The largest database of trusted experimental protocols

Acquity uplc beh c18 1.7 m column

Manufactured by Waters Corporation
Sourced in United States

The ACQUITY UPLC BEH C18 1.7 µm column is a high-performance liquid chromatography column developed by Waters Corporation. It features a 1.7 µm particle size and a C18 stationary phase, designed for use in ultra-performance liquid chromatography (UPLC) systems.

Automatically generated - may contain errors

13 protocols using acquity uplc beh c18 1.7 m column

1

SCFA Quantification by LC-MS/MS

Check if the same lab product or an alternative is used in the 5 most similar protocols
In total, 10 µL of the diluted SCFA derivatives was injected into a RSLC UltiMate 3000® system (ThermoFisher, Waltham, MA, USA) coupled on-line with a QTRAP 5500® mass spectrometer (AB Sciex, Framingham, MA, USA). Chromatographic separation of SCFAs was performed on an Acquity UPLC BEH C18 column (1.7 µm; Waters, Eschborn, Germany) using A: H2O (0.01% formic acid, FA) and B: acetonitrile (0.01% FA) as the mobile phases. The flow rate was set to 0.35 mL/min, the column temperature held at 40 °C. The gradient elution was performed as follows: 2 min at 15% B, 15–50% B in 15 min, then held at 100% B for 1 min. Finally, the column was equilibrated for 3 min at 15% B. For identification and quantitation, a scheduled MRM method was used, with specific transitions for every SCFA. Peak areas were determined in Analyst® Software (v1.6.2, AB Sciex) and areas for single SCFAs were exported. Metabolite quantification with calibration curves, normalization to OD600 and statistical analysis using an ANOVA followed by a Tukey’s HSD were performed with in-house written R scripts in RStudio (Version 4.0.2).
+ Open protocol
+ Expand
2

Quantification of Prostanoids and CIII

Check if the same lab product or an alternative is used in the 5 most similar protocols
Extracted samples were reconstituted in 50 µL of 20% acetonitrile in Milli-Q water. The samples were centrifuged at 3000 g for 5 minutes at 4°C and the supernatants were transferred to glass vials. Analytes of interest were quantified using a triple quadrupole mass spectrometer (ACQUITY TQ detector, Waters) equipped with an ACQUITY H-class ultraperformance (UP)LC (Waters). Separation was performed on a 50 × 2.1 mm ACQUITY UPLC BEH C18 column 1.7 µm (Waters) with a 12 minutes stepwise linear gradient (20%-95%) at a flow rate of 0.6 mL/min with 0.05% FA in acetonitrile as mobile phase B and 0.05% FA in Milli-Q as mobile phase A. Prostanoids were quantified in negative mode and CIII in positive mode 23 with multiple reaction monitoring methods. Data were analyzed using MassLynx software, version 4.1, with internal standard calibration and quantification to external standard curves.
+ Open protocol
+ Expand
3

Proteomic Analysis of Ultrasound-Stimulated Exosomes

Check if the same lab product or an alternative is used in the 5 most similar protocols
HAs were grown to 80% confluency in Falcon T-125 flasks. Cells were treated by ultrasound stimulation as described. Exosomes were isolated from the supernatants by ultracentrifugation. Samples were stored at -80 °C until required. Proteins were extracted, the concentration was measured by the BCA assay, and the samples were analyzed by SDS-PAGE. Samples were reductively alkylated, and trypsin was used for enzymatic hydrolysis. The peptides were labeled with iTRAQ 8PLEX (SCIEX) and the equally mixed labeled peptides were pre-separated by UHPLC (Thermo SCIENTIFIC Vanquish) with C18 reversed-phase column (ACQUITY UPLC BEH C18 Column 1.7 µm, 2.1 mm×150 mm, Waters, USA). Subsequently, liquid chromatography-tandem mass spectrometry (LC-MS-MS) was performed for the identification and quantification of proteins. After searching for the Sequence or Mascot module in Proteome Discoverer Software 2.2, statistical and bioinformatic analyses of the results were performed.
+ Open protocol
+ Expand
4

