The largest database of trusted experimental protocols

Acquity uplc m class

Manufactured by Waters Corporation
Sourced in United States, United Kingdom

The ACQUITY UPLC M-Class is a high-performance liquid chromatography (HPLC) system designed for advanced analytical applications. It features a modular design, delivering enhanced separation performance and robustness. The system is capable of generating high pressures, enabling the use of sub-2-micron particle columns for improved resolution and sensitivity.

Automatically generated - may contain errors

28 protocols using acquity uplc m class

1

Label-free quantification of recombinant proteins

Check if the same lab product or an alternative is used in the 5 most similar protocols
Data used for label-free quantification was acquired on a nanoACQUITY UPLC system coupled to SYNAPT G2-Si mass spectrometer (Waters). A C 18 trapping column (Waters Acquity UPLC M-Class, 0.18×20 mm, 5 μm, 100 Å) and a HSS T3 analytical column (Waters Acquity UPLC M-Class, 75 μm×150 mm, 1.8 μm, 100 Å) were employed. Mobile phases A and B consist of 0.1% FA in water and 0.1% FA in ACN, respectively. A 60 min and 120 min length gradient of 1-40 % acetonitrile at a flow rate of 300 nL/min was used for separation of recombinant protein digests and cell lysates samples, respectively. MS scan range was set to m/z 350-1500 with a scan time of 0.2 s, and MS/MS scan range was set to m/z 50-2000 using data-dependent acquisition (DDA). The top 10 abundant precursors were subjected to MS/MS fragmentation with a ramp CE set between low energy (14-19 eV) and elevated energy (60-90 eV) using a scan time of 0.15 s per function.
+ Open protocol
+ Expand
2

LC-MS Protein Desalting Protocol

Check if the same lab product or an alternative is used in the 5 most similar protocols
The mass analysis was performed by LC-MS using an M-class ACQUITY UPLC (Waters UK, Manchester, UK) interfaced to a Synapt G2S Q-IMT-TOF mass spectrometer (Waters UK, Manchester, UK). 1 µL of 5 µM sample was loaded onto a MassPREP protein desalting column (Waters UK, Manchester, UK) washed with 10% solvent B (0.1% formic acid in acetonitrile) in solvent A (0.1% formic acid in water) for 5 min at 25 µL min−1. After valve switching, the bound protein was eluted by a wash of 95% solvent B in A for 3 min before re-equilibration at 10% solvent B in A ready for the next injection. The column eluent was directed in to the mass spectrometer via a Z-spray electrospray source. The MS was operated in positive TOF mode using a capillary voltage of 3kV, sample cone of 40V and source offset of 80V. Backing pressure was 7.9 mbar and trap bias 2V. The source temperature was 80°C and desolvation was 150°C. Argon was used as the buffer gas at a pressure of 9.1 × 10−3 mbar in the trap and transfer regions of the TriWave device. Mass calibration was performed by a separate injection of [Glu]-fibrinopeptide b at a concentration of 250 fmol µl−1 in MS/MS mode and a CID voltage (trap region) of 28V. Data processing was performed using the MassLynx v4.1 suite of software (Waters UK, Manchester, UK) supplied with the mass spectrometer.
+ Open protocol
+ Expand
3

Protein Desalting and Mass Analysis

Check if the same lab product or an alternative is used in the 5 most similar protocols
Protein desalting and mass analysis was performed by LC-MS using an M-class ACQUITY UPLC (Waters UK, Manchester, UK) interfaced to a Xevo QToF G2-XS mass spectrometer (Waters UK, Manchester, UK). Samples were diluted to 5 µM using 0.1% TFA. 1 µL of the 5 µM sample was loaded onto a MassPREP protein desalting column (Waters UK, Manchester, UK) washed with 10% solvent B in A for 5 min at 25 µL min. After valve switching, the bound protein was eluted by a gradient of 2–40% solvent B in A over 1 min at 25 µL min. The column was subsequently washed with 95% solvent B in A for 6 min before re-equilibration at 5% solvent B in A ready for the next injection. Solvent A was 0.1% formic acid in water, solvent B was 0.1% formic acid in acetonitrile.
The column eluent was directed into the mass spectrometer via a Z-spray electrospray source. The MS was operated in positive TOF mode using a capillary voltage of 3.2 kV, sample cone of 20 V and source offset of 80 V. The source temperature was 100 °C and desolvation was 250 °C. Mass calibration was performed by a separate injection of [Glu]-fibrinopeptide b at a concentration of 250 fmol µL. Data processing was performed using the MassLynx v4.1 suite of software supplied with the mass spectrometer.
+ Open protocol
+ Expand
4

