The largest database of trusted experimental protocols

Acquity uplc beh shield rp18 column

Manufactured by Waters Corporation
Sourced in United States

The ACQUITY UPLC BEH Shield RP18 column is a reversed-phase liquid chromatography column designed for high-performance liquid chromatography (HPLC) and ultra-high-performance liquid chromatography (UHPLC) applications. It features a hybrid particle technology that combines the advantages of both silica and polymer-based stationary phases, providing enhanced stability and performance.

Automatically generated - may contain errors

29 protocols using acquity uplc beh shield rp18 column

1

Algae Compound Analysis by UPLC-PDA-ESI-QTOF

Check if the same lab product or an alternative is used in the 5 most similar protocols
Dried extract (3 mg) of algae was dissolved in 1 mL of MeOH/H2O 1/1 v/v. The analysis of compounds from algae was carried out with the use of an ACQUITY Ultra Performance LC system equipped with a photodiode array detector with a binary solvent manager (Waters Corporation, Milford, MA, USA) series with a mass detector Q/TOF micro mass spectrometer (Waters) equipped with an electrospray ionization (ESI) source operating in negative mode at the following conditions: capillary voltage, 2300 kV; source temperature, 100 °C; cone gas flow, 40 L/h; desolvation temperature, 500 °C; desolvation gas flow, 11,000 L/h; and scan range, m/z 50–1500. Separation of individual compounds was carried out using an ACQUITY UPLC BEH Shield RP18 column (1.7 μm, 2.1 mm × 100 mm; Waters Corporation, Milford, MA, USA) at 40 °C. The elution gradient test was carried out using water containing 1% acetic acid (A) and acetonitrile (B), and applied as follows: 0 min, 1% B; 2.3 min, 1% B; 4.4 min, 7% B; 8.1 min, 14% B; 12.2 min, 24% B; 16 min, 40% B; 18.3 min, 100% B, 21 min, 100% B; 22.4 min, 1% B; 25 min, 1% B. The sample volume injected was 2 μL and the flow rate used was 0.6 mL/min. The compounds were monitored at 280 nm. Integration and data elaboration were performed using MassLynx 4.1 software (Waters Corporation, Milford, MA, USA) [54 (link)].
+ Open protocol
+ Expand
2

Untargeted Metabolite Profiling by UPLC-HRMS

Check if the same lab product or an alternative is used in the 5 most similar protocols
An Acquity UPLC system (Waters Corporation, Milford, MA, USA) combined with a Q-Exactive high-resolution mass spectrometer (Thermo Fisher, Waltham, MA, USA) with an Orbitrap mass analyzer was used. Samples (5 μL) were injected onto an ACQUITY UPLC BEH Shield RP18 column (150 × 2.1 mm, particle size 1.7 μm) (Waters, Manchester, MA, USA), with a flow rate of 0.35 μL min−1 at 50 °C. Mobile phases contained 0.1% (v/v) formic acid in water (A) (LC-MS grade, Merck, Darmstadt, Germany) and acetonitrile (B) (LC–MS grade, Merck). A multi-step linear gradient was as follows: 5% B—1.5 min, 80% B—10.5 min, 98% B—11.5 min, 5% B—13 min.
Mass spectrometry analysis was performed using heated electrospray ionization (H-ESI) in positive and negative modes. A 3.5 kV and 2.5 kV ion spray voltage was applied for positive and negative ionization, respectively. Ion source temperature was 320 °C. Data were acquired in Full MS/data-dependent MS2 mode in the 100–1500 m/z range. The resolution of Full MS was 70,000 and of ddMS2 17,500. Normalized collision energy in the ddMS2 experiment was set to 30%. Xcalibur software (ThermoFisher Scientific, Waltham, MA, USA) was used for system operation, data acquisition, and data analysis [58 (link)].
+ Open protocol
+ Expand
3

UPLC-ESI-TOF MS Analysis of Orange By-products

Check if the same lab product or an alternative is used in the 5 most similar protocols
The analyses of the orange by-products using the optimized conditions obtained with the Box–Behnken design were carried out in duplicate on an ACQUITY Ultra Performance LC system (Waters Corporation, Milford, MA, USA) coupled to an electrospray ionization (ESI) source operating in the negative mode and a time-of-flight (TOF) mass detector (Waters Corporation, Milford, MA, USA). The compounds of interest were separated on an ACQUITY UPLC BEH Shield RP18 column (1.7 µm, 2.1 mm × 100 mm; Waters Corporation, Milford, MA, USA) at 40 °C using a gradient previously stated by Verni et al. [21 (link)] using water containing 1% acetic acid as mobile phase A and acetonitrile as mobile phase B. The data were elaborated using MassLynx 4.1 software (Waters Corporation, Milford, MA, USA).
+ Open protocol
+ Expand
4