Rapid Protein Characterization by HDX-MS

Check if the same lab product or an alternative is used in the 5 most similar protocols
Protein samples were rapidly thawed and injected onto an integrated fluidics system containing a HDX-3 PAL liquid handling robot and climate-controlled chromatography system (LEAP Technologies), a Dionex Ultimate 3000 UHPLC system, as well as an Impact HD QTOF Mass spectrometer (Bruker). Proteins were run over two immobilized pepsin columns (Applied Biosystems; Poroszyme Immobilized Pepsin Cartridge, 2.1 mm × 30 mm; Thermo-Fisher 2‐3131‐00; at 10 and 2°C, respectively) at 200 μL/min for 3 min. The resulting peptides were collected and desalted on a C18 trap column (Acquity UPLC BEH C18 1.7 µm column [2.1 × 5 mm]; Waters 186002350). The trap was subsequently eluted in line with a C18 reverse-phase separation column (Acquity 1.7 μm particle, 100 × 1 mm2 C18 UPLC column, Waters 186002352), using a gradient of 5–36% B (buffer A 0.1% formic acid; buffer B 100% acetonitrile) over 16 min. Lipids were directly captured on the LC system, and eluted off at the 100% acetonitrile step, with no interference on mass spectrometer or LC performance. Full details of the LC setup and gradient for lipid samples are in Stariha et al., 2021 (link). Mass spectrometry experiments were performed on an Impact II QTOF (Bruker) acquiring over a mass range from 150 to 2200 m/z using an electrospray ionization source operated at a temperature of 200°C and a spray voltage of 4.5 kV.
+ Open protocol
+ Expand
5

HPLC-MS Analysis of Botanical Extracts

Check if the same lab product or an alternative is used in the 5 most similar protocols
Dried extracts were dissolved in methanol prior to the mass spectrometry (MS) analysis. MS experiments were carried out on a Dionex Ultimate 3000 UPLC system (ThermoFisher Scientific, Waltham, MA, USA) coupled to PDA detector (stationary phase 30 or 100 mm ACQUITY UPLC BEH C18 1.7 µm column (Waters Corporation, Milford, MA, USA), mobile phase: Linear gradient of [A] ddH2O + 0.1% formic acid/[B] acetonitrile + 0.1% formic acid, 5% to 95% at flow rate of 0.6 mL/min). Further mass detection was performed coupling either an amaZon speed (Bruker, Billerica, MA, USA) or LTQ Orbitrap XL mass spectrometer (ThermoFisher Scientific, Waltham, MA, USA) using positive ionization mode and mass range detection of m/z 200 to 2000. Data analysis was performed using software Compass Data Analysis v. 4.1 (Bruker) and Xcalibur v. 3.0 (ThermoFisher Scientific).
+ Open protocol
+ Expand
6

Polyamine and Amino Acid Analysis

Check if the same lab product or an alternative is used in the 5 most similar protocols
Polyamines and amino acids were analyzed after derivatization with 6-Aminoquinolyl-N-hydroxysuccinimidyl carbamate (AccQ-Tag Ultra Derivitization Kit, Waters, Mildorf, USA) according to the manufacturer’s instructions. Chromatographic separation was achieved using an Acquity UPLC BEH C18 1.7 µm column (2.1 x 50 mm, Waters, Mildorf, USA) with a pre-column. One microliter of sample was injected before running the solvent gradient: 0.01% B for 0.54 min, then up to 9.1% B in 6.5 min, 2 min at 21.2% B followed by 0.4 min at 59.6% B and then back to initial conditions in 0.6 min (total run time 10.1 min). The column was maintained at 55°C with a flow rate of 0.7 mL/min. The source temperature was set to 150°C and the desolvation temperature was set to 650°C. The capillary voltage was set to 3.0 kV. Nitrogen was used as the drying and nebulizing gas, with 600 L/h gas flow and 1200 L/h desolvation gas flow.
+ Open protocol
+ Expand
7

