The largest database of trusted experimental protocols

Solarix xr 9.4t

Manufactured by Bruker
Sourced in Germany

The SolariX XR 9.4T is a high-performance Fourier Transform Ion Cyclotron Resonance (FT-ICR) mass spectrometer. It offers a magnetic field strength of 9.4 Tesla, providing high resolving power and mass accuracy for advanced analyses.

Automatically generated - may contain errors

7 protocols using solarix xr 9.4t

1

High-Resolution Mass Spectrometry Analytical Protocol

Check if the same lab product or an alternative is used in the 5 most similar protocols
Positive ion mode: The LC methods are as described in Section 3.4.1. The Bruker solariX XR 9.4T was operated over the m/z range 57.75–2000.00 with an electrospray ionisation source. Analytes were detected in the positive ion mode using the following MS parameters: the dry gas flow was 7.0 L/min; the dry gas temperature was 200 °C, the source voltage was 4000 V and the nebuliser gas pressure was 2.0 bar.
Negative ion mode: The LC methods are as described in Section 3.4.1. The Bruker solariX XR 9.4T was operated in the negative ion mode over the m/z range 57.75–2000.00 with an electrospray ionisation source. Analytes were detected in the negative ion mode using the following MS parameters: dry gas flow was 7.0 L/min; the dry gas temperature was 200 °C and the source voltage was 4000 V. To optimise ionisation, the nebuliser gas pressure was 2.0 bar (0–6 min), 1.5 bar (6–12 min), 1.3 bar (12–21 min), 1.0 (21–25 min), 1.5 (15–26 min), 1.8 (26–27.5 min) and 2.0 bar (27.5–30 min).
+ Open protocol
+ Expand
2

Mass Spectrometry Analysis of Peptide Purity

Check if the same lab product or an alternative is used in the 5 most similar protocols
Once the purity of each peptide had been determined, the set of seven‐point calibration standards was then analysed using two different mass spectrometers and two ionisation sources. The two instruments used were the Bruker ultraflex III (TOF) and the Bruker solariX XR 9.4 T (FT‐ICR). The ultraflex was used with its fixed MALDI source in positive ion mode with 800 laser shots per sample acquisition. The data were acquired using flexcontrol software version 3.0 (Bruker Daltonics). Each spot was analysed in reflector mode using a smartbeam™ Nd:YAG laser (355 nm). Spectra were analysed using flexAnalysis software version 3.0 (Bruker Daltonics). The solariX was operated with either an ESI (direct infusion) or MALDI source in positive ion mode (800 laser shots per sample acquisition) with a smartbeam™ Nd:YAG laser (355 nm). Spectra were acquired using the solariXcontrol software and processed with DataAnalysis version 4.2 (Bruker Daltonics). Each sample was analysed in triplicate and the average values of the peak intensities are reported.
+ Open protocol
+ Expand
3

MALDI-FT-ICR MS Imaging of Microbial Colonies

Check if the same lab product or an alternative is used in the 5 most similar protocols
Mass spectrometry images were obtained as recently described (76 (link)) using an FT-ICR mass spectrometer (SolariX XR 9.4T, Bruker Daltonics, Bremen, Germany) mass calibrated from 200 m/z to 2,300 m/z to reach a mass accuracy of 0.5 ppm. A region of interest from agar microbial colonies was directly collected from the petri dish and transferred onto an indium tin oxide (ITO) glass slide (Bruker, Bremen, Germany) previously covered with double-sided conductive carbon tape. The samples were dried under vacuum and covered with an α-cyano-4-hydroxycinnamic acid matrix solution at 5 mg/mL (70:30 acetonitrile:water [vol/vol]). In total, 60 layers of α-cyano-4-hydroxycinnamic acid matrix were sprayed using a SunCollect instrument (SunChrom, Friedrichsdorf, Germany). FlexImaging 5.0 (Bruker Daltonics, Bremen, Germany) software was used for MALDI-FT-ICR MS imaging acquisition, with a pixel step size for the surface raster set to 100 μm.
+ Open protocol
+ Expand
4

