The largest database of trusted experimental protocols

1290 infinity 2 lc system

Manufactured by Agilent Technologies
Sourced in United States, Germany

The Agilent 1290 Infinity II LC System is a high-performance liquid chromatography (HPLC) platform designed for efficient and reliable separation and analysis of a wide range of samples. The system features advanced technology and components to deliver high-quality data and results.

Automatically generated - may contain errors

54 protocols using 1290 infinity 2 lc system

1

Quantifying Mycotoxin Metabolites in Cells

Check if the same lab product or an alternative is used in the 5 most similar protocols
Sample analysis was performed with a 1290 Infinity II LC System (Agilent Technologies, Waldbronn, Germany) coupled to a QTrap 6500+ LC-MS/MS system (AB Sciex, Redwood City, CA, USA). For the analysis of DON and its biotransformation products and CER, two tailored methods were applied. This ensured highest sensitivities for CER and DON and its metabolites which required the presence or absence of ammonium acetate in the eluents. The LC-MS/MS system was operated by Analyst 1.7.0 software, whereas data analysis and evaluation were performed with Sciex OS 1.5.0 and Microsoft Excel 2016. Results are expressed as detected analyte in ng/mL medium for basolateral and apical medium and pg/µg cell protein for cell lysate samples.
+ Open protocol
+ Expand
2

Analytical LC-MS/MS Method for Compounds

Check if the same lab product or an alternative is used in the 5 most similar protocols
We used an Agilent Technologies 1290 Infinity II LC system, including an autosampler, binary pumps and a thermostatted column compartment with 6470 Triple Quad (Agilent Technologies, Santa Clara, CA, USA). Other necessary equipment were MS 40+ Vacuum Products (Agilent Technologies, Santa Clara, CA, USA) and a nitrogen generator NM32LA (Peak Scientific, Inchinnan, UK). The sample separation was carried out on a reversed phase column SB-C8, 1.8 µm, 2.1 × 100 mm (Agilent Technologies, Santa Clara, CA, USA).
+ Open protocol
+ Expand
3

Quantitative LC-MS/MS Peptide Analysis

Check if the same lab product or an alternative is used in the 5 most similar protocols
LC-MS/MS analyses were performed using an Acquity UPLC CSH C18 1.7 μm, 2.1 mm × 50 mm column (Waters, Milford, MA) at 40 °C on a 1290 Infinity II LC System (Agilent, Santa Clara, CA) coupled with an Agilent 6495 triple quadrupole mass spectrometer. Mobile phase A consisted of 0.1% formic acid (FA) in water, and mobile phase B was composed of 0.1% FA in acetonitrile. Tryptic peptides were separated at a flow rate of 0.4 mL/min with a linear gradient from 5% mobile phase B to 25% mobile phase B over 8 min. The column was then washed at 90% mobile phase B for 1 min and re-equilibrated to 5% mobile phase B for 2 min with a total analysis time of 15 min. The MRM mass spectrometer settings were as follows: polarity was positive mode, MS1 and MS2 resolution was wide unit, dwell time was 50 ms, and 12 quantifying and 16 confirming MRM transitions were monitored at various collision energy as described in Table I and Table S1, respectively. The MRM-MS data was analyzed using Skyline (v4.1, MacCoss Lab, the University of Washington in Seattle, WA). Total peak areas from two fragment ions for each quantifying peptide were used for quantitation.
+ Open protocol
+ Expand
4

LC-MS/MS Proteomics Analysis of Peptides

Check if the same lab product or an alternative is used in the 5 most similar protocols
Peptide samples were analyzed using Agilent 1290 Infinity II LC system coupled with Agilent 6545XT Q-TOF mass spectrometer. LC separation was conducted on AdvanceBio Peptide Mapping column (120 Å, 2.1 × 150 mm, 2.7 µm) at 60 °C. Injection volume was 10 µl. Mobile phase A was 0.1% FA in water and mobile phase B was 0.1% FA in acetonitrile. LC gradient was set as follows; 0% B for 2 min, 0–20% B in 33 min, 20–30% B in 20 min, 30–50% B in 10 min, 50–90% B in 5 min, 90% B for 5 min, 90–0% B in 5 min and 0% B for 5 min, with constant flow rate of 0.4 ml min−1. MS analysis was conducted in positive mode with a mass range of 100–1700 m/z. MS parameters were set as follows; gas temperature at 325 °C, nebulizer at 35 psi, dying gas at 13 L min−1, sheath gas temperature at 350 °C, sheath gas flow at 12 L min−1, capillary voltage at 4000 V, nozzle voltage at 500 V, fragmentor voltage at 175 V and skimmer voltage at 65 V. Acquisition time was 1 spectrum per s. Maximum 2 precursor ions per cycle were selected for MS/MS fragmentation. Collision energy (CE) was varied according to the charge state of the peptide. For peptides with charge + 1 and + 2, the CE was calculated using a formula of (3.1 × ((m/z)/100) + 1), while peptides with charge ≥  + 3, the CE was calculated using a formula of (3.6 × ((m/z)/100) − 4.8).
+ Open protocol
+ Expand
5

