The largest database of trusted experimental protocols

Sequant zic philic column

Manufactured by Merck Group
Sourced in Germany, United States

The SeQuant ZIC-pHILIC column is a hydrophilic interaction chromatography (HILIC) column designed for the separation of polar and hydrophilic analytes. It utilizes a zwitterionic stationary phase to facilitate the retention and separation of these compounds. The column's core function is to provide efficient chromatographic separation of polar and hydrophilic compounds.

Automatically generated - may contain errors

58 protocols using sequant zic philic column

1

Metabolomic Profiling of siRNA-Mediated RRM2 Knockdown

Check if the same lab product or an alternative is used in the 5 most similar protocols
After transfection with siRNAs targeting RRM2 for 72 h, the intracellular metabolites were extracted using 1 mL of cold MeOH:ACN:H2O (2:2:1, v/v) solvent mixture in liquid nitrogen for 1 min. The dried supernatant was stored at −80 °C, followed by centrifugation at 14,000×g for 15 min at 4 °C. LC-MS/MS analysis was performed using a triple quadrupole mass spectrometry (QqQ-MS) coupled to a Dionex Ultimate 3000 UHPLC system (Thermo Fisher Scientific, USA). The chromatographic separation of samples was performed on an Agilent 1260 HPLC system (Agilent Technologies) with a SeQuant ZIC pHILIC column (Merck) and a SeQuant ZIC pHILIC guard column (Merck) at 250 µL/min flow rate. The mass spectrometer was operated in a reaction monitoring (SRM) mode referring to the literature.17 (link)
+ Open protocol
+ Expand
2

Quantitative LC-MS Metabolite Profiling

Check if the same lab product or an alternative is used in the 5 most similar protocols
LC-MS analysis was performed using a UHPLC system (1290 series, Agilent Technologies) coupled to a quadruple time-of-flight mass spectrometer (6550 series, Agilent Technologies). Merck SeQuant ZIC-pHILIC column (particle size, 5 μm; 100 mm (length) ×2.1 mm (i.d.)) was used. Mobile phases A = 25 mM ammonium acetate and 25 mM ammonium hydroxide in 100% water, and B = 100% acetonitrile, were used for both ESI positive and negative modes. And the linear gradient eluted from 80% B (0.0–2.0 min, 0.2 mL/min), 80% B to 20% B (2.0–17.0 min, 0.2 mL/min), 20% B to 80% B (17.0–17.1 min, 0.2 mL/min), 80% B (17.1–22.1 min, 0.4 mL/min), 80% B to 80% B (22.1–22.2 min, 0.2 mL/min). The sample injection volume was 2 μL.
ESI source parameters on QTOF 6550 were set as followings: sheath gas temperature, 300°C; dry gas temperature, 250°C; sheath gas flow, 12 L/min; dry gas flow, 16 L/min; capillary voltage, 2500 V or −2500 V in positive or negative modes, respectively; nozzle voltage, 0 V; and nebulizer pressure, 20 psi. The MS1 data acquisition frequency was set as 4 Hz, and the TOF scan range was set as m/z 60–1200 Da.
+ Open protocol
+ Expand
3

Metabolite Extraction and Liquid Chromatography Protocol

Check if the same lab product or an alternative is used in the 5 most similar protocols
Metabolites were extracted from mid-exponential-phase culture pellets (five replicates) with a cold methanol method (40 (link)) and analyzed using an Agilent 6550 Q-TOF apparatus (Agilent Technologies, Santa Clara, CA). Liquid chromatography (LC) separation was conducted on an Agilent 1290 ultrahigh-performance LC (UHPLC) system (Agilent Technologies, Santa Clara, CA) with a 150- by 2.1-mm, 5-µm, 200-Å SeQuant ZIC-pHILIC column (EMD Millipore, Billerica, MA) and guard column of 20 by 2.1 mm, 5 µm (EMD Millipore, Billerica, MA) (41 (link)). MassHunter (Agilent) was used for initial data analysis, and Metabolite Atlas was used for targeted data analysis. Compounds were confirmed using authentic standards.
+ Open protocol
+ Expand
4

