The largest database of trusted experimental protocols

39 protocols using flexcath advance

1

Cryoballoon Ablation for Pulmonary Vein Isolation

Check if the same lab product or an alternative is used in the 5 most similar protocols
One long sheath (FlexCath Advance; Medtronic and SL0; Abbott Medical) was introduced into the LA and RA after the transseptal puncture. A 10-polar circular mapping catheter (Inquiry; AF Division, Abbott Medical, Minneapolis, MN, USA) was positioned at the SVC. All patients underwent PV isolation using a second-generation 28-mm CB catheter inserted through the FlexCath Advance (Medtronic). Balloon positioning and the degree of PV occlusion were evaluated by the injection of a contrast medium diluted at a 1:1 ratio with 0.9% saline. All cryo applications were performed for 180 or 240 s per application alongside one or more additional cryo applications, as determined by the Achieve catheter (Medtronic). During cryo applications to the right PV, continuous phrenic nerve stimulation (10 V for 2.9 ms) was performed using the circular mapping catheter. The delivery of cryo energy was immediately terminated during the loss of phrenic capture [9] (link). Subsequently, PVs were carefully mapped with an Achieve catheter (Medtronic). In the event PV isolation was not achieved, touch-up ablation was performed using an RF catheter (Thermocool, Biosense Webster; Flexability, Abbott Medical; CoolPath, Abbott Medical; or Ablaze, Japan Lifeline, Tokyo, Japan).
+ Open protocol
+ Expand
2

Cryoballoon Ablation for PV Isolation

Check if the same lab product or an alternative is used in the 5 most similar protocols
All the included studies performed PV isolation using 28-mm CB-A inserted through a 15-Fr steerable sheath (FlexCath Advance; Medtronic Inc, Minneapolis, MN) over a 20-mm diameter inner lumen mapping catheter (Achieve; Medtronic Inc). Optimal PV occlusion was achieved by “proximal-seal” technique. The CB-A was inflated and advanced toward the ostium of each PV. When the balloon nadir temperature exceeded −60 °C, the ablation was terminated. During CB ablation of the right-sided PVs, high output right phrenic stimulation (1500 ms, 20 mA) was performed using a quadripolar catheter within the superior vena cava. If loss of phrenic capture occurred, the cryoapplication was immediately terminated. For LCPVs the following ablation strategies were applied: if an antral occlusion of the LCPV, determined by contrast injection and fluoroscopy, was achieved, the freeze cycle was initiated. If an antral LCPV occlusion could not be obtained, the operators adapted a sequential ablation approach. At first the first superior branch of the LCPV was targeted, followed by ablation of the first inferior branch. In all the studies an antral freezing cycle was applied as much as possible to LCPV.
+ Open protocol
+ Expand
3

Cryoballoon Ablation for Atrial Fibrillation

Check if the same lab product or an alternative is used in the 5 most similar protocols
Cryoballoon ablation was performed under conscious or deep sedation. Femoral venous access was not guided by ultrasound. Access to the left atrium was achieved with a single transseptal puncture with an SL-0 sheath (St. Jude Medical, Minneapolis, Minnesota) using intracardiac echocardiography. The SL-0 sheath was then replaced by a steerable sheath (FlexCath Advance, Medtronic), after which left atrial and pulmonary vein (PV) anatomy were visualised by angiography. In order to record PV signals prior to ablation, an innerlumen mapping catheter (Achieve catheter, Medtronic) was placed through the steerable sheath proximal to each PV ostium. The ablation was performed using the second-generation 28 mm cryoballoon (Arctic Front Advance Cardiac CryoAblation Catheter System; Medtronic, Minneapolis, Minnesota). Until 2017, at least two cryothermal applications (lasting 240 s) were delivered to isolate each vein. In 2017, the approach proposed by Aryana et al was adopted.10 (link) The endpoint of the procedure was electrical isolation of PV signals. This was evaluated using the circular Achieve mapping catheter to verify entrance block of all isolated veins. Pacing within the PV’s was performed to assess exit block if there was doubt of entrance block. Diaphragmatic stimulation was performed during ablation of the right side veins to avoid phrenic nerve injury.
+ Open protocol
+ Expand
4

