The largest database of trusted experimental protocols

Chromeleon version 6

Manufactured by Thermo Fisher Scientific
Sourced in United States

Chromeleon version 6.8 is a chromatography data system software developed by Thermo Fisher Scientific. The core function of Chromeleon 6.8 is to provide a comprehensive platform for data acquisition, processing, and reporting for various chromatography techniques, including HPLC, UHPLC, and IC.

Automatically generated - may contain errors

12 protocols using chromeleon version 6

1

Lysozyme Aggregation Assessment via SEC

Check if the same lab product or an alternative is used in the 5 most similar protocols
Size exclusion chromatography (SEC) was performed using a Dionex UltiMate 3000 HPLC System (Thermo Scientific, Sunnyvale, CA) coupled with a TSKgel G3000SWXL column (Tosoh Bioscience LLC, Tokyo, Japan). A previously published method was utilized,21 (link) which consisted of an injection volume of 75 μL and isocratic elution at a rate of 0.5 mL/min using a buffer of 100 mM sodium phosphate and 250 mM chloride adjusted to pH 6.8. UV absorbance at 280 nm was measured over the course of 35 min. Prior to injection, samples were diluted to approximately 1.2 mg/mL of lysozyme. To determine if aggregation of lysozyme occurred, the retention time of the peak in the sample chromatogram was compared against a known standard. Relative retention time (RRT) was calculated by dividing the lysozyme standard main peak retention time by the sample main peak retention time. The percent area of each peak in the chromatogram represented the area as a percentage of the total area of all peaks. A decrease in RRT or the appearance of earlier eluting peaks (represented by a decrease in percent peak area of the peak corresponding to the standard peak) was indicative of protein aggregation. Chromeleon Version 6.80 software (Thermo Scientific, Sunnyvale, CA) was used to process all chromatography data.
+ Open protocol
+ Expand
2

HPLC Analysis of CSP7 Compound

Check if the same lab product or an alternative is used in the 5 most similar protocols
CSP7 was chemically analyzed with a Thermo Scientific™ Dionex™ UltiMate™ 3000 HPLC System (Thermo Scientific, Sunnyvale, CA, USA) with a wavelength of 220 nm. The two mobile phases were designated Phase A (0.1% trifluoroacetic acid in water) and Phase B (0.09% trifluoroacetic acid in a mixture of 20:80 water and acetonitrile). The HPLC gradient that operated at 1.0 mL·min−1 ran from 25% to 35% for mobile Phase B for 20 min then equilibrated at 25% for 5 min. An Ultimate 3000 Autosampler was utilized to inject 20 μL samples. Injections were passed through a Phenomenex® Luna 5 μm C18(2) 100 Å, 150 mm × 4.6 mm (Phenomenex, Torrance, CA, USA), maintained at ambient temperature. The retention time of CSP7 was approximately 8.3 min. The standard curve of the CSP7 (concentrations of 0.015625–1.00 mg·mL−1) were prepared. All analyses maintained linearity in the range tested. Chromeleon Version 6.80 software (Thermo Scientific) was used to process all chromatography data.
+ Open protocol
+ Expand
3

UHPLC-Orbitrap MS Analysis Protocol

Check if the same lab product or an alternative is used in the 5 most similar protocols
The analysis was carried out using a UHPLC (Dionex Ultimate 3000, Thermo Fisher Scientific, Waltham, MA, USA) equipped with an autosampler (PAL, CTC Analytics, Zwingen, Switzerland) and coupled to a mass spectrometer Orbitrap Q Exactive quadrupole Focus (Thermo Fischer Scientific, Waltham, MA, USA). The UHPLC was equipped with an Ascentis Express C18 column (150 × 2.1 mm, 2.7 µm, Supelco, Sigma-Aldrich, St. Louis, MO, USA) maintained at 35 °C. Column flow was set at 0.3 mL/min. Gradient elution started with 2% B for 2 min, changed to 98% B in 20 min, and then 98% B was kept for 6 min. Eventually, the mobile phase composition was changed to the starting conditions and kept for 4 min. The measurements were acquired in negative and positive ionization in the full scan data-dependent mode (ddMS2) in the m/z 50–1000 Da range. Details about the mass spectrometer settings are reported in Table S2. Xcalibur™ version 3.0 (Thermo Scientific, Waltham, MA, USA) and Chromeleon™ version 6.8 (Thermo Scientific, Waltham, MA, USA) software were used for instrument control.
+ Open protocol
+ Expand
4

