The largest database of trusted experimental protocols

Prominence series uflc system

Manufactured by Shimadzu
Sourced in France

The Prominence series UFLC system is an ultra-fast liquid chromatography system manufactured by Shimadzu. It is designed for high-performance and efficient separation of analytes in various applications. The system features a compact design and advanced technology to provide rapid and precise analysis.

Automatically generated - may contain errors

5 protocols using prominence series uflc system

1

Gradient HPLC Analysis of Samples

Check if the same lab product or an alternative is used in the 5 most similar protocols
Gradient HPLC analysis was done by using Shimadzu Prominence series UFLC system with a CBM-20A controller bus module, a LC-20 AD liquid chromatograph, a CTO-20A column oven and a SPD-20A UV-visible detector. UV-visible absorption was measured at 295 nm. 20 μL of sample were loaded in the solvent injection ratio: 95% solvent A – 5% solvent B (A = Milli-Q water/TFA 99.9:0.1 v/v; B = CH3CN/Milli Q water/TFA 90:9.9:0.1 v/v/v) onto a Jupiter C4 column (150 × 4.60 mm, 5 μm, 300 Å, Phenomenex) at a flow rate of 1 mL/min over 5 min. In a second step, samples were eluted by a gradient developed from 5 to 90% of solvent B in solvent A over 15 min. The concentration of solvent B was maintained over 5 min. Then, the concentration of solvent B was decreased to 5% over a period of 5 min to re-equilibrate the system, followed by additional 5 min at this final concentration. Before each sample measurement, a baseline was performed following the same conditions by loading Milli-Q water into the injection loop.
+ Open protocol
+ Expand
2

Multi-Instrumental Analysis of Gadolinium Complexes

Check if the same lab product or an alternative is used in the 5 most similar protocols
HPLC-UV/Vis system: A Shimadzu Prominence series UFLC system with a CBM-20A controller bus module, an LC-20 AD liquid chromatograph, a CTO-20A column oven and an SPD-20A UV-visible detector. UV-visible absorption was measured at 295 nm.
HPLC-ICP/MS system: A Nexion 2000B (Perkin-Elmer, Villebon Sur Yvette, France), coupled with a Flexar LC system (Perkin-Elmer). Gd signal was monitored at m/z 152. Syngistix software version 2.3 was used to control the ICP-MS. The Gd signal was acquired through Empower software version 7.3.
HPLC-ESI/MS system: The ESI/MS measurements were carried out on a triple quadrupole spectrometer Xevo TQ-S (Waters, Milford, MA, USA), coupled with UHPLC chain Acquity H-Class (Waters). The analyses were performed on positive mode (ESI+) and detection SCAN mode set on 300–1500 uma range.
+ Open protocol
+ Expand
3

HPLC Analysis of Compounds

Check if the same lab product or an alternative is used in the 5 most similar protocols
Shimadzu Prominence series UFLC system (Lyon, France), equipped with a CBM-20A controller bus module, a LC-20 AD liquid chromatograph, a CTO-20A column oven and an SPD-20A UV-visible detector was used for Gradient HPLC analysis. Wavelength was fixed at 295 nm for UV-VIS detection. The separation was performed using a C4 reverse-phase column (Jupiter®, 5 µm, 300 A, 150 × 4.6 mm) at a flow rate of 1 mL min−1. The gradient initial solution is 95% solvent A − 5% solvent B (A = H2O/ACN/TFA: 98.9 v%/1 v%/0.1 v%, B = ACN/H2O/TFA: 89.9 v%/10 v%/0.1 v%) over 5 min. In a second step, the samples were eluted by a gradient developed from 5 to 90% of solvent B in solvent A over 15 min. The concentration of solvent B was maintained over 5 min. Then, the concentration of solvent B was decreased to 5% over a period of 5 min to re-equilibrate the system, followed by an additional 5 min at this final concentration.
+ Open protocol
+ Expand
4

Gradient HPLC Analysis of Biomolecules

Check if the same lab product or an alternative is used in the 5 most similar protocols
Gradient HPLC analysis was carried out using the Shimadzu® Prominence series UFLC system equipped with a CBM-20A controller bus module, an LC-20AD liquid chromatograph, a CTO-20A column oven, a SPD-20A UV–Visible detector allowing absorption measurement at two chosen wavelengths and a RF-20A fluorescence detector allowing fluorescence measurements at specific excitation and emission wavelengths. For all measurements, the UV–visible detectors were set respectively at 295 and 350 nm. Sample aliquots of 20 µL were injected in a 95 % solvent A—5 % solvent B (A = Milli-Q water/TFA 99.9:0.1 v/v; B = CH3CN/Milli-Q water/TFA 90:9.9:0.1 v/v/v) into a Jupiter C4 column (150 × 4.60 mm, 5 µm, 300 Å, Phenomenex®) at a flow rate of 1 mL min−1 over 5 min. In a second step, samples were eluted by a gradient from 5 to 90 % of solvent B in solvent A over 15 min. Finally, the concentration of solvent B was maintained over 5 min. Then, the concentration of solvent B was decreased to 5 % over a period of 5 min to re-equilibrate the system, followed by additional 5 min at this final concentration. Before each sample measurement, a baseline was recorded following the same conditions by loading Milli-Q water onto the injection loop.
+ Open protocol
+ Expand
5

Gradient HPLC Analysis of Biomolecules

Check if the same lab product or an alternative is used in the 5 most similar protocols
Gradient HPLC analysis was performed using the Shimadzu Prominence series UFLC system with a LC-20 AD liquid chromatograph, a CBM-20A controller bus module, a CTO-20A column oven, and an SPD-20A UV-visible detector. UV-visible absorption was measured at 295 nm. Separation was performed at a flow rate of 1 mL min-1 using C4 reverse phase HPLC column (Jupiter®, 5 µm, 300 A, 150 x 4.6 mm). The gradient initial solution was 95% solvent A - 5% solvent B (A = H2O / ACN / TFA: 98.9 / 1 / 0.1% v, B = ACN / H2O / TFA: 89.9 / 10 / 0.1% v) over 5 min. During the next step, samples were eluted over 15 mins by a gradient developed from 5 to 90% of solvent B in solvent A. The concentration of solvent B was maintained over 5 min and then the concentration of solvent B was decreased to 5% over a period of 5 min to re-equilibrate the system, followed by an additional 5 min at this final concentration.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!