The largest database of trusted experimental protocols

Agilent 1290 series uhplc system

Manufactured by Agilent Technologies
Sourced in United States, Germany

The Agilent 1290 series UHPLC system is a high-performance liquid chromatography (HPLC) instrument designed for ultra-high-performance liquid chromatography (UHPLC) analysis. The system is capable of operating at pressures up to 1200 bar, enabling the use of sub-2 micron particle size columns for improved resolution, sensitivity, and analysis speed.

Automatically generated - may contain errors

19 protocols using agilent 1290 series uhplc system

1

Serum Metabolomics Analysis of Feeding and Parity

Check if the same lab product or an alternative is used in the 5 most similar protocols
Details concerning the determination of the serum metabolome and the results related to feeding and parity are given in Pacífico et al. (4 (link)). Briefly, 10 μl of aliquots of serum samples were processed using a targeted metabolomics approach based on the Biocrates MxP® Quant 500 kit (Biocrates Life Sciences AG, Innsbruck, Austria). Analysis of serum metabolome as well as of reference standards and quality controls (provided by the manufacturer) was carried out by ultra-high-performance liquid chromatography (uHPLC) and flow injection analysis (FIA), both coupled to tandem mass spectrometry. An Agilent 1290 series UHPLC system coupled to a 6500+ QTrap mass spectrometer equipped with an Ion-Drive Turbo V® ESI source (both Sciex, Foster City, CA, USA) was used for the analysis. Chromatographic and mass spectrometric parameters were set according to manufacturer's instructions. Data analysis was carried out in Analyst 1.6.3 (Sciex) for LC-MS/MS data and in the Biocrates MetIDQ software for FIA-MS/MS data.
+ Open protocol
+ Expand
2

Mycotoxin Analysis in Wheat via LC-MS/MS

Check if the same lab product or an alternative is used in the 5 most similar protocols
Determination of mycotoxins with the LC-MS/MS method in wheat grains and wheat malt was performed as previously described [31 (link)]: In brief, 5.00 g of ground wheat was extracted with 20 mL of extraction solvent composed of acetonitrile (AcN):water (W):acetic acid (HAC) = 79:20:1 (v:v:v) on a rotary shaker (GFL 3017, GFL; Burgwedel, Germany) for 90 min at room temperature in a horizontal position. After extraction, 500 µL of the extract was diluted with 500 µL of dilution solvent composed of AcN:W:HAC = 20:79:1 in vials. Finally, 5 µL was injected into an LC-MS/MS system composed of a QTrap 5500 MS/MS (Sciex, Foster City, CA, USA) coupled with an Agilent 1290 series UHPLC system (Agilent Technologies, Waldbronn, Germany). The separation of analytes was performed on a Gemini C18 column (150 × 4.6 mm i.d., 5 µm particle size) with a 4 × 3 mm precolumn with the same characteristics (Phenomenex, Torrance, CA, USA). The eluents used were composed of methanol (MeOH):W:HAC = 10:89:1 (v:v:v) as eluent A, and MeOH:W:HAC = 97:2:1 (v:v:v) as eluent B. The analysis was performed on the fully validated method described in detail by Sulyok et al. (2020) for measurement of 500+ mycotoxins and other secondary metabolites.
+ Open protocol
+ Expand
3

Quantification of Bile Acids by UHPLC-Q-TOF-MS

Check if the same lab product or an alternative is used in the 5 most similar protocols
We performed qualitative and quantitative analyses of the plasma concentration of BA on an Agilent 6530 Accurate-Mass Q-TOF LC/MS mass spectrometer coupled to an Agilent 1290 series UHPLC system (Agilent Technologies, USA), as described elsewhere.45 (link) Sample separation was performed using a Zorbax Eclipse-Plus C18 column (2.1 × 100 mm, 1.8 µm; Agilent Technologies, USA) heated at 50°C.5 mM ammonium acetate (pH 6) in water was used as eluent A and acetonitrile as eluent B in a binary gradient system. The detection in the mass spectrometer was operated in negative ionization mode using the Dual AJS Jet stream ESI Assembly. During acquisition the internal calibration was performed by infusing continuously the reference mass solution 5 mM purine, 1 mM HP-921 (Agilent reference mass kit, Agilent Technologies USA) in 95% MeOH acidified with 0.1% formic acid and allowed to permanently achieve a mass accuracy better than 5 ppm. HR mass spectra were acquired over the range of m/z 300–700 at an acquisition rate of three spectra/s. The data was processed using the Mass Hunter Workstation (Agilent Technologies, USA). Extracted ions chromatograms (XIC) were based on a retention time (RT) window of ±0.5 min with a mass extraction-windows (MEW) of ±30 ppm centered on m/ztheor of each bile acid.
+ Open protocol
+ Expand
4

Serum Metabolomics by UHPLC-MS

Check if the same lab product or an alternative is used in the 5 most similar protocols
The serum samples were measured on an Agilent 1290 series UHPLC system (Agilent Technologies, Santa Clara, USA) coupled with a ZORBAX RRHD 300 SB-C18 column (100 × 2.1 mm, 1.8-μm, Agilent Technologies, Santa Clara, USA) for chromatography and separation. During the analysis, the setting conditions were set as follows: sample maintaining temperature, 4 °C; injection volume: 4 μL; column temperature: 30 °C; flow rate, 0.30 mL/min. The mobile phases were composed as solvent A (0.1% formic acid in acetonitrile), and solvent B (0.1% formic acid in water). The gradient elution was setting as Table 1.

