The largest database of trusted experimental protocols

Ultraplus field emission

Manufactured by Zeiss
Sourced in Germany

The UltraPlus Field Emission is a high-performance field emission scanning electron microscope (FE-SEM) designed for advanced imaging and analysis applications. It provides high-resolution imaging capabilities, enabling detailed examination of a wide range of samples. The core function of the UltraPlus is to capture high-quality, high-resolution images of specimens using a focused beam of electrons.

Automatically generated - may contain errors

8 protocols using ultraplus field emission

1

Comprehensive Nanoparticle Characterization

Check if the same lab product or an alternative is used in the 5 most similar protocols
NP size, polydispersity, and superficial
charge are measured by dynamic light scattering (DLS, Zetasizer Nano
ZS, Malvern U.K.). Typically, 1 mL of the diluted sample is put in
12 mm square glass cuvettes for 90° sizing (Optical Cuvette,
Sarstedt) and measured at least in triplicate. ζ potential measurements
are performed at room temperature on a Zetasizer Nano ZS (Malvern,
U.K.), fitted with a high-concentration ζ potential cell. Nanoparticle
morphology is assessed by scanning and transmission electron microscopy,
SEM and TEM, respectively. For SEM observations (Carl Zeiss Ultraplus
Field Emission), 100 μL of purified samples are dropped on a
polycarbonate isopore membrane filter (cutoff 0.05 μm) under
vacuum ultrafiltration and let dry overnight. Before the observation,
5–7 nm of Au is deposited on the sample. For cryo-TEM observation
of EVs (Cryo-TEM TECNAI by FEI), 5 μL of the purified sample
is dropped on a Formvar/Carbon 200 mesh grid (Agar scientific) and
observed at an 80 kV accelerating voltage. The concentration of cHANPs
and EVs is quantified by nanoparticle tracking analysis NTA (Nanosight
NS300, Malvern Instruments Ltd., U.K.). Both samples are observed
at a dilution of 1:100, at 25 °C for 300 s with a manual shutter
and gain adjustment. Results are analyzed with NP Tracking Analysis
software.
+ Open protocol
+ Expand
2

Characterization of Polymer Nanostructure Arrays

Check if the same lab product or an alternative is used in the 5 most similar protocols
The scanning electron microscope (SEM) images were obtained using Zeiss Ultra Plus field emission scanning electron microscope with an electron energy of 5 kV. Atomic force microscope (AFM) images were obtained using a nanoscope scanning probe microscope (Dimension fastscan, Bruker, Germany) under ambient conditions. Au coating was sputtered on the surface of the polymer nanostructure arrays using an ion sputtering apparatus (E-1045, Hitachi, Japan). The light absorption performance of substrates were measured using a spectrophotometer (Shimadzu UV3600, Shimadzu, Japan) combined with an integrating sphere for the 300−800 nm wavelength range. For Raman scattering measurements, 10 μL R6G aqueous solution (10−6 M) was dropped onto the substrates and then dried in the dark. Also, the same amount of 10−1 M R6G solution was drop-casted on glass to get reference Raman spectra. SERS spectra were collected using a Dispersive Raman Microscope (Senterra R200-L, Bruker Optics, Germany).
+ Open protocol
+ Expand
3

Characterization of Conjugated Nanoparticles

Check if the same lab product or an alternative is used in the 5 most similar protocols
Size and polydispersity of the bare and conjugated cHANPs are measured by dynamic light scattering (DLS) by Zetasizer Nano, Malvern Panalytical (UK). The size measurement is performed dropping 1 mL of suspension in a square glass cuvette (Optical Cuvette, Sarstedt) at room temperature in triplicate.
Scanning electron microscopy (SEM), Carl Zeiss Ultraplus Field Emission, Zeiss (Germany) and transmission electron microscopy (TEM), Cryo-TEM TECNAI G2-20, and FEI (OR, USA) observations are performed to investigate the nanoparticle morphology.
For SEM observations, 200 μL of suspension is deposited by ultrafiltration on polycarbonate Isopore™ membrane filters with a 50 nm cut-off by Merck KGaA (Germany) then left to dry. A 7 nm Au layer is deposited via sputter coating on the membrane filter that is observed at 10 kV. 20 μL of suspension is dropped on copper grids with carbon film by Agar Scientific Ltd. (Stansted, UK) and left to dry prior to observation at 120 kV.
+ Open protocol
+ Expand
4

