The largest database of trusted experimental protocols

In10 mx

Manufactured by Thermo Fisher Scientific
Sourced in United States

The IN10 MX is a compact infrared microscope designed for the analysis of small samples. It provides high-quality infrared spectroscopic imaging capabilities for a variety of applications.

Automatically generated - may contain errors

Lab products found in correlation

5 protocols using in10 mx

1

Characterization of Synthesized Graphene Oxide

Check if the same lab product or an alternative is used in the 5 most similar protocols
Dynamic light scattering (DLS) measurement was performed on a Zetasizer (BIC Omni) to determine the zeta potential of synthesized GO in the solution. The surface morphology of the deposited sample was observed by scanning electron microscopy (SEM, MIRA3 TESCA). The X-ray photoelectron spectra (XPS) were obtained on Nexsa ThermoFisher scientific using a focused monochromatic aluminium Kα X-ray (1486.6 eV) source and a spherical section analyzer. The X-ray beam used was 150 W and the sample area was 300 μm × 700 μm. Wide-scan data were collected using a pass energy of 160 eV. The infrared spectrum was performed using iN10 MX (Thermo Fisher) and GO film was tested by ATR-FTIR (Zr single crystal). Raman spectrum was obtained with a confocal Raman microscope (Renishaw inVia Qontor, England). Materials were irradiated with light from a solid-state laser at 532 nm and the power of the excitation laser is 10 mW to avoid lase-induced heating. The surface roughness was measured by atomic force microscope (AFM, FastScan Bio Bruker America tip is AC240, 160). Contact angle test was performed by using diluted GO colloidal to detect the change of surface affinity by digital camera. Finally, the square resistance was measured by a multifunction digital four-probe tester (ST2258C, Suzhou Jingge, China). Detailed steps are listed in the ESI.
+ Open protocol
+ Expand
2

FTIR Mapping of Organic Molecules

Check if the same lab product or an alternative is used in the 5 most similar protocols
FTIR measurements were carried out at the Institut de Physique du Globe de Paris, France with a Thermo Scientific iN10 MX® (Ever-Glo infrared source) equipped with a ×15 objective (NA = 0.7) and a liquid nitrogen cooled MCT-A detector in transmission and reflection mode. We acquired spectra in the range 4000–710 cm-1 by mapping over areas of several mm2 with individual spot sizes of 30×30 μm and steps of 30 μm. Each spectrum corresponded to 128 summed scans collected with a spectral resolution of 4 cm-1. The raw data was converted into–ASCII files with the Omnic Atlus® imaging software by Thermo ScientificTM and afterwards normalized and background-reduced using the ChemoSpec package for the software R [38 ].
The vibrational bands summarized in Table 1 were considered as indicative of bonds or functional groups in organic molecules, which potentially originate from biological material [39 ;40 (link)].
+ Open protocol
+ Expand
3

Quantifying PCBs in Microplastics

Check if the same lab product or an alternative is used in the 5 most similar protocols
Polyethylene (PE, green
fluorescent UVPMS-BG,
spherical, diameter 10–180 μm, density 0.94 kg/L)24 (link) was used in the bioassay. PE polymer identity
was confirmed by FTIR (ThermoFisher, iN10 MX). For microscope images
and particle size distributions of the PE the reader is referred to
the publication by Velzeboer et al.24 (link) Polyoxymethylene
sheets (POM, 76 μm thickness) were employed as passive samplers.23 (link),24 (link),29 (link) The selected PCB congeners were
28, 31, 44, 52, 101, 118, 138, 153, 170, and 180. Further details
are provided in the Supporting Information (SI).
+ Open protocol
+ Expand
4

Spectroscopic Analysis of Compound 1

Check if the same lab product or an alternative is used in the 5 most similar protocols
The mid infrared spectral information of Compound 1 was collected by a microscopic imaging infrared spectrometer (iN10 MX, Thermo, USA) with the spectral range of 4000-450 cm−1. The spectral data of 1H-NMR, 13C-NMR, 1H-1H shift correlation spectroscopy (COSY), heteronuclear single quantum coherence (HSQC) and heteronuclear multiple bond connectivity (HMBC) were obtained by a 700MHz nuclear magnetic resonance spectrometer (AVANCE NEO, Bruker, Billerica, MA, USA) with the scanning range 1H: −1–13 ppm, 13C: −12–230 ppm.
+ Open protocol
+ Expand
5

FTIR Analysis of Avian Amyloid Fibrils

Check if the same lab product or an alternative is used in the 5 most similar protocols
A FTIR microscope iN10MX (Thermo Scientific, Waltham, MA, USA) was used for the analysis. Molecular structures were identified by infrared absorption bands. The spectra obtained from five analyzed areas (50 x 50 μm) of the AAF were compiled and compared with each other. The analyses were carried out at room temperature directly on the sample surface using the reflection technique at the spectral range 650–4000 cm-1 with a resolution of 4 cm-1.
The FTIR spectra of the AAF and egg white lysozyme (Fluka) were compared using a FTIR Nicolet 8700A spectrometer (Thermo Scientific, Waltham, MA, USA). The analyses were carried out at room temperature directly on the surface of the sample by ATR with a diamond crystal at a spectral range of 400–4000 cm-1 with a resolution of 4 cm-1.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!