The largest database of trusted experimental protocols

Q exactive hf orbitrap mass spectrometer

Manufactured by Thermo Fisher Scientific
Sourced in Germany, United States

The Q Exactive HF Orbitrap mass spectrometer is a high-resolution, accurate-mass (HRAM) instrument designed for advanced proteomics, metabolomics, and small molecule analysis. It features a high-field Orbitrap mass analyzer that provides high resolving power and mass accuracy. The instrument is capable of performing both full-scan and data-dependent acquisition modes.

Automatically generated - may contain errors

142 protocols using q exactive hf orbitrap mass spectrometer

1

Label-free Quantification of EVs and WCL

Check if the same lab product or an alternative is used in the 5 most similar protocols
For label-free quantification, 2 μg of each EV and WCL digest were analyzed by nanoLC-MS/MS on an Orbitrap Q-Exactive HF Mass Spectrometer (ThermoFisher Scientific, Bremen) coupled to an Agilent 1290 Infinity Ultra-High Pressure Liquid Chromatography (UHPLC) system (Agilent Technologies), operating in reverse phase (C18) equipped with a Reprosil pur C18 trap column (100 µm × 2 cm, 3 µm, Dr. Maisch) and a Poroshell 120 EC C18 analytical column (75 µm × 50 cm, 2.7 µm, Agilent Technologies). After trapping for 5 min in a flow rate of 0.05 ml/min in 100% solvent A (0.1% FA in H2O), peptides were eluted with a 160 min LC gradient from 10 to 36% solvent B (0.1% FA, 80% ACN) at a flow rate of 300 nL/min. The mass spectrometer was operated in data-dependent acquisition mode, automatically switching between MS and MS2. Full scan MS spectra were acquired using the following settings: full-scan automatic gain control (AGC) target 3e6 at 60,000 resolution; scan range 375–1600 m/z; Orbitrap full-scan maximum injection time 20 ms. HCD MS2 spectra were generated for up to 12 precursors with a normalized collision energy of 27%. The fragment ions were acquired at a resolution of 30,000 (isolation window of 1.4 m/z) with an AGC target value of 1e5 charges and a maximum injection time of 100 ms. The dynamic exclusion was set to 24 s.
+ Open protocol
+ Expand
2

Quantitative Liquid Chromatography-Mass Spectrometry

Check if the same lab product or an alternative is used in the 5 most similar protocols
Each sample was analyzed as technical duplicate on a LC–ESI–MS-MS quadrupole Orbitrap QExactive-HF mass spectrometer (Thermo Fisher Scientific, Milan, Italy). Separation of peptides was achieved on a linear gradient from 93% solvent A (2% ACN, 0.1% formic acid) to 60% solvent B (80% acetonitrile, 0.1% formic acid) over 110 min, and from 60% to 100% solvent B in 10 min at a constant flow rate of 0.25 µL/min on UHPLC Easy-nLC 1000 (Thermo Fischer Scientific, Milan, Italy) connected to a 23 cm fused-silica emitter of 75 µm ID (New Objective, Inc. Woburn, MA, USA), packed in-house with ReproSil-Pur C18-AQ 1.9 µm beads (Dr Maisch Gmbh, Ammerbuch, Germany) using a high-pressure bomb loader (Proxeon, Odense, Denmark). MS data were acquired using a data-dependent top 20 method for HCD fragmentation. Survey full scan MS spectra (300–1650 Th) were acquired in the Orbitrap with resolution 60,000, AGC target 3e6, IT 20 ms. For HCD spectra, resolution was set to 15,000 at m/z 200, AGC target 1e5, IT 80 ms; NCE 28%, dynamic exclusion 20 s and isolation width 1.2 m/z [63 (link)].
+ Open protocol
+ Expand
3

High-Throughput Proteomic Analysis by Nano-LC-MS/MS

Check if the same lab product or an alternative is used in the 5 most similar protocols
For mass spectrometry analysis, tryptic peptides were loaded on a reverse phase column with an inner diameter of 75 μm packed with 1.9 μm C18 beads using the autosampler of the Thermo Easy LC system (Thermo Scientific). Temperature of the column was kept at a constant 50oC in a column oven (Sonation). Peptides were loaded in buffer A (0.1% formic acid), separated with a 130 min gradient of 5%–30% buffer B (80% ACN and 0.1% formic acid) at a flowrate of 300 (or 250) nl/min and directly applied onto a benchtop Orbitrap Q Exactive HF mass spectrometer (Thermo Scientific) via electrospray. The Q-Exactive HF was operated in data-dependent mode with survey scans at mass range of 300 to 1650 m/z. Up to the 10 or 15 most abundant precursor patterns from the survey scan were selected and fragmented by higher energy collisional dissociation. MS/MS spectra were acquired with a resolution of 15,000 (FWHM), at a maximum injection time of 50 ms, and a target value of 1e5 charges.
+ Open protocol
+ Expand
4

