The largest database of trusted experimental protocols

13 protocols using rslc ultimate 3000 nano uplc

1

Multinotch-MS3 for Accurate Quantitation

Check if the same lab product or an alternative is used in the 5 most similar protocols
In order to obtain superior quantitation accuracy, we employed multinotch-MS3 (McAlister GC, ref below) which minimizes the reporter ion ratio distortion resulting from fragmentation of co-isolated peptides during MS analysis [23 (link)]. Orbitrap Fusion (Thermo Fisher Scientific) and RSLC Ultimate 3000 nano-UPLC (Dionex) was used to acquire the data. Two μl of the sample was resolved on a PepMap RSLC C18 column (75 μm i.d. × 50 cm; Thermo Scientific) at the flow-rate of 300 nl/min using 0.1% formic acid/acetonitrile gradient system (2–22% acetonitrile in 150 min; 22–32% acetonitrile in 40 min; 20 min wash at 90% followed by 50 min re-equilibration) and directly spray onto the mass spectrometer using EasySpray source (Thermo Fisher Scientific). Mass spectrometer was set to collect one MS1 scan (Orbitrap; 120K resolution; AGC target 2 × 105; max IT 100 ms) followed by data-dependent, “Top Speed” (3 s) MS2 scans (collision induced dissociation; ion trap; NCE 35; AGC 5 × 103; max IT 100 ms). For multinotch-MS3, top 10 precursors from each MS2 were fragmented by HCD followed by Orbitrap analysis (NCE 55; 60K resolution; AGC 5 × 104; max IT 120 ms, 100–500 m/z scan range).
+ Open protocol
+ Expand
2

Improved Peptide Quantitation via Multinotch-MS3

Check if the same lab product or an alternative is used in the 5 most similar protocols
In order to obtain superior quantitation accuracy, we employed multinotch-MS3 which minimizes the reporter ion ratio distortion resulting from fragmentation of co-isolated peptides during MS analysis (72 (link)). Orbitrap Fusion (Thermo Fisher Scientific) and RSLC Ultimate 3000 nano-UPLC (Dionex) was used to acquire the data. Two μl of the sample was resolved on a PepMap RSLC C18 column (75 μm i.d. x 50 cm; Thermo Scientific) at the flow-rate of 300 nl/min using 0.1% formic acid/acetonitrile gradient system (2-22% acetonitrile in 150 min;22-32% acetonitrile in 40 min; 20 min wash at 90% followed by 50 min re-equilibration) and directly sprayed onto the mass spectrometer using EasySpray source (Thermo Fisher Scientific). Mass spectrometer was set to collect one MS1 scan (Orbitrap; 120K resolution; AGC target 2×105; max IT 100 ms) followed by data-dependent, “Top Speed” (3 seconds) MS2 scans (collision induced dissociation; ion trap; NCE 35; AGC 5x103; max IT 100 ms). For multinotch-MS3, top 10 precursors from each MS2 were fragmented by HCD followed by Orbitrap analysis (NCE 55; 60K resolution; AGC 5x104; max IT 120 ms, 100-500 m/z scan range).
+ Open protocol
+ Expand
3

Multinotch-MS3 Quantitation for Accuracy

Check if the same lab product or an alternative is used in the 5 most similar protocols
In order to obtain superior quantitation accuracy, we employed multinotch-MS3 (McAlister GC) which minimizes the reporter ion ratio distortion resulting from fragmentation of co-isolated peptides during MS analysis. Orbitrap Fusion (Thermo Fisher Scientific) and RSLC Ultimate 3000 nano-UPLC (Dionex) was used to acquire the data. 2 μl of the sample was resolved on a PepMap RSLC C18 column (75 μm i.d. × 50 cm; Thermo Scientific) at the flowrate of 300 nl/min using 0.1% formic acid/acetonitrile gradient system (2–22% acetonitrile in 150 min;22–32% acetonitrile in 40 min; 20 min wash at 90% followed by 50 min re-equilibration) and directly spray onto the mass spectrometer using EasySpray source (Thermo Fisher Scientific). Mass spectrometer was set to collect one MS1 scan (Orbitrap; 120K resolution; AGC target 2×105; max IT 100 ms) followed by data-dependent, “Top Speed” (3 seconds) MS2 scans (collision induced dissociation; ion trap; NCE 35; AGC 5×103; max IT 100 ms). For multinotch-MS3, top 10 precursors from each MS2 were fragmented by HCD followed by Orbitrap analysis (NCE 55; 60K resolution; AGC 5×104; max IT 120 ms, 100–500 m/z scan range).
+ Open protocol
+ Expand
4

