The largest database of trusted experimental protocols

Nanoacquity beh c18 column

Manufactured by Waters Corporation
Sourced in United States

The NanoAcquity BEH C18 column is a high-performance liquid chromatography (HPLC) column designed for the separation and analysis of a wide range of analytes. It features a bonded C18 stationary phase, which provides effective retention and separation of both polar and non-polar compounds. The column is compatible with aqueous and organic mobile phases, making it suitable for a variety of analytical applications.

Automatically generated - may contain errors

7 protocols using nanoacquity beh c18 column

1

Quantitative Proteomics by LC-MS/MS

Check if the same lab product or an alternative is used in the 5 most similar protocols
Fractions were analyzed using an LC-MS/MS system comprised of an ACQUITY UPLC M-Class System (Waters, Milford, MA, USA) directly coupled to a QExactive mass spectrometer (Thermo Fisher Scientific, Waltham, MA, USA). Peptides were trapped on a C18 pre-column (180 µm × 20 mm; Waters, Milford, MA, USA) using 0.1% FA in water as a mobile phase and transferred to a nanoAcquity BEH C18 column (75 µm × 250 mm, 1.7 µm; Waters, Milford, MA, USA) using ACN gradient (0–35% ACN in 160 min) in the presence of 0.1% FA at a flow rate of 250 nL/min. Data acquisition was carried out using a data-dependent method with the top 12 precursors selected for MS2 analysis after collisional induced fragmentation (CID) with normalized collision energy (NCE) of 27. Full MS scans covering the mass range of 300–1600 m/z (mass-to-charge ratio) were acquired at a resolution of 70,000 with a maximum injection time of 60 ms and an AGC target value of 1e6. MS2 scans were acquired with a maximum injection time of 120 ms and an AGC target value of 5e5 with a resolution of 35,000. The isolation window was set to 1.2 m/z and a dynamic exclusion was set to 30 s.
+ Open protocol
+ Expand
2

Saliva Proteome Analysis by UPLC-MS/MS

Check if the same lab product or an alternative is used in the 5 most similar protocols
Here, 2 ug of each saliva sample was analyzed using a nanoAcquity UPLC (Waters) directly coupled to a QExactive mass spectrometer (Thermo Scientific, Bremen, Germany). Peptides were trapped on a C18 precolumn (180 µm × 20 mm, Waters) with 0.1% FA in water as a mobile phase and transferred to a nanoAcquity BEH C18 column (75 µm × 250 mm, 1.7 µm, Waters) using ACN gradient (0–35% ACN in 160 min) in the presence of 0.1% FA at a flow rate of 250 nL/min. Measurements were performed in data-dependent mode with top 12 precursors selected for MS2. Full MS scans covering the range of 300–1650 m/z were acquired at a resolution of 70,000, with a maximum injection time of 60 ms and an AGC target value of 1e6. MS2 scans were acquired at a resolution of 17,500 and an AGC target value of 5e5. Dynamic exclusion was set to 30 s.
+ Open protocol
+ Expand
3

Gad8 Protein Identification by Mass Spectrometry

Check if the same lab product or an alternative is used in the 5 most similar protocols
Large-scale Gad8 immunoprecipitations were run on a Nupage Bis–Tris 4–12% SDS–PAGE gel (Invitrogen), The Gad8 Coomassie-stained band was excised and digested with either 20 ng sequencing-grade trypsin (Sigma-Aldrich), 400 ng LysN (Associates of Cape Cod) or 350 ng Elastase (Calbiochem) in 100 µl 40 mM ammonium bicarbonate, 9% (v/v) acetonitrile at 37°C for 18 h. The peptides were separated using a Nano-Acquity UPLC system (Waters) using a Waters NanoAcquity BEH C18 column (75 µm ID, 1.7 µm, 25 cm) with a gradient of 1–25% (v/v) of acetonitrile, 0.1% formic acid over 30 min at a flow rate of 400 nl min−1. The LTQ-Orbitrap XL mass spectrometer was operated in parallel data-dependent mode where the MS survey scan was performed at a nominal resolution of 60 000 (at m/z 400) resolution in the Orbitrap analyser between m/z range of 400–2000. The top six precursors were selected for CID in the LTQ at normalized collision energy of 35% using multi-stage activation at m/z 98.0, 49.0 and 32.7 Da.
+ Open protocol
+ Expand
4