Hydrogen-Deuterium Exchange Mass Spectrometry

Check if the same lab product or an alternative is used in the 5 most similar protocols
Hydrogen-deuterium exchange was carried out using the commercial Waters HDX system. Briefly: samples were mixed and injected in a labeling time-randomized order by a PAL3 Autosampler followed by UPLC separation and mass spectrometry analysis using a Waters M-Class ACQUITY UPLC with peptides detected on a Waters Select Series Cyclic IMS Mass Spectrometer. Labeling times of 1, 10, and 60 min at RT were used, followed by quenching (7.5 M guanidine hydrochloride and 0.5M TCEP) at 0°C and digestion using an Enzymate BEH Pepsin column. The peptides were reverse-phase separated using an ACQUITY UPLC BEH C18 column. Peptide identification was carried out using HDMSe with CID fragmentation in the transfer cell and ProteinLynx Global Server (PLGS), followed by HDX analysis using DynamX. Structures were visualized using PyMOL 2.5.
+ Open protocol
+ Expand
5

TIMS-TOF Analysis of Bacterial-Fungal Interactions

Check if the same lab product or an alternative is used in the 5 most similar protocols
The samples for TIMS-TOF analysis were prepared as follows: The agar plugs containing bacteria, confrontation zone, and fungus were immersed in 1 mL of acetonitrile/water/trifluoroacetic acid (70/30/0.1 v/v/v). The mix was stirred at 850 rpm and 20 °C for 5 h (Thermomix.). The supernatant was diluted 1:10 in water before analysis. The chromatographic separation was performed on a M-ClassACQUITY UPLC (Waters, Milford, MA, United States). A 3-min-long sample trapping step was first achieved on a reversed-phase (RP) ACQUITY UPLC M-Class Trap Column (Symmetry C18, 100 Å, 5 μm, 180 μm × 20 mm, Waters, Milford, USA) prior to releasing on a ACQUITY UPLC M-Class BEH C18 analytical column (100 Å, 1.8 μm spherical silica, 75 μm × 100 mm, Waters, United States). Water and acetonitrile both supplemented with 0.1% (v:v) of formic acid (FA) were used as eluents and mixed according to a 32 min-long gradient method. The flow rate was set at 0.6 μL·min−1. The mass detection was performed on a timsTOF spectrometer (Bruker, Bremen, Germany) with the following parameters: timsTOF, scan range m/z, 100–2200. The untargeted profiling data were acquired using Auto MS/MS. Experimental data were processed using Bruker Compass DataAnalysis 6.0.
+ Open protocol
+ Expand
6

Quantitative Proteomics by UPLC-MS/MS

Check if the same lab product or an alternative is used in the 5 most similar protocols
LC-MS/MS was performed using an ACQUITY UPLC M-Class (Waters)
paired with a Q Exactive Plus (Thermo). Peptides were separated on a
65-cm-long, 75-µm-internal-diameter PicoFrit column (New
Objective) packed in-house to a length of 50 cm with 1.9 µm
ReproSil-Pur 120 Å C18-AQ (Dr. Maisch) using methanol as the
packing solvent. Peptide separation was achieved using a non-linear
90-min gradient from 1% ACN 0.1% formic acid to 90% ACN 0.1% formic acid
with a flowrate of 250 nl/min. Approximately 3–4 µg of
peptides were run for each of the six biological replicates with at
least one blank run between samples to avoid peptide carryover.
+ Open protocol
+ Expand
7

Protein Extraction and Digestion from E. coli

Check if the same lab product or an alternative is used in the 5 most similar protocols
Protein extraction and digestion from E. coli was performed using cell pellets from 200 μL stationary-phase cultures. Lysis, reduction, alkylation, trypsin digest, and acid cleavage were performed as in (Lajoie et al., 2013 (link)). All resulting peptides were purified and desalted using a C18 MacroSpin column (The Nest Group), dried in a rotary vacuum centrifuge, and resuspended in 6 μL 70% formic acid and 16 μL 0.1% trifluoroacetic acid. Following A280 peptide quantification, samples were diluted to a concentration of 0.5 μg/μL in the same buffer, and 4 μL of sample (2 μg total) were injected onto an analytical column using ACQUITY UPLC M-Class (Waters) for mass spectrometry. Column specifications, solvent gradients, and mass spectrometry parameters for the Q Exactive Plus (Thermo) were performed as described in (Ferdaus et al., 2016 (link)).
+ Open protocol
+ Expand
8