HPLC-MS Analysis of Metabolites

Check if the same lab product or an alternative is used in the 5 most similar protocols
HPLC–MS was carried out using an Agilent 1200 HPLC fitted with an Atlantis® T3 column (Waters, cortecs 2.7 µm, 3 × 150 mm); the column temperature was maintained at 25 °C. The HPLC was coupled to an electrospray FT-ICR mass spectrometer (Bruker solariX XR 9.4T). The mobile phase was composed of water (A) and acetonitrile (B) both with 0.1% (v/v) formic acid, and the following gradient was used: 5% B increasing linearly to 95% over 22 min and held for 2 min. B was then returned to 5% over 0.33 min and held for 5.33 min to allow column equilibration. The flow rate was 300 µL/min and the injection volume was 5 µL.
+ Open protocol
+ Expand
5

MALDI-FT-ICR MS Imaging of Microbial Colonies

Check if the same lab product or an alternative is used in the 5 most similar protocols
Mass spectrometry images were obtained as recently described (76 (link)) using an FT-ICR mass spectrometer (SolariX XR 9.4T, Bruker Daltonics, Bremen, Germany) mass calibrated from 200 m/z to 2,300 m/z to reach a mass accuracy of 0.5 ppm. A region of interest from agar microbial colonies was directly collected from the petri dish and transferred onto an indium tin oxide (ITO) glass slide (Bruker, Bremen, Germany) previously covered with double-sided conductive carbon tape. The samples were dried under vacuum and covered with an α-cyano-4-hydroxycinnamic acid matrix solution at 5 mg/mL (70:30 acetonitrile:water [vol/vol]). In total, 60 layers of α-cyano-4-hydroxycinnamic acid matrix were sprayed using a SunCollect instrument (SunChrom, Friedrichsdorf, Germany). FlexImaging 5.0 (Bruker Daltonics, Bremen, Germany) software was used for MALDI-FT-ICR MS imaging acquisition, with a pixel step size for the surface raster set to 100 μm.
+ Open protocol
+ Expand
6

High-Resolution Mass Spectrometry Analysis Protocol

Check if the same lab product or an alternative is used in the 5 most similar protocols
The high-resolution mass spectrometry analysis was performed using a SolarixXR 9.4 T (Bruker, Germany), equipped with an electrospray ionization (ESI, Bruker) source, set in negative ionization mode. The instrument was externally calibrated with Tuning Mix from Agilent. Samples were infused directly into the ESI source. The parameters were optimized to obtain a stable ion current with a minima ion injecting time into the mass analyser. The infusion flow rate was 2.0 µL min−1, the drying gas temperature was 200 °C, and the drying and nebulizing gas flow rate were 4.0 L min−1 and 1 bar, respectively. The ESI capillary voltage was 3.9 kV. Three hundred scans were accumulated for each spectrum. Methanol was injected prior to the injection of each sample, and acquisition was performed to evaluate the potential presence of residual pollutants. The acquisition size was set to 8 M, resulting in a mass resolving power of up to (6.6 ± 1.6) × 105 for the full mass range.
+ Open protocol
+ Expand
7

HPLC-MS Analysis of Compounds

Check if the same lab product or an alternative is used in the 5 most similar protocols
Analysis was carried out with a Dionex UltiMate 3000 high performance liquid chromatograph (HPLC) fitted with a Dionex Acclaim 120 C18 column (3 μm, 120 Å, 2.1 × 150 mm) and Phenomenex SecurityGuard system with a C18 (4 × 2.00 mm) cartridge, coupled to a Bruker HCTultra ETD II ion trap MS, a Bruker solariX XR 9.4 T Fourier transform ion cyclotron resonance (FTICR) MS (full scan modes) or an Applied Biosystems/ MDS Sciex API 3000 triple quadrupole MS (selected reaction monitoring (SRM) mode), using electrospray ionisation (ESI, positive mode). Separation used a water : acetonitrile gradient (10%-90% acetonitrile over 5 min). The ion trap instrument was used for method development experiments before moving onto the triple quadrupole in order to gain optimal sensitivity, while the FTICR was used to obtain full scan data with high mass accuracy in order to try to detect compounds for which standards could not be obtained.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!