SEM Ultrastructural Examination of BMDMs

Check if the same lab product or an alternative is used in the 5 most similar protocols
BMDMs were washed with PBS and post-fixed with 2.5% glutaraldehyde in 0.1 M phosphate buffer overnight and further washed with PBS. Cells were fixed in 1% osmium tetroxide in double distilled water for 1 hr and dehydrated in a series of alcohol. Dehydrated samples were dried using liquid carbon dioxide using critical point drying. Samples were then sputter-coated with platinum (3 nm thickness) at 15 mA for 2 min using the EMI TECH K550 Sputter coater and visualized under a Zeiss UltraPlus Field emission scanning electron microscope at 5 kV.
+ Open protocol
+ Expand
5

SEM Analysis of Peptide-Treated Bacteria

Check if the same lab product or an alternative is used in the 5 most similar protocols
Mid-logarithmic phase bacteria were washed and resuspended in PBS before peptide treatment at 100 µg/mL for 12 h or protein treatment at 1.84 µM for 6 h on coverslips. Treated bacteria were washed with PBS and post-fixed with 2.5% glutaraldehyde in 0.1 M phosphate buffer for 3 h and further washed with PBS. Cells were stained in 1% osmium tetroxide in distilled water for 1 h and dehydrated in a series of alcohol. Dehydrated samples were dried using liquid carbon dioxide critical point drying. Samples were then sputter-coated with platinum (3 nm thickness) at 15 mA for 2 min using the EMI TECH K550 Sputter coater and visualized under a Zeiss UltraPlus Field emission scanning electron microscope at 5 kV.
+ Open protocol
+ Expand
6

Nanoparticle Characterization by SEM and TEM

Check if the same lab product or an alternative is used in the 5 most similar protocols
NP morphological and structural surface features are analyzed with a scanning electron microscope (SEM, Ultraplus Field Emission, Carl Zeiss, Oberkochen, Germany). Examined samples are dropped on glass directly from the microfluidic platform outlet or filtered, after collection and dialysis, on a membrane of 50 nm pore size. Nanoparticles are coated with 5.5 nm Au or Pt/Pd prior observation. Another characterization is carried out with a TEM (Tecnai FEI® transmission electron microscope, Hillsboro, OR, USA) that allows examining NPs’ internal features. Samples are collected on Formvar/Carbon 200 mesh Cu Agar® small net from platform outlet or dropping off 20–50 μL of solution on it, before or after dialysis.
+ Open protocol
+ Expand
7

SEM Analysis of Glutaraldehyde-Fixed BMDMs

Check if the same lab product or an alternative is used in the 5 most similar protocols
BMDMs were washed with PBS and post-fixed with 2.5% glutaraldehyde in 0.1 M phosphate buffer overnight and further washed with PBS. Cells were fixed in 1% osmium tetroxide in double distilled water for 1 h and dehydrated in a series of alcohol. Dehydrated samples were dried using liquid carbon dioxide using critical point drying. Samples were then sputter-coated with platinum (3 nm thickness) at 15 mA for 2 min using the EMI TECH K550 Sputter coater and visualized under a Zeiss UltraPlus Field emission scanning electron microscope at 5 kV.
+ Open protocol
+ Expand
8

Scanning Electron Microscopy of BMDMs

Check if the same lab product or an alternative is used in the 5 most similar protocols
BMDMs were washed with PBS and postfixed with 2.5% glutaraldehyde in 0.1 M phosphate buffer overnight and further washed with PBS. Cells were fixed in 1% osmium tetroxide in double-distilled water for 1 hour and dehydrated in a series of alcohol. Dehydrated samples were dried using liquid carbon dioxide using critical point drying. Samples were then sputter-coated with platinum (3 nm thickness) at 15 mA for 2 min using the Emitech K550 sputter coater and visualized under a Zeiss Ultra Plus Field emission scanning electron microscope at 5 kV.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!