UPLC-MS/MS Quantitative Peptide Analysis

Check if the same lab product or an alternative is used in the 5 most similar protocols
Peptide separation was performed using an ACQUITY UPLC M-class (Waters). Mobile phase A was 0.1% formic acid in the water, and mobile phase B was 0.1% formic acid in acetonitrile. Samples were reconstituted with 40 µL of SIS peptide mixture in mobile phase A, injected with a full sample loop injection of 5 µL, and separated in an ACQUITY UPLC peptide CSH C18 column (1 mm i.d., 10 cm length, pore size 130 Å, particle size 1.7 µm; Waters). A gradient with a flow rate of 20 µL/min and 5% B for 3 min, 5–25% B for 22 min, 25–60% for 1 min, 60–60% for 1 min, 60–5% B for 1 min, and 5% B for 5 min, followed by column washing with 80% B for 5 min, 80% to 5% for 6 min, 5–80% for 6 min, and re-equilibration with 5% B for 13 min. MRM analysis was performed using a 5500 Qtrap mass spectrometer. The MS detection was carried out in positive mode with the following parameters: ion spray voltage of 5500 V, curtain gas at 20 psi, nebulizer gas at 25 psi, heating gas at 20 psi, resolution at 0.7 Da (unit) for Q1/Q3, interface temperature at 400 °C, and scan mass range 300–1250 m/z QQQ mode. Quantification experiments were performed using the scheduled MRM mode. The MRM detection window time was 120 s, and the cycle time was 1.5 s. The mass spectrometer was operated with Analyst software (Version 1.6.2, SCIEX), which generated MRM-MS data with the file name *.wiff.
+ Open protocol
+ Expand
9

UPLC-MS Analysis of Phenolic Compounds

Check if the same lab product or an alternative is used in the 5 most similar protocols
Free and bound phenolic compounds were determined using an ACQUITY UPLC M-Class liquid chromatograph from Waters (Waters Corp., Milford, MA, USA) that included a degasser, a binary pump, an oven and an automatic liquid sampler, and it was coupled with mass spectrometry. Phenolic compounds were separated on HPLC column Poroshell 120, SB-C18 (3.0 mm × 100 mm, 2.7 µm) from Agilent Technologies (Agilent Technologies, Palo Alto, CA, USA). Oven temperature was set at 25 °C and separation was conducted according to the conditions previously proposed by Gómez-Caravaca et al. [23 (link)].
+ Open protocol
+ Expand
10

Phosphorylation Site Mapping of Sirt1 by LKB1

Check if the same lab product or an alternative is used in the 5 most similar protocols
To prepare samples for mass spectrometric analysis of phosphorylation site(s) of Sirt1 by LKB1, Sirt1 protein purified from E. coli was incubated with recombinant LKB1-STRADα-Mo25α complex in the presence of ATP and then separated by SDS-PAGE and depicted with colloidal Coomassie blue staining. Following reduction with dithiothreitol and alkylation with iodoacetamide, in-gel digestion of Sirt1 was performed with sequencing-grade Glu-C (Promega, V1651) at 1/50 (Glu-C/protein, w/w) ratio at 37 °C for 16 h in 50 mM NH4HCO3 (pH 8.0). The samples were then dried and dissolved in 0.1% formic acid (FA) water. LC-MS analysis was performed on ACQUITY UPLC M-Class (Waters) and XEVO G2-XS QTof Mass Spectrometer System (Waters). The Sirt1 protein samples were separated by a 120-min gradient with the linear gradient from 3% to 80% B (A = 0.1% FA in H2O, B = 0.1% FA in acetonitrile) at a flow rate of 300 nl/min and the Sirt1-669-peptide samples were separated by a 30-min gradient with the linear gradient from 3% to 80% B (A = 0.1% FA in H2O, B = 0.1% FA in acetonitrile) at a flow rate of 300 nl/min. The QToF mass spectrometer was operating in MSE mode or target MSMS mode, and the raw data were searched with Progenesis QI for proteomics V4.1 Software against the Human uniprot database.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!