The largest database of trusted experimental protocols

7 protocols using acquity uplc sample manager

1

Sensitive LC-MS/MS Quantification of Triptolide

Check if the same lab product or an alternative is used in the 5 most similar protocols
The LC/MS/MS method was conducted as described in Reference [12 ]. The method was carried out using a Waters Acquity UPLC Sample Manager and a Waters Acquity UPLC Binary Solvent Manager connected to a Waters Quattro Premier XE triple-quadruple mass spectrometer and Mass Lynx 1.5.2 software (Waters, Milford, MA, USA). An Agilent Zorbax XDB-C18 (3.5 mm, 2.1 mm × 50 mm) was also used. The gradient mode was used to achieve triptolide and internal standard using mixtures of 40 mM ammonium formate (mobile phase A) and methanol with 0.1% (v/v) formic acid (mobile phase B) at a flow rate of 0.5 mL/min. Mobile phase A was 98% at 0.1–0.6 min, then a linearly programmed gradient from 98% to 5% at 3–4.5 min, at last, mobile phase A ramp to 98% at 4.5–6 min toward the end of the analysis. The system was auto-injected with 10 μL of each analyte. Multiple reaction monitoring (MRM) mode was applied to achieve quantification with monitoring precursor-product ion transitions of m/z 361.0→105.0 for triptolide using electrospray ionization mass spectrometry (ESI-MS) on an API-4000, with Turbo Ionspray. The system was in positive ESI-mode during the run. The desolvation temperature was set at 550 °C and the nebulizer gas pressure was 60 psi.
+ Open protocol
+ Expand
2

Sensitive UPLC-MS/MS Quantitation of Fentanyl

Check if the same lab product or an alternative is used in the 5 most similar protocols
Fentanyl in plasma was quantitated using a validated UPLC-MS/MS method consisting of a Waters Acquity UPLC sample manager coupled to a triple quadruple mass spectrometer operating in the multiple reaction monitoring mode (MRM) with positive ion electrospray ionization (Waters, Etten-Leur, The Netherlands). The multiple reaction monitoring transition was set at 337→188.
Chromatographic separation of fentanyl was achieved on an Acquity UPLC® BEH C18 1.7 μm 2.1 x 100 mm column eluted at a flow-rate of 0.350 mL/min on a gradient of methanol. The overall cycle time of the method was 6 minutes. The calibration curves were linear over the range of 0.1 to 10 ng/mL with the lower limit of quantitation validated at 0.1 ng/mL for fentanyl. The within and between-run precisions at five tested concentrations, including the LLQ, were ≤ 5.5% and ≤ 6.1%, respectively, while the average accuracy ranged from 86.2% to 97.5%. The extraction of 200 μL of plasma involved a deproteinization step with acetone followed by a simple liquid extraction with ethyl acetate. The organic phase was evaporated and subsequently dissolved in 100 μL methanolic solutions, from which aliquots of 10 μL were injected into the UPLC-MS/MS system.
+ Open protocol
+ Expand
3

Quantification of Oxycodone and Metabolites

Check if the same lab product or an alternative is used in the 5 most similar protocols
Plasma concentrations of oxycodone and its metabolites, oxymorphone, nor-oxycodone and nor-oxymorphone were quantified using a validated UPLC-MS/MS method consisting of a Waters Acquity UPLC sample manager coupled to a triple quadrupole mass spectrometer operating in the multiple reaction monitoring mode (MRM) with positive ion electrospray ionization (Waters, Etten-Leur, The Netherlands). Chromatographic separations were achieved on an Acquity UPLC® (Waters, Etten-Leur, The Netherlands) BEH C18 1.7 µm 2.1 × 100 mm column eluted at a flow rate of 0.350 mL/min on a gradient of methanol combined with ammonium formate. The overall cycle time was 10 min. The calibration curves were linear with the lower limit of quantitation validated at 0.200 ng/mL for oxycodone and oxymorphone and 1.00 ng/mL for nor-oxycodone and nor-oxymorphone.
+ Open protocol
+ Expand
4