Proteomic Analysis of Cilia Samples

Check if the same lab product or an alternative is used in the 5 most similar protocols
Three µg of peptides from each total cilia proteome sample and one third of proximity labeling assay and immunoprecipitation samples were analysed on an LC–MS system composed of an UPLC chromatograph (nanoAcquity, Waters, Milford, MA, USA) directly coupled to a Q Exactive or Elite mass spectrometer (Thermo Scientific, Rockford, IL, USA). Data acquisition for analysis of total cilia proteome was solely performed on Q Exactive system. Peptides were trapped on C18 pre-column (180 µm × 20 mm ,Waters, Milford, MA, USA) using water containing 0.1% FA as a mobile phase and then transferred to a nanoAcquity BEH C18 column (75 µm × 250 mm, 1.7 µm, Waters, Milford, MA, USA) using acetonitrile gradient (0–35% ACN in 160 min) in the presence of 0.1% formic acid at a flow rate of 250 nl/min. Data acquisition was carried out using a data-dependent method with top 12 precursors selected for MS2 analysis after collisional induced fragmentation (CID) with an NCE of 27. Full MS scans covering the mass range of 300–2000 were acquired at a resolution of 70,000 with a maximum injection time of 60 ms and an automatic gain control (AGC) target value of 1e6. MS2 scans were acquired with a maximum injection time of 60 ms and an AGC target value of 5e5 with an isolation window of 3.0 m/z. Dynamic exclusion was set to 30 s.
+ Open protocol
+ Expand
5

Peptide Identification by Orbitrap-Fusion Mass Spectrometry

Check if the same lab product or an alternative is used in the 5 most similar protocols
The peptides were reconstituted in 0.1% formic acid and peptide estimation was carried out using a NanoDrop. Peptides (1 μg) were analysed on LTQ‐Orbitrap Fusion mass spectrometer (Thermo Scientific, Bremen, Germany) interfaced with a nanoAcquity UHPLC (Waters, MA, USA). The peptide samples were first loaded onto a trap column (Waters, Milford, MA, USA) at a flow rate of 5 μl/min and then resolved on a BEH C18 nanoAcquity column (Waters, Milford, MA, USA). The peptides were resolved using a gradient of 8% to 70% solvent B (0.1% formic acid in acetonitrile) for 45 min using a flow rate of 300 nl/min. The total run time for each sample was 60 min. The mass spectrometry settings were as follows: MS1 Resolution – 60,000; Mass Range – 350–1800 m/z; AGC Target – 1e6, Maximum injection time 22 ms. Include charge state – 2–6; Dynamic exclusion – 30 s. MS2: Isolation mode – Quadrupole; Isolation window – 1.2; Activation type – HCD, Collision energy – 30%; AGC target – 50,000, Maximum injection time – 40 ms.
+ Open protocol
+ Expand
6

Tryptic Peptide Analysis by Triple TOF Mass Spectrometry

Check if the same lab product or an alternative is used in the 5 most similar protocols
The
tryptic peptides were analyzed using a Triple TOF 6600 mass spectrometer
(Sciex). Samples were reconstituted in 50 μL of 0.1% formic
acid, and 2 μL of sample was delivered into the instrument using
an Eksigent Nano-LC system mounted with a nanoACQUITY UPLC Symmetry
C18 Trap Column and an analytical BEH C18 nanoACQUITY Column (Waters,
MA, USA). A NanoSpray III source was fitted with a 10 μm inner
diameter PicoTip emitter (New Objective). Samples were loaded in 0.1%
formic acid onto the trap, which was then washed with 2% ACN/0.1%
FA for 10 min at 2 μL/ min before switching in-line with the
analytical column. A gradient of 2–50% (v/v) ACN/0.1% (v/v)
FA over 90 min was applied to the column at a flow rate of 300 nL/min.
Spectra were acquired automatically in positive ion mode using information-dependent
acquisition, using mass ranges of 400–1600 Da in MS and 100–1400
amu in MS/MS. Up to 25 MS/MS spectra were acquired per cycle (approximately
10 Hz) using a threshold of 100 counts per s, with dynamic exclusion
for 12 s and rolling collision energy.
+ Open protocol
+ Expand
7

Peptide Analysis using Nano-UPLC-MS

Check if the same lab product or an alternative is used in the 5 most similar protocols
Samples were reconstituted in 2% ACN, 0.1% FA (v/v) prior to analysis using a Triple TOF 6600 mass spectrometer (Sciex, UK) delivered into the instrument using an Eksigent NanoLC Ultra HPLC system. Samples were injected onto a nanoACQUITY UPLC Symmetry C18 Trap Column (P/N Waters, MA, USA) and washed for 10 min at 2 µL/min with 0.1% FA. A gradient from 1.6% ACN/0.1% FA to 95% ACN/0.1% FA was applied over 95 min at a flow rate of 300 nL/min through a Peptide BEH C18 nanoACQUITY Column (Waters, MA, USA). MS was operated as described in previous methods (Meng et al., 2016 (link)).
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!