The largest database of trusted experimental protocols

14 protocols using elute uhplc

1

Testosterone Analysis by LC-MS/MS

Check if the same lab product or an alternative is used in the 5 most similar protocols
MS/MS analysis was performed on an Elute UHPLC interfaced with a Compact Q-TOF (Bruker Daltonics). The nonpolar extract was prepared in MeOH + 0.1% FA (Formic acid) at 10 μg/mL, and a commercial standard of testosterone (T1500, Sigma-Aldrich, St. Louis, MO, USA) was prepared in MeOH + 0.1% FA to 10 mM. Samples (5 µL) were analyzed on a C18 column (C18 Poroshell UPLC 1.9 µm (2.1 × 1.5 mm)) using an isocratic gradient at 60% B (A: H20 + 0.1% FA, B: MeOH + 0.1% FA) over 10 min. The instrument was internally calibrated using ESI Tune Mix (Bruker Daltonics) and set to the following parameters: Mass range: m/z 50–2000, Positive mode, Auto MS/MS with 9 precursor ions per cycle. Testosterone was fragmented at 35 eV.
+ Open protocol
+ Expand
2

CbFic2 Automodification Quantification

Check if the same lab product or an alternative is used in the 5 most similar protocols
In 100 µl total volume, 0.2 mg ml−1 (approximately 4.5 µM) of CbFic2 or CbFic2E66G were incubated in 20 mM HEPES pH 7.5, 150 mM NaCl, 1 mM MgCl2, 1 mM TCEP, 1 mM ATP each in the presence and absence of 5 µM 20 bp dsDNA at 37 °C in the autosampler of LC-MS. Samples were analyzed hourly with maXis II ETD ESI-qTOF LC-MS (Bruker Daltonics) coupled to Elute UHPLC (Bruker) using a ProSwift™ RP-4H 1 ×50 mm column (Thermo Fisher Scientific). Data were evaluated using DataAnalysis (Version 5.1, Bruker Daltonics). The degree of automodification was detected by the specific mass gain of AMPylation of 329 Da. AMPylation was quantified by the proportion of the sum of the signal intensity of all AMPylated signals to the total intensity of all CbFic2 signals. Measurements were performed in biological triplicates. Data were processed using GraphPad Prism 8.0 and represent the mean with standard deviation as error bars.
+ Open protocol
+ Expand
3

DLT Leachate Composition Analysis

Check if the same lab product or an alternative is used in the 5 most similar protocols
Analysis of DLT leachate solution composition and identification of Tinuvin 292 mixture was performed using a Bruker Impact II O-TOF High Resolution Time of Flight Mass Spectrometer connected to a Bruker Elute UHPLC (Bruker Corp., Billeric, MA) in combination with Compass HyStar 4.1 Data Acquisition software for instrument operation and Compass Data Analysis for data analysis and processing. Samples were run using an ACQUITY UPLC HSS C18SB column (50 mm length, 2.1 mm ID, particle size = 1.8 μm) maintained at 40 °C. A two solvent gradient containing HPLC-grade H2O with 0.1% formic acid (Solvent A) and HPLC-grade acetonitrile with 0.1% formic acid (Solvent B) was used with a flow rate of 0.3 mL/min. The resulting gradient with respect to time was used for all analyses: 95% A and 5% B at 0.0 min, 95% A and 5% B at 1.0 min, 0% A and 100% B at 7.5 min, 0% A and 100% B at 9.0 min, 95% A and 5% B at 9.1 min, and 95% A and 5% B at 10.0 min. MS/MS was performed utilizing the Auto MS/MS function of the Q-TOF operating at 25 eV for collision energy.
+ Open protocol
+ Expand
4