Quantitative Proteome Analysis by LC-MS/MS

Check if the same lab product or an alternative is used in the 5 most similar protocols
Peptides were analyzed using a Q ExactiveTM Hybrid Quadrupole-OrbitrapTM Mass Spectrometer (Thermo Fisher Scientific), which was connected to a 1290 Infinity II LC System (Agilent). The trap column was made of C18 (Dr. Maisch Reprosil) material and the analytical column was a 50 cm, 50 μm inner diameter Poroshell C18 (Agilent) column. Both the trap and analytical columns were packed in-house. Solvent A consisted of 0.1% formic acid (Merck) in deionized water (Merck) and Solvent B of 0.1% formic acid in 80% acetonitrile (Biosolve). Peptides were first trapped at 50 μl/min with solvent A and then eluted with solvent B in a 120 min gradient at 100 nl/min: 0–10 min, 100% solvent A; 10.1–105 min, 13–40% solvent B; 105–108 min, 40–100% solvent B; 108–109 min, 100% solvent B; 109–110 min, 0–100% solvent A; 110–120 min, 100% solvent A. The Orbitrap was operated in a data-dependent manner, with the following settings: ESI voltage, 1700 V; inlet capillary temperature 320°C; full-scan automatic gain control (AGC) target, 3 × 106 ions at 35000 resolution; scan range, 350–1500 m/z; Orbitrap full-scan maximum injection time, 250 ms; MS2 scan AGC target, 5 × 104 ions at 17500 resolution; maximum injection, 120 ms; normalized collision energy, 25; dynamic exclusion time, 30; isolation window 1.5 m/z; 10 MS2 scans per full scan.
+ Open protocol
+ Expand
6

Caffeine Quantification by HPLC

Check if the same lab product or an alternative is used in the 5 most similar protocols
The caffeine concentration was determined by using a 1290 infinity II LC System (AGILENT Santa Clara California) equipped with a C18 reversed-phase column (ACQUITY UPLC BEH-C18, 1.7 μm, 100 x 2.1 mm, WATERS CORPORATION, Massachusetts) set at 25°C. A 5 μL sample was injected for isocratic elution at 0.25 mL/min. The composition of the eluent was 80% water/20% methanol. The wavelength was set at 273 nm. Standard calibration curves for caffeine (0.1 and 1000 mg/L) were used.
+ Open protocol
+ Expand
7

Heterodimer Formation Analysis by LC-MS

Check if the same lab product or an alternative is used in the 5 most similar protocols
Samples for each of the heterodimer constructs were analyzed via LC-MS to determine the percentage of heterodimer formation. The samples were set up in 96-well plates with a final volume of 20 μl and a concentration of 5 μM. Each expressed protein was analyzed in the absence and presence of EndoS2 (final concentration, 50 nM). EndoS2 is an IgG-specific endoglycosidase from Streptococcus pyogenes, which removes the glycans from the conserved Asn297 site and simplifies data analysis. The samples were analyzed using an Agilent 1290 Infinity II LC System equipped with a 50-mm polymeric reverse-phase column from Agilent with 1000-Å pore size. The LC system is attached to an Agilent 6560 Ion Mobility (IM) Quadrupole Time-Of-Flight mass spectrometer (Agilent, Santa Clara, CA). Relative amounts of each of the deglycosylated peaks were quantified after deconvolution of the raw data and identification of the corresponding peaks using BioConfirm (Agilent, Santa Clara, CA).
+ Open protocol
+ Expand
8

