The largest database of trusted experimental protocols

Agilent 1200 series diode array detector

Manufactured by Agilent Technologies
Sourced in Germany, United States

The Agilent 1200 series diode array detector is a high-performance liquid chromatography (HPLC) detector that utilizes a diode array to measure the absorption of light by analytes in a sample. It provides simultaneous detection across a wide range of wavelengths, allowing for the identification and quantification of multiple compounds in a single analysis.

Automatically generated - may contain errors

7 protocols using agilent 1200 series diode array detector

1

LC-MS Analysis of Phytochemical Compounds

Check if the same lab product or an alternative is used in the 5 most similar protocols
LC-MS was used to identify the major compounds in the FAA extract. Chromatographic analysis of FAA was carried out by reverse phase elution (Waters Symmetry LC-18 column 250 × 4.6 mm, 5 μm) on Agilent 6500 Series Accurate-Mass Quadrupole Time-of-Flight (Q-TOF; Agilent Santa Clara, CA, USA) LC/MS system with Agilent 1200 Series Diode Array Detector (module G1315B; detection type: 1024-element photodiode array; light source: deuterium and tungsten lamps; wavelength range 190–950 nm). The mobile phase consisted of (A) formic acid (0.1%, v/v); (B) acetonitrile + 0.1% formic acid; gradient (in solvent B): (i) 20%, from 0 to 20 min, (ii) 95%, from 20 to 27 min, and (iii) 35%, at 27–30 min of total run time; flow rate: 0.2 ml/min; injection volume 3 L; ESI parameters: both negative and positive ion mode; mass range 100–1200 m/z; spray voltage 4 kV; gas temperature 325°C; gas flow 10 L/min; Nebulizer 40 psi and the mass was analysed by using Agilent technologies Mass-Hunter software.
+ Open protocol
+ Expand
2

Phytochemical Profiling of P. juliflora Leaves

Check if the same lab product or an alternative is used in the 5 most similar protocols
Liquid chromatography-mass spectrometry (LC-MS) was used to identify the major chemical compounds in the crude ethyl acetate extract of P. juliflora leaves. Chromatographic analysis was carried out by reverse phase elution (Waters Symmetry LC18 column 250 × 4.6 mm, 5 μm) on Agilent 6500 Series Accurate-Mass Quadrupole Time-of-Flight (Q-TOF; Agilent Santa Clara, CA, USA) LC/MS system with Agilent 1200 Series Diode Array Detector (module G1315B; detection type: 1024-element photodiode array; light source: deuterium and tungsten lamps; wavelength range 190-950 nm). The mobile phase consisted of (A) formic acid (0.1 %, v/v); (B) acetonitrile + 0.1 % formic acid; gradient (in solvent B): (i) 20 %, from 0 to 20 min, (ii) 95 %, from 20 to 27 min, and (iii) 35 %, at 27-30 min of total run time; flow rate: 0.2 ml/min; injection volume 3 L; ESI parameters: both negative and positive ion mode; mass range 100-1200 m/z; spray voltage 4 kV; gas temperature 325 °C; gas flow 10 L/min; Nebulizer 40 psi and the mass was analyzed by using Agilent technologies Mass-Hunter software.
+ Open protocol
+ Expand
3

HPLC Analysis of Synthesized Compounds

Check if the same lab product or an alternative is used in the 5 most similar protocols
HPLC analyses were conducted with an Agilent 1200 HPLC system equipped with a quaternary pump and an Agilent 1200 series diode array detector. The HPLC analyses were used to determine the purities of the synthesised compounds. HPLC-grade acetonitrile (Merck) and Milli-Q water (Millipore) were used for the chromatography. A Venusil XBP C18 column (4.60 × 150 mm, 5 μm) was used for the separation, and the mobile phase consisted at the start of each run of 30% acetonitrile and 70% Milli-Q water. The flow rate was set to 1 ml/min. At the start of each run, a solvent gradient program was initiated. The composition of the acetonitrile in the mobile phase was increased linearly to 80% over a period of 5 min. Each HPLC run lasted 15 min, and a time period of 5 min was allowed for equilibration between runs. A volume of 20 µL of solutions of the test compounds (1 mM dissolved in acetonitrile) was injected into the HPLC system, and the eluent was monitored at a wavelength of 254 nm.
+ Open protocol
+ Expand
4