LCMS Analysis of Lipid Mediators

Check if the same lab product or an alternative is used in the 5 most similar protocols
The LCMS lipid mediators analysis was performed based on previous publications [26 (link),36 (link)]. Briefly, high-performance liquid chromatography (HPLC) was performed using an Agilent 1290 series chromatographer (Agilent, Santa Clara, CA, USA). Reversed-phase separation was achieved by Acquity UPLC BEH shield RP18 column (2.1 × 100 mm; 1.7 m; Waters) and maintained at 40 °C. The mobile phase was a gradient of solvents A (ACN/water/acetic acid (60/40/0.02, v/v)) and B (ACN/IPA (50/50, v/v) with a flow rate of 0.5 mL/min. The stepwise gradient conditions were carried out for 10 min as follows: 0–5.0 min, 1–55% of solvent B; 5.0–5.5 min, 55–99% of solvent B, and finally 5.5–6.0 min, 99% of solvent B. Injection volume was 10 μL, and all samples were kept at 4 °C throughout the analysis.
The HPLC system was coupled to Agilent 6495 triple-quad mass spectrometer (Agilent, Santa Clara, CA, USA). The electrospray ionization was conducted in negative mode, and the dynamic MRM option was used and performed for all compounds with optimized transitions and collision energies. The determination and integration of all peaks were manually performed using the Mass Hunter Workstation software (Agilent, Santa Clara, CA, USA).
+ Open protocol
+ Expand
5

UPLC-MS/MS Analysis of Bile Salts

Check if the same lab product or an alternative is used in the 5 most similar protocols
Sample supernatants (5 µL) were injected using a Waters FTN Sample manager onto an ACQUITY UPLC BEH Shield RP18 column (Waters; particle size: 1.7 µm, length: 50 mm, internal diameter: 2.1 mm). The initial solvent was 81% A (solvent A: 100% water with 5 mmol/L ammonium acetate and 0.01% formic acid; solvent B: 100% methanol with 5 mmol/L ammonium acetate and 0.01% formic acid) at a flow rate of 0.4 mL/min at 45 °C. Gradient elution (Table 8) was performed using a Waters ACQUITY UPLC I-class pump. Mass spectrometric detection employed a Xevo TQ-XS (Waters) with Unispray ionization in negative ion mode. Individual bile salts were monitored in MRM windows, either as pseudo-transitions (for unconjugated species) or as glycine/taurine daughter ions.
+ Open protocol
+ Expand
6

Capillary Column Preparation for Chromatography

Check if the same lab product or an alternative is used in the 5 most similar protocols
A fused silica capillary tube with an inner diameter of 100 μm and outer diameter of 360 μm from Polymicro Technology, LLC (Phoenix, AZ, USA) was used to prepare the capillary columns. Watchers® ODS-P C-18 particles (3 μm and 100 Å) from Isu Industry Corp. (Seoul, Korea) and BEH Shield C18 particles (1.7 μm and 130 Å), unpacked from an ACQUITY UPLC BEH Shield RP18 column (2.1 mm × 100 mm) purchased from Waters (Milford, MA, USA), were used as packing materials for capillary columns.
+ Open protocol
+ Expand
7

UPLC-MS Analysis of Compounds

Check if the same lab product or an alternative is used in the 5 most similar protocols
UPLC-MS analysis was performed using an Agilent Technology 1290 Infinity UPLC equipped with an Agilent Technology 6460 Triple Quad liquid chromatography/mass spectrometer instrument (Agilent, Santa Clara, CA, USA). Chromatographic separations were performed using an Acquity UPLC BEH shield RP18 column (2.1 mm × 100 mm, 1.7 mm particle size; Waters, Milford, MA, USA) maintained at 25 °C. The mobile phase consisted of 0.3% formic acid (solvent A) and acetonitrile (solvent B) with a gradient as follows: 2% B for 0–2 min, 2%–5% B for 2–5 min, 5%–15% B for 5–10 min, and 15%–35% B for 10–32 min. The flow rate was kept at 0.3 mL/min, and the injection volume was 2 μL. The UV detection wavelength was set at 235 nm.
The MS analyses were performed using an electrospray ionization ion source under a positive ion mode with the full scan mass from 100 to 1000 m/z. The voltage of the capillary was set at 7 kV. The gas temperature was 350 °C, and the gas flow was 13 L/min. The nebulizer was maintained at 60 psi.
+ Open protocol
+ Expand
8

