The largest database of trusted experimental protocols

Db ffap column

Manufactured by Agilent Technologies
Sourced in United States, Japan, Germany, Canada

The DB-FFAP column is a gas chromatography column designed for the separation and analysis of polar compounds. It features a polyethylene glycol stationary phase, which provides high thermal stability and excellent peak shape for a wide range of polar analytes.

Automatically generated - may contain errors

102 protocols using db ffap column

1

Quantification of Acetate in Biological Samples

Check if the same lab product or an alternative is used in the 5 most similar protocols
Acetate was detected by GC-MS/MS. Wuhan MetWare Biotechnology was in charge of extracting the sample and quantifying metabolites, as previously described [40 (link)]. Briefly, samples of BAA6-CM, serum, and feces were thawed and vortexed for 1 min prior to analysis. Samples (50 μL) were added to 100 μL of 36% phosphoric acid solution. The mixture was vortexed for 3 min and 150 μL methyl tert-butyl ether (containing 2 μg/mL 2-methylvaleric acid as an internal standard) solution was added. After that, centrifugal force at 12,000× g was used to centrifuge the mixture at 4 °C for 10 min. The supernatant was used for further analysis. Agilent 7890B instrument, coupled to a 7000D Triple Quadrupole gas chromatography mass spectrometry system with a DB-FFAP column (30 m × 0.25 mm × 0.25 μm, J&W Scientific, Folsom, CA, USA), was utilized. The carrier gas was helium, and the flow rate was 1.2 mL/min. The oven temperature was held at 90 °C for 1 min, risen to 100 °C at a rate of 25 °C/min, risen to 150 °C at a rate of 20 °C/min, held on 0.6 min, risen to 200 °C at a rate of 25 °C/min, and kept for 0.5 min after running for 3 min. The quantitation of acetate in BAA6-CM, serum, and feces was counted using constructing calibration curves.
+ Open protocol
+ Expand
2

HS-GC-IMS Analysis of Volatile Compounds

Check if the same lab product or an alternative is used in the 5 most similar protocols
For the HS-GC-IMS analysis, a Shimadzu GC-2010 gas chromatograph (Shimadzu, Kyoto, Japan) equipped with a Perkin Elmer TurboMatrix16 autosampler (Perkin Elmer, MA, USA.) was coupled to an IMS module from GAS (Dortmund, Germany).
The analytes were separated in a DB-FFAP column (60 m × 0.25 mm × 0.25 μm film thickness; J & W Scientific; Folsom, CA, USA.) using nitrogen gas (>99.999%) at a constant flow rate of 0.8 mL/min. The temperature of the column was maintained at 40 °C for 3 min, then was increased to 150 °C at 4 °C/min, and held at 150 °C for 5 min (total of 35.5 min).
After the analytes were separated in the column, they were driven into the IMS module. First, the volatile organic compounds were ionized by the tritium source in positive mode. Then, the ions were placed into a 9.8 cm long drift tube operating at 500 V/cm and 45 °C. Next, nitrogen drift gas (>99.999%) was introduced at 150 mL/min. An average of 12 scans was performed at a repetition rate of 30 ms and a grid pulse width of 150 μs to build each spectrum. HS-GC-IMS data were obtained by Standalone (GAS, Dortmund, Germany), and the raw data were analyzed using VOCal (GAS, Dortmund, Germany) software to reveal information regarding the composition of the samples.
+ Open protocol
+ Expand
3