UHPLC-MS/MS Analysis of Metabolites

Check if the same lab product or an alternative is used in the 5 most similar protocols
Ultra high performance liquid chromatography tandem-mass spectrometry (UHPLC-MS/MS) analysis was performed on the Vanquish UHPLC system (Thermo Fisher Scientific, Massachusetts, USA) coupled with an Orbitrap Q Exactive™ HF mass spectrometer (Thermo Fisher Scientific, Massachusetts, USA). In detail, chromatographic separation was conducted in Vanquish UHPLC system equipped with a Hypersil GOLD columns (100 mm × 2.1 mm, 1.9 μm) using a 17-min linear gradient at a flow rate of 0.2 mL/min. The eluents for the positive polarity mode were eluent A (0.1% FA in water) and eluent B (methanol). The eluents for the negative polarity mode were eluent A (5 mM ammonium acetate, pH 9.0) and eluent B (methanol). The solvent linear gradient was set as follows: 2% B, 1.5 min; 2%−100% B, 12.0 min; 100% B, 14.0 min; 100%−2% B, 14.1 min; 2% B, 17 min. The ESI-MSn experiments were executed on a Q ExactiveTM HF mass spectrometer in positive/negative polarity mode with spray voltage of 3.2 kV. Sheath gas and auxiliary gas flow e were set at 40 and 10 arbitrary units, respectively, and the capillary temperature was 320°C. The analyzer scanned over a mass range of 100 m/z–15,000 m/z for full scan at a mass resolution of 70,000. Finally, data-dependent acquisition MS/MS experiments were performed.
+ Open protocol
+ Expand
5

UHPLC-MS/MS Metabolite Profiling Protocol

Check if the same lab product or an alternative is used in the 5 most similar protocols
In Novogene Co., Ltd. (Beijing, China), UHPLC–MS/MS analyzes were carried out utilizing a Vanquish UHPLC system (Thermo Fisher, Germany) paired with an Orbitrap Q Exactive™HF mass spectrometer (Thermo Fisher, Germany). A 17-min linear gradient was used to inject samples onto a HypesilGoldcolumn (100 × 2.1 mm, 1.9 ) at a flow rate of 0.2 mL/min. Eluents A (0.1% formic acid (FA) in water) and B (methanol) were used in the positive polarity mode. Eluents A (5 mM ammonium acetate, pH 9.0) and B (Methanol) were used in the negative polarity mode. The following settings were made for the solvent gradient: 2% B, 1.5 min; 2–85% B, 3 min; 85–100% B, 10 min; 100–2% B, 10.1 min; 2% B, 12 min. With a spray voltage of 3.5 kV, capillary temperature of 320°C, sheath gas flow rate of 35 psi, aux gas flow rate of 10 L/min, S-lens RF level of 60, and aux gas heater temperature of 350°C, the QExactiveHF mass spectrometer was operated in positive/negative polarity mode.
+ Open protocol
+ Expand
6

Quantitative Proteomics by LC-MS/MS

Check if the same lab product or an alternative is used in the 5 most similar protocols
LC-MS/SM analysis of digested peptides was performed on an Orbitrap Q Exactive HF mass spectrometer (Thermo Fisher Scientific, Bremen) coupled to an EASY-nLC 1200 (Thermo Fisher Scientific). Peptides were loaded and separated at 250 nl/min on a home-made C18 50 cm capillary column picotip silica emitter tip (75 μm diameter filled with 1.9 μm Reprosil-Pur Basic C18-HD resin, (Dr. Maisch GmbH, Ammerbuch-Entringen, Germany)) equilibrated in solvent A (2% ACN, 0.1% FA). Peptides were eluted using a gradient of solvent B (80% ACN, 0.1% FA) from 3% to 6% in 5 min, 6% to 29% in 130 min, 29% to 56% in 26 min, 56% to 90% in 5 min (total length of the chromatographic run was 180 min including high ACN level steps and column regeneration). Mass spectra were acquired in data-independent acquisition mode with the XCalibur 4.1.31.9 software (Thermo Fisher Scientific, Bremen).
Each cycle was built up as follows: one full MS scan at resolution 30 000 (scan range between 400 and 1200 m/z), AGC was set at 3*106, ion trap was set at 50 ms. All MS1 was followed by 40 isolation windows of 20 m/z, covering the MS1 range from 400 m/z to 1200 m/z. The AGC target was 2*105, and NCE was set to 27. All acquisitions were done in positive and profile mode.
+ Open protocol
+ Expand
7