Fentanyl Quantification in EDTA Plasma

Check if the same lab product or an alternative is used in the 5 most similar protocols
We quantified fentanyl in EDTA plasma. A validated UPLC-MS/MS method. This method consisted of a Waters Acquity UPLC sample manager, coupled to a triple quadruple mass spectrometer operating in the multiple reaction monitoring mode (MRM) with positive ion electro spray ionization (Waters, Etten-Leur, The Netherlands). The multiple reaction monitoring transitions was set at 337→188.
Chromatographic separations for fentanyl were achieved on an Acquity UPLC® BEH C18 1.7 μm 2.1 × 100 mm column eluted at a flow-rate of 0.350 mL/min on a gradient of methanol. A cycle time for this method was about 6 minutes. Calibration curves were linear over a wide range (0.100 to 10.0 ng/mL) with at lower limit of quantitation (LLQ) of 0.100 ng/mL for fentanyl. The within and between-run precisions, including the LLQ, were ≤ 5.52 % and ≤ 6.12 %, respectively, while the average accuracy ranged from 86.2 % to 97.5%. The extraction of 200 μL of plasma involved a deproteinization step with acetone, followed by a simple liquid extraction with ethyl acetate. The organic phase was evaporated and subsequently dissolved in 100 μL methanolic solutions, from which aliquots of 10 μL were injected into the UPLC-MS/MS system.
+ Open protocol
+ Expand
5

UPLC/MS/MS Analysis of Metabolites

Check if the same lab product or an alternative is used in the 5 most similar protocols
The UPLC/MS/MS method was conducted using a Waters Acquity UPLC Sample Manager and a Waters Acquity UPLC Binary Solvent Manager connected to a Waters Quattro Premier XE triple-quadrupole mass spectrometer equipped with a combined ESI probe and Mass Lynx 1.4 software (Waters, MA, USA). An Acquity UPLC BEH C18 column (1.7 µm, 2.1 × 100 mm2) was used. Instrument settings were as follows: ESI+; source temperature, 110 °C; desolvation temperature, 350 °C; capillary voltage, 3.2 kV; desolvation N2, 600 L/h; cone N2, 50 L/h; LM resolution 1, 13.5; and HM resolution 1, 13.5. The data acquisition was carried out at m/z 50–500 Da in the total ion scan mode. The MS/MS spectra were acquired at m/z 50–300 Da in the daughter ion scan mode. The solvent system consisted of solvent A (0.1% CH3COOH) and solvent B (100% acetonitrile). The flow rate was 0.2 mL/min. The use of different HPLC gradients of 10–80% B for 16 min for imperatorin and isoimperatorin after exposure to human liver microsomes was necessary to separate the metabolites from the parent compound.
+ Open protocol
+ Expand
6

UPLC Separation and Detection of Vitamin K

Check if the same lab product or an alternative is used in the 5 most similar protocols
The UPLC system consisted of the ACQUITY UPLC sample manager and a binary solvent manager (Waters). Sample separation was performed using a VanGuard BEH C18 (1.7 µm) as the precolumn (Waters) and an ACQUITY UPLC BEH C18 (1.7 µm, 2.1 mm × 100 mm) column (Waters) as the main column. Vitamin K1 and MK-7 were detected with the ACQUITY UPLC fluorescent detector (Waters) after post-column reduction using CQ-R 2.0 × 20 mm column (OSAKA SODA, Osaka, Japan). The column temperature was maintained at 50 °C during analysis. The mobile phase was a mixture of Milli-Q water (solvent A) and liquid chromatography-grade methanol (solvent B). The extraction and emission wavelengths of the fluorescence detector were set to 244 nm and 430 nm, respectively.
+ Open protocol
+ Expand
7

UPLC-MS/MS Analysis of Vitamin K

Check if the same lab product or an alternative is used in the 5 most similar protocols
The UPLC system consisted of the ACQUITY UPLC sample manager and binary solvent manager (Waters, MA, USA). Separations were performed with a VanGuard BEH C18 (1.7 μm) as the precolumn (Waters) and an ACQUITY UPLC BEH C18 (1.7 μm, 2.1×100 mm) Column (Waters) as the main column. Vitamin K1 and menaquinone 7 (as internal standard) were detected with the XEVO TQ-XS tandem quadrupole mass spectrometer (Waters). The sample temperature was kept at 4°C, and the column temperature was kept at 50°C. The mobile phase was a mixture of MilliQ water (solvent A) and liquid chromatography-grade methanol (Nacalai Tesque) (solvent B). The UPLC conditions are shown in Supplementary Table 1 and the mass spectrometer conditions are shown in Supplementary Table 2. Data analyses were performed using MassLynx NT software version 4.1 (Waters).
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!