The largest database of trusted experimental protocols

Flash 3d

Manufactured by Siemens
Sourced in Germany

Flash 3D is a laboratory equipment product designed for rapid three-dimensional (3D) scanning and imaging. It captures high-resolution 3D data using advanced optical technology. The core function of Flash 3D is to provide accurate and detailed 3D scans of various objects or samples within a laboratory setting.

Automatically generated - may contain errors

11 protocols using flash 3d

1

Quantitative SPECT Imaging with Tissue Segmentation

Check if the same lab product or an alternative is used in the 5 most similar protocols
SPECT data were reconstructed with Flash3D with attenuation and scatter correction (Siemens, Germany) as the standard reconstruction for clinical use. Quantitative uptake maps were reconstructed with xSPECT Quant and xSPECT Bone (both Siemens, Germany), and include attenuation and scatter correction and standardized calibration for absolute quantification. xSPECT Bone uses the CT information to provide images with improved tissue boundary resolution. It uses a zone map derived from the CT data to segment the anatomical image into five specific tissue classes “zones”: cortical bone, spongiosa, soft tissue, fat tissue and air [23 (link), 24 ]. For attenuation correction, CT density data with and without metal artefact reduction [25 , 26 (link)] were used.
Planar scintigraphies were obtained with a triple phase protocol using the same SPECT/CT scanner system: perfusion phase immediately after tracer injection, blood pool phase ~ 3 min post-injection and delayed phase ~ 3 h post-injection.
+ Open protocol
+ Expand
2

Quantitative Lutetium-177 SPECT/CT Imaging

Check if the same lab product or an alternative is used in the 5 most similar protocols
For each patient, whole-body (WB) SPECT/CT scans were acquired (three bed positions from the top of the head to the upper thighs; 90 projections and 25 s per projection) on a Symbia T2 scanner (Siemens Healthineers, Erlangen, Germany) at 1.5 ± 0.5 h, 6 ± 1 h, 24 ± 3 h, 48 ± 3 h, and at 7 days p.i. with a lutetium-177 reference-standard of approximately 10 MBq within the field of view. The scanner was equipped with a medium-energy low-penetration collimator. Three energy windows were acquired and used for further processing, a peak window of 20% width centered around the 208 keV energy peak and two adjacent corresponding lower and upper scatter energy windows of 10% width.
The SPECT images were stitched and quantitatively reconstructed using a commercial 3D ordered-subset expectation maximization (OSEM) algorithm (Flash 3D, Siemens Medical Solution, Germany) using 8 iterations and 9 subsets applying uniformity correction, CT-based attenuation correction, energy window-based scatter correction, and collimator-detector response modeling.
To yield quantitative images in units of Bq/mL, a calibration factor was determined from a phantom experiment using an IEC NEMA body phantom filled with 765 MBq lutetium-177 and applied to each patient SPECT dataset.
+ Open protocol
+ Expand
3

Scintigraphic Data Reconstruction Algorithms

Check if the same lab product or an alternative is used in the 5 most similar protocols
Scintigraphic data were reconstructed using 3 different algorithms:

The reference ordered subset expectation maximization (OSEM) 3D iterative algorithm (FLASH3D, Siemens) (OSEM-3D), 8 iterations, and 15 subsets with a 128 × 128 matrix (pixel size 4.8 × 4.8 × 4.8 mm), and a 12 mm full width at half maximum (FWHM) Gaussian postfilter.

The ordered subset conjugate gradient minimization (OSCGM) xSPECT algorithm (Siemens), allowing to perform SUV quantification thanks to the xSPECT Quant tool, 8 iterations and 6 subsets with a 256 × 256 matrix (pixel size 1.9 × 1.9 × 1.9 mm), and a 10 mm FWHM Gaussian postfilter.

The OSCGM-enhanced (OSCGM-e) xSPECTbone algorithm (Siemens), which uses CT data to constrain uptakes to bone structures, also allowing SUV quantification thanks to the xSPECT Quant tool, 8 iterations and 6 subsets with a 256 × 256 matrix (pixel size 1.9 × 1.9 × 1.9 mm), and a 10 mm FWHM Gaussian postfilter. Attenuation and scatter corrections were applied in the 3 reconstructions. Figure 1 shows an example of images obtained with the 3 reconstructions.

