The largest database of trusted experimental protocols

Gpc system

Manufactured by Agilent Technologies
Sourced in United States

The GPC system is a lab equipment product that is used for gel permeation chromatography. It separates molecules based on their size and molecular weight, providing information about the molecular weight distribution of polymers and other macromolecules.

Automatically generated - may contain errors

5 protocols using gpc system

1

Characterization of F-POSS and Copolymers

Check if the same lab product or an alternative is used in the 5 most similar protocols
Fourier transform infrared (FT-IR) spectrum of F-POSS and F-POSS-Acrylate copolymer were measured by a TENSOR 27 FT-IR spectrometer (Bruker, Karlsruhe, Germany) from 400 to 4000 cm−1. The samples were prepared by following the KBr pellet method. 29Si NMR and 1H NMR spectrum of 7F-MAP-POSS before and after copolymerization process were recorded on a 500MHz AVANCE III spectrometer (Bruker, Karlsruhe, Germany) using CDCl3 as a solvent. A gel permeation chromatography (GPC) system (Agilent, Palo Alto, CA, USA) was calibrated with narrow molecular weight polystyrene standards. The resulting calibration curves were used to estimate the molecular weights and polydispersity (PD) of 7F-MAP-POSS-Acrylate copolymers. THF was introduced as the mobile phase at a flow rate of 1.0 mL/min. The concentrations of samples were 5.0 mg × mL−1, and the injection volume was 100 μL for each analysis. The columns and detector were maintained at 40 °C. Contact angles (CA) were measured using SL200KS contact angle analyzer (KINO, Boston, MA, USA) that had a CCD camera equipped for image capture.
Scanning electron microscopy (SEM) images and energy dispersive spectrum (EDS) were taken using a JSM-6700F scanning electron microscope (JEOL, Tokyo, Japanese). The surface morphology of the films was observed after gold was sputtered onto the surfaces.
+ Open protocol
+ Expand
2

Molecular Weight Characterization of Polymers

Check if the same lab product or an alternative is used in the 5 most similar protocols
Size-exclusion chromatograms were recorded using a GPC system from Agilent Technologies with its corresponding Agilent software, dedicated to multi-detector GPC calculation. The system used two PL1113-6300 ResiPore 3 µm 300 × 7.5 mm columns with DMF as the eluent with a flow rate of 1 mL/min and toluene as flow rate marker. The detector suite comprised a PL0390-06034 capillary viscometer, and a 390-LC PL0390-0601 refractive index detector. The entire SEC-HPLC system was thermostated at 35 °C. Low dispersity PMMA standards were used for the calibration. The typical sample concentration was 10 mg/mL. The molar masses and dispersity of the polymers were measured using universal calibration.
+ Open protocol
+ Expand
3

Oligomer Molecular Weight Analysis

Check if the same lab product or an alternative is used in the 5 most similar protocols
Molecular weights of the oligomers were measured using the GPC system (Agilent Technology, Santa Clara, CA, USA) and were calculated in comparison to polystyrene standards (PSS). The molecular weights were stated as number average molecular weight (Mn), weight average molecular weight (Mw), and dispersity value (Ð) [8 (link)].
+ Open protocol
+ Expand
4

Characterization of Polyethylene Glycol Amphiphiles

Check if the same lab product or an alternative is used in the 5 most similar protocols
The 1H and 13C NMR spectra were recorded on JEOL 400 MHz and JEOL 100.5 MHz spectrometers (Tokyo, Japan) respectively; the solvent residual peak was used for referencing. The chemical shift values are on a δ scale, and the coupling constant values (J) are in Hertz. The HRMS data were collected on a Waters (Micromass) LCT1 using the ESI-TOF MS technique and Agilent 6210 ESI-TOF, Agilent Technologies, Santa Clara, CA, USA. Infrared spectra (IR) were recorded on a Perkin-Elmer FT-IR model 9 spectrometer. Molecular weight analysis M¯w , M¯n and the Polydispersity Index (PDI) of polyethyleneglycol (PEG) unit containing amphiphiles was done using an Agilent GPC system equipped with an Agilent 1100 pump and a refractive index detector, using PLgel columns, with THF as an eluent at a flow rate of 1.0 mL min−1 using polystyrene standards.
+ Open protocol
+ Expand
5

Molecular Weight Determination via GPC

Check if the same lab product or an alternative is used in the 5 most similar protocols
Molecular weight (MW) was measured with an Agilent GPC system equipped with a Sephadex LH-20 column. The column was equilibrated with running buffer (0.1 M citrate pH 4.5 and 0.025% sodium azide). For the measurement, the column was loaded with 200 μL sample, at a concentration of 1 mg/mL in running buffer, and run for 20 minutes at constant flowrate of 5 mL/h at 20 °C. The change in refractive index was monitored with a Wyatt MALS laser light scattering instrument, and the values were used to generate the dn/dc standard curve. The MW was calculated with Astra software.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!