DNPM Production Optimization in S. aurantiacus

Check if the same lab product or an alternative is used in the 5 most similar protocols
S. aurantiacus LU19075 was grown in 20 mL of TSB in a 100 mL baffled flask for one day, and 1 mL of the culture was inoculated into 100 mL of DNPM production medium in a 500 mL baffled flask. Cultures were grown for 6 days at 28 °C and 180 rpm in an Infors multitron shaker (Infors AG, Basel, Switzerland). Metabolites were extracted from the supernatant with butanol, concentrated to dryness, and dissolved in 1 mL methanol (MeOH). One microliter was separated on a Dionex Ultimate 3000 UPLC system (Thermo Fisher Scientific, Waltham, MA, USA) equipped with an ACQUITY UPLC BEH C18 1.7 µm column (30, 50, or 100 mm, Waters Corporation, Milford, MA, USA) using a linear gradient of 5–95 vol% aqueous acetonitrile (ACN) with 0.1% formic acid (FA) at a flow rate of 0.6 mL/min and a column oven temperature of 45 °C. Sample analysis was carried out on a coupled PDA detector followed by an amaZon speed (Bruker, Billerica, MA, USA) for production control. High-resolution masses were obtained from either an LTQ Orbitrap XL mass spectrometer (Thermo Fisher Scientific, Waltham, MA, USA) or a MaXis high-resolution LC-QTOF system (Bruker, Billerica, MA, USA) using positive ionization mode and mass range detection of m/z 200 to 2000. Data analysis was performed using Compass Data Analysis v. 4.1 (Bruker) and Xcalibur v. 3.0 (Thermo Fisher Scientific).
+ Open protocol
+ Expand
8

UPLC-QTOF Quantification of VRC and Omeprazole

Check if the same lab product or an alternative is used in the 5 most similar protocols
The concentrations of VRC and omeprazole were determined using a Waters ACQUITY UPLC- QTOF system (Waters, Milford, MA, USA). An ACQUITY UPLC BEH C18 1.7 µm column (Waters) was used at a temperature of 40°C. Chromatographic separation was achieved using a gradient mobile phase consisting of 0.01% formic acid in water (A) and acetonitrile (B) at a flow rate of 0.3 mL/min as follows: VRC, 5–99% B (0–5.0 min); 99% B (5.0–7.0 min); 99–5% B (7.0–7.1 min); 5% B (7.1–10.0 min); omeprazole, 10–85% B (0–1.5 min); 85% B (1.5–6.0 min); 85–10% B (6.0–7.0 min), and 10% B (7.0–10.0 min). The analytes were quantified by electrospray ionization in the MS scan mode. Detail method of VRC and N-oxide UPLC-MS/MS analysis can refer to the Supplementary File 1.
+ Open protocol
+ Expand
9

Quantification of DAF-FM-T in Cells

Check if the same lab product or an alternative is used in the 5 most similar protocols
To detect DAF-FM-T, a theoretical product of NO and DAF-FM, in the DAF-FM DA-treated cells, an LC–MS/MS analysis was performed. The cell extracts were obtained as described above. Methanol was added to each cell extract to produce a 60% final concentration. The cell extracts were filtered and centrifuged at 12,000 g for 5 min to remove proteins, and the supernatant was used as the LC–MS/MS sample. LC–MS/MS analysis using an Acquity UPLC BEH C18 1.7-µm column (Waters, Milford, MA) was performed with an LCMS™-8045 (Shimadzu Corp., Kyoto, Japan). A gradient elution of 5–100% acetonitrile was used as the mobile phase, with a flow rate of 0.3 mL/min. The positive electrospray ionization mode was selected with fragments of m/z 413.0 > 369.0 (DAF-FM), 424.0 > 332.0 (DAF-FM-T), and 437.0 > 373.0 (UNK436). The standard DAF-FM-T was prepared in vitro as described previously using DAF-FM (Goryo Chemical Inc.) and a NO-donor38 (link).
+ Open protocol
+ Expand
10

UPLC-MS Analysis of Crude Extracts

Check if the same lab product or an alternative is used in the 5 most similar protocols
Dry crude extracts were dissolved in methanol prior to MS analysis. Sample compounds were separated on a Dionex UltiMate 3000 UPLC (Thermo Fisher Scientific) coupled to a PDA detector (stationary phase: 100 mm ACQUITY UPLC BEH C18 1.7 µm column (Waters Corporation, Milford, MA, USA), mobile phase: linear gradient of [A] ddH2O + 0.1% formic acid/[B] acetonitrile + 0.1% formic acid, 5% to 95% at flow rate 0.6 mL/min). Mass detection was carried out by coupling to an amaZon speed (Bruker, Billerica, MA, USA) or a MaXis II (Bruker) mass spectrometer, which provides a mass accuracy of 5 ppm. Spectra were aquired using positive ionization mode and with a mass selection range of m/z 100 to 2000. Compass Data Analysis 4.2 was used for mass spectrometry data analysis.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!