The largest database of trusted experimental protocols

18 protocols using keithley 2400 source meter

1

Measuring External Quantum Efficiency

Check if the same lab product or an alternative is used in the 5 most similar protocols
The measurement of the External Quantum Efficiency (EQE) vs. photon energy of completed devices were measured with a pre-calibrated Bentham PVE300 system in the 300–1200 nm wavelength range, and calibrated using Si and Ge photodiodes. Reversed voltage-biased EQE curves were collected by connecting a Keithley 2400 source meter (Keithley Instruments Inc., Cleveland, Ohio, USA) directly to the primary coil of the transformer and biasing the device at the desired potential, referenced against the cell voltage.
+ Open protocol
+ Expand
2

Fabrication and Characterization of MoS2/SWNT Hybrid Films

Check if the same lab product or an alternative is used in the 5 most similar protocols
Equal amount of composite dispersions were filtered onto a nitrocellulose membrane and dried at room temperature. These films were cut into 0.5×2 cm strips and subsequently transferred onto a glass slide using the transfer method of Wu et al. 72 . Electrical conductivity values were calculated from resistivity measurement using a four-point probe technique with a Keithley 2400 source meter (Keithley Instruments, Inc.). It was controlled by a Lab View program (National Instruments, Inc.). The films were also used for Raman spectroscopy and scanning electron microscopy (SEM) . The as prepared MoS2/SWNT hybrid dispersions (with various MoS2:SWNT ratios) were vacuum filtered onto polyester filter membrane pore size 0.45 µm.
The membranes were dried at room temperature and the free standing hybrid films were peeled off. Free standing hybrid films were cut into strips of width ~2.25 mm. Films thicknesses were in the range of 70-80 µm measured using a digital micrometer. N.B. we limited the composites prepared to mass fractions of 6 wt% or less due to the high nanotube masses required to make such thick films. Mechanical measurements were performed using Zwick tensile tester at a strain rate of 0.5 mm/minute. Each data point is an average of 4 measurements.
+ Open protocol
+ Expand
3

4-Point Probe Sheet Resistance Measurement

Check if the same lab product or an alternative is used in the 5 most similar protocols
A Loresta-GX MCP-T700 (Nittoseiko Analytech Co., Ltd., Kanagawa, Japan) resistivity meter was used to measure the sheet resistance (Rs) of all samples via the 4-point probe method after deposition of 400 μg of each sample on coated paper, drying at room temperature (RT), and “polishing” using mild pressure so that films with an average thickness of about 17μm were formed. The Rs values of the printed substrates were measured using a 4-point probe system (Lucas Labs Pro4 Resistivity System, Lucas Signatone Corp., Gilroy, CA, USA) and a Keithley 2400 Source Meter (Keithley Instruments, Cleveland, OH, USA).
+ Open protocol
+ Expand
4

Dye-Sensitized Solar Cell Fabrication

Check if the same lab product or an alternative is used in the 5 most similar protocols
The resulting paste was then used to fabricate the working electrode using the well-known doctor blade method as reported in [25 (link)]. The synthesized paste was employed on a cleaned and blocking layer-coated fluorine-doped tin oxide (FTO) glass and calcinated at 315°C for 12 h at a heating rate of 1°C/min. The calcinated working electrode was then soaked in (0.5 mM in ethanol) dye solution overnight and washed with ethanol to remove excess of the dye adsorbed on the surface of the electrode. The solid polymer electrolyte was prepared by dissolving the polyethylene oxide (PEO)/PEG blend in acetonitrile and adding NaI and I2 (75:25 w/w%) and iodine (10 w/w%), respectively. The platinum counter electrode was prepared by spin coating the H2PtCl6 solution on a clean FTO glass and heated at 400°C for 30 min. A drop of the above prepared electrolyte was sandwiched between the fabricated working electrode and counter electrode and used to record the current density-voltage (J-V) curve using a Keithley 2400 sourcemeter (Keithley Instruments Inc., Cleveland, OH, USA).
+ Open protocol
+ Expand
5

Low-Impedance Platinum Coating Electrodeposition

Check if the same lab product or an alternative is used in the 5 most similar protocols
In order to demonstrate a low-impedance surface coating, a nano-porous layer of platinum was electrodeposited on the microelectrodes using a current of (−4 µA) per electrode in a solution of chloroplatinic acid diluted with HCl and lead acetate (1% chloroplatinic acid; 0.005% lead acetate; 0.01 M HCl, all from Sigma Aldrich, St. Louis, MO, USA). A platinum wire (Sigma Aldrich, St. Louis, MO, USA) was used as the counter electrode and every electrode was electroplated under DC conditions for 40 s using a current source (Keithley 2400 Source Meter, Keithley Instruments, Cleveland, OH, USA).
+ Open protocol
+ Expand
6

