The largest database of trusted experimental protocols

Hp 1100 series liquid chromatograph

Manufactured by Agilent Technologies
Sourced in Germany, United States

The HP 1100 series liquid chromatograph is a high-performance liquid chromatography (HPLC) system designed for analytical and preparative applications. It features a modular design, allowing for the configuration of various components such as pumps, detectors, and autosamplers to meet specific analytical needs. The HP 1100 series provides reliable and precise performance for the separation, identification, and quantification of a wide range of chemical compounds.

Automatically generated - may contain errors

3 protocols using hp 1100 series liquid chromatograph

1

Size-Exclusion Chromatography for Protein Analysis

Check if the same lab product or an alternative is used in the 5 most similar protocols
Size-exclusion chromatography (SEC) was performed under isocratic elution conditions with a HiLoad 16/600 Superdex 200 preparative grade column (10,000–600,000 relative molecular mass (Mr) range, GE Healthcare, Chicago, IL, USA) and a HiLoad 16/600 Superdex 75 preparative grade column (3,000–70,000 Mr range, GE Healthcare) for the analysis of SDP and EHPP, respectively, using an HP 1100 series liquid chromatograph with an ultraviolet absorption diode array detector (Agilent Technologies, Waldbronn, Germany). Instrument control, data acquisition and data processing were performed using the Chemstation LC3D software (Agilent Technologies). The mobile phase consisted of phosphate buffered saline (PBS, pH = 7.50) and absorbance was monitored at 214 nm. After column equilibration with mobile phase (two column volumes), a calibration curve was constructed by injecting 1 mL of a 500 mg/L mixture of protein standards of known Mr according to the manufacturer instructions. SDP and EHPP were analyzed by injecting volumes of 1 mL, and fractions (for EHPP only) were collected using an Agilent 1200 series analytical fraction collector (Agilent Technologies).
+ Open protocol
+ Expand
2

Liquid Chromatography Analysis of Water Pollutants

Check if the same lab product or an alternative is used in the 5 most similar protocols
The quantification
of the target
pollutants in water samples was performed by liquid chromatography
coupled to a UV detector (HP1100 series liquid chromatograph (Agilent,
Palo Alto, CA), which provides good sensitivity levels as it is indicated
in Table S1. The chromatographic separation
took place on an Agilent Eclipse XDB-C18 column (150 mm × 4.6
mm i.d. × 5.0 μm) (Waltham, MA, USA) using a flow rate
of 1 mL min–1 for the mobile phase. A gradient elution
mode was used. The initial mobile phase’s composition consisted
of 70% component A (water) and 30% component B (ACN). The percentage
of component B was linearly increased from 30 to 70% (from 4 to 7
min) and then decreased again to 30% (from 7 to 8 min) for column
re-equilibration. 20 μL of solutions was automatically injected
by an autosampler, and the pollutants were measured at 280 nm.
+ Open protocol
+ Expand
3

Quantitative Analysis of Organic Acids

Check if the same lab product or an alternative is used in the 5 most similar protocols
Malic acid, ascorbic acid, and citric acid determination was carried out according the method described by Scherer et al. (15 ). An HP 1100 series liquid chromatograph (Agilent Technologies, Santa Clara, CA, USA) equipped with a degasser, a quaternary pump, an autosampler adjusted to deliver 20 μL injections, an RP-C18 column (5 μm particle size, 250×4.6 mm I.D., kept at 25°C), and a UV-visible diode array detector. The mobile phase consisted of a 25 mM KH2PO4 buffer solution (pH=2.60 adjusted with o-phosphoric acid). An isocratic elution procedure with a flow rate of 0.8 mL/min was used. Malic acid and citric acid were detected at 210 nm, and ascorbic acid was detected at 250 nm. Peaks of interest were identified by their retention times. Peaks observed on the UV-visible spectra of each sample were compared to peaks observed on the UV-visible spectra of known standards. A five-point external standard curve was used to quantify each analyte. The concentrations of the standard curves ranged from the quantification limits and 0.25 mg/mL, 0.5 mg/mL, 0.01 mg/mL, and 0.8 mg/mL for malic, ascorbic, and citric acids, respectively. The linearity range of each analyte was evaluated by plotting the peak area, as a function of the known concentration of each standard.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!