Cryoballoon Ablation for Atrial Fibrillation

Check if the same lab product or an alternative is used in the 5 most similar protocols
After obtaining a single transseptal puncture (similar as above), the transseptal sheath was exchanged over a guidewire for a 15Fr steerable sheath (FlexCath Advance, Medtronic, Minneapolis, USA). A spiral inner lumen mapping catheter (Achieve, Medtronic) preceded the balloon and brought it to each pulmonary vein. A 28 mm second-generation CB (Arctic Front Advance, Medtronic) was inflated and positioned at the ostium. The degree of sealing was confirmed using contrast-medium injected from the distal tip of the balloon. Ablation was performed with a single 240-second freeze for the left upper pulmonary vein and 180-second for other veins. An additional application was done for another 180 seconds, if necessary. If electrical isolation was not achieved after 3 applications per vein, touch-up ablation for 150 seconds was performed using a cryo-catheter (Freezer Max, Medtronic) for each application. As in the RF group, entrance and exit block was also confirmed. However, because of the low incidence of dormant conduction after CB isolation in previous studies (24 (link),25 ), only a high-dose isoproterenol challenge was performed, and dormant conduction was not confirmed by ATP.
+ Open protocol
+ Expand
5

Cryoballoon Pulmonary Vein Isolation Protocol

Check if the same lab product or an alternative is used in the 5 most similar protocols
Cryoballoon PVI was performed as described before in detail [14 (link)]. A 12 F steerable sheath (Flexcath advance, Medtronic, Minneapolis, MN, USA) was positioned in the left atrium (LA) by a single transseptal puncture, followed by PV angiography. A 28 mm cryoballoon (Arctic Front Advance and Arctic Front Advance ST, Medtronic, Minneapolis, MN, USA) was advanced to the LA and guided to the target PV over a 10–25-mm spiral mapping catheter (Achieve or Achieve Advance, Medtronic, Minneapolis, MN, USA). In addition, 45 patients were treated with a fixed protocol with application of one or more freeze cycles of 240 s at each PV and a bonus freeze for each PV after documented PVI by performing entry- and exit-block testing. A total of 174 patients were treated with a time-to-isolation (TTI) guided protocol; the single freeze duration was adapted by the observed TTI. The freeze duration was shortened to 120 s by a TTI of <30 s. A TTI of <60 s resulted in a freeze duration of 180 s without a bonus freeze; a TTI of ≥60 s resulted in a single freeze of 180 s and a bonus freeze of 180 s. If PVI was not achieved following a single freeze application, an additional freeze was administrated until electrical isolation was verified.
+ Open protocol
+ Expand
6

Cryoballoon Ablation for Pulmonary Vein Isolation

Check if the same lab product or an alternative is used in the 5 most similar protocols
LA access was obtained via femoral venous access and a trans-septal approach. A dedicated 15 Fr delivery sheath (FlexCath Advance, Medtronic, Inc., Minneapolis, MN, USA) was introduced into the LA over the wire along with the cryoballoon catheter (Arctic Front Advance, Medtronic). A dedicated inner lumen mapping catheter (Achieve, Medtronic) was commonly used for delivery of the cryoballoon and sheath. The Achieve mapping catheter/guidewire was maneuvered into a pulmonary vein. The inflated cryoballoon was advanced toward the antral surface of the PV and cryoapplication was initiated. PV isolation was confirmed using a spiral mapping catheter, and when the temperature dropped from −40 °C to −60 °C and time to isolation (TTI) was achieved within 60 s, cryoablation was applied for 180 s. If the temperature did not reach an acceptable level or PVI was not achieved within 60 s, the cryoablation lesion was interrupted and additional applications were performed after rewarming and repositioning the balloon catheter (Figure 2B). Phrenic nerve pacing was conducted using a diagnostic catheter at the level of the right subclavian vein during right-sided ablation and diaphragmatic movement was monitored. Catheter ablation was immediately terminated upon weakened diaphragmatic response.
+ Open protocol
+ Expand
7