+ Open protocol
+ Expand
4

SPECT Imaging of 166Ho Activity

Check if the same lab product or an alternative is used in the 5 most similar protocols
SPECT images were reconstructed using a 3D OSEM algorithm (Flash 3D; Siemens) with 10 iterations, 8 subsets, incorporating attenuation correction. To correct for scatter during the reconstruction of the 166Ho activity distribution, downscatter in the 81 keV photopeak window due to higher energy emissions of both 166Ho and 99mTc was estimated from the 118 keV energy window by applying a single combined k-factor of 1.15 (see Additional file 1: Supplemental material for details). Photopeak scatter, i.e., scattered photons originating from the 81 keV primary photopeak, was not accounted for.
+ Open protocol
+ Expand
5

Lung Perfusion SPECT/CT Imaging for Radiotherapy

Check if the same lab product or an alternative is used in the 5 most similar protocols
Lung perfusion SPECT/CT was performed with patients positioned supine in the same position as RT simulation using a breast board allowing for direct registration of the SPECT/CT scans and RT treatment planning scans. The lung SPECT/CT scans were performed at baseline prior to RT and at one year following completion of RT. SPECT/CT was performed after intravenous injection of 5 mCi of [99mTc]macroaggregated albumin (MAA). SPECT/CT imaging consisted of acquisition of 64 projection views covering 360° with a dual-head, hybrid tomograph (Siemens Symbia T6, Siemens Medical, Hoffman Estates, IL, USA). CT imaging of the field-of-view employed 70 mAs, 130 kVp, pitch 0.8, 2.5 mm axial reconstruction. The radiotracer projection views were reconstructed tomographically employing the CT data for attenuation correction using an iterative algorithm (OSEM, with 8 iterations and 4 subsets; FLASH 3D, Siemens Medical, Hoffman Estates, IL, USA). Use of attenuation correction in SPECT reconstructions renders the data linearly proportional to the voxel-wise concentration of radiotracer. Co-registered MAA perfusion images and X-ray CT images were provided for further analyses. All pre-RT and one year post-RT lung perfusion SPECT/CT scans were read by the same expert nuclear medicine physician. These two scans were compared for each patient and qualitative perfusion changes were recorded.
+ Open protocol
+ Expand
6

Phantom and Patient SPECT/CT Imaging Protocol

Check if the same lab product or an alternative is used in the 5 most similar protocols
Planar and SPECT/CT imaging for both the phantom and patients were performed on Siemens Symbia systems (Intevo or T series) equipped with low-energy high-resolution (LEHR) collimators. The standard clinic acquisition protocol was used. The acquisition window was set at 15% with an adjacent 15% low-energy scatter window for SPECT. For SPECT, a 128 × 128 matrix, 60 views/head, non-circular continuous orbit was used. The patient acquisition time was 70 s each for the two (chest and abdomen) anterior/posterior planar scans and 10 s for each SPECT projection (total scan time 10 min). The CT for patients was performed in low-dose mode (130 kVp; 80 mAs) during free breathing. The phantom acquisition times were chosen to mimic count levels typical in patient imaging: 500,000 counts on the anterior planar view and 6 million total counts for the SPECT projections. The planar and SPECT/CT phantom acquisitions at each LSF were repeated three times without re-positioning to assess precision (repeatability).
Phantom and patient SPECT data were reconstructed using eight iterations four subsets of OS-EM (Siemens Flash 3D) with and without energy window-based scatter correction (SC) and CT-based attenuation correction (AC). Collimator-detector response modeling and an 8.4 mm Gaussian post-filter were used in all reconstructions.
+ Open protocol
+ Expand
7

SPECT/CT Imaging Protocol Reconstruction

Check if the same lab product or an alternative is used in the 5 most similar protocols
One 38.7 cm length SPECT/CT centered on the desired region (as requested by the interpreting physician based on planar scintigraphy findings) was reconstructed by an independent person into transaxial, coronal, and sagittal slices using e.soft reconstruction software (Siemens Healthcare, Germany). The reconstruction parameters were the same as for whole-body SPECT/CT: iterative reconstruction was performed using ordered-subsets expectation maximization with 8 iterations and 8 subsets (Flash3D, Siemens Healthcare, Germany), and images were smoothed with a 3D spatial Gaussian filter (1 mm full width at half maximum).
+ Open protocol
+ Expand
8

