The largest database of trusted experimental protocols

Au2700 automatic biochemical analyzer

Manufactured by Olympus
Sourced in Japan

The AU2700 is an automatic biochemical analyzer designed for clinical laboratory use. It is capable of performing a variety of biochemical tests on patient samples to aid in medical diagnosis and treatment. The core function of the AU2700 is to automatically analyze and measure the levels of different biochemical compounds in a sample, providing healthcare professionals with valuable data to support their clinical decision-making.

Automatically generated - may contain errors

22 protocols using au2700 automatic biochemical analyzer

1

Assessment of Blood Pressure, Uric Acid, and Kidney Function

Check if the same lab product or an alternative is used in the 5 most similar protocols
Blood pressure was measured in all participants by trained professionals using electronic sphygmomanometry. Blood pressure was measured in the right arm in the seated position with elbow and forearm resting on the armrest after the participant rested for 10 minutes31 (link),32 (link). Blood pressure was measured 2 times at 2-min intervals in all participants and mean systolic and diastolic blood pressure were recorded. Hypertension was defined as systolic blood pressure ≥ 140 mm Hg or diastolic blood pressure ≥ 90 mm Hg33 (link).
Venous blood samples were collected after an overnight fast (≥ 12 h). Serum uric acid concentration was measured by the uricaseperoxidase method34 (link). Hyperuricemia was defined as serum uric acid concentration ≥ 7 mg/dL in men or ≥ 6 mg/dL in women35 (link). The estimated glomerular filtration rate (eGFR) was measured using an Olympus AU2700 automatic biochemical analyzer. eGFR ≥ 60 mL/min/1.73 m2 was regarded as conserved kidney function and eGFR < 60 mL/min/1.73 m2 was regarded as reduced renal function36 (link). Total cholesterol, triglycerides, and fasting plasma glucose were measured using an Olympus AU2700 automatic biochemical analyzer.
+ Open protocol
+ Expand
2

Quantitative Analysis of Uric Acid in Biological Samples

Check if the same lab product or an alternative is used in the 5 most similar protocols
To demonstrate the application of the proposed method, the ratiometric fluorescent system comprising the N-CDs was utilized to quantitatively analyze UA in human serum and urine samples. Human blood samples provided by the Fourth Affiliated Hospital of Nanchang University (Nanchang, China) were centrifuged at 4000 rpm for 15 min, after which the supernatants (serum) were extracted and diluted 5-fold using PBS solution. To verify the results of the N-CDs-based fluorescence system, the human serum samples were also analyzed by an Olympus AU2700 automatic biochemical analyzer. Before biochemical analysis, the serum samples were centrifuged at 4000 rpm for 3 min. Each of the resulting supernatants (5.6 μL) was supplemented with aliquots of uricase, 4-aminoantipyrine (4-AAP), and 3-diphenylamine disodium salt (MADB). After that, the mixtures were analyzed by the biochemical analyzer at 660/800 nm.
The urine samples were collected from two volunteers in the morning and filtered through a 0.22-μm Millipore filter. The resulting filtrates were centrifuged at 8000 rpm for 5 min and diluted 50-fold with PBS solution for further use.
+ Open protocol
+ Expand
3

Plasma Lipid and Cytokine Profiling

Check if the same lab product or an alternative is used in the 5 most similar protocols
Blood sample was collected from the abdominal vena cava and the plasma supernatants were collected for further study. The levels of total cholesterol (TC), low-density lipoprotein cholesterol (LDL-C), triglycerides (TG), and high-density lipoprotein cholesterol (HDL-C) were detected using an AU-2700 automatic biochemical analyzer (Olympus, Tokyo, Japan). The serum cytokine levels, including tumor necrosis factor-α (TNF-α) and interleukin-1β (IL-1β), were detected using a Mcytomag-70 K-3 Mouse Cytokine/Chemokine Magnetic Bead Panel (Merck Millipore Co. LTD).
+ Open protocol
+ Expand
4