Quantification of Methyl Gallate by HPLC

Check if the same lab product or an alternative is used in the 5 most similar protocols
The EtOAc fraction was dissolved in 10 mL of 50% MeOH with multi vortexing, and filtered through a Dismic-13 JP membrane filter (Advantec Toyo, Tokyo, Japan; pore diameter: 0.2 µm). We injected 10 µL of the sample into a reverse-phase HPLC using a YMC pack ODS-AM (4.6 × 250 mm, 5 µm pore size), with a column temperature of 35 °C. Mobile phase: A = 0.5% formic acid (aq.), B = acetonitrile. The gradient conditions were as follows: 0 min, 0% B; 4 min, 12% B; 20 min, 18% B; 24 min, 22% B; 34 min, 26% B; 44 min, 30% B; 54 min, 100% B. The flow rate was 0.8 mL/min. The UV absorbance at 254 nm was monitored using an Agilent 1200 series diode array detector (Agilent Technologies, Waldbronn, Germany). The peak was assigned by carrying out co-injection tests with an authentic sample and comparing with the UV spectral data. The measurement was repeated three times for each sample. Representative HPLC results are illustrated in Figure 2. Quantification of methyl gallate was conducted by peak area measurement. The calibration curve of the standard compound (methyl gallate) was made over a range of 1–8 µg/mL. The detector response was linear over the concentration range used. The regression coefficient (r2) of standard compound was higher than 0.996 (Table 4). The amount of methyl gallate is shown in Table 5.
+ Open protocol
+ Expand
5

HPLC Analysis of Compounds Using Agilent 1200 Series

Check if the same lab product or an alternative is used in the 5 most similar protocols
The HPLC system used in the present study was Agilent 1200 Series equipped with Quaternary Pump, Agilent 1200 Series Vacuum Degasser, Agilent 1200 Series Autosampler and Agilent 1200 Series Diode Array Detector Purchased from Agilent Technologies (Hewlett-Packard Strasse Waldbronn, Germany). Chromatographic separation of the compounds was performed on a C18 analytical column (Techsphere 5ODS, 25 cm × 4.6 mm ID; HPLC Technology, Macclesfield, Cheshire, UK). Data acquisition and processing were accomplished with LC Chemstation software (Agilent Technologies). Adwa pH meter (AD1020 pH/mV/ISE/Temperature, Hungary) was used for the determination of the sample pH and A 800 model centrifuge, China, was used to speed up the phase separation. An electronic balance (Adam Equipment Company, UK) was utilized for weighing the different chemicals involved in the experiments. For the measurement of total dissolved solids (TDS) and electrical conductivity, conductivity meter (Postfach 24 80, Germany) was used.
+ Open protocol
+ Expand
6

HPLC Analysis of Synthesized Compounds

Check if the same lab product or an alternative is used in the 5 most similar protocols
Purity of the synthesised compounds was determined by HPLC using an Agilent 1200 HPLC system equipped with a quaternary pump and an Agilent 1200 series diode array detector. HPLC grade acetonitrile (Merck) and Milli-Q water (Millipore) were used for the chromatography. A Venusil XBP C18 column (4.60 × 150 mm, 5 µm) was used with acetonitrile (30%) and Milli-Q water (70%) as the initial mobile phase at a flow rate of 1 mL/min. At the start of each injection, a solvent gradient programme was initiated by linearly increasing the composition of the acetonitrile in the mobile phase to 85% acetonitrile over a period of 5 min. Each run lasted 15 min and a time period of 5 min was allowed for equilibration between runs. A volume of 20 µL of solutions of the test compounds in acetonitrile (1 mM) was injected into the HPLC system, and the eluent was monitored at wavelengths of 210, 254 and 300 nm.
+ Open protocol
+ Expand
7

HPLC Analysis of Safflower Seed Extracts

Check if the same lab product or an alternative is used in the 5 most similar protocols
An aliquot of 10 mg of the 70% ethanol extract of safflower seed was dissolved in 10 mL of 50% methanol by repeated vortexing, and passed through a Dismic-25 JP membrane filter (Advantec Toyo, Tokyo, Japan; pore diameter: 5 µm). Next, 5 µL of the sample was injected into a reverse-phase HPLC system with an INNO C18 column (4.6 × 250 mm, 5 µm pore size, YoungJin Biochrom, Korea) at 35°C. The mobile phase comprised (A) a 0.1% aqueous solution of formic acid and (B) acetonitrile, and the following gradient conditions were used: 0 min, 0% B; 5 min, 6% B; 10 min, 22% B; 50 min, 28% B; 66 min, 50% B; and 70 min, 100% B. The flow rate was 0.8 mL/min. The UV absorbance at 254 nm was monitored by using an Agilent 1200 series diode array detector (Agilent Technologies, Waldbronn, Germany). All peaks were assigned through co-injection tests with authentic samples and comparison of the UV spectral data; through this process, compounds including serotonin and its derivatives were detected from safflower seed extract. All measurements were repeated in triplicate. Representative HPLC results are illustrated in Figure 1. The quantification of serotonin, N-(p-coumaroyl)serotonin, N-feruloyl serotonin, luteolin, and luteolin-7-O-glucoside was achieved by peak area measurement. The quantity of each identified compound is shown in Table 1.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!