Inova 300
The Inova 300 is a nuclear magnetic resonance (NMR) spectrometer manufactured by Agilent Technologies. It is designed to analyze the structure and composition of chemical compounds through the detection and measurement of nuclear magnetic resonances.
Lab products found in correlation
14 protocols using inova 300
Characterization of DNA-Conjugated Agents
Synthesis and Characterization of OGR1 Probe
Polycation Characterization by SEC-MALS
weight (Mn) and dispersity (Mw/Mn) for each polycation
was determined by size exclusion chromatography (SEC) equipped with
Eprogen columns [CATSEC1000 (7 μm, 50 × 4.6), CATSEC100
(5 μm, 250 × 4.6), CATSEC300 (5 μm, 250 × 4.6),
and CATSEC1000 (7 μm, 250 × 4.6)], a Wyatt HELEOS II light
scattering detector (λ = 662 nm), and an Optilab rEX refractometer
(λ = 658 nm). The columns were maintained at 30 °C. An
aqueous eluent (0.1 M Na2SO4/1 v/v % acetic
acid) was utilized at a flow rate of 0.4 mL/min. The dn/dc values for each of the polymers were determined
offline with the Optilab rEX refractometer.
1H NMR
measurements were performed with a Varian Inova 300 at 70 °C.
Samples were dissolved in D2O (HOD used as the internal
standard), and the block copolymer compositions were determined by
calculating the ratio between the integrals of resonances of the PMAG
block and those of the PAEMA or PDMAPMA block.
Synthesis and Characterization of SpiroZin1
Synthesis of Triblock Copolymer PLGA-PEG-PLGA
Non-invasive Intracellular pH Monitoring
to a cell density of 108 cells/mL in a total volume of
500 mL in a 4 L Erlenmeyer
flask with constant shaking at 37 °C in Difco Middlebrook 7H9
media supplemented with oleic acid/albumin/dextrose and 0.05% Tween
80. Cells were harvested by centrifugation, and the pellet was washed
twice with 5 mM phosphate buffer, pH 6.8. The cell pellet was then
resuspended in 200 μL of the same buffer and 500 μL of
the resulting cell slurry transferred to a 5 mm NMR tube. Chemical
shifts were referenced with respect to 85% phosphoric acid in D2O in a coaxial capillary. 31P NMR spectra were
obtained using a Varian INOVA 300 (at 121.5 MHz) using 60° pulse
excitation, proton decoupling, and a 1 s recycle time. A total of
1024 scans were accumulated corresponding to approximately a 60 min
total data acquisition time (without aeration). Spectra were analyzed
as described elsewhere.46 (link) The peak corresponding
to the α-phosphate of ATP (at ∼−10.5 ppm) and
the inorganic phosphate peaks of interest (in the region of 0–1.5
ppm) were used to calculate the internal and external pH using the
following equation, where d is the distance between
the α-phosphate of ATP and the inorganic phosphate peak, in
ppm.
NMR Spectroscopy and HPLC Analysis
Example 4
General Methods:
1H NMR spectra were recorded on a Varian Inova 300 or 500 MHz NMR instrument. Chromatographic purities were determined on an Agilent 1200 Series, 1100 Series or 6130 Series LC/MS system using a Merck Chromolith RP-18e analytical HPLC column (monolithic, 50×2 mm) and the following analytical HPLC method: injection volume 5 μL; flow rate 1 mL/min; 5→95% acetonitrile in water with 0.05% AcOH over 5 mins; Agilent diode array detector at λ=254, 220 or 195 nm; room temperature.
NMR and HPLC Characterization Methods
Example 1
General Methods:
1H NMR spectra were recorded on a Varian Inova 300 or 500 MHz NMR instrument. Chromatographic purities were determined on an Agilent 1200 Series or 1100 Series LC/MS system using a Merck Chromolith RP-18e analytical HPLC column (monolithic, 50×2 mm) and the following analytical HPLC method: injection volume 5 μL; flow rate 1 mL/min; 5→95% acetonitrile in water with 0.05% AcOH over 5 mins; Agilent diode array detector at λ=254, 220 or 195 nm; room temperature.
Spectroscopic and Colloidal Characterization of Nanostructures
Transmission electron microscopy (TEM) images were observed on a JEM-2100 (HR) electron microscope, using an accelerating voltage of 200 kV. TEM samples were prepared by dropping solutions onto copper grids coated with Formvar films, until the solvents were evaporated in a dust protected atmosphere. The dynamic laser light scattering (DLS) was performed on a light scattering goniometer (ALV/CGS-8F, ALV, Hessen, Germany) with a wavelength 632.8 nm from a He–Ne laser. The scattering angle (θ) was set at 90°. All of the sample solutions were filtered into the light-scattering bottles through the 0.45 μm filter (NYL, Whatman, UK).
In-vivo M. smegmatis Phosphate Analysis
About PubCompare
Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.
We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.
However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.
Ready to get started?
Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required
Revolutionizing how scientists
search and build protocols!