with the pass energies of 100 eV and 20 eV, respectively. X-ray beam size was approximately 100 μm. On-GQDs and off-GOQDs samples for XPS measurement were prepared via a spin coating technique. Silicon (Si) substrate was used for spin coating. The rotation speed was adjusted to 3000 rpm. The samples were dried at room temperature for 2 h before the measurement. Room temperature PL spectra of on-GQDs and off-GOQDs were collected using a photoluminescence spectrophotometer (Horiba, Fluoromax-plus with 150 W xonon arc lamp) in the wavelength range from 300 nm to 700 nm. Raman spectra were obtained for both samples using Raman spectrometer (LabRam Aramis, Horiba jobin Yvon, 514 nm Ar-ion laser used with a power of 1 mW). A time-correlated single photon counting (TCSPC) spectrometer (Horiba Jobin Yvon, FluoroCube) was used to measure nanosecond lifetime.
+ Open protocol
+ Expand
9

Photophysical Characterization of Organic B

Check if the same lab product or an alternative is used in the 5 most similar protocols
UV–vis measurements were performed in transmission mode using an AvaLight-DHS-Bal (Balanced Deuterium-Halogen light source (200−2000 nm)). Extinction spectra of B dissolved in toluene were recorded on a jasco UV/Vis V-670 spectrophotometer. Photoluminescence spectra including the derived quantum yields were measured under current nitrogen flow using a PTI QuantaMaster 40 spectrofluorometer equipped with an integrated sphere (Labsphere) and a Hamamatsu R928P photomultiplier. The samples were excited at wavelengths of 355, 375, and 410 nm with a Xenon lamp using a monochromator. The TRPL spectra were recorded under current nitrogen flow using a time-correlated single photon counting (TCSPC) system (Horiba Fluorocube). TRPL samples were excited at 375 nm with a pulsed laser diode (Horiba NanoLED Model N-375L).
+ Open protocol
+ Expand
10

Synthesis and Characterization of 1-Pyrenebutyric Acid N-Hydroxysuccinimide Ester

Check if the same lab product or an alternative is used in the 5 most similar protocols
Unless noted otherwise, all reagents were obtained from commercial sources and used without further purification. The 1-pyrenebutyric acid N-hydroxysuccinimide ester was obtained from Tokyo Chemical Industry Co., Ltd. (Tokyo, Japan). Dimethyl sulfoxide was obtained from Fujifilm Wako Pure Chemical Corporation (Osaka, Japan). DMSO-d6, was obtained from Sigma-Aldrich Inc. (St. Louis, MO, USA). UV absorption spectroscopy was performed using a DU-800 spectrophotometer (Beckman-Coulter, Brea, CA, USA). Fluorescence spectra were measured using a LS50B fluorescence spectrometer (Perkin Elmer, Waltham, MA, USA). Time-resolved fluorescence measurement was performed using a single photon counter FluoroCube (Horiba, Kyoto, Japan). 1H NMR spectra were recorded using an Avance III-500 NMR instrument (Bruker, Billerica, MA) at the Biomedical Research Institute, National Institute of Advanced Industrial Science and Technology (AIST). HRMS was performed by the Open Research Facilities Station, TIA Central Office, AIST.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!