The largest database of trusted experimental protocols

27 protocols using 717 plus

1

Quantitative HPLC Analysis of Metabolic Transformations

Check if the same lab product or an alternative is used in the 5 most similar protocols
All metabolic transformations were analyzed using RP-HPLC with UV-vis detection with 5 mm Suplex pKb-100 analytical column (0.46 cm × 25 cm, C18) (Supelco, Bellefonte, PA, USA) with Waters HPLC system equipped with 1525 binary pump, 7725i Rheodyne injector, 2487 dual λ absorbance detector and 717 plus autosampler controlled with Breeze software (Waters Co., Milford, MA, USA). Aliquots of 200 µL of the prepared samples were analyzed by HPLC-UV-vis at a flow rate of 1 mL/min in 50 mM ammonium formate buffer, pH = 3.2, with a linear gradient from 15% to 80% methanol for 25 min, followed by a linear gradient from 80% to 100% for 3 min. The column was then re-equilibrated at initial conditions for 20 min between runs. The elution of sample components was monitored at 254 nm (testosterone transformation), 330 nm (7-OH-TFC glucuronidation) or 420 nm (C-1305 and C-1311 metabolism).
+ Open protocol
+ Expand
2

HPLC Analysis of α-Tocopherol in Oils

Check if the same lab product or an alternative is used in the 5 most similar protocols
Samples of 0.1 g oil were dissolved in 1.9 mL acetone, shaken, and filtered through 0.22 –4 μ m syringe cellulose filters. An aliquot of 20 μ L of this solution was injected onto the HPLC system (Waters, Milford, MA, USA), which was equipped with a pump unit (600 Controller; Waters) and an auto-sampler (717 plus; Waters). The chromatography column (Spherisorb ODS2; 250 × 4.6 × 5 μ m) was kept at 25 °C, with a pre-column used (Phenomenex cartridge, Torrance, CA, USA, C18 AJO-4287). The mobile phase was acetonitrile and methanol (1:1; v/v) at a flow rate of 1 mL/min. α -Tocopherol was detected using a photodiode array detector (996; Waters) at a wavelength of 295 nm, using a run time of 18 min. An Enpower 2 Work Station was used for the data processing. The α -tocopherol concentrations were initially in mg/L based on the calibration curve, with α -tocopherol (Sigma-Aldrich) as the external standard. From the oil weight in the 2 mL sample (see above), the α -tocopherol was finally expressed as mg per kg oil (mg/kg).
+ Open protocol
+ Expand
3

HPLC Analysis of Phloroglucinol Adducts

Check if the same lab product or an alternative is used in the 5 most similar protocols
Phloroglucinolysis products were separated on a Cadenza CD-C18 (75 mm × 4.6 mm, 3 μm) column (Imtakt, Kyoto, Japan). The liquid chromatograph was a Waters (Milford, MA, USA) system equipped with diode array, scanning fluorescence detectors (Waters 474), and an autosampler (Waters 717 plus). Solvent A (25 mL of acetic acid and 975 mL of water) and solvent B (acetonitrile) were used in the following gradients: initial, 5 mL/100 mL B; 0–15 min, to 10 mL/100 mL B linear; 15–25 min to 60 mL/100 mL B linear; followed by washing and reconditioning of the column. A flow rate of 1 mL/min and an oven temperature of 15 °C were used with the injection of the filtrate (20 μL) into the HPLC system. The fluorescence detection was recorded at an excitation wavelength of 278 nm and an emission wavelength of 360 nm. The calibration curves were established using (+)-catechin and (−)-epicatechin (phloroglucinol adduct standards). All data were obtained in triplicate. The results were expressed as mg/L.
+ Open protocol
+ Expand
4

Peptide Purification by Semi-Preparative RP-HPLC

Check if the same lab product or an alternative is used in the 5 most similar protocols
Peptide purification was performed by semi-preparative RP-HPLC. Separation of peptides was performed on an HPLC system (Waters Corp., Mildford, MA, USA) equipped with two pumps (module Delta 600), a pump controller (module 600), an autosampler (module 717 plus), and a diode array detector (module 996). The data-processing software was Empower 2 (Waters Corp.). A 250 × 21.5 mm Hi-Pore 318 reverse phase column (BioRad, Hercules, CA, USA) was used. Peptide fractions were dissolved in solvent A (water: trifluoroacetic acid (TFA), 1000:1, v/v) at a concentration of 30 mg/mL. Fractions (400 μL) were injected and eluted at a flow rate of 10 mL/min, with a linear gradient of solvent B (acetonitrile:TFA, 1000:0.8, v/v) in A, going from 5% to 45% B in 50 min. Each chromatographic run was repeated 15–20 times and the subfractions were collected automatically with a Fraction Collector (Model II, Waters Corp.). The collection times were from min 6 to 12 for F1, min 12 to 16 for F2, min 16 to 22 min for F3, and min 22 to 35 for F4. The collected fractions were pooled, freeze-dried, and stored at −20 °C until further analysis. Quantification of peptides in each subfraction was performed by the Quantitative Colorimetric Peptide Assay, according to the manufacturer’s protocol.
+ Open protocol
+ Expand
5