Mobile phases for serum metabolomics analysis

T (min)A (v/v)%B (v/v)%
0–1.0955
1.0–9.095–605–40
9.0–19.060–1040–90
19.0–21.010–090–100
21.0–25.00100

A, 0.1% formic acid in acetonitrile; B, 0.1% formic acid in water

+ Open protocol
+ Expand
5

Intracellular Metabolic Profiling of H9c2 Cardiomyocytes

Check if the same lab product or an alternative is used in the 5 most similar protocols
In this study, an Agilent 1290 series UHPLC system (Agilent Technologies, United States) was used to analyse the intracellular metabolic profile of H9c2 cardiomyocytes. A ZORBAX RRHD 300 SB‐C18 column (100 mm × 2.1 mm, 1.8 μm, Agilent Technologies) was applied for chromatographic separation. Mobile phase A was 0.1% formic acid in acetonitrile, and phase B was 0.1% formic acid in water. The column temperature was kept at 30°C, and the flow rate was 0.30 mL/min. The gradient elution of mobile phase A was set as follows: 0‐1.0 minutes, 95%; 1.0‐9.0 minutes, 95%‐60%; 9.0‐19.0 minutes, 60%‐10%; 19.0‐21.0 minutes, 10%‐0%; 21.0‐25.0 minutes, 0%. All samples were preserved at 4°C with a 4 μL injection volume during the analysis.
Then, mass spectrometry was carried out on an Agilent 6550A Q‐TOF/MS (Agilent Technologies, United States) with electrospray ionization (ESI) mode. The mass range was set from 80 to 1000 m/z, and the gas flow was set as 11 L/min. The conditions in positive ionization mode were set as follows: electrospray capillary voltage, 4.0 kV; gas temperature, 225°C; nebulizer, 45 pisg. And conditions in negative ionization mode were set as follows: electrospray capillary voltage, 3.0 kV; gas temperature, 200°C; nebulizer, 35 pisg. The other parameters were as follows: sheath gas temperature, 350°C; sheath gas flow rate, 12 L/min; nozzle voltage, 500 V.
+ Open protocol
+ Expand
6

UHPLC-MS/MS Characterization of Organic Compounds

Check if the same lab product or an alternative is used in the 5 most similar protocols
HPLC experimental conditions: the atomized gas and dry gas are nitrogen, and the collision gas is helium. The capillary voltage is 3500 eV, the atomization temperature is 350 °C, the dry gas is 10.0 L / min, and the atomized gas is 206.85 kPa. The MS scanning range is m/z 80 ~ 1000, and the data storage mode is the centroid. The ion source was double ESI spray, and the data were collected by mass spectrometry through 2 known standards [Hexakis (1H, 1H, 3H-tetrafluoropropoxy) phosphazine And 7H-purine, corresponding to m/z 922.00980 and 121.0509] for real-time correction. The reference solution was injected into the mass spectrum at a rate of 0.01 ml/min by an Agilentisonic pump. The AutoMS/MS experiment uses CID collision. Liquid chromatography is an Agilent 1290 series UHPLC system. The liquid phase is separated by a reversed-phase column ( Atlantis, T3, 150 m m × 2.1 mm, 3 μm,Waters,IRELAND), Mobi le phase A phase is 0.1% formic acid–water, and phase B is 0.1 percent formic acid- (methanol: acetonitrile = 1: 1), Gradient elution conditions under positive ion conditions: 0–40 min, 92% B; 17–32 min, flow rate 0.3 ml/min, column temperature 35 °C, injection volume 10 μl; Negative ion gradient elution conditions: 0–5 min, 33–35% B; 5–40 min, 35% -65% B, The column temperature was 35 °C and the injection volume was 10 μL.
+ Open protocol
+ Expand
7

