Within 1 month after diagnosis and treatment initiation, 57 subjects were scanned on a Siemens Skyra 3 T and 12 subjects on a General Electric Signa 3 T. Parameters for the whole brain 3D sagittal T1 weighted and fat saturated sequence are provided in Table 1. Images for this study were acquired 10 min after intravenous injection of 0.2 ml/kg of 0.5 mmol/ml gadoterate meglumine (Guerbet, France).

Magnetic resonance imaging parameters for whole brain 3D sagittal contrast enhanced T1 weighted and fat saturated sequence used to rate enhancement of cranial nerves III–XII

Siemens Skyra 3 T (n = 57)General Electric Signa 3 T (n = 12)
DescriptionT1_sag_space_FS_CESagCUBET1FatsatK + 
Coil configuration64 channel head32 channel head
Slice thickness (mm)0.931.0
FOV read (mm)256256
FOV phase100%100%
TR (ms)500650
TE (ms)3.8Minimum
Fat suppressionFat satFat sat
Band width (Hz/Px)630625
Turbo factor4030
Averages/NEX1.01.0
ShimB0 standard/B1 TrueformAuto
+ Open protocol