Lipid Profiling using LC-QTOF/TQ-MS

Check if the same lab product or an alternative is used in the 5 most similar protocols
Lipid extracts were analysed on an Agilent 1290 Infinity II LC System coupled to the Agilent 6546 LC/Q-TOF system using a ZORBAX Eclipse Plus, C18, 95 Å, 1,8 µm, 2,1 × 50 mm (Agilent). Samples for MRM targeted analysis were measured with the same chromatographic system connected to the Agilent 6495 LC/TQ. The Agilent Data Acquisition Software Version 10.1 was used for all LC–MS analyses.
For the chromatographic runs, Water/Acetonitrile (6:4, v/v) + 10 mM Ammonium Formate was used as Mobile Phase A and Isopropanol/Acetonitrile (9:1, v/v) + 10 mM Ammonium Formate was used as Mobile Phase B, with an elution gradient starting at 0 min with 80% A and including: 2 min 40% A, 12 min 0% A, 14 min 80% A. The stop time was set at 15.8 min with a flow rate of 0.4 mL/min and a column temperature of 40 °C. An injection volume of 4 µL was used in positive mode and 8 µL in negative mode. For the dilution series, injection volumes of 0.25, 0.5, 1, 2, 4 and 8 µL were used.
The ion source parameters for DDA and DIA were as follows: Gas temperature 325 °C, Drying gas 10L/min, Nebulizer 20 psi, Sheath gas temperature 350 °C, Sheath gas flow 12L/min, Capillary voltage 3500 V, Nozzle voltage 1000 V, Fragmentor voltage 180 V, Skimmer voltage 45 V and Octopole RF 750 V.
+ Open protocol
+ Expand
9

Comprehensive Untargeted Metabolomics of Plasma

Check if the same lab product or an alternative is used in the 5 most similar protocols
Aliquots (40 µL for (positive mode) and 120 µL for (negative mode)) of thawed plasma (NIST SRM 1950 Metabolites in Frozen Plasma, Sigma, St. Louis USA) were each extracted using a modified Folch extraction procedure [36 (link)] and reconstituted in 100 µL of a methanol/chloroform mixture (9:1, v/v). LC separation was performed on an Agilent 1290 Infinity II LC System, with a 19 min gradient time on a reverse phase C18 column (Agilent InfinityLab Poroshell 120 EC-C18, 3.0 × 100 mm, 2.7 µm). Mobile phase consisted of 10 mM ammonium acetate and 0.2 mM ammonium fluoride in 9:1 water/methanol, while mobile phase B consisted of 10 mM ammonium acetate and 0.2 mM ammonium fluoride in 2:3:5 acetonitrile/methanol/isopropanol. Negative and positive polarity data was acquired on the Agilent 6546 LC/Q-TOF using iterative MS/MS acquisition mode on 6 injections of extracted plasma for each polarity [37 ]. Detailed experimental methods for chromatography and mass spectrometry can be found in Supplemental Table S1 and Table S2, respectively, and in the Agilent application note 5994-0775en [37 ]. Two methods were used, a high-load and a low-load method, to determine the effect of high injection volumes/concentration on the number of annotations using the Agilent 6546 LC/Q-TOF.
+ Open protocol
+ Expand
10

Cryopreserved Hepatocyte Incubation Assay

Check if the same lab product or an alternative is used in the 5 most similar protocols
Cryopreserved hepatocytes from humans (mixed gender pool of five donors) and Wistar rats were incubated with 1 µM [3H]asundexian in William’s E medium at 37 °C in suspension culture at a cell density of 106 viable cells/mL. Aliquots taken at 0 hours and 4 hours of incubation time were mixed with acetonitrile to a final content of about 30% (v/v), and samples were stored at −20 °C. Before analysis, the samples were thawed and centrifuged for 5 min. To control for the recovery of radioactivity, an aliquot of 10 μL of the supernatant was analyzed by LSC using a Tri‑Carb 2900TR counter (PerkinElmer, Turku, Finland) with automatic quench correction (workup recoveries were > 88%). Further aliquots of the supernatants were analyzed by HPLC-HRMS using a 1290 Infinity II LC system (Agilent, Waldbronn, Germany) and a Q Exactive Plus with a heated electrospray ionization source used in positive ionization mode (Thermo Fisher Scientific, Bremen, Germany) and a 1:10 split to a fraction collector (Gilson robotic liquid handler 222 XL, Abimed Analysen-technik, Langenfeld, Germany). Fractions were mixed with 150 µL Ultima-Flo AP scintillation cocktail, and offline radioactivity monitoring was conducted using a 1450 MicroBeta2 counter (PerkinElmer, Turku, Finland).
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!