Targeted Metabolomics Profiling Using UHPLC-MS

Check if the same lab product or an alternative is used in the 5 most similar protocols
For negative ion mode, LC separation was achieved with Elute UHPLC (Bruker Daltonics, Germany) using mobile phase A: methanol/water (1:1; v/v) and mobile phase B: methanol/isopropanol (2:8; v/v), both containing 0.1% formic acid, 7.5 mM ammonium formate, and 0.1% triethylamine (TEA), respectively. For positive ion mode, the same solvent system was used without TEA. The LC method consisted of a 20 min run time with the following gradient: T0 to T1 – mobile phase A at 60% and mobile phase B at 40% which was then elevated to 90% mobile phase B and 10% mobile phase by T16 and up to 99% mobile phase B by T16.5 and then escalated back to 40% mobile phase B by T20. A constant flow rate of 0.2 mL/min was maintained throughout the runtime. Throughout the analysis, the autosampler was maintained at 4 °C. The samples were injected onto the column with a partial loop mode with an injection volume of 20 µL and 10 µL corresponding on the column to 1 µL plasma in negative ion mode and to 0.5 µl plasma in positive ion mode, respectively. For the separation, a C18 Luna Omega column (100 × 2.1 mm × 1.6 μm) (Phenomenex, Germany) was used and thermostated at 45 °C. Total run time including column wash was 20 min. No carry-over was observed in negative and positive ion mode when using wash-run protocol and analysis batch structure inferred from previous work22 (link),59 .
+ Open protocol
+ Expand
5

Targeted Metabolomic Analysis by LC-MS

Check if the same lab product or an alternative is used in the 5 most similar protocols
LC-MS was performed as previously described [21 (link)]. Chromatographic separation was performed using an Elute UHPLC (Bruker Daltonics GmbH, Bremen, Germany) equipped with a Millipore Chromolith Performance/PR-18e, C18 analytical column (100 mm × 2 mm, Merck, Darmstadt, Germany) with Chromolith® RP-18 endcapped 5-3 guard cartridges (Merck, Darmstadt, Germany). Mass spectra were obtained on a Maxis Impact Q-TOF mass spectrometer (Bruker Daltonics GmbH, Bremen, Germany) equipped with an electrospray ionization source (Bruker Daltonics GmbH, Bremen, Germany) and operated in positive ionization mode. Mass spectra were analyzed using DataAnalysis® and TASQ® software (Bruker Daltonics GmbH, Bremen, Germany).
+ Open protocol
+ Expand
6

HPLC-ESI-MS Analysis of Unknown Compounds

Check if the same lab product or an alternative is used in the 5 most similar protocols
HPLC-ESI-MS analysis was performed on Impact II QqTOF high-resolution mass spectrometer (Bruker Daltonik, Germany) with Elute UHPLC (Bruker Daltonik, Germany) on Intensity Solo 1.8 C18-2 2.1 × 100 mm 1.8 μm 90 Å reverse-phase column (Bruker Daltonik, Germany) with the following conditions: column flow 0.25 ml/min, gradient elution from 5 to 70% B in 25 min (A: 0.1% formic acid in water, B: 0.1% formic acid in AcN), column temperature 40°C, injection volume 5 μl, ESI source in positive mode, HV capillary at 4.5 kV, spray gas–nitrogen at 2.1 bar, dry gas–nitrogen at 8 l/min 220°C, scan range m/z 50–2200, 2-Hz scan rate for full scan, automatic MS/MS mode (CID) with dynamic scan rate 2–8 Hz, nitrogen as collision gas, collision energy from 23 eV at m/z 300 to 65 eV at m/z >1300, and automatic internal calibration with ESI-L low concentration tuning mix (Agilent Technologies, USA). Spectra were processed with BioPharma Compass 3.1.1 (Bruker Daltonik, Germany).
+ Open protocol
+ Expand
7

HPLC-MS/MS Analysis of Compounds

Check if the same lab product or an alternative is used in the 5 most similar protocols
HPLC-MS/MS analysis was performed as described earlier [3 (link)]. Impact II QqTOF high-resolution mass-spectrometer (Bruker Daltonik, Germany) equipped with Elute UHPLC (Bruker Daltonik, Germany) and Intensity Solo 1.8 C18-2 2.1*100 mm 1.8 μm 90 Å reverse-phase column (Bruker Daltonik, Germany) was used. Chromatography was performed in an acetonitrile gradient with 0.1% (v/v) formic acid as additive.
+ Open protocol
+ Expand
8