HPLC Analysis of Colon Content in Piglets

Check if the same lab product or an alternative is used in the 5 most similar protocols
Colon content (25 mg) from piglets of two breeds (n = 7) was extracted with 1 mL of 0.3 M HClO4, and sonicated once for 5 min after vortex mixing, followed by centrifugation for 5 min (16,090 g, 4°C). The supernatant (160 μL) was then transferred to a new tube containing 2 M KOH solution and equilibrated at 4°C for 3 h. After centrifugation (5 min, 16,090 g, 4°C), the supernatant was passed through a membrane filter (0.22 μm), and 150 μL was transferred into a 2 mL glass vial for HPLC analysis. HPLC analysis was performed using methanol: 50 mM KH2PO4 buffer solution (9:91, v:v) (pH 6.5) as the mobile phase with a flow rate of 0.5 mL/min, and detected using a UV detector at a wavelength of 259 nm. Finally, data processing was performed using Chromeleon version 6.8 (Thermo Fisher).
+ Open protocol
+ Expand
5

Ammonium Ion Analysis by Cation Exchange Chromatography

Check if the same lab product or an alternative is used in the 5 most similar protocols
Samples were run on a Dionex ICS-3000 analytical system controlled by Chromeleon version 6.8 software (Thermo Scientific, Sunnyvale, CA). The autosampler injected 25μL of sample onto a 4 x 250 mm IonPac CS12A cation exchange column preceded by an IonPac CG12A guard column (4 x 50 mm) (Thermo Scientific, Sunnyvale, CA). Isocratic separation of the ammonium ion was achieved using a 20mM methanesulphonic acid (MSA) eluent prepared by the EG40 eluent generator (Thermo Scientific, Sunnyvale, CA) that required only a deionized water source. The column temperature was maintained at 25°C. The pumps and eluent generators were turned on at least 60 min before the first injection to allow baseline stabilization. Ammonium ions were detected using a conductivity detector. Prior to detection by the conductivity detector, the MSA eluent conductance was suppressed by the cation self-regenerating suppressor (CSRS). All ammonium values generated in Chromeleon were transferred to Microsoft Excel 2010 for further analysis. Statistical evaluations were done using JMP Software (SAS Institute Inc., Cary NC).
+ Open protocol
+ Expand
6

UHPLC-Fluorescence and MS Analysis of Oligosaccharides

Check if the same lab product or an alternative is used in the 5 most similar protocols
Labelled OS were analysed using a Dionex Ultimate 3000RS Ultra High Performance Liquid Chromatography (UHPLC) system coupled to a Dionex RF2000 fluorescence detector (both Thermo, Sunnyville, USA) and a Q-Trap 4000 Mass spectrometer (AB Sciex, Framingham, MA). Separation was achieved on a BEH Glycan analytical column (2.1×150 mm, 1.7 µm, Waters, Milford, MA) using one of three different methods described below. For two of the methods (LC methods 1 and 2) a Van Guard BEH Amide guard column (1.7 µm; 2.1×5 mm, Waters) was used as a trapping column to capture the 2AB labelled OS and remove excess labelling reagents [23] (link). For the third method (LC method 3) excess reagents were not removed and a guard column was not employed. For methods 1 and 2 the system was controlled by Analyst version 1.5 (AB Sciex) with DCMS-link (Thermo) to control the chromatography system. Method 3 was used without the mass spectrometer and in this case the system was controlled by Chromeleon version 6.8 (Thermo).
+ Open protocol
+ Expand
7