Cecal Metabolomic Profiling via LC-MS/MS

Check if the same lab product or an alternative is used in the 5 most similar protocols
The cecal samples were prepared and the supernatant was injected into a liquid-mass spectrometry (LC–MS/MS) system for analysis (Want et al., 2012 (link); Yuan et al., 2012 (link)). Furthermore, the ultra-high-performance liquid Chromatography (UHPLC)-MS/MS analyses were performed using a Vanquish UHPLC system (Thermo Fisher, Germany) coupled with an Orbitrap Q Exactive™ HF mass spectrometer (Thermo Fisher, Germany) at Novogene Co. Ltd. (Beijing, China). The cecal samples were then injected onto a Hypersil Gold column (100 × 2.1 mm, 1.9 μm) using a 17 min linear gradient at a flow rate of 0.2 ml/min. The eluents for the positive polarity mode were eluent A (0.1% FA in Water) and eluent B (Methanol). The eluents for the negative polarity mode were eluent A (5 mM ammonium acetate, pH 9.0) and eluent B (Methanol), and the solvent gradient was set as follows: 2% B, 1.5 min; 2–100% B, 12.0 min; 100% B, 14.0 min; 100–2% B, 14.1 min; and 2% B, 17 min. Then, the Q ExactiveTM HF mass spectrometer was operated in positive/negative polarity mode with spray voltage of 3.2 kV, capillary temperature of 320°C, sheath gas flow rate of 40 arb, and aux gasflow rate of 10 arb.
The raw data files generated by UHPLC–MS/MS were processed using Compound Discoverer 3.1 (CD3.1, Thermo Fisher) to perform peak alignment, peak picking, and quantitation for each metabolite.
+ Open protocol
+ Expand
8

DESI-MS Imaging of Tissue Sections

Check if the same lab product or an alternative is used in the 5 most similar protocols
All DESI-MS analyses were performed using a laboratory-built DESI sprayer fitted to a Thermo Orbitrap Q-Exactive HF mass spectrometer at a resolving power of 60,000 in the negative ion mode in the mass range m/z 100 to 1500. Tissue sections were thawed and dried in a chemical fume hood 10 min prior to DESI-MS analysis. DESI-MS imaging of tissue sections was performed at a spatial resolution of 150 μm using a spray flow rate of 1.5 μl min−1 with a histologically compatible solvent system, acetonitrile:dimethylformamide 3:1 (v/v). The pressure of the nebulizing N2 gas was set to 180 psi. DESI-MS ion images were constructed using FireFly (Thermo Scientific) and BioMAP (Novartis) software. Ion identifications were made with high mass accuracy measurements (≤5 ppm mass error) and/or tandem MS experiments (Figs. S4 and S9) using high-energy collision dissociation. METASPACE and LIPID MAPS databases were used to aid in molecular identification.
+ Open protocol
+ Expand
9

UHPLC-MS/MS Metabolomics Analysis

Check if the same lab product or an alternative is used in the 5 most similar protocols
All analyses were performed on a Vanquish UHPLC system (Thermo Fisher) coupled with an Orbitrap Q Exactive HF mass spectrometer (Thermo Fisher). The raw data generated by UHPLC-MS/MS were processed using the Compound Discoverer (Thermo Fisher) to perform peak alignment, peak picking, and quantitation for each metabolite. After that, peak intensities were normalized to the total spectral intensity, and the normalized data was further used to predict the molecular formula based on additive ions, molecular ion peaks, and fragment ions. Then, peaks were matched with the Lipidmaps and Lipidblast database to obtain accurate qualitative and relative quantitative results. We applied univariate analysis (t test) to calculate the statistical significance (P value). The metabolites with VIP > 1 and P value < 0.05 and fold change ≥ 2 or ≤ 0.5 were considered to be differential metabolites between LT and LW.
+ Open protocol
+ Expand
10

Targeted Proteomics Validation of Key Proteins

Check if the same lab product or an alternative is used in the 5 most similar protocols
All the hub proteins determined above and the proteins of the top 25 combinations with the smallest root mean squared error (RMSE) values and AUC value of 1 were validated by PRM in independent samples.
First, the proteins were extracted, digested and mixed samples were prepared, and the full spectrum was scanned by the “label-free” method using the EASY-nLC1200 connected to the Orbitrap Q-Exactive HF mass spectrometer (Thermo, Scientific, USA). Second, the Proteome Discoverer 2.2 software was used to search the library. The search results were imported into Skyline(version 20.1.0.155) software [33 ] to obtain the target protein peptide information. Then, the PRM method can be established, and the obtained data were imported into Skyline software for quantification. The parameters of PRM were set as follows: the primary resolution was 12,000 (at 300–1400 m/z) with an automatic gain control (AGC) target value of 3e6, a maximum injection time of 80 ms, and a Normalized Collision Energy (NCE) of 27%; the secondary resolution was 15,000 with an AGC target value of 2e4, a maximum injection time of 19 ms. The mass tolerances for precursor and fragment ions were also set at 10 ppm and 0.02 Da, respectively.
The ROC curve analyses of the validated combinations were examined for PDR, and the AUC of each ROC curve was calculated.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!