UPLC-PDA Analysis of FBEZF in Water

Check if the same lab product or an alternative is used in the 5 most similar protocols
The detection of FBEZF in water was performed on a Waters ACQUITY UPLC H-class system fitted with a sample manager, a quaternary solvent manager, a PDA detector, and an ACQUITY UPLC BEH Shield RP18 column (50.0 mm × 2.1 mm i. d., 1.7 μm film thickness) (Waters Corporation, Milford, MA, USA). The column temperature was at 40 °C. A total of 2 µL sample solution was injected, and the chromatography was run with acetonitrile/water (30/70, v/v) at a flow rate of 0.2 mL min−1. The chromatographic conditions were determined from the trial experiments for optimal results in terms of peak shape, column efficiency, chromatographic analysis time, selectivity, and resolution. FBEZF was detected at 212 nm. Retention time of FBEZF was 3.9 min under the optimized chromatographic conditions.
+ Open protocol
+ Expand
9

Phenolic Profiling of Avocado Leaf Extracts

Check if the same lab product or an alternative is used in the 5 most similar protocols
Phenolic compounds present in the fermented and non-fermented avocado leaf extracts were analysed using an Acquity Ultra Performance Liquid Chromatography (UPLC) system (Waters Corporation, Milford, MA, USA) coupled to an electrospray ionization (ESI) source operating in the negative mode and a mass detector time of flight (TOF) micro mass spectrometer (Waters). The compounds of interest were separated on an ACQUITY UPLC BEH Shield RP18 column (1.7 μm, 2.1 × 100 mm; Waters Corporation, Milford, MA, USA) at 40 °C using the conditions and gradient previously stated [31 (link)]. H2O acidified with 1% of acetic acid and acetonitrile were used as phase A and B, respectively. Analyses were performed in triplicate. The identification of the phenolic compounds was made according to the literature. For ensuring the mass accuracy, the tolerances chosen had a score higher than 90% and error lower than 5 ppm. To quantify the phenolic compounds identified in the avocado leaf extracts, calibration curves were used for vanillic acid (y = 8.1947x + 122.91; R² = 0.9976), chlorogenic acid (y = 85.138x + 135.16; R² = 0.9978), ferulic acid (y = 16.507x + 92.06; R² = 0.9980), quercetin (y = 112.8x + 287.12; R² = 0.9957), catechin (y = 41.108x + 335.6; R² = 0.9959) and rutin (y = 26.176x + 403.46; R² = 0.9924). The results are expressed as µg/g d.w.
+ Open protocol
+ Expand
10

Accurate Quantitation of Phosphatidylcholine Metabolites

Check if the same lab product or an alternative is used in the 5 most similar protocols
The analysis of selected reaction intermediates was performed by the method of Vo Duy et al. (35 (link)). LC instrumentation included an ACQUITYTM Ultra Performance Liquid Chromatography integrated system from Waters (Milford, MA). Mass spectrometry experiments were performed using a triple quadrupole mass spectrometer TSQ Quantum UltraTM (Thermo Fisher Scientific Inc., Waltham, MA). Metabolites were separated using an Acquity UPLC BEH Shield RP18 column (150 mm × 2.1 mm, 1.7 µm; Waters, Saint-Quentin en Yvelines, France) in the positive ionization mode. Heptafluorobutyric acid was added to the mobile phase as an ion pair reagent. Quantitation was performed using one internal standard. Each metabolite was quantified against a calibration curve prepared by spiking RBC cell pellets (3 × 106 cells) with appropriate working solutions of all metabolites and the internal standard. Using the method Vo Duy et al. (35 (link)) the quantities of P-Cho were overestimated because of a lack of specificity. The introduction of a second product ion for P-Cho allowed us to increase the specificity of the method and to obtain correct values.
For the water-soluble metabolites and the final PLs, the area of peaks were determined and reported as percent of the total contents of the corresponding metabolite or PL.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!