Identification of VOCs in Distillates using GC-MS/O

Check if the same lab product or an alternative is used in the 5 most similar protocols
For identification of the volatile organic compounds (VOCs) present in the distillates, GC-MS/O was performed using a Trace GC Ultra and a Trace dual stage quadrupole (DSQ) mass spectrometer (both Thermo Fisher Scientific GmbH) equipped with a DB-FFAP column (30 m × 0.32 mm, 0.25 μm film thickness, J&W Scientific, Waldbronn, Germany). Measurements were performed following the method described in detail by Bickel Haase et al. (2021) (link). Samples were evaluated by three trained panellists.
Further GC-MS/O determinations were performed using an Agilent GC-MSD (Agilent Technologies, Waldbronn DE). Aliquots (2 μL) of the samples were injected in pulsed splitless mode at 250 °C by means of an autosampler PAL RSI 85 (CTC Analytics AG, Zwingen, Switzerland). At the end of the column (DB-FFAP, 30 m × 0.25 mm, 0.25 μm film thickness, J&W Scientific), the effluent was split 1:1 by a Y-splitter. Two deactivated fused silica capillaries led either to an odor detection port (235 °C) or a mass spectrometer. The oven temperature was 40 °C at the beginning, raised at 6.0 °C/min after 2 min to 235 °C, and held for 5 min. The flow rate of the helium carrier gas was 2.0 mL/min in constant flow. Mass spectra were generated in the positive electron impact (EI) ionization mode at 70 eV (Scan 35–400 m/z).
+ Open protocol
+ Expand
4

Quantitative Analysis of Methional

Check if the same lab product or an alternative is used in the 5 most similar protocols
Methional was enriched by SPE and quantified by GC–PFPD. DB-FFAP column (30 m × 0.32 mm i.d., 1 μm film thickness, J&W Scientific Inc., Folsom, CA, USA) was used. The flow of the helium was 2 mL/min. Temperature of oven was programmed at 35 °C for 3 min and was increased to 150 °C at 10 °C/min, and held for 5 min. Then, it was increased at 20 °C/min to a final temperature of 220 °C and held for 3 min. Temperature of the GC injection and the detector was 250 °C. Sulfur gate time was 6–24.9 ms, and pulse frequency was approximately three pulses/s. Isopropyl disulfide (48.62 mg/L, 8 μL) was used as an internal standard (IS6) [12 (link)].
+ Open protocol
+ Expand
5

Carbonyl Compounds Quantitation by GC-MS

Check if the same lab product or an alternative is used in the 5 most similar protocols
Quantitation of carbonyl compounds after derivatization was performed by GC−MS on a DB-FFAP column (60 m × 0.25 mm i.d., 0.25 μm film thickness, J&W Scientific) modified from literature. A sample of 8 mL saturated with NaCl was placed in a 20-mL standard headspace vial. A total of 10 μL of p-Fluorobenzaldehyde (internal standard, 100 mg/L) and 300 μL of PFBHA (20 g/L in water) were added. A 50/30-μm DVB/CAR/PDMS fiber (Supelco, Inc.,Bellefonte, PA) was used. The sample was equilibrated at 65 °C for 10 min, then extracted for 45 min under stirring at 250 rpm and desorption at 250 °C for 300 s. Sample was injected in a splitless mode. Carrier gas for the column was helium at a constant flow rate of 1 mL/min. Temperature for the injector was 280 °C. The oven temperature was 50 °C for 2 min and was increased to 100 °C at 6 °C/min for 0.1 min, then was increased to 160 °C at a rate of 2 °C/min for 0.1 min, and finally at 5 °C/min to 230 °C for 10 min. The electron impact energy was 70 eV with SIM. The ion monitored for p-Fluorobenzaldehyde after derivatization was m/z 319. The standard curve, LOD and recovery of carbonyl compounds, were measured by the method mentioned above.
+ Open protocol
+ Expand
6

FAME Analysis of D. neapolitana

Check if the same lab product or an alternative is used in the 5 most similar protocols
Methyl esters of fatty acids (FAME) of D. neapolitana (~50 mg) were prepared following the method described by Aued-Pimentel et al. [38 (link)] (through transmethylation of FA using a mixture of methanolic solution KOH (2 M) and saturated NaCl). The resulting FAME were analyzed in a QP2010 Ultra Shimadzu gas chromatography–mass spectrometry, equipped with an auto-sampler a DB-FFAP column with 30 m length, 0.32 mm internal diameter, and 0.25 µm film thickness (J&W Scientific, Folsom, CA, USA). The column was initially programmed to 80 °C, increasing 25 °C min−1 until 160 °C, 2 °C min−1 from 160 to 220 °C, and 30 °C min−1 from 220 to 250 °C, using helium as the carrier gas, at a flow of 1.8 mL min−1. FAME identification was accomplished through comparison of retention times with those mixed FAME standards (C4–C24, Supelco 37 Component Fame Mix) and by comparison of the mass spectrum of each relative to standard spectra from the library “AOCS Lipid Library” (http://lipidlibrary.aocs.org/ (accessed on 5 January 2024)).
+ Open protocol
+ Expand
7