Ultrahigh Pressure Nanoflow LC-TIMS-TOF MS

Check if the same lab product or an alternative is used in the 5 most similar protocols
LC was performed on an ultrahigh pressure nanoflow chromatography system (Elute UHPLC; Bruker Daltonics). Peptides were separated on a reversed-phase column (40 cm × 75 μm i.d.) at 50 °C packed with Magic C18 AQ 2.5-μm 200-Å resin with a pulled emitter tip. A solution is 0.1% FA in water, and B solution is 0.1% FA in ACN. The gradient time is 60 min, and the total run time is 75 min including washes and equilibration. Peptides were separated with a linear gradient from 0 to 5% B within 5 min, followed by an increase to 30% B within 55 min and further to 35% B within 5 min, followed by a washing step at 95% B and re-equilibration. LC was coupled online to a hybrid trapped ion-mobility spectrometry quadrupole time-of-flight mass spectrometer (Bruker timsTOF Pro) via a CaptiveSpray nanoelectrospray ion source. We performed data-dependent acquisition (DDA) in parallel accumulation—serial fragmentation (PASEF) mode with ten PASEF scans per topN acquisition cycle. Singly charged precursors were excluded by their position in the m/z-ion mobility plane and precursors that reached a “target value” of 20,000 au were dynamically excluded for 0.4 min. For data-independent acquisition, we adopted the isolation scheme of 25 Da × 32 windows to cover 400 to 1200 mz.
+ Open protocol
+ Expand
9

UHPLC-MS/MS Analysis of Metabolites

Check if the same lab product or an alternative is used in the 5 most similar protocols
The UHPLC–MS/MS
analysis
was performed in a Maxis II ETD Q-TOF mass spectrometer (Bruker Daltonics,
Germany) using an electrospray ionization (ESI) source with either
an Elute UHPLC (Bruker Daltonics, Germany) or a 1260 Infinity II Binary
Pump (Agilent Technologies, USA) system. The separation was performed
on an Acquity UPLC HSS T3 column (1.8 μm, 100 × 2.1 mm)
from Waters Corporation. Milli-Q water with 0.1% formic acid was used
as mobile phase A, and LC–MS grade methanol with 0.1% formic
acid was used as mobile phase B. The column temperature was kept at
40 °C, and the autosampler temperature was kept at 4 °C.
The flow rate was set to 0.22 mL/min with an injection volume of 5
μL. The gradient used was as follows: 0–2 min, 0% B;
2–15 min, 0–100% B; 15–16 min, 100% B; 16–17
min, 100–0% B; 17–23 min, 0% B. The system was controlled
using the Compass HyStar software package from Bruker (Bruker Daltonics,
Germany). High-resolution mass spectra were acquired in negative mode
at a mass range of m/z 50–1200.
Data acquisition was performed in AutoMSMS mode (data-dependent acquisition,
DDA) with a cycle time of 0.5 s and a ramped collision energy from
20 to 50 eV. A solution of sodium formate [10 mM in a mixture of 2-propanol/water
(1/1, v/v)] was used for internal calibration at the beginning of
each run, in a segment between 0.10 and 0.31 min.
+ Open protocol
+ Expand
10

High-Resolution HPLC-MS/MS Analysis

Check if the same lab product or an alternative is used in the 5 most similar protocols
HPLC-MS/MS analysis was performed on Impact II QqTOF high-resolution mass-spectrometer (Bruker Daltonik, Germany) equipped with Elute UHPLC (Bruker Daltonik, Germany) on Intensity Solo 1.8 C18-2 2.1*100 mm 1.8 µm 90 Å reverse-phase column (Bruker Daltonik, Germany) with following conditions: column flow 0.25 mL/min, gradient elution from 30% to 100% B in 60 min (A: 0.1% (v/v) formic acid in water, B: 0.1% (v/v) formic acid in ACN), column temperature 40°C, injection volume 15 μL (1 µg/µL of dry sample in 60% (v/v) ACN), ESI source in positive mode, HV capillary at 4.5 kV, spray gas – nitrogen at 2.1 bar, dry gas – nitrogen at 8 L/min 220°C, scan range m/z 50-2200, 2 Hz scan rate for full scan, automatic MS/MS mode (CID) with dynamic scan rate 2-8 Hz, nitrogen as collision gas, collision energy from 23 eