Metal-free plastic centrifuge tubes are laboratory equipment designed for sample processing, particularly in centrifugation applications. These tubes are made of plastic material that does not contain any metal components, providing an inert environment for sensitive samples.

Automatically generated - may contain errors

2 protocols using metal free plastic centrifuge tubes

1

Trace Metal Analysis of Microbial Growth Media

Check if the same lab product or an alternative is used in the 5 most similar protocols
Media samples (1.0 mL) were boiled in 2.0 mL metal-free plastic centrifuge tubes (Eppendorf). Dried sample from 1 mL of media (MRS or CDM) was digested with 280 μL Milli-Q water and 15 μL trace metal concentrated nitric acid at 80°C for at least 12 h. Digested samples from a total of 8 mL media were diluted with 2% trace metal HNO3 to bring to a total volume of 3 mL. Sample was stored at 4°C until analysis by ICP-OES.
Medium samples were analyzed by an Agilent Simultaneous 725 ICP-OES equipped with VistaChip II CCD detector and image mapping technology (I-MAP) to provide complete wavelength coverage from 167 to 785 nm. Trace metal HNO3 (2%) was used as blank. Metal concentrations in media were measured by ICP-OES for Ca, V, Cr, Mn, Fe, Co, Ni, Cu, Zn, Se, Mo, and Cd. ICP-OES calibration standards contain 5,000, 2,000, 1,000, 750, 500, 250, 100, 40, 25, 16, 10, 5, 2.5, 1, and 0.5 ppb for each metal and were all prepared by diluting commercially available Inorganic Ventures' ICP-MS Complete Standard 71A (10 ppm) and Inorganic Ventures' ICP-MS Refractory Elements Standard 71B (10 ppm) with 2% trace metal HNO3.
+ Open protocol
+ Expand
2

Trace Metal Analysis of Microbial Growth Media

Check if the same lab product or an alternative is used in the 5 most similar protocols
Media samples (1.0 mL) were boiled in 2.0 mL metal-free plastic centrifuge tubes (Eppendorf). Dried sample from 1 mL of media (MRS or CDM) was digested with 280 μL Milli-Q water and 15 μL trace metal concentrated nitric acid at 80°C for at least 12 h. Digested samples from a total of 8 mL media were diluted with 2% trace metal HNO3 to bring to a total volume of 3 mL. Sample was stored at 4°C until analysis by ICP-OES.
Medium samples were analyzed by an Agilent Simultaneous 725 ICP-OES equipped with VistaChip II CCD detector and image mapping technology (I-MAP) to provide complete wavelength coverage from 167 to 785 nm. Trace metal HNO3 (2%) was used as blank. Metal concentrations in media were measured by ICP-OES for Ca, V, Cr, Mn, Fe, Co, Ni, Cu, Zn, Se, Mo, and Cd. ICP-OES calibration standards contain 5,000, 2,000, 1,000, 750, 500, 250, 100, 40, 25, 16, 10, 5, 2.5, 1, and 0.5 ppb for each metal and were all prepared by diluting commercially available Inorganic Ventures' ICP-MS Complete Standard 71A (10 ppm) and Inorganic Ventures' ICP-MS Refractory Elements Standard 71B (10 ppm) with 2% trace metal HNO3.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!