Targeted Metabolomics Analysis Protocol

Check if the same lab product or an alternative is used in the 5 most similar protocols
Targeted metabolomics analyses were performed as described.57 Briefly, extraction solution used was 50% methanol, 30% ACN and 20% water. The volume of extraction solution added was calculated from the weight of powdered tissue (60 mg/mL) or the volume of the plasma (33ul/ml). After addition of extraction solution, samples were vortexed for 5 min at 4°C, and then centrifuged at 16 000 × g for 15 min at 4°C. The supernatants were collected and analysed by liquid chromatography–mass spectrometry using SeQuant ZIC‐pHilic column (Merck) for the liquid chromatography separation. Mobile phase A consisted of 20 mM of ammonium carbonate plus 0.1% ammonia hydroxide in water. Mobile phase B consisted of ACN. The flow rate was kept at 200 ml/min, and the gradient was 0 min, 80% of B; 30 min, 20% of B; 31 min, 80% of B; and 45 min, 80% of B. The mass spectrometer (QExactive Orbitrap, Thermo Fisher Scientific) was operated in a polarity switching mode, and metabolites were identified using TraceFinder Software (Thermo Fisher Scientific). The specific peak area is presented.
+ Open protocol
+ Expand
5

Targeted LC-MS Metabolomics Profiling

Check if the same lab product or an alternative is used in the 5 most similar protocols
The liquid chromatography separation was performed with an Accela ultra-high-performance LC (UHPLC) system. A polymeric SeQuant® ZIC®-pHILIC column (5 μm, 150 mm × 2.1 mm) (150,460, Merck) and a SeQuant® ZIC®-pHILIC Guard Kit (20 × 2.1 mm) (50,438, Merck) were used operating at 45 °C. The injection mode was set at 5 μL, and the mobile phase flow rate was set at 0.3 mL/min. Mobile phase solvents were A (95% H2O, 5% methanol, 0.1% formic acid) and B (100% methanol). The eluting gradient program in both positive and negative ion mode was the following: 0–1.0 min (95% A, 5% B), 1.0–4.0 min (45% A, 55% B), 4.0–9.0 min (45% A, 55% B), 9.0–10.0 min (20% A, 80% B), 10.0–10.1 min (20% A, 80% B), 10.1–15.0 min (0% A, 100% B), 15.0–15.1min (0% A, 100% B), and 15.1–20.0 min (95% A, 5% B).
+ Open protocol
+ Expand
6

pHILIC Separation of Metabolites

Check if the same lab product or an alternative is used in the 5 most similar protocols
The samples were analysed using a Thermo Scientific Ultimate 3000 RSLCnano system (Thermo Scientific, CA, USA). The pHILIC separation was performed with a SeQuant ZIC-pHILIC column (150 × 4.6 mm, 5 µm) equipped with the corresponding pre-column (Merck KGaA, Darmstadt, Germany)—the column temperature was maintained at 25 °C. A linear biphasic LC gradient was conducted from 80% B to 20% B over 15 min, followed by a 2 min wash with 5% B, and 8 min re-equilibration with 80% B, where solvent B is acetonitrile and solvent A is 20 mM ammonium carbonate in water. The flow rate was 300 µl/min, column temperature was held at 25 °C, injection volume was 10 µl, and samples were maintained at 4 °C in the autosampler (Creek et al. 2011 (link)).
+ Open protocol
+ Expand
7

Comprehensive Metabolite Analysis by GC-MS and LC-MS

Check if the same lab product or an alternative is used in the 5 most similar protocols
GCMS analysis was performed on a Shimadzu GCMS‐QP2010 with a Rtx‐5MS (30 m × 0.25 μm × 0.25 mm i.d.) capillary column. The injection temperature was 200°C for metabolites, samples were run on splitless mode, and pressure was held at 65.2 kPa. The column flow was 1.00 ml/min and the ion source temperature was 200°C.
LCMS analysis was conducted on Acuity I‐Class UPLC (Waters) coupled to a Q‐ExactiveTM HF mass spectrometer (Thermo Fisher Scientific). Cellular metabolites were separated on a SeQuant ZIC‐pHILIC column (2.1 × 100 mm, 5 µm) (EMD Millipore). The column temperature was set at 25°C and the flow rate was 0.15 ml/min. A 25‐min gradient was used for metabolites separation. Metabolites were detected in negative mode full scan analysis with resolution at 120,000.
+ Open protocol
+ Expand
8

