Lct premier xe tof ms

Manufactured by Waters Corporation
Sourced in United States

The LCT-Premier XE TOF-MS is a high-performance time-of-flight mass spectrometer designed for accurate mass measurement and analysis. It features a high-resolution time-of-flight analyzer and provides precise mass determination for a wide range of applications.

Automatically generated - may contain errors

Lab products found in correlation

2 protocols using lct premier xe tof ms

1

Characterizing Pathogen Box Compounds

Check if the same lab product or an alternative is used in the 5 most similar protocols
The molecules were provided diluted in DMSO at 10mM and were stored at −20 °C. The full data on the Pathogen Box compounds are available at https://www.pathogenbox.org (accessed on 30 May 2021). A LCT-Premier XE TOF-MS (Waters, Milford, MA, USA) was employed for registering MS spectra in the 100–1500 m/z range of the selected molecules.
+ Open protocol
+ Expand
2

High-Resolution Mass Spectrometry Analysis

Check if the same lab product or an alternative is used in the 5 most similar protocols
MS data were acquired by using a Waters LCT Premier XE TOF-MS (Waters, Mircomas, MA, USA). Positive and negative ionization modes were acquired via DuoSpray electrospray ionization (ESI). Optimal conditions for high-resolution MS analysis were as follows: Negative ion mode: capillary voltage 1.5 kV, cone voltage 50 V, ion source temperature 110 °C, dissolvent gas temperature 350 °C, dissolvent gas flow rate 750 L/h. Positive ion mode: capillary voltage 1.3 kV, cone voltage 45V, ion source temperature 110 °C, dissolvent gas temperature 350 °C, dissolvent gas flow rate 750 L/h. For accurate mass calibration application, Waters Lockspary calibration system was used for online real-time calibration with leucine-enkephalin solution ([M+H]+ = 556.2771, [M−H] = 554.2620), with a calibration solution injection rate of 100 µL/min and calibration frequency of 5 s. The scan mode was full scan, mass data acquisition range of m/z 100–1500. The scan time was 0.2 s, and data acquisition was performed in centroid mode.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!