The largest database of trusted experimental protocols

34 protocols using dsc q10

1

Thermal Analysis of Material Samples

Check if the same lab product or an alternative is used in the 5 most similar protocols
The DSC thermograms were obtained on a scanning calorimeter (DSC Q10, TA Instruments). The instrument was calibrated at various temperatures before starting the procedure. Each sample weighed 5 mg and was heated in sealed aluminum pans under a nitrogen environment at 50 mL/min. The limit of detection was controlled from 10 °C to 300 °C and the ramp was 10 C/min.
+ Open protocol
+ Expand
2

Thermal Analysis of Compound Glass Transitions

Check if the same lab product or an alternative is used in the 5 most similar protocols
Differential scanning calorimetry (DSC) was performed on a TA Instruments DSC Q10 instrument with a scanning range from room temperature to 300 °C. The sample was first scanned at a heating rate of 10 °C min−1 and was cooled down to room temperature rapidly using liquid N2. The second and third scans were performed at a heating rate of 5 °C min−1. The glass transition temperatures were determined from either the second or the third scan for each compound.
+ Open protocol
+ Expand
3

Characterization of GO/PVDF Nanocomposites

Check if the same lab product or an alternative is used in the 5 most similar protocols
Field-emission scanning electron microscopy (S-4800, Hitachi) was used to observe the morphology of the GO/PVDF composite films. Differential scanning calorimetry (DSC) measurements (DSC Q10, TA instrument) of the GO/PVDF nanocomposites were conducted under nitrogen gas in the 193 to 473 K temperature range with cooling/heating speeds of 20 K/min. The crystalline structure of GO/PVDF nanocomposites was determined using XRD (Rigaku Ultima IV) with CuKα radiation. FTIR (NICOLET 380, Thermo Scientific) measurements were conducted to examine the changes in the bonding structure. The electrical properties of the rGO/PVDF nanocomposites under controlled temperatures were measured by a source meter (S-2400, Keithley). For the TCR measurements, the temperature of the hot stage was controlled by the Advanced Peltier System attached to ARES-G2 (TA Instruments).
+ Open protocol
+ Expand
4

Thermal Behavior Analysis of Samples

Check if the same lab product or an alternative is used in the 5 most similar protocols
The thermal behavior of the samples was measured using differential scanning calorimetry (DSC). DSC thermograms were recorded using a thermal analysis device (DSC Q10, TA Instruments, New Castle, DE, USA) from room temperature to 300 °C with a heating rate of 10 °C/min under a nitrogen gas atmosphere.
+ Open protocol
+ Expand
5

Glass Transition Temperature Measurement

Check if the same lab product or an alternative is used in the 5 most similar protocols
Tg was confirmed using a DSC (DSC-Q10, TA Instruments, New Castle, DE, USA). DSC measurements were performed at temperatures ranging between −100 and 100 °C at a heating rate of 10 °C/min. After the first heating-cooling cycle, Tg was obtained during the second heating.
+ Open protocol
+ Expand
6

Thermal Analysis of E-LqIR Polymers

Check if the same lab product or an alternative is used in the 5 most similar protocols
The glass-transition temperatures (Tg) of the E-LqIRs were measured by differential scanning calorimetry (DSC; DSC-Q10, TA Instruments, New Castle, DE, USA). The E-LqIR samples (3–6 mg) were analyzed from −80 to 100 °C at a heating rate of 10 °C/min.
+ Open protocol
+ Expand
7

Thermal Analysis of PECoVA Composites

Check if the same lab product or an alternative is used in the 5 most similar protocols
Differential scanning calorimetry (DSC) model DSCQ-10 by TA instruments (New Castle, Delaware, United States) was used to measure thermal behaviors of the virgin PECoVA, PECoVA/DOL and PECoVA/OMCD composites. The weight of the sample was about 5 mg. Next, the heating rate was set at 10 °C/min and the atmosphere was purged with nitrogen gas, with the gas flow rate of 50 mL/min, in order to get the melting temperature (Tm) and the melting enthalpy (ΔH) of the samples. The samples were then heated from room temperature (30 °C) to 150 °C. The degree of crystallinity of the samples, XC, was calculated using Equation (2) below: XC %=ΔHfΔH0×100%
where ΔHf represents the melting enthalpy of the sample, which is obtained by calculating the second heating endothermic peak, and ΔH0 is the theoretical melting enthalpy of the 100% crystalline polymer. In this case, the ΔH0 of PECoVA is approximately 23.47 J/g.
+ Open protocol
+ Expand
8

Thermal and Phase Transition Analysis of PCM Microcapsules

Check if the same lab product or an alternative is used in the 5 most similar protocols
The thermal and phase transition characteristics of the unmodified and shell-modified PCM microcapsules, as well as of the textile fabric containing these microcapsules were determined by standard method EN 16806-1 [43 ], using DSC module (DSC Q10, TA Instruments, New Castle, DE, USA) equipment under a nitrogenous atmosphere. The mass of test specimens was about 10 mg. A lid was pressed on the crucible to ensure good contact between the specimen and the bottom of the crucible. The micro-encapsulated PCM samples were dried at 60 °C for 24 h to remove all water. The samples underwent a heating-cooling-heating cycle from −20 °C to +60 °C at a heating and cooling rate of 5 °C/min. Based on the recorded second heating cycle, enthalpy of fusion in J/g was determined by measuring the area under the peak to the baseline constructed, using software of DSC Q10. In the same way, the enthalpy of crystallization in J/g was determined based on the recorded cooling cycle.
The transition temperatures—peak melting and peak crystallization temperatures—extrapolated onset and end melting as well as crystallization temperatures, were also determined based on respectively the recorded second heating or based on the recorded cooling cycle, as defined in EN ISO 11357-3 [44 ].
+ Open protocol
+ Expand
9

Thermal Characterization of CAP-NCs

Check if the same lab product or an alternative is used in the 5 most similar protocols
DSC was used to characterize the thermal properties of the components in the optimized formulations, i.e., unloaded CAP-NCs and CHX-CAP-NCs. The samples were weighed (3–5 mg) directly in hermetic aluminium pans and scanned in a temperature range of 0–400 °C at a heating rate of 10 °C/min under a nitrogen flux of 50 mL/min utilizing a previously calibrated and adjusted calorimeter (DSC Q10, TA Instruments, New Castle, DE, USA).
+ Open protocol
+ Expand
10

Determination of Glass Transition Temperature

Check if the same lab product or an alternative is used in the 5 most similar protocols
The determination of glass-transition temperature (Tg) was carried out on a differential scanning calorimeter (DSC-Q10, TA Instruments, Delaware, USA). The curves for samples (3–6 mg) were obtained by heating sample from −120 to −20 °C at a heating rate of 10 °C/min under nitrogen atmosphere.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!