The largest database of trusted experimental protocols

Tracegold tg 5silms column

Manufactured by Thermo Fisher Scientific

The TraceGold TG-5SilMS column is a gas chromatography (GC) column designed for the separation and analysis of trace-level organic compounds. It features a 5% phenyl-95% methylpolysiloxane stationary phase coated on a fused silica capillary column. The column is intended for general-purpose GC applications requiring high inertness and low bleed.

Automatically generated - may contain errors

4 protocols using tracegold tg 5silms column

1

GC-MS Analysis of Derivatized Samples

Check if the same lab product or an alternative is used in the 5 most similar protocols
MOX + TMS derivatized samples were analyzed by GC-MS. 1 μL of derivatized sample was injected by autosampler into a Trace 1300 GC (ThermoFisher Scientific) fitted with a TraceGold TG-5SilMS column (ThermoFisher Scientific) operating under the following conditions: split ratio = 20:1, split flow = 24 μL/min, purge flow = 5 mL/min, carrier mode = Constant Flow, and carrier flow rate = 1.2 ml/min. The GC oven temperature gradient was as follows: 80 °C for 3 min, increasing at a rate of 20 °C/min to 280 °C, and holding at a temperature at 280 °C for 8 min. Between sample runs, the autosampler injection syringe was washed 3 times with methanol and 3 times with pyridine. Ion detection was performed by an ISQ 7000 mass spectrometer (ThermoFisher Scientific) operated from 3.90 to 21.00 min in EI mode (-70eV) using select ion monitoring (SIM). The mass spectrometer was tuned and calibrated daily. Samples were analyzed in a randomized order, and the pooled sample detailed above was analyzed at the beginning of, at a set interval (about every eight injections) during, and the end of the analytical run.
+ Open protocol
+ Expand
2

Metabolite Derivatization and GC-MS Analysis

Check if the same lab product or an alternative is used in the 5 most similar protocols
The dried samples, QC sample, and processing blank sample, were derivatized using methoxyamine hydrochloride (MOX) and N,O-Bis(trimethylsilyl)trifluoroacetamide (TMS) (Sigma). Briefly, dried extracts were reconstituted in 30 μL of 11.4 mg/ml MOX in anhydrous pyridine (VWR), vortexed for 10 minutes, and heated at 60°C for 1 hour. Next, 20 μL TMS was added to each reconstituted extract, vortexed for 1 minute, and heated at 60°C for 30 min. The derivatized samples, QC samples and processing blank samples were immediately analyzed using GC/MS.
GC chromatographic separation was conducted on a Thermo Trace 1300 GC with a TraceGold TG-5SilMS column (0.25 μM film thickness; 0.25mm ID; 30 m length). 1 μL of derivatized sample, QC, or blank was injected. The GC was operated in split mode with the following settings: 20:1 split ratio; split flow: 24 μL/min, purge flow: 5 mL/min, Carrier mode: Constant Flow, Carrier flow rate: 1.2 mL/min). The GC inlet temperature was 250°C. The GC oven temperature gradient was as follows: 80°C for 3 min, ramped at 20°C/min to a maximum temperature of 280°C, which was held for 8 min. The injection syringe was washed 3 times with pyridine between each sample. Metabolites were detected using a Thermo ISQ single quadrupole mass spectrometer. The data was acquired from 3.90 to 21.00 min in EI mode (70eV) by single ion monitoring (SIM).
+ Open protocol
+ Expand
3

Gas Chromatography-Mass Spectrometry Metabolite Profiling

Check if the same lab product or an alternative is used in the 5 most similar protocols
Gas chromatographic separation was conducted on a ThermoFisher Scientific Trace 1300 GC with a TraceGold TG-5SilMS column (0.25 μm film thickness; 0.25 mm ID; 30 m length; ThermoFisher Scientific), as described (Tompkins et al., 2019 (link)). The injection volume of 1 μL was used for all samples. The gas chromatograph was operated in split mode with the following settings: 20:1 split ratio; split flow: 24 μL/min, purge flow: 5 mL/min, Carrier mode: Constant Flow, Carrier flow rate: 1.2 mL/min). The inlet temperature was 250°C. The oven temperature gradient was: 80°C for 3 minutes, ramped at 20°C/minute to a maximum temperature of 280°C, held for 8 minutes. The injection syringe was washed 3 times with methanol and 3 times with pyridine between samples. Metabolites were detected using a ThermoFisher Scientific Q-Exactive Plus mass spectrometer. The data were acquired from 3.90 to 21.00 minutes in EI mode (−70eV) in full scan (m/z 56.7–850) at 60K resolution. Metabolite profiling data were analyzed using Tracefinder 4.1, utilizing standard verified peaks and retention times.
+ Open protocol
+ Expand
4

GC-MS Sample Derivatization with MOX and MSTFA

Check if the same lab product or an alternative is used in the 5 most similar protocols
GC–MS sample derivatization with MOX + MSTFA was accomplished as follows: 30 μL of pyridine containing 11.4 mg/mL of MOX was added to the dried metabolite extracts. Samples were then vortexed for 10 min and heated at 60 °C for 60 min. Next, 20 uL of MSTFA was added to the pyridine/MOX derivatized samples, and they were vortexed for 5 min, and heated at 60 °C for an additional 30 min. GC–MS analysis was performed using a Trace 1300 GC (Thermo) coupled to an ISQ-LT single quadrupole mass spectrometer (Thermo). For each sample, 1 μL of sample was injected into the GC by an autosampler in split mode (split ratio: 20:1; split flow: 24 μL/min, purge flow: 5 mL/min, Carrier mode: Constant Flow, Carrier flow rate: 1.2 mL/min). Separation was accomplished using a standard fused silica TraceGold TG-5SilMS column (Thermo). The temperature gradient was as follows: 80 °C for 3 min, ramped at a rate of 20 °C/min to 280 °C, and held for 8 min. Between sample runs, the injection syringe was washed 3 times with methanol and 3 times with pyridine. The MS was operated from 3.90 to 21.00 min in EI mode (−70eV) using select ion monitoring (SIM). The mass spectrometer was tuned and calibrated daily. The pooled QC sample was analyzed at the beginning and at the end of the GC/MS run, as well as about every eight injections throughout.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!