The largest database of trusted experimental protocols

4800 maldi tof tof mass spectrometer

Manufactured by AB Sciex
Sourced in United States, France

The 4800 MALDI-TOF/TOF mass spectrometer is a high-performance analytical instrument designed for the analysis of biomolecules. It utilizes matrix-assisted laser desorption/ionization (MALDI) and tandem time-of-flight (TOF/TOF) mass spectrometry techniques to provide accurate and sensitive measurements of molecular masses and structural information.

Automatically generated - may contain errors

20 protocols using 4800 maldi tof tof mass spectrometer

1

Nanoflow LC-MALDI-TOF/TOF Peptide Analysis

Check if the same lab product or an alternative is used in the 5 most similar protocols
MS and MS/MS analysis of nanoLC-off-line spotted peptide samples was performed using the 4800 MALDI-TOF/TOF mass spectrometer (Sciex, Les Ulis, France) controlled by the 4000 Series Explorer software v. 3.5. The mass spectrometer was operated in a positive reflector mode. Each spectrum was externally calibrated using the Peptide Calibration Standard II (Bruker Daltonics, Bremen, Germany), and the peptide mass tolerance was set to 50 ppm. MS spectra were acquired in a m/z 700–4000 range. Up to 30 of the most intense ions per spot position characterized by a S/N (signal/noise) ratio higher than 40 were chosen for MS/MS analysis. Selected ions were activated by using CID (collision-induced dissociation) in order to obtain the corresponding MS/MS spectrum.
+ Open protocol
+ Expand
2

Mass Spectrometry-based Proteomics Workflow

Check if the same lab product or an alternative is used in the 5 most similar protocols
Spectra were acquired on a 4800 MALDI TOF/TOF mass spectrometer (Sciex) fitted with a UV laser (wavelength: 337 nm) in reflectron mode. All MS spectra of the tryptic digests were externally calibrated using a mixture of peptide standards. MS/MS spectra were acquired on selected ions of interest under the following conditions: precursor isolation set to resolution of 200, collision energy of 2 kV, CID cell pressure of 2 × 10–5 torr, air as collision gas. The instrument was calibrated in MS/MS mode using five daughter ions (m/z: 175.119, 684.346, 813.389, 1056.475 and 1441.634) generated from the fragmentation of Glu-fibrino peptide (m/z: 1570.6774). Data were analysed comparing the monoisotopic peaks with the theoretical values corresponding to the expected peptide digestion products. Peak lists were searched with Mascot Server using the following parameters: enzyme = trypsin; maximum missed cleavages = 2; variable modifications = carbamidomethylation of cysteine, oxidation of methionine; mass tolerance = 50 ppm. The SwissProt database (homo sapiens, human) was used as reference protein sequence database.
+ Open protocol
+ Expand
3

MALDI-TOF MS Analysis of Venom Peptides

Check if the same lab product or an alternative is used in the 5 most similar protocols
MALDI-TOF MS analyses were performed using a 4800 MALDI TOF/TOF™ mass spectrometer (Sciex, Les Ulis, France) in a linear positive middle mass mode (m/z 2000-14,000, mass tolerance 5 m/z units). The reduced/alkylated venom peptide was mixed with 0.5 µL of sinapinic acid at 10 mg/mL and spotted in duplicates onto MALDI Opti-TOF 384-well plate (Sciex, Les Ulis, France).
+ Open protocol
+ Expand
4

MALDI-TOF/TOF Mass Spectrometry Protocol

Check if the same lab product or an alternative is used in the 5 most similar protocols
MS and MS/MS analyses were performed using the 4800 MALDI-TOF/TOF mass spectrometer (Sciex, Les Ulis, France) controled by the 4000 Series Explorer software v. 3.5. For LMW color pI marker analysis, MS spectra were acquired in positive reflector ion mode in an m/z range of 250-650. MS/MS spectra were performed manually for each LMW color pI marker ion by using collision-induced dissociation (CID) as the activation mode. For peptide analysis, MS spectra were acquired in positive reflector ion mode in an m/z range of 700-4000. The 30 most intense ion signals per spot position having an S/N >20 were selected as precursor ions for MS/MS analysis. MS/MS analyses were performed automatically byusing CID as the activation mode.
+ Open protocol
+ Expand
5

Identifying Co-eluted Compound from RBC AChE

Check if the same lab product or an alternative is used in the 5 most similar protocols
The goal was to identify a brown substance that co-eluted with RBC AChE from Hupresin in 1% TFA. A 1 μL aliquot was applied to a MALDI plate with 2,5-dihydroxybenzoic acid (DHB) matrix. Spectra were acquired in positive mode using a laser voltage of 5,000 on a 4,800 MALDI-TOF/TOF mass spectrometer (AB Sciex).
+ Open protocol
+ Expand
6