Catheter Ablation Techniques and Monitoring

Check if the same lab product or an alternative is used in the 5 most similar protocols
All the patients were treated with RF ablation or cryo‐balloon ablation. Mainly, three sheaths were introduced from the femoral vein for ablation, mapping, and insertion of ultrasonography catheters. We used an 8.5‐Fr steerable sheath, Agilis™ NxT (Abbott) or Destino™ (Oscor) for RF ablation and a 12‐Fr steerable sheath, FlexCath Advance™ (Medtronic), for cryo‐balloon ablation. The Swartz™ Braided Transseptal Guiding Introducer SL1 8.5 F (Abbott) was used for the mapping catheter. The Radifocus Introducer IIH® 9‐ or 10‐Fr sheath (Terumo) was used for intracardiac echocardiography. The puncture site was decided on the discretion of the operator, but we mainly inserted the sheaths for ablation and mapping catheter from the right femoral vein and the sheath for the ultrasonography catheter from the left femoral vein from April 2012 to March 2015. Thereafter, we inserted the three sheaths only from the right femoral vein.
We monitored the arterial blood pressure in all the cases with a 20‐G monitoring catheter or the Radifocus Introducer IIH® 4‐Fr sheath (Terumo) for the patients who received concurrent coronary angiography. We inserted these catheters from the right femoral artery.
+ Open protocol
+ Expand
8

Cryoballoon Ablation Procedure Protocol

Check if the same lab product or an alternative is used in the 5 most similar protocols
Cryoballoon ablation procedure was performed either 2nd or 4th generation cryoballoon and has been described in detail before (17 (link)). Briefly, a steerable sheeth (Medtronic, Minneapolis, MN, Flexcath Advance) was used to introduce the cryoballoon in the left atrium after single transseptal puncture. After balloon inflation at the PV ostia a 20-mm spiral mapping catheter (Achieve, Medtronic, Minneapolis, MN) was positioned in the PV at the closest achievable proximity to enable real-time observation of PV potentials during PV isolation. PV occlusion was documented by injection of contrast medium.
Phrenic nerve integrity was monitored by stimulation via the diagnostic catheter placed in the superior vena cava and palpation of the right-sided diaphragm and with additional recording of compound motor action potentials (CMAP) of the right sided diaphragm (18 (link), 19 (link)). Cryoenergy application was aborted in case of reduced palpational feedback and/or reduced CMAP. In general, cryoenergy application was stopped by a single-stop-technique.
+ Open protocol
+ Expand
9

Cryoballoon Ablation Procedure for Atrial Fibrillation

Check if the same lab product or an alternative is used in the 5 most similar protocols
The cryoballoon ablation procedure was performed as previously described.4 (link) In brief, after voltage mapping of the LA and PV, a 15F steerable sheath (FlexCath Advance; Medtronic Inc, Minneapolis, MN) was placed. A 28-mm second-generation cryoballoon (ArcticFront Advance, Medtronic Inc; Cryoballoon) was inflated at the orifice of each PV anchored using a 20-mm circular mapping wire catheter with 10-pole electrodes (Achieve; Medtronic Inc). An optimal occlusion was confirmed using the pooling and leakage maneuver (proximal seal technique). The double-freezing protocol (180 seconds and subsequent 120 seconds) was initially applied to each PV in the following order: left superior (LSPV), left inferior, right inferior (RIPV), and right superior. When the esophageal luminal temperature reached <20°C, the application of cryoenergy was stopped.
+ Open protocol
+ Expand
10

Radiofrequency vs Cryoballoon Ablation

Check if the same lab product or an alternative is used in the 5 most similar protocols
During all analyzed periods, RF ablation was performed using double transseptal puncture for ablation catheter and circular mapping catheter. Changes in RF ablation methodology over time are described below. The CB ablation was performed in standard manner, using a long 8.5 F sheath and Brockenburgh needle for transseptal puncture, then exchanging it for a flexible 15F sheath (FlexCath Advance, Medtronic, USA) and inserting a balloon with diagnostic Achieve electrode (Medtronic, USA). One important difference between patients treated with RF vs CB was the presence of the common trunk of left PV. When this type of anatomy was detected on the pre-procedural cardiac computed tomography (CT) or intra-procedural intracardiac echocardiography (ICE), we used RF.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!