Quantitative SPECT Imaging Protocol

Check if the same lab product or an alternative is used in the 5 most similar protocols
Directly, post-injection planar images were acquired in dynamic mode (128 × 128 matrix, 20 frames of 1 min) in anterior-posterior projection followed by static mode (256 × 256 matrix, during 4 min) in anterior-posterior and lateral projections (30 min and 2 h post-injection), on a Siemens Symbia T16 SPECT-CT scanner, using ‘low- and medium energy’ (LME) collimators to limit septal penetration (reducing shine through) [16 (link)]. In addition to the planar imaging 2 h post-injection, SPECT-CT scans were acquired on a 128 × 128 matrix (pixel spacing, 3.9 × 3.9 mm), with 128 angles, 20 s per projection, over a non-circular 360° orbit (CT: 110 kV, 40 mAs eff., 16 × 1.2 mm). SPECT images were reconstructed using clinical reconstruction software (Siemens Flash3D), with attenuation and scatter correction (6 iterations, 8 subsets, 5-mm Gaussian filter). Additionally, quantitative SPECT reconstructions were generated using the Utrecht Monte Carlo System (UMCS), a dedicated SPECT reconstructor [17 (link), 18 (link)] which includes Monte Carlo modelling of scatter and collimator-detector interactions. During lymphoscintigraphy, a source with known radioactivity was scanned in the same frame as the patient, acting as a verification of quantitative accuracy.
+ Open protocol
+ Expand
9

Y-90 Quantitative SPECT Imaging with Scatter Correction

Check if the same lab product or an alternative is used in the 5 most similar protocols
Test data, including the patient data, were reconstructed using an in-house developed OS-EM algorithm with the scatter estimate included as an additive term in the following statistical model:
Yi~Poissonj=1npaijxj+si,
where Yi denotes the number of counts measured in the ith detector pixel, aij denotes elements of the system matrix A that models effects of depth-dependent attenuation and collimator/detector blur for a photon leaving the jth voxel towards the ith detector pixel, si denotes the estimated scatter component for the ith detector pixel and x=(x1,,xJ) denotes the vector of unknown 90Y activity voxel values. The current implementation used the same CNN-generated scatter estimate in all iterations, without updating. OS-EM reconstruction according to above equation was performed 1) without SC (NO-SC) using clinic software (Siemens Flash 3D) 2) with our previously reported [3 (link)] MC scatter model (MC-SC) 3) with the scatter estimate from the trained DCNN (DCNN-SC) 4) with the true scatter estimate from SIMIND (TRUE-SC), which was available only in the case of the simulated test data set. All reconstructions included CT-based attenuation correction, 3D collimator-detector response compensation and no post-filtering. MC-SC used two updates of the SIMIND generated scatter estimate based on our prior findings on convergence [3 (link)].
+ Open protocol
+ Expand
10

Intratherapeutic Radioiodine Imaging Protocol

Check if the same lab product or an alternative is used in the 5 most similar protocols
We prefer using the term intratherapeutic imaging instead of posttherapeutic imaging to avoid confusion with follow-up imaging. The intratherapeutic WBS was performed 5-10 d after radioiodine therapy (mean 131 I activity, 4.4 GBq; median, 3.0 GBq; range, 1.0-10.0 GBq) using a doublehead g-camera (Symbia S; Siemens) equipped with a high-energy, parallelhole collimator. The table speed was 15 cm/min. The matrix was 256 • 1,024, resulting in 2.4 • 2.4 mm pixels.
All patients underwent SPECT/CT of the neck on a scanner (Symbia T2; Siemens) equipped with a high-energy, parallel-hole collimator. Low-dose CT for attenuation correction was performed without a contrast agent (tube voltage, 130 kVp; tube current-time product, 17 mAs; beam pitch, 1.5; slice width, 5 mm). The SPECT scan was acquired using 128 angles over 360°a nd 25 s per stop. Images were iteratively reconstructed and corrected for attenuation and scatter (Flash 3D [Siemens], 4 subsets and 8 iterations; gaussian intersliced smoothing filter; attenuation coefficient, 0.15 cm 21 ). The image matrix was 128 • 128, resulting in a cuboid voxel length of 4.8 mm.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!