The largest database of trusted experimental protocols

Frontier ftir nir spectrometer

Manufactured by PerkinElmer
Sourced in United States

The Frontier FTIR/NIR spectrometer is a versatile laboratory instrument designed for spectroscopic analysis. It utilizes Fourier-transform infrared (FTIR) and near-infrared (NIR) spectroscopy techniques to collect and analyze data. The core function of this spectrometer is to measure the absorption or transmission of infrared light by a sample, providing information about its chemical composition and molecular structure.

Automatically generated - may contain errors

7 protocols using frontier ftir nir spectrometer

1

Comprehensive Characterization of Material Samples

Check if the same lab product or an alternative is used in the 5 most similar protocols
The morphology images of samples were taken with a scanning electron microscopy (SEM) (S-4800; Hitachi Ltd., Tokyo, Japan) and dimensioned using ImageJ software (ImageJ; version 1.52a, National Institutes of Health, Bethesda, MD, USA). Crystallinity analysis and phase identification were carried out by infrared spectrum (FTIR) (Frontier FTIR/NIR spectrometer; PerkinElmer Inc., Waltham, MA, USA) in the 4000–400 cm−1 wavenumber range using the KBr pellet method, and X-ray diffraction (XRD) (D8 Advance-Bruker X-Ray Diffractometer; Bruker AXS GmbH, Karlsruhe, Germany) with Cu Kα radiation (λ = 1.54178Å), at a scanning rate of 5°/min in the 2θ range from 5° to 70°. Thermogravimetric analysis (TGA) was performed to determine the purity of the phases using a thermal analysis system (TGA-DSC 3+; Mettler Toledo Inc., Columbus, OH, USA) from 0–300 °C at a heating rate of 10 °C /min in the air.
+ Open protocol
+ Expand
2

FTIR Analysis of Protein Structure

Check if the same lab product or an alternative is used in the 5 most similar protocols
FTIR samples were prepared the same as aforementioned in LV-SEM. Their secondary
structures were determined with a Fourier transform infrared spectroscope
(Frontier FT-IR/NIR Spectrometer, PerkinElmer, Waltham, MA, USA). Scanning
wavelength ranged from 4,000 to 400 cm−1.
+ Open protocol
+ Expand
3

Fourier Transform Infrared Spectroscopy of Powders

Check if the same lab product or an alternative is used in the 5 most similar protocols
FT-IR spectra were recorded using the Diffuse Reflectance Infrared Fourier Transform (DRIFT) technique by means of a Perkin Elmer Frontier FT-IR/NIR spectrometer (Perkin Elmer, Waltham, MA, USA).
Approximately 2 mg of powder was weighed and ground with 200 mg of KBr in an agate mortar and then transferred to the sample holder cup. KBr and samples must be perfectly dry (since water can also be measured by IR, and then we would obtain false peaks, which might mask the peaks of other functional groups), so they must be dried at 105 °C. The homogeneous pale gray samples and pure KBr were placed in the sample holder, and then placed in the device. First, the spectrum of KBr was measured, and then the given sample was scanned. All spectra were recorded over the range of 4000–600 cm−1 at a resolution of 1 cm−1.
+ Open protocol
+ Expand
4

Oat Flour NIR Spectroscopy Protocol

Check if the same lab product or an alternative is used in the 5 most similar protocols
The NIR spectrum of the oat flour sample was measured by using Frontier FT-IR/NIR Spectrometer (PerkinElmer Inc., Waltham, MA, USA) using a NIR infrared reflectance accessory—NIRA (PerkinElmer Inc., Waltham, MA, USA) equipped with a spinning sample module and Spectrum-V.7.3.1.1023 (PerkinElmer Inc., Waltham, MA, USA) software. The spectrum was recorded in diffuse reflectance mode from 700 to 2500 nm using 32 scans at 1 nm intervals. The measurements were performed in duplicate, and the average spectrum was used for further analysis.
+ Open protocol
+ Expand
5

Protein Structural Analysis via FTIR

Check if the same lab product or an alternative is used in the 5 most similar protocols
Fourier-transform infrared spectroscopy (Frontier-FTIR/NIR Spectrometer, PerkinElmer, MA, USA) was used to quantify secondary structures in the protein backbone. Scanning was conducted at wavelengths ranging from 4000 to 400 cm−1. Graphs were created by arranging the wavelength range from 1450 to 1750 cm−1, representing the secondary structure of the protein matrix.
+ Open protocol
+ Expand
6

FTIR and NIR Spectroscopic Characterization of CNT Films

Check if the same lab product or an alternative is used in the 5 most similar protocols
The FTIR measurements were performed
using a PerkinElmer Frontier FTIR/NIR spectrometer. All of the measurements
were performed in the mid-IR spectral range (7800–400 cm–1, the optimum scan range) in transmission mode. Thin-film
NIR (15000–2000 cm–1, the optimum scan range)
was also performed for stitching to give a broader picture of the
CNT film optical density or −log(T). Stitching
in Figure 3a was simply
done by matching absorbance at the edge of each respective spectrum
by normalization.
+ Open protocol
+ Expand
7

ATR-FTIR Characterization of Powder Samples

Check if the same lab product or an alternative is used in the 5 most similar protocols
ATR-FTIR spectra of all powder samples were measured directly without any pretreatment. A Frontier FT-IR/NIR spectrometer with a universal ATR sampling accessory (PerkinElmer, Waltham, MA, USA) was used to collect the spectra. The internal reflection element (IRE) of the ATR accessory was a diamond/ZnSe composite crystal and the angle of incidence was 45°. The powder sample was placed on the top surface of the IRE crystal and pressed by a lever to ensure tight and consistent contact. Each spectrum was the average of 16 scans in the range of 4000–650 cm−1 with a spectral resolution of 4 cm−1. The influence of water vapor and carbon dioxide was subtracted automatically.
The software Spectrum v10.4.3 (PerkinElmer, USA) was used to control the spectrometer and process the spectra. First, the spectral ordinate was transformed into absorbance. Next, ATR correction with a contact factor of zero and automatic baseline correction were applied. Finally, the spectrum was normalized to set the maximum and minimum absorbance in the range of 1800–800 cm−1 at 1 and 0, respectively. Triplicates of each sample were measured and the average spectrum was used.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!