Measuring Serum Albumin and Prealbumin

Check if the same lab product or an alternative is used in the 5 most similar protocols
The levels of serum albumin and serum prealbumin were measured from fasting venous blood within 24 h after admission. Analyses were performed by Olympus AU2700 automatic biochemical Analyzer.
+ Open protocol
+ Expand
5

Liver Function and Histology Assessment

Check if the same lab product or an alternative is used in the 5 most similar protocols
Activities of serum aspartate transaminase (AST) and alanine transaminase (ALT) were measured using an AU2700 automatic biochemical analyzer (Olympus, Tokyo, Japan). For histopathological analysis, liver tissue was fixed in 10% phosphate buffered formalin and paraffin embedded, and cut into 4 μm sections. Sections were stained with hematoxylin and eosin (H&E) and analyzed by light microscopy.
+ Open protocol
+ Expand
6

Serum Biomarker Assessment of Renal Function

Check if the same lab product or an alternative is used in the 5 most similar protocols
Serum was centrifuged at 3600 g for 10 minutes and the supernatant was then used to assess the renal function. The levels of BUN and Scr were determined on an AU2700 automatic biochemical analyzer (Olympus, Japan). The levels of TNF-α, IL-1β and AGEs were measured according to the manufacturer’s instructions of the immunoassay kit. The assay employs the ELISA techniques.
+ Open protocol
+ Expand
7

Streptozotocin-Induced Diabetic Kidney Hypertrophy

Check if the same lab product or an alternative is used in the 5 most similar protocols
Body weights and fasting blood glucose levels were recorded each week before and after STZ induction, including drug intervention period. FBG was measured by a blood glucose meter (LifeScan, CA, USA) in samples obtained from the tail vein. It was about 2 mins from blood collection to obtaining results. The renal hypertrophy index (mg/g) was calculated as follows: renal index = kidney weight (mg)/body weight (g). 24-h urine protein was determined on an AU2700 automatic biochemical analyzer (Olympus, Japan).
+ Open protocol
+ Expand
8

Serum Biomarker Measurement Protocol

Check if the same lab product or an alternative is used in the 5 most similar protocols
Serum AST and ALT activities were measured using an AU2700 automatic biochemical analyzer (Olympus, Tokyo, Japan). Serum TNF-α, IL-1β and IL-6 levels were determined using enzyme-linked immunosorbent assay (ELISA) kits (Boster Biological Technology Ltd, Wuhan, China) according to the manufacturer's protocol.
+ Open protocol
+ Expand
9

Evaluating Diabetic Nephropathy Biomarkers

Check if the same lab product or an alternative is used in the 5 most similar protocols
The levels of FBG were measured before and after each STZ injection. Samples obtained from the tail vein were tested for FBG using a blood glucose meter (LifeScan, CA, United States). Metabolic cages were used to collect 24-h urine from rats. The levels of 24 h U-Alb, blood urea nitrogen (BUN), and serum creatinine (Scr) were measured using an AU2700 automatic biochemical analyzer (Olympus). AGE level was measured according to the ELISA kit’s instructions (Cusabio, China).
+ Open protocol
+ Expand
10

Lipid Biomarker Measurement Protocol

Check if the same lab product or an alternative is used in the 5 most similar protocols
Fasting venous blood samples were collected to measure glucose, TC. HDL-C, triglyceride and creatinine levels using the Olympus AU2700 Automatic Biochemical Analyzer (Japan). LDL-C was calculated using the Friedewald equation. Hs-CRP was determined by the high-sensitivity nephelometric method. Plasma ox-LDL was measured by sandwich enzyme-linked immunosorbent assay procedure (Mercodia, Uppsala, Sweden)16 (link). The interassay coefficient of variation was 6.8%.
After centrifugation at 1500 × g for 10 min at 4 °C, plasma samples were stored at −80 °C for future measurements of Lp-PLA2 mass, as long-term stability of Lp-PLA2 has been demonstrated35 (link). Briefly, Lp-PLA2 mass was measured using a commercial enzyme-linked immunosorbent assay kit (Tianjin Kangerke Bioscience, Tianjin, China). The interassay coefficient of variation was 4.8%. All measurements were performed according to the manufacturers’ instructions.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!