Fecal Short-Chain Fatty Acids Quantification

Check if the same lab product or an alternative is used in the 5 most similar protocols
Fecal SCFAs extraction was performed following De Baere et al.,66 (link) protocol. SCFAs concentrations were measured using high-performance liquid chromatography UltiMate 3000 UHPLC system (HPLC-UV) (Thermo Scientific, Breda, The Netherlands) with external calibration standards curve method as previously described in details.67 (link)Briefly, calibration standards were prepared at concentrations ranging from 0.5 mM to 50 mM for acetic acid (AA), butyric acid (BA), propionic acid (PA), and succinic acid (SA) as internal standard (all purchased from Sigma-Aldrich, St. Louis, MO, USA). After chromatographic separation testing on a Hypersil Gold aQ column (150 × 4.6 mm i.d.) with a 3 μm particle (Thermo Scientific, Breda, The Netherlands), HPLC-UV was performed on thermostated and guard column protected HPLC columns using UV detection at 210 nm. The mobile phase consisted of 20 mM of sodium dihydrogen phosphate (Merck, Darmstadt, Germany) in HPLC water (pH 2.2) (Merck) (A) and HPLC grade acetonitrile (Sigma-Aldrich, St. Louis, MO, USA) (B).
HPLC-UV data were processed using Chromeleon version 6.8 software (Thermo Fisher Scientific, MA, USA). SCFAs concentration were calculated using mathematical equation: SCFA (AA, BA, PA) = (organic acid in fecal sample × 6 × 10−3)/(succinic acid in fecal sample × fecal sample mass) × 1000 [mmol/kg]. All measurements were done in triplicate.
+ Open protocol
+ Expand
8

Monosaccharide Analysis by HPAEC-PAD

Check if the same lab product or an alternative is used in the 5 most similar protocols
The monosaccharides released by enzymatic hydrolysis were analyzed using a high-performance anion-exchange chromatography system coupled with a pulsed amperometric detector (HPAEC-PAD) (Dionex ISC-3000, Sunnyvale, CA, USA). A CarboPac PA1 analytical column (250 mm × 4 mm) and a guard column (50 mm × 4 mm) were used. Prior to analysis, the samples were boiled for 10 min to denature proteins, filtered through a 0.2 µm nylon filter, and the samples were diluted with Milli-Q water. Water was used as eluent at 1 ml min−1, and 300 mM NaOH was added at 0.5 ml min−1 before the detector. The column was cleaned with 200 mM NaOH dissolved in 170 mM sodium acetate. Sample injection was 25 µl, and the column was maintained at 30 °C during analysis. Calculations were performed using Chromeleon version 6.8 (Dionex).
+ Open protocol
+ Expand
9

HPLC Analysis of Glucose and Organic Acids

Check if the same lab product or an alternative is used in the 5 most similar protocols
The glucose and organic acids were measured by High Performance Liquid Chromatography—HPLC, Ultimate 3000 (Dionex, Sunnyvale, CA, USA). Samples were diluted five or ten times in sulfuric acid (50 mM) and filtrated in the membrane at 0.45 μm before being loaded automatically in the aminex column HPX—87H (300 mm × 7.8 mm; Bio Rad, Hercules, CA, USA) at 60 °C, with an UV (210 nm) and RI detector; the mobile phase was sulfuric acid at 5 mM, and a flow rate of 0.6 mL/min was used. The integration of chromatographic peaks to determine the concentration determination was calculated through the software Chromeleon, version 6.8 (Dionex, Sunnyvale, CA, USA).
+ Open protocol
+ Expand
10

Carotenoid and Vitamin Separation by HPLC

Check if the same lab product or an alternative is used in the 5 most similar protocols
Separation of the carotenoids and vitamins from all the extracts was accomplished using a high-performance liquid chromatography (HPLC) system (DionexTM, Thermo Fisher Scientific, CA, USA) equipped with a Diode Array Detector UltiMate 3000 (Dionex, Thermo Fischer Scientific, CA, USA) monitored at 450 nm for carotenoids and 290 and 220 nm for vitamins. Separation of the carotenoids was achieved using a YMC Carotenoid C30 column (250 × 4.6 mm, 3 μm, YMC America, Devens, MA, USA) at ambient temperature. Two mobile phases of A (MeOH/MTBE/H2O = 81/15/4) and B (MeOH/MTBE/H2O = 6/90/4) were used at elution gradient of 1–100% B for 90 min and at a flow rate of 1 mL/min. Separation of the vitamins was achieved using a XTerra MS C18 column (250 × 4.6 mm, 5 μm, Waters Limited, Wilmslow, Cheshire, England) at 40 °C. Phosphate buffer (10 mM TBAOH and 10 mM KH2PO4) with a pH of 5.2 and acetonitrile with the ratio of 90:10 v/v were used as mobile phases for the water-soluble vitamins at a flow rate of 0.8 mL/min. For the fat-soluble vitamins, acetonitrile and MeOH with a ratio of 60:40 v/v were used at a flow rate of 1 mL/min. Dionex Chromeleon version 6.8 software was used for data interpretation.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!