Proteomic Identification of Murine Sera

Check if the same lab product or an alternative is used in the 5 most similar protocols
The protein identification and characterization analysis of the murine sera have been done according45 (link),84 (link). Briefly, the mass spectra were recorded in the positive reflectron mode on a 4800 MALDI-TOF/TOF mass spectrometer (AB Sciex, Foster City, CA, USA) and operated in its delayed extraction mode. Internal calibration of mass spectra was conducted utilizing the tryptic autolytic peptides. Acquired data were evaluated using GPS Explorer Software v.3.6 (AB Sciex, Australia) (https://sciex.com/products/software), which integrates the Mascot search algorithm against the Francisella tularensis subsp. holarctica OSU18 genome database. The following parameters were used for protein identification: 100 ppm error, optional oxidation of methionine, fixed carbamidomethylation of cysteine residues, and one possible missed cleavage site allowed. Proteins were considered to be identified with confidence when a minimum of two peptide sequences per protein were identified and the GPS protein score confidence interval was equal to 100%. Trypsin was selected as the proteolytic enzyme. The identities of all proteins were determined according to the UniProtKB database85 (link) (https://www.uniprot.org).
+ Open protocol
+ Expand
7

Quantifying OP-Peptide Conjugation to Proteins

Check if the same lab product or an alternative is used in the 5 most similar protocols
The number of OP-peptides crosslinked to the carrier proteins BSA, human serum albumin (HSA), ovalbumin (OVA) and lysozyme was quantified from the shift in mass of the carrier protein observed in the MALDI-TOF mass spectrometer, as described by Belabassi et al.7 Proteins were spotted onto a MALDI target plate with sinapinic acid. Mass data were acquired in a 4800 MALDI-TOF/TOF mass spectrometer (AB Sciex) set to linear high mass positive mode, range 30,000 to 100,000 Da, bin size 20 ns, detector voltage multiplier 0.95, delayed extraction 1650 nsec, low mass gate enabled, low mass gate offset 100 Da, collected 3000 shots with laser voltage set to 7800 V. The list of OP-tyrosine peptides conjugated to BSA, HSA, OVA and lysozyme is in Table S1 in the Supporting Information section. Samples are in 0.1 M MES buffer pH 6.2 at a concentration of 2.5 mg protein per mL.
+ Open protocol
+ Expand
8

Cross-Linking Mass Spectrometry of Chromatin Proteins

Check if the same lab product or an alternative is used in the 5 most similar protocols
Protamine- and histone-NPM2 complexes, prepared as described in Section 2.5, were extensively dialyzed against 100 mM NaCl, 10 mM PIPES (pH 7.3), and 0.1 mM EDTA buffer for 4 h at 4 °C. A minimum concentration of 30 μg of protein was used for each CXMS experiment. Two samples, NPM2+histone octamers (NPM2-histone run) and NPM2+protamines (NPM2-protamine run) were subjected to CXMS in two independent runs. The pH of each mixture was adjusted to 8.0-8.5 by the addition of 0.2 M Na2HPO4, and proteins were crosslinked with 0.05 mM cyanurobiotindipriopyonylsuccinimide (CBDPS) (Creative Molecules Inc.) for 30 min at 25 °C [17] . Reactions were quenched with ammonium bicarbonate at a final concentration of 30 mM, and cross-linked proteins were then digested with sequencing-grade trypsin (Promega, Madison, WI) overnight at 37 °C at a 10:1 substrate:enzyme ratio. Resulting peptide mixtures were affinity-purified using monomeric avidin-agarose beads (Thermo Scientific, Rockford, IL) and analysed using nano-LC-MALDI-MS/MS via the Eksigent® 1D nano-LC system (AB/Sciex, Dublin, CA), Dionex® Probot spotter (Thermo Scientific, Rockford, IL) and the 4800 MALDI-TOF/TOF mass spectrometer (AB/Sciex, Foster City, CA). Data was analyzed with the DXMSMS Match program [22] of ICC-CLASS [23] .
+ Open protocol
+ Expand
9

Protein Identification by Mass Spectrometry

Check if the same lab product or an alternative is used in the 5 most similar protocols
mAb150-immunoprecipitated proteins were resolved in 12.5% SDS-PAGE gel stained with mass-spectroscopy compatible Coomassie stain (Pierce, CA, USA). Individual protein bands (based on molecular weight) were excised, digested, desalted and purified for acquisition of spectra a 4800 MALDI-TOF/ TOF mass spectrometer (AB Sciex, Framingham, MA) and protein identification performed using MASCOT (version 2.1; http://www.martixscience.com) on search engine against the SwissProt database as described earlier [16] (link).
+ Open protocol
+ Expand
10

Protein Molecular Weight Determination

Check if the same lab product or an alternative is used in the 5 most similar protocols
Molecular weight was determined with sulphate-polyacrylamide gel electrophoresis (SDS-PAGE) and mass spectrometry (MS) as described previously [20 (link)]. Briefly, cell lysate or purified fusion protein was mixed with loading buffer and boiled for 3–5 min, then loaded onto a 10% (w/v) polyacrylamide gel (Sigma, St. Louis, MO, USA) and stained using Coomassie Brilliant Blue. Molecular weight was qualitatively analysed by referring to protein molecular weight standards. Quantitative analysis of molecular weight was performed using a 4800 MALDI-TOF/TOF mass spectrometer (AB SCIEX, Foster City, CA, USA).
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!