The largest database of trusted experimental protocols

Acquity uplc beh glycan column

Manufactured by Waters Corporation
Sourced in United States

The ACQUITY UPLC BEH-Glycan column is a high-performance liquid chromatography (HPLC) column designed for the separation and analysis of glycan structures. It features a bonded ethylene-bridged hybrid (BEH) stationary phase that provides efficient separation and resolution of a wide range of glycan species.

Automatically generated - may contain errors

7 protocols using acquity uplc beh glycan column

1

Glycan Profiling by UPLC-Fluorescence

Check if the same lab product or an alternative is used in the 5 most similar protocols
The AB- or ProA-labeled O-glycans were profiled on an ACQUITY UPLC H-Class System (Waters, Milford, MA, USA) equipped with a Quaternary Solvent Manager and Sample Manager coupled to an ACQUITY UPLC fluorescence detector (Waters). LC separations were performed on an ACQUITY UPLC BEH-Glycan column (1.7 μm, 2.1 × 150 mm; Waters) at 40 °C. The fluorescence was observed at excitation and emission wavelengths of 330 and 420 nm (for AB), and 310 and 370 nm (for ProA). The flow rate was 200 μL/min. The mobile phase consisted of 50 mM ammonium formate, pH 4.4 (A), and 100% acetonitrile (B), and the column was equilibrated with 75% B before sample injection. The samples were separated over a 25%–50% gradient of A for 50 min before washing the column with 100% A for 9 min and re-equilibrating the column with 75% B at a constant flow rate of 200 μL/min.
+ Open protocol
+ Expand
2

Enzymatic Release and Fluorescent Labeling of N-Glycans

Check if the same lab product or an alternative is used in the 5 most similar protocols
N-glycans were released at 5 mg/mL in 50 mM Tris HCl, pH 7.8 with 7.5 units of N-glycosidase F by overnight incubation at 37 °C. Released glycans were labelled with 2-AB using a Prozyme 2AB labelling kit. Excess 2AB was removed using HILIC cartridges and a clean up 96-well plate (Prozyme) operated via a vacuum manifold. Analysis of the 2AB-labelled N-glycans was performed on a Waters Acquity UPLC BEH Glycan column (2.1 mm × 150 mm, 1.7 μm particle) and an Acquity UPLC with a fluorescence detector (Waters, Milford, MA, U.S.A.). 2AB-labelled dextran ladder standard containing glucose unit (GU) oligomers (Prozyme) was used for glycan retention time normalisation, converting retention times into glucose units.
+ Open protocol
+ Expand
3

Quantifying N-Linked Glycans in Eculizumab

Check if the same lab product or an alternative is used in the 5 most similar protocols
The identification and relative percentage of N-linked glycan in eculizumab was determined by hydrophilic interaction UPLC (HILIC-UPLC). Eculizumab (SB12 and RP) samples were treated with PNGase-F enzyme to release N-glycan after denaturation. The released N-glycan was separated from proteins using cold ethanol, and dried completely. 2-aminobenzamide (2-AB)-labeled N-glycan was gradually separated using an Acquity UPLC BEH glycan column (Waters), and quantified on a fluorescence detector connected to a Waters UPLC system. For identification, N-glycan was labeled with procainamide and identified on LC-ESI-MS connected to a Waters UPLC system.
+ Open protocol
+ Expand
4

N-Glycan Profiling of Recombinant IgG1 Fc

Check if the same lab product or an alternative is used in the 5 most similar protocols
N-glycan analysis of rhIgG1 Fc derived from ALB::hIgG1 Fc chicken was performed by UPLC/MS-MS. Briefly, purified rhIgG1 Fc was incubated with PNGase F for 16 h at 37 °C. Deglycosyated rhIgG1 Fc was precipitated using ethanol and centrifuged at 10,000 g for 10 min. The supernatant containing released N-glycan was transferred to a new tube and dried completely using Speed-Vac concentrator. The dried sample was labelled with procainamide for fluorescence analysis. The labeled N-glycan sample was analyzed and quantified using UPLC/MS-MS. An ACQUITY UPLC BEH Glycan column (1.2 × 150 mm, 1.7 μm; Waters, New Castle, DE) with a fluorescence detector (Waters iClass UPLC) was used for the separation and detection of N-glycans. The LC conditions were as follows: flow rate (0.5 mL/min), column temperature (60 °C), mobile phase buffer A (100 mM ammonium formate, pH 4.5), buffer B (100% acetonitrile), injection volume (8 mL), linear gradient (75–60% B for 46.5 min, 60–0% B for 1.5 min, 0% B for 1 min, 0–75% B for 1 min, and 75% B for 13 min). A high-resolution mass spectrometry, triple-TOF MS (AB SCIEX, Concord, Ontario, Canada), was used for N-glycan identification. The N-glycan distribution was analyzed with Empower (Waters).
+ Open protocol
+ Expand
5

Glycan Analysis by UPLC-Fluorescence

Check if the same lab product or an alternative is used in the 5 most similar protocols
Glycan release and derivatization were performed according to Mittermayr et al. [15 (link)]. Chromatographic separation was carried out using ACQUITY UPLC H-Class Bio System with a linear gradient from 22 to 50% acetonitrile using 100 mM ammonium formate solution at pH 4.50 (mobile phase A). Fluorescence detection was set at 250 nm and 420 nm excitation wavelength using 150 × 2.1 mm, 1.7 μm ACQUITY UPLC BEH glycan column coupled to 1.7 μm VanGuard BEH glycan precolumn from Waters Corp. (Milford, MA).
+ Open protocol
+ Expand
6

O-Glycan Profiling by LC-MS/MS

Check if the same lab product or an alternative is used in the 5 most similar protocols
The AB- or ProA-labeled O-glycans were subjected to LC on an Ultimate 3000 LC system (Thermo Fisher Scientific, Waltham, MA, USA) equipped with an ACQUITY UPLC BEH-Glycan column (1.7 μm, 2.1 × 150 mm; Waters). A Q Exactive Hybrid Quadrupole Orbitrap mass spectrometer (Thermo Fisher Scientific) was operated in ESI-positive mode. The mobile phase and separation conditions were the same as those used in UPLC. The O-glycans were identified by the elution time of the UPLC chromatogram and the presence of the characteristic oxonium ions of O-glycans in the MS spectrum. The quantity (%) of each glycan relative to total O-glycans (as 100%) was determined from the extracted ion chromatogram (EIC) areas for all observed charge states in LC–ESI–HCD–MS/MS.
+ Open protocol
+ Expand
7

Glycan Analysis by UPLC-Fluorescence

Check if the same lab product or an alternative is used in the 5 most similar protocols
Glycan release and derivatization were performed as previously described [20 (link)]. Chromatographic separation was carried out using an ACQUITY UPLC H-Class Bio System with a linear gradient from 22 to 50% of acetonitrile using 100 mM ammonium formate aqueous solution at pH 4.50 as mobile phase A. Fluorescence detection was set at an excitation wavelength of 250 nm and 420 nm for emission, using a 150 × 2.1 mm, 1.7 μm ACQUITY UPLC BEH glycan column coupled with a 1.7 μm VanGuard BEH Glycan Precolumn from Waters Corp. (Milford, MA).
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!