Metabolomic Analysis of PDT-Treated Cells

Check if the same lab product or an alternative is used in the 5 most similar protocols
SK-ChA-1 cells were seeded in 6-wells plates and cultured until confluence. Cells were treated using the PDT protocol as described in “PDT protocol” (n = 3 per group). After 90 min, the cells were washed with 1 mL cold PBS and the cells were lysed in 1 mL lysis buffer (40% acetonitrile, 40% methanol, 20% water). The cells were scraped and transferred to 2-mL centrifuge tubes that were shaken for 10 min at 4 °C. Next, the samples were centrifuged for 15 min at 20,000×g (4 °C), after which the supernatant was aspirated and stored at −80 °C. LC-MS analysis was performed on an Exactive mass spectrometer (Thermo Scientific) coupled to a Dionex Ultimate 3000 autosampler and pump (Thermo Scientific). The MS operated in polarity-switching mode with spray voltages of 4.5  and −3.5  kV. Metabolites were separated using a Sequant ZIC-pHILIC column [2.1 × 150 mm, 5 µm, guard column 2.1 × 20 mm, 5 µm (Merck)] using a linear gradient of acetonitrile and eluent A (20 mM (NH4)2CO3, 0.1% NH4OH in ULC/MS grade water [Biosolve, Valkenswaard, the Netherlands)]. The flow rate was set to 150 µL/min. Metabolites were identified and quantified using LCquan software (Thermo Scientific) on the basis of exact mass within 5 ppm and further validated in accordance with the retention times of standards. Peak intensities were normalized based on total ion count.
+ Open protocol
+ Expand
9

Metabolite Profiling of Murine Tissues

Check if the same lab product or an alternative is used in the 5 most similar protocols
Equal amounts of wet weight murine tissue were lysed in 250 ml extraction solution (30% acetonitrile, 50% methanol and 20% water) per 10 mg tissue in a Bullet Blender (Next Advance) following the manufacturer’s instructions. The suspension was immediately centrifuged (16,000g, 15 min at 4°C) and the supernatant analysed by liquid chromatography-mass spectrometry. Sample extracts were run twice on a liquid chromatography system fitted with a Sequant ZIC-HILIC column (5 μm, 4.6 x 150 mm) and afterwards with a Sequant ZIC-pHILIC column (5 μm, 2.1 x 150 mm), with the corresponding guard columns (both 2.1 x 20 mm, 5 μm) (all from Merck), and according to previously described gradient elution methods [1 (link)]. The mass spectrometer (Thermo Q Exactive) was operated in full scan mode with polarity switching. Samples were randomized to avoid bias due to machine drift and the operator was blinded to the sample key. Spectra were analysed using Xcalibur Quan Browser software (Thermo Fisher Scientific) by referencing to an internal library of compounds.
+ Open protocol
+ Expand
10

LC-MS/MS analysis of metabolites

Check if the same lab product or an alternative is used in the 5 most similar protocols
LC-MS/MS analysis was performed using a Q Exactive Quadrupole-Orbitrap mass spectrometer coupled to a Vanquish UHPLC system (Thermo Fisher Scientific). The liquid chromatography system was fitted with a Sequant ZIC-pHILIC column (150 mm × 2.1 mm) and guard column (20 mm × 2.1 mm) from Merck Millipore and temperature was maintained at 35°C. The sample (2μL) was separated at a flow rate of 0.1 mL/minute. The mobile phase was composed of 10 mM ammonium carbonate and 0.15% ammonium hydroxide in water (solvent A) and acetonitrile (solvent B). A linear gradient was applied by increasing the concentration of solvent A from 20 to 80% within 22 minutes and then maintained for 7 minutes. The mass spectrometer was operated in full MS and polarity switching mode, in the range of 70-1000m/z and resolution 70000. Major ESI source settings were: spray voltage 3.5 kv, capillary temperature 275°C, sheath gas 35, auxiliary gas 5, AGC target 3e6, and maximum injection time 200 minutes. For the targeted analysis, the acquired spectra were analyzed using XCalibur Qual Browser and XCalibur Quan Browser software (Thermo Scientific). The compound discoverer 3.1 (CD) (Thermo Scientific)) was used for untargeted and novel feature detection and annotation with library scoring. Features with the fold change >2 and p <0.05 were selected as discriminating markers. Samples were analysed by quintuplicate.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!