Gas Chromatography for SCFA Analysis

Check if the same lab product or an alternative is used in the 5 most similar protocols
Procedures for SCFA analysis were described by a previous study (Williams et al., 2011 (link)). The gas chromatography GC-17A (Shimadzu, Kyoto, Japan) with a flame ionization detector (FID) equipped with a DB-FFAP column (30 m × 0.53 mm) (J&W Scientific) was applied to detect SCFA.
+ Open protocol
+ Expand
8

Extraction and Quantification of β-Damascenone

Check if the same lab product or an alternative is used in the 5 most similar protocols
β-damascenone was extracted using CH2Cl2 as the organic solvent, and L-menthol (100 mg/L) was used as the internal standard. Solid samples (10 g) were added to 25 mL sterile saline (0.85% NaCl and 1% CaCl2), ultrasonically treated for 30 min in an ice-water mixture, and then centrifuged at 8,000 × g for 5 min (4°C) to obtain supernatants. For liquid samples, yeast fermentation broths were centrifuged at 8,000 × g for 5 min (4°C). For the quantitation of β-damascenone, CH2Cl2 (2 mL), internal standard (20 μL), and NaCl (7 g) were added to the supernatants (20 mL). The mixture was stirred for 10 min at ambient temperature and sonicated in an ice bath for 10 min. The above obtained sample was centrifuged and refrigerated overnight to separate the organic phase. Next, the lower organic phase was collected and dried over anhydrous sodium sulfate. The extracted organic phase was analyzed using a gas chromatography-mass spectrometry (GC-MS) system (Agilent Technologies Inc., Santa Clara, CA, USA) equipped with a DB-FFAP column (30.0 m × 0.25 mm × 0.25 μm, J&W Scientific, Santa Clara, CA, USA). The standard solution was made from aqueous alcohol solution and applied to construct a standard curve. All standard solutions were analyzed in the same manner as described for samples. The conditions for GC-MS were based on a previously reported method (Sun X. Z. et al., 2022 (link)).
+ Open protocol
+ Expand
9

Serum Metabolite Profiling by GC-MS

Check if the same lab product or an alternative is used in the 5 most similar protocols
Serum samples were thawed and well mixed prior to analysis. Each sample (50 uL) was vortex mixed with 100 μL of phosphoric acid solution (36% v/v) for three minutes, followed by adding 150 μL of methyl tert-butyl ether analytical standard (MTBE standard solution). The mixture was vortexed for 3 min again and ultrasonicated for 5 min. After that, the mixture was centrifuged, and the supernatant was collected for GC-MS analysis. An Agilent 7890B GC system coupled to a 7000D GC/MS with a DB-FFAP column (30 m length × 0.25 mm inner diameter × 0.25 μm film thickness; J&W Scientific, USA) was used to detect the SCFAs and neurotransmitters in the serum samples.
+ Open protocol
+ Expand
10

Volatile Compounds Identification in Distillates

Check if the same lab product or an alternative is used in the 5 most similar protocols
For identification of the VOCs present in the distillates, GC-MS/O was performed using a Trace GC Ultra and a Trace dual stage quadrupole (DSQ) mass spectrometer (both Thermo Fisher Scientific GmbH) equipped with a DB-FFAP column (30 m × 0.32 mm, 0.25 μm film thickness, J&W Scientific, Waldbronn, Germany). The distillates (2 µL) were injected automatically by a multipurpose autosampler MPS 2 (Gerstel GmbH & Co. KG, Mülheim an der Ruhr, Germany) using the cold on-column technique. The initial temperature of 40 °C was held for 2 min, raised at 6.0 °C/min to 235 °C, and held for 5 min. The flow rate of the helium carrier gas was 2.2 mL/min. At the end of the capillary column, the effluent was split between an ODP and the MS using deactivated fused silica capillaries (50 cm × 0.2 mm). Mass spectra were recorded in positive electron ionization (EI) mode at 70 eV (m/z range 35–250).
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!