The largest database of trusted experimental protocols

Xdb c18 guard column

Manufactured by Agilent Technologies
Sourced in United States

The XDB-C18 guard column is a protective device used in high-performance liquid chromatography (HPLC) systems. Its primary function is to extend the lifetime of the analytical column by trapping particulates, strongly retained compounds, and other contaminants that could otherwise damage the analytical column. The XDB-C18 guard column is designed to be easily replaced, allowing the analytical column to maintain optimal performance for longer periods.

Automatically generated - may contain errors

3 protocols using xdb c18 guard column

1

Quantification of HBAMP in Human Saliva

Check if the same lab product or an alternative is used in the 5 most similar protocols
The protocols in this section were approved by the Ethics Committee of Zhongshan Hospital, Xiamen University. The methods employed were performed in accordance with the approved guidelines. After acquiring informed consent, human saliva was collected from individual volunteers (two males and two females; Minimum age: 22; maximum age: 28) in the morning prior to oral cleaning. The collected saliva samples were centrifuged at 12,000 rpm for 20 min. The supernatant was collected and filtered through the 0.45 mm membrane filter to remove any debris. HBAMP was added into the human saliva to achieve the final concentration of 500 μg/ml. The saliva with HBAMP was incubated at 37 °C. At 0, 5, 20 and 60 min, 1 ml samples were taken from the total sample and served in the specific sample bottles, the machine will automatically extract samples from the bottles to detect. The samples were analysed by reversed-phase HPLC using Zorbax Eclipse XDB-C18 analytical column (150 × 4.6 mm, Agilent Technologies, Inc., Santa Clara, CA, USA) protected by a XDB-C18 guard column (4 × 4 mm). For the elution of HBAMP, a flow rate of 1.2 mL/min and a linear gradient from 88:12 to 65:35 (0.1% TFA in water: 0.1% TFA in acetonitrile) for 10 min were employed. Total run time of HPLC-UV (215 nm) analysis was 15 min and the injection volume was set at 40 μL.
+ Open protocol
+ Expand
2

Detailed Analytical Protocols for Natural Product Characterization

Check if the same lab product or an alternative is used in the 5 most similar protocols
Medium-pressure liquid chromatography was performed on a Buchi Sepacore chromatography system (Flawil, Switzerland). Semi-preparative HPLC was carried out on an Agilent 1100 series with an Agilent DAD spectrophotometer and Agilent XDB-C18 reversed-phase column (5 µm, 250 ×10 mm) with an Eclipse XDB-C18 guard column. Sephadex LH-20 (GE Healthcare Bio-Sciences, Uppsala, Sweden) and ODS (S-50 and 75 µm, YMC Co., Ltd., Kyoto, Japan) were used for regular and medium-pressure column chromatography fractionations. Optical rotations: PE-341 polarimeter. IR spectra: Nicolet-Magna-FT-IR 750 spectrometer. UV spectra: Shimadzu UV-2450 spectrophotometer. HR-ESI-MS: Thermo Fisher Scientific LTQ Orbitrap XL mass spectrometer. 1H and 13C NMR were recorded on a Varian VNMRS-600MHz spectrometer (Agilent Technologies, Santa Clara, CA) operating at 600 MHz for 1H NMR and 150 MHz for 13C NMR at room temperature. The chemical shifts (δ) were reported in ppm and were referenced to the residual solvent peak. The coupling constants (J) were quoted in hertz. NMR data processing was performed using standard Agilent, and ACD/Labs software.
+ Open protocol
+ Expand
3

Metabolite Profiling by RRLC-QTOF

Check if the same lab product or an alternative is used in the 5 most similar protocols
Profiling of tissue/serum metabolites was performed on a RRLC 1260 system (Agilent Technologies, Santa Clara, CA, USA) coupled to a QTOF 6520 detector (Agilent Technologies, Santa Clara, CA, USA), equipped with an electrospray source operating in full scan mode, from 50 to 1000 Da for both positive and negative ionization modes. The gas temperature was set at 350 °C, gas flow of 12 L/min, capillary voltage at 3.5 kV, and fragmentor voltage at 120 V. Two reference masses were used to maintain the mass accuracy during analysis: m/z 121.050873 and m/z 922.009798 in positive mode, and m/z 112.985587 and m/z 980.016375 in negative mode. Sample aliquots of 10 mL were injected on a Sb-Aq column (100 × 2.1 mm, particle size 1.8 mm, Agilent Technologies, Santa Clara, CA, USA), protected by a XDB-C18 guard column (5 × 2.1 mm, particle size 1.8 mm, Agilent Technologies) and heated at 40 °C. The gradient mobile phase consisted of 0.2% acetic acid (v:v in water) (A) and acetonitrile (B). The flow rate was set at 0.3 mL/min. The initial condition was set as 98% phase A and 2% phase B, and the gradient changes as follows: from 2% to 95% phase B in 7 min, 95% phase B for 3 min, and equilibration with 2% phase B for 3 min. The autosampler was kept at 4 °C. Profiling data were treated as described below.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!