Measuring Surface Electrical Conductivity

Check if the same lab product or an alternative is used in the 5 most similar protocols
The surface electrical conductivity, per square, ( σs ) of the pellets, filaments and 3D-printed parts were measured by using the ASTM D257 standard method. Two silver paint rings were used in pellets and filaments sections to measure the electrical resistance in order to minimize the electrical contact resistance (Figure 2(b1,b2)). In the case of 3D-printed parts, two lines of silver paint were used as contact (Figure 2(d1,d2)). The distance between the electrodes in all the cases was ~10 mm. The surface electrical conductivity (per square) was calculated from the electrical resistance, which was measured with a Keithley 2400 Source Meter (Keithley Instruments, Cleveland, OH, USA), following Equation (2).
σv=1R·lPc,
where R is the measured electrical resistance, l is the length between electrical contacts and Pc is the electrical contact perimeter.
+ Open protocol
+ Expand
7

Multifaceted Characterization of Thin-Film Samples

Check if the same lab product or an alternative is used in the 5 most similar protocols
The thin layer deposited samples were examined using both optical and confocal microscopes. The optical microscope Nikon MM-400 (Nikon Metrology, Brighton, USA) was used to visualize the sample and predetermine damage at the surface. Morphological observations of the samples were undertaken by means of a Carl Zeiss Ultra Plus field emission scanning electron microscope (Zeiss, Oberkochen, Germany). Both the surface and the cross section after coating fracture were evaluated. The conductivity of the samples was determined via four-probe method. An electric current (5–30 mA) was passed through collinear outer metal electrodes by a Keithley 2400 source meter (Keithley Instruments, Cleveland, USA) and the voltage drop was measured between two inner electrodes with a HP 34401 A multimeter (Agilent Technologies, Santa Clara, USA). Considering that the distance between adjacent electrodes (s) was 2.5 mm and the thickness (t) of the films was close to 200 nm, the necessary condition for employing four-point probe method for conductivity measurements (t < 
+ Open protocol
+ Expand
8

Photovoltaic Device Characterization

Check if the same lab product or an alternative is used in the 5 most similar protocols
Photocurrent-voltage (J-V) characteristics were performed using a Keithley 2400 SourceMeter (Keithley Instruments Inc., Cleveland, OH, USA) under simulated AM 1.5G illumination (100 mW/cm2) provided by a solar light simulator (94043A, Newport Corp., Irvine, CA, USA). Cyclic voltammetry (CV) and the electrochemical impedance spectroscopy (EIS) were measured with an Autolab electrochemical workstation (PGSTAT 302 N, Metrohm AG, Utrecht, The Netherlands). CV was carried out in a three-electrode system with different counter electrodes as working electrodes, a Pt foil as counter electrode, and a Ag/Ag+ electrode as reference electrode at a scan rate of 50 mV/s. The electrodes were immersed into an anhydrous acetonitrile solution containing 0.1 M LiClO4, 10 mM LiI, and 1 mM I2. EIS was actualized with a symmetric cell assembled with two identical counter electrodes at open-circuit voltage (Voc) bias under dark condition. The measured frequency ranged from 10 mHz to 1 MHz and the magnitude of the alternative signal was 10 mV.
+ Open protocol
+ Expand
9

Fabrication and Characterization of OLED Devices

Check if the same lab product or an alternative is used in the 5 most similar protocols
The ITO glass substrates with a sheet resistance of 15 Ω per square were firstly cleaned with acetone, ethanol, and deionized water. They are then dried in an oven at 100 °C for 5 h and treated by ultraviolet ozone for 15 min. All the layers were thermally evaporated at a vacuum of 2 × 10−6 Torr. The deposition rates of HIL, HTL, EML, ETL, EIL, and Al were 0.2, 2 (for HTL, EML, ETL), 0.2, and 6 A/s, respectively, and were monitored by quartz crystals. The emissive area of the devices was 3 × 3 mm2. The EL spectra, luminance, and CIE coordinates of devices were measured by using Spectrascan PR655 photometer (JADAK, Syracuse, NY, USA) and Keithley 2400 source meter (Tektronix, Beaverton, OR, USA). EQEs were calculated from the corresponding current density, luminance, and EL spectra.
+ Open protocol
+ Expand
10

Characterizing SiC Diode Isolation

Check if the same lab product or an alternative is used in the 5 most similar protocols
Since p-n junctions are formed between the n+ and p epitaxial films, back-to-back diodes are present between adjacent traces, which provides isolation. This isolation was evaluated by measuring the forward and reverse blocking voltages of test structures consisting of p-n diodes and n-p-n junctions formed between adjacent traces that were built on the 3C-SiC wafer. A Keithley 2400 SourceMeter (Tektronix, Inc., Beaverton, OR, USA) was used to generate current-voltage (I-V) plots for adjacent traces to observe these voltages. The voltage was increased from −10 V to +10 V at a rate of 0.1 V/s for the diodes and n-p-n junctions, and the observed currents recorded. The forward voltage was estimated using a semi-logarithmic current scale I-V plot [40 (link)]. The breakdown voltage occurs when the current rapidly increases during application of negative voltage. The root mean square (rms) of the current amplitude between breakdown and forward potentials for the diodes was defined as reverse leakage current [33 (link)]. The threshold current for defining the breakdown voltages was 10 µA.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!