Multinotch-MS3 for Accurate Quantitation

Check if the same lab product or an alternative is used in the 5 most similar protocols
In order to obtain superior quantitation accuracy, we employed multinotch-MS3 (McAlister GC, ref below) which minimizes the reporter ion ratio distortion resulting from fragmentation of co-isolated peptides during MS analysis [23 (link)]. Orbitrap Fusion (Thermo Fisher Scientific) and RSLC Ultimate 3000 nano-UPLC (Dionex) was used to acquire the data. Two μl of the sample was resolved on a PepMap RSLC C18 column (75 μm i.d. × 50 cm; Thermo Scientific) at the flow-rate of 300 nl/min using 0.1% formic acid/acetonitrile gradient system (2–22% acetonitrile in 150 min; 22–32% acetonitrile in 40 min; 20 min wash at 90% followed by 50 min re-equilibration) and directly spray onto the mass spectrometer using EasySpray source (Thermo Fisher Scientific). Mass spectrometer was set to collect one MS1 scan (Orbitrap; 120K resolution; AGC target 2 × 105; max IT 100 ms) followed by data-dependent, “Top Speed” (3 s) MS2 scans (collision induced dissociation; ion trap; NCE 35; AGC 5 × 103; max IT 100 ms). For multinotch-MS3, top 10 precursors from each MS2 were fragmented by HCD followed by Orbitrap analysis (NCE 55; 60K resolution; AGC 5 × 104; max IT 120 ms, 100–500 m/z scan range).
+ Open protocol
+ Expand
5

Multinotch-MS3 Quantification Pipeline

Check if the same lab product or an alternative is used in the 5 most similar protocols
An Orbitrap Fusion (Thermo Fisher Scientific) and RSLC Ultimate 3000 nano-UPLC (Dionex) were used to acquire the data. For superior quantitation accuracy, we used multinotch-MS3 59 (link). Two microliters of each fraction was resolved on a nanocapillary reverse-phase column (75 μm internal diameter by 50 cm; PepMap RSLC C18 column, Thermo Fisher Scientific) at a flowrate of 300 nL/min using 0.1% formic acid/acetonitrile gradient system (2 to 22% acetonitrile in 110 min; 22 to 40% acetonitrile in 25 min; 6-min wash at 90% acetonitrile; 25 min re-equilibration) and directly sprayed onto the Orbitrap Fusion using EasySpray source (Thermo Fisher Scientific). The mass spectrometer was set to collect one MS1 scan [Orbitrap; 120,000 resolution; AGC target 2 × 105; max IT (maximum ionization time) 50 ms] and data-dependent, “Top Speed” (3 s) MS2 scans [collision-induced dissociation; ion trap; NCE (normalized collision energy) 35; AGC (automatic gain control) 5 × 103; max IT 100 ms]. For multinotch-MS3, the top 10 precursors from each MS2 were fragmented by high energy collisional dissociation (HCD), followed by Orbitrap analysis (NCE 55; 60,000 resolution; AGC 5 × 104; max IT 120 ms, 100 to 500 mass/charge ratio scan range).
+ Open protocol
+ Expand
6

High-Sensitivity Orbitrap Fusion Proteomics

Check if the same lab product or an alternative is used in the 5 most similar protocols
An Orbitrap Fusion (Thermo Fisher Scientific) and RSLC Ultimate 3000 nano-UPLC (Dionex) were used to acquire the data. For superior quantitation accuracy, we used multinotch-MS3 (57 (link)). Two microliters of each fraction was resolved on a nanocapillary reverse-phase column (75 μm internal diameter by 50 cm; PepMap RSLC C18 column, Thermo Fisher Scientific) at a flow rate of 300 nl/min using 0.1% formic acid/acetonitrile gradient system (2–22% acetonitrile in 110 min; 22–40% acetonitrile in 25 min; 6-min wash at 90% acetonitrile; and 25 min reequilibration) and directly sprayed onto the Orbitrap Fusion using EasySpray source (Thermo Fisher Scientific). The mass spectrometer was set to collect one MS1 scan [Orbitrap; 120,000 resolution; automatic gain control (AGC) target 2 × 105; max IT (maximum ionization time) 50 ms] and data-dependent, “Top Speed” (3 s) MS2 scans [collision-induced dissociation; ion trap; normalized collision energy 35; AGC 5 × 103; max IT 100 ms]. For multinotch-MS3, the top 10 precursors from each MS2 were fragmented by high energy collisional dissociation, followed by Orbitrap analysis (normalized collision energy 55; 60,000 resolution; AGC 5 × 104; max IT 120 ms, 100–500 mass/charge ratio scan range).
+ Open protocol
+ Expand
7

Multinotch-MS3 for Quantitative Accuracy

Check if the same lab product or an alternative is used in the 5 most similar protocols
To obtain superior quantitative accuracy, we employed multinotch-MS3 (McAlister GC), as described by McAlister et al. [23 (link)]. This technique minimizes the reporter ion ratio distortion resulting from fragmentation of co-isolated peptides during MS analysis. Orbitrap Fusion (Thermo Fisher Scientific) and RSLC Ultimate 3000 nano-UPLC (Dionex) were used to acquire the data. Two μL of the sample was resolved on a PepMap RSLC C18 column (75 μm i.d. × 50 cm; Thermo Scientific) at the flow-rate of 300 nL/min using 0.1% formic acid/acetonitrile gradient system (2–22% acetonitrile in 150 min; 22–32% acetonitrile in 40 min; 20 min wash at 90% followed by 50 min re-equilibration) and directly sprayed onto the mass spectrometer using EasySpray source (Thermo Fisher Scientific). The mass spectrometer was set to collect one MS1 scan (Orbitrap; 120 K resolution; AGC target 2 × 105; max IT 100 ms) followed by data-dependent, “Top Speed” (3 s) MS2 scans (collision induced dissociation; ion trap; NCE 35; AGC 5 × 103; max IT 100 ms). For multinotch-MS3, top 10 precursors from each MS2 were fragmented by HCD followed by Orbitrap analysis (NCE 55; 60 K resolution; AGC 5 × 104; max IT 120 ms, 100–500 m/z scan range).
+ Open protocol
+ Expand
8