Measuring Chlorophyll-a in Seawater

Check if the same lab product or an alternative is used in the 5 most similar protocols
For measuring the photosynthetic pigment Chl-a one liter of seawater samples were taken from Niskin bottles, immediately filtered on GF/F filters, frozen in liquid nitrogen, and stored at -80°C until further analyses by high-performance liquid chromatography (HPLC) at the home laboratory of the Alfred Wegener Institute Helmholtz Centre for Polar and Marine Research after the cruise. The samples were measured using a Waters 600 controller equipped with an auto sampler (717 plus), a photodiode array detector (2996) and the EMPOWER software. Chl-a was analyzed by reverse-phase HPLC using a VARIAN Microsorb-MV3 C8 column (4.6 3 100 mm) and HPLC-grade solvents (Merck). The solvents gradient and routine of analysis are fully described in Taylor et al. [54 ]. Chl-a concentrations were quantified based on peak area of the external standard, which was spectrophotometrically calibrated using extinction coefficients published in Jeffrey et al. [55 ].
+ Open protocol
+ Expand
6

High-Performance Size-Exclusion Chromatography Analysis

Check if the same lab product or an alternative is used in the 5 most similar protocols
Conventional HP-SEC analyses were performed with a Waters 515 pump (Water Inc., Milford, MA, USA), equipped with an auto-injector (Waters717plus), ultraviolet (UV) detector (Waters 2487) and differential refractometer (RI) detector (Waters 2410). Both polymeric gel based columns and silica based columns were used. In the first case, TSK G3000 PWXL (7.8 × 300 mm) and G2500 PWXL (7.8 × 300 mm) columns (Tosoh Corp.), connected in series, were eluted with 0.1 M NaNO3 at a flow rate of 0.6 mL/min, according to HP-SEC/TDA procedure. In the second case, TSK SW guardcolumn, TSK G3000 SWXL (7.5 × 300 mm) and TSK G2000 SWXL (7.5 × 300 mm) (Supelco), connected in series, were eluted with 0.2 M Na2SO4, at a flow of 0.5 mL/min, according to the EP indications. In both cases, processes were aided by Chromatography Manager Software Millennium–32 (Waters), with its GPC Empower option.
+ Open protocol
+ Expand
7

HPLC Analysis of Drug-Loaded Nanoparticles

Check if the same lab product or an alternative is used in the 5 most similar protocols
Drug content (assay), entrapment efficiency (EE) and in vitro permeability samples were analyzed using high performance liquid chromatography (HPLC) method. The HPLC system comprising of Waters 717 plus auto sampler, Waters 2487 dual absorbance detector, 600 Waters controller pump, and an Agilent 3395 Integrator. Luna® C18 (4.6 mm × 250 mm) column, with a mobile phase of 1:1 isocratic solution of water and acetonitrile, at a flow rate of 1 mL/min and detection wavelength (λmax) of 254 nm, and AUFS of 1, was used for analysis of TA [26 ].
+ Open protocol
+ Expand
8

HPLC Analysis of Phyllanthin, Hypophyllanthin, and Corilagin in P. amarus

Check if the same lab product or an alternative is used in the 5 most similar protocols
The HPLC system used was from Breeze with a dual λ absorbance detector (Waters 2487, Massachusetts, USA) isocratic HPLC Pump (Waters 1515 Massachusetts, USA), autosampler (Waters 717 plus Massachusetts, USA). The analysis of phyllanthin (1), hypophyllanthin (2), niranthin (3), and corilagin (4) in P. amarus extract was obtained from Wu, et al. [31 (link)] and Murugaiyah and Chan [32 (link)] with modification.
Isocratic mode with a flow rate of 1.0 mL/min and an injection volume of 20 µL using Column Waters X Bridge C18 (4.6 mm × 250 mm, 5 um particle size) was used for the quantification of P. amarus. Acetonitrile:water in the ratio of 55:45 was used as a solvent system to quantify phyllanthin (1), hypophyllanthin (2), and niranthin (3) using UV wavelength of 344 nm, whereas, 1% acetic acid in acetonitrile was used as a solvent in quantifying corilagin (4) with a UV wavelength of 274 nm.
+ Open protocol
+ Expand
9

HPLC Analysis of PAH Mixtures

Check if the same lab product or an alternative is used in the 5 most similar protocols
Qualitative and quantitative changes in PAH mixtures have been achieved using high-performance liquid chromatography (Waters HPLC 600E) for aromatic fractions. PAH standards purchased from Supelco CRMs for chromatography solutions (https://www.sigmaaldrich.com/analytical-chromatography/analytical-products.html?TablePage=120241095) have been also subjected successively to be analyzed by high-performance liquid chromatography (HPLC). The apparatus consists of an autosampler, Waters 717 plus, attached to a computerized system with Millennium 3.2 software and a dual UV absorbance detector, Waters 2487. The separation has been performed according to the conditions mentioned in [19 (link)].
+ Open protocol
+ Expand
10

HPLC Analysis of FITC-Ficoll

Check if the same lab product or an alternative is used in the 5 most similar protocols
A HPLC system (Waters, Milford, MA) was applied to quantify the concentration and size distribution of the FITC‐Ficoll samples. Size separation of the plasma and urine samples was accomplished using an Ultrahydrogel 500 column (Waters) connected to a guard column (Waters). A phosphate buffer (0.15 M NaCl, pH 7.4) was used as the mobile phase, driven by a pump (Waters 1525). Fluorescence in the samples was detected at an excitation wavelength 492 nm and emission wavelength of 518 nm. Samples were loaded onto the system using an autosampler (Waters 717 plus). As is described at some length elsewhere (Asgeirsson et al., 2006), the system was calibrated using protein standards and narrow Ficoll standards. The urine FITC‐Ficoll concentration for each size (ae) was multiplied by the urine‐to‐plasma FITC‐Inulin concentration ratio to approximate the concentration of Ficoll in primary urine (Cu). The glomerular sieving coefficient was calculated by dividing Cu by the plasma concentration of Ficoll.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!