The largest database of trusted experimental protocols

Uv 1600pc spectrometer

Manufactured by Avantor
Sourced in Canada

The UV-1600PC spectrometer is a laboratory instrument designed for ultraviolet-visible spectroscopy. It measures the absorption or transmittance of light by a sample within the ultraviolet and visible light spectrum.

Automatically generated - may contain errors

6 protocols using uv 1600pc spectrometer

1

Characterization of Synthesized Nanostars

Check if the same lab product or an alternative is used in the 5 most similar protocols
The synthesized nanostars were characterized by transmission electron microscopy (TEM), ultraviolet-visible (UV-Vis) extinction spectroscopy, and nanoparticle tracking analysis (NTA). TEM measurements were obtained using a Tecnai F20 electron microscope in the Center for Electron Microscopy and Analysis (CEMAS) at The Ohio State University. UV-Vis extinction measurements were obtained on VWR UV-1600PC spectrometer. NTA analysis was performed using a Malvern NanoSight (NS300).
+ Open protocol
+ Expand
2

Bacterial growth, dilution, and enumeration

Check if the same lab product or an alternative is used in the 5 most similar protocols
Lysogeny broth was used to grow S. aureus bacteria at 37 °C overnight. The grown solution was serially diluted from 1/10 to 1/109 times in PBS. Each solution was spotted onto agar plates (3 × 10 μL), as well as measured using a UV-1600PC spectrometer (VWR International, Mississauga, ON, Canada) to measure the optical density at 600 nm (OD600). The plates were incubated overnight at 37 °C and spot counted to calculate CFU per OD600 (see Supplementary Material, Section S3).
For TSM measurements, S. aureus, E. coli, or P. aeruginosa bacteria was grown overnight at 37 °C. The next day, 1 mL of the bacteria solution was centrifuged at 14,500× g RPM (5 min). A total of 1 mL of PBS was used to resuspend the bacteria pellet. The OD600 of the solution was used to calculate the base CFU, and then the solution was diluted to the desired CFU. To prepare milk samples, the diluted PBS bacteria solution was centrifuged at 14,500× g RPM (5 min) to pellet the bacteria, and then the bacteria was resuspended in milk.
+ Open protocol
+ Expand
3

Photosensitizer Characterization and Quantification

Check if the same lab product or an alternative is used in the 5 most similar protocols
Characterization of the different synthesized photosensitizers was achieved by 1H NMR spectroscopy recorded on a 600 MHz Bruker AVANCE NEO spectrometer and HR-MS (ESI positive, direct injection) recorded with a Bruker Impact II spectrometer. All UV-Vis spectra were recorded using a UV-1600PC spectrometer from VWR with a resolution of 1 nm. Surface analysis of the npAu supports and the final hybrid materials was performed on a scanning electron microscope (SEM, Supra 40, Zeiss) operated at 15.0 kV acceleration voltage, 300 pA probe current and 10 mm working distance and equipped with a Bruker XFalsh 6/30 EDX detector. For the quantification of immobilized photosensitizer, the hybrid material (10 mg) was dissolved in ultrapure aqua regia (2 mL) prior to determination by ICP-MS using an iCAP-Q spectrometer form Thermo Fisher Scientific.
+ Open protocol
+ Expand
4

SN38 Quantification and Formulation

Check if the same lab product or an alternative is used in the 5 most similar protocols
HSA (25% solution) and SN38 were purchased from Octapharma USA and GLPbio Technology, respectively. Methanol, ethanol, other chemicals and suppliers were purchased from VWR. UV spectrum measurement and quantitation were conducted on a UV-1600 PC spectrometer (VWR). Both SPESN38–5 and SPESN38–5 were prepared following similar protocols for making SPEDOX-6 [32 ].
+ Open protocol
+ Expand
5

Aqueous UV-Vis Spectroscopy Protocol

Check if the same lab product or an alternative is used in the 5 most similar protocols
UV-vis measurements were obtained using a VWR UV-1600PC spectrometer. Samples were prepared in aqueous solution and measured in 1 cm quartz cuvettes. Spectra were acquired over the wavelength range of 200 to 400 nm with a spectral resolution of 1 nm.
+ Open protocol
+ Expand
6

Quantum Yield Characterization of Nanoparticles

Check if the same lab product or an alternative is used in the 5 most similar protocols
All absorbance measurements were obtained with a VWR UV-1600PC spectrometer at 1 nm resolution. Quantum yields (QY) were determined via comparison with a fluorescein isothiocyanate (FITC, Sigma) standard using a NIST calibrated Photon Technologies International fluorimeter for each sample, with emission measured from 485 to 800 nm using 475 nm excitation. Each QY was calculated using Equation (3) (Grabolle et al., 2009 (link)). High-resolution transmission electron micrographs (Figures S4, S5) were obtained in the CAMCOR facility at the University of Oregon, and with a Philips CM 100 at 80 kV. Cores exhibited an average diameter of 2.9 ± 0.3 nm, with the core-shells being slightly larger with CdTe/ZnS and CdTe/Cd averaging 3.2 ± 0.5 and 3.1 ± 0.5 nm, which matches the ICP observations of single monolayer-regime coverage.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!