Multinotch-MS3 for Accurate Peptide Quantification

Check if the same lab product or an alternative is used in the 5 most similar protocols
In order to obtain superior quantitation accuracy, we employed multinotch-MS3 which minimizes the precursor ion interference that occurs from fragmentation of co-isolated peptides during MS analysis62 (link). Orbitrap Fusion (Thermo Fisher Scientific) and RSLC Ultimate 3000 nano-UPLC (Dionex) was used to acquire the data. 2 μL of the sample was resolved on a PepMap RSLC C18 column (75 μm i.d. × 50 cm; Thermo Fisher Scientific) at the flow-rate of 300 nl/min using 0.1% formic acid/acetonitrile gradient system (2–22% acetonitrile in 150 min;22–32% acetonitrile in 40 min; 20 min wash at 90% followed by 50 min re-equilibration) and directly spray onto the mass spectrometer using EasySpray source (Thermo Fisher Scientific). Mass spectrometer was set to collect one MS1 scan (Orbitrap; 120 K resolution; AGC target 2 × 105; max IT 100 ms) followed by data-dependent, “Top Speed” (3 s) MS2 scans (collision induced dissociation; ion trap; NCE 35; AGC 5 × 103; max IT 100 ms). For multinotch-MS3, top 10 precursors from each MS2 were fragmented by HCD followed by Orbitrap analysis (NCE 55; 60 K resolution; AGC 5 × 104; max IT 120 ms, 100–500 m/z scan range). Since the TMT modification is used as a “fixed modification”, unlabeled peptides are not identified or used.
+ Open protocol
+ Expand
9

Orbitrap Fusion Multinotch-MS3 Quantification

Check if the same lab product or an alternative is used in the 5 most similar protocols
For data acquisition, an Orbitrap Fusion (ThermoFisher) and RSLC Ultimate 3000 nano-UPLC (Dionex) was used to obtain raw data. To increase accuracy and confidence in protein abundance measurements, a multinotch-MS3 method was employed for MS data analysis. Two microliters from each fraction were resolved in 2D on a nanocapillary reverse phase column (Acclaim PepMap C18, 2 micron, 75 μm i.d. × 50 cm, ThermoFisher) using a 0.1% formic/acetonitrile gradient at 300 nl/m (2–22% acetonitrile in 150 m, 22–32% acetonitrile in 40 m, 20 min wash at 90% followed by 50 min reequilibration) and directly sprayed onto Orbitrap Fusion with EasySpray (ThermoFisher; Spray voltage (positive ion) = 1900 V, Spray voltage (negative ion) = 600 V, method duration = 180 min, ion source type = NSI). The mass spectrometer was set to collect the MS1 scan (Orbitrap; 120 K resolution; AGC target 2 × 105; max IT 100 ms), and then data-dependent Top Speed (3 s) MS2 scans (collision induced dissociation; ion trap; NCD 35; AGC 5 × 103; max IT 100 ms). For multinotch-MS3, the top 10 precursor ions from each MS2 scan were fragmented by HCD followed by Orbitrap analysis (NCE 55; 60 K resolution; AGC 5 × 104; max IT 120 ms; 100-500 m/z scan range).
+ Open protocol
+ Expand
10

Multinotch-MS3 for Protein Quantification

Check if the same lab product or an alternative is used in the 5 most similar protocols
To increase the accuracy and confidence in measurements of protein abundance, a multinotch-MS3 method was employed for analyzing MS data [33 (link)]. Raw data were acquired by using RSLC Ultimate 3000 nano-UPLC (Dionex) coupled to Orbitrap Fusion MS (Thermo Fisher Scientific, MA). Two microliters from each fraction were resolved on an Acclaim PepMap C18 reverse phase column (2 microns, 75 μm i.d. × 50 cm) using a 0.1% formic/acetonitrile gradient at 300 nl/min. The mass spectrometer was set to collect one MS1 scan (Orbitrap; 120 K resolution; AGC target 2×105; max IT 100 ms) followed by data-dependent, “Top Speed” (3 s) MS2 scans (collision induced dissociation; ion trap; NCD 35; AGC 5×103; max IT 100 ms). For multinotch-MS3, the top ten precursors from each MS2 scan were fragmented by high-energy collisional dissociation (HCD) followed by Orbitrap analysis (NCE 55; 60,000 resolution; AGC 5×104; max IT 120 ms, 100–500 m/z scan range).
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!