UHPLC-MS Metabolite Profiling Protocol

Check if the same lab product or an alternative is used in the 5 most similar protocols
Chromatography was performed using an Agilent 1290 series UHPLC system (Agilent Technologies, Santa Clara, CA, USA) equipped with quaternary pump, online degasser, autosampler, and thermostatted column compartment. The volume of injection samples was fixed at 4 μL. All samples were maintained at 4°C during the analysis. The separation was performed on a ZORBAX RRHD 300 SB-C18 column (2.1 mm × 100 mm, 1.8 μm, Agilent Technologies, Santa Clara, CA, USA). The column temperature was set to 30°C. The mobile phases consisted of 0.1% formic acid in acetonitrile (solvent A) and 0.1% formic acid in water (solvent B). The flow rate was set to 0.30 mL/min. The following gradient was used: a linear gradient of 95% A for the first minute, 95–60% A from 1.0 to 9.0 min, 60–10% A from 9.0 to 19.0 min, 10–0% A from 19.0 to 21.0 min, and 0% A from 21.0 to 25.0 min. The eluent was directly introduced to the mass spectrometer. After the injection of 10 samples, a pooled sample, the QC sample, followed by a blank was injected in order to ensure the stability and repeatability of the LC–MS systems.
+ Open protocol
+ Expand
8

Quantitative Cannabinoid Analysis in Serum

Check if the same lab product or an alternative is used in the 5 most similar protocols
For cannabinoid analysis, 150 µl of ice-cold methanol and 10 µl of isotope-labeled internal standard mixture were added to 50 µl of each individual sample (PBS as blank sample, calibration standards, QC standards, and serum samples) for a 1-h protein precipitation. All the samples were then centrifuged at 13,000 × g for 20 min. For each sample, 180 µl of the supernatant was loaded into the center of corresponding wells of a 96-well deep well plate, followed by the addition of 90 µl of water to each well. The plate was then shaken at 600 rpm for 15 min before LC–MS/MS analysis. Mass spectrometric analysis was performed on an AB Sciex 5500 QTRAP® tandem mass spectrometry instrument (Applied Biosystems/MDS Analytical Technologies, Foster City, CA) equipped with an Agilent 1290 series UHPLC system (Agilent Technologies, Palo Alto, CA). Data analysis was done using Analyst 1.6.3.
+ Open protocol
+ Expand
9

Quantitative Analysis of Mycotoxins by UHPLC-MS/MS

Check if the same lab product or an alternative is used in the 5 most similar protocols
An Agilent 1290 series UHPLC system coupled to a 6460 Triple Quadrupole (QqQ) mass spectrometer (both Agilent Technologies, Waldbronn, Germany) was used to analyze the samples. Precursor and product ion selection as well as the optimization of collision energies were performed with flow injection of single-analyte solutions. UHPLC separations were performed in a reversed-phase C18 analytical column of 50 mm × 2.1 mm and 1.8 μm particle size (ZORBAX RRHD Eclipse Plus C18) by Agilent Technologies (Waldbronn, Germany).
The chromatographic solvents were water 0.1% formic acid solution (eluent A) and acetonitrile 0.1% formic acid (eluent B). The gradient program was as follows: 0.0 min, 10% B; 2.4 min, 42% B; 6.0 min, 51% B; 6.2 min, 95% B; 7.0 min, 10% B. A subsequent re-equilibration time (5 min) should be performed before next injection. The constant flow rate was 0.3 mL/min while the injection volume was 2 μL. Moreover, the column temperature was maintained at 30 °C.
MS/MS analyses of mycotoxins were performed on an 6460 QqQ mass spectrometer with Agilent Jet Stream Technology under the dynamic multiple reaction monitoring (DMRM) conditions in ESI+. The following settings were used: nebulizer, 40 psi; drying gas temperature, 350 °C; drying gas flow, 10 L/min; capillary voltage, 4000 V.
+ Open protocol
+ Expand
10

Quantitative UHPLC-MS/MS Determination of DHAS

Check if the same lab product or an alternative is used in the 5 most similar protocols
The LC separations were performed on an Agilent 1290 series UHPLC system (Agilent Technologies, USA). A Waters ACQUITY UPLC® BEH-C18 column (2.1 mm × 100 mm, 1.7 μm) at 40°C was used to separate DHAS and IS with a mobile phase gradient of acetonitrile (A) and 0.01% formic acid in ultra-pure water (B): 10% A (0–1.0 min), 10–80% A (1.0–4.0 min), and 80–10% A (4.1–6.5 min). The total run time was 6.5 min at a flow rate of 0.4 mL/min and the injection volume was 1 μL.
The API 4000 triple-stage quadrupole mass spectrometer (AB Sciex, USA) equipped with a Turboionspray™ interface was detected using electrospray ionization (ESI) in the negative ionization mode, and the multiple reaction monitoring mode was used to quantify. The optimized conditions were as follows: Ion source gas 1 (GS 1), 60 psi; ion source gas 2 (GS 2), 60 psi; curtain gas, 25 psi; source temperature, 500°C; capillary voltage, −4500 V; entrance potential, −10 V; and collision cell exit potential, −11 V. The precursor–product ion pairs were: m/z 428.9→96.0 (DHAS) and m/z 320.9→151.8 (chloramphenicol, IS). The de-clustering potential and collision energy were −80 and −80 eV for DHAS and −71 and −23 eV for IS. The instruments controlled and data collected were acquired using Analyst® software (version 1.5.2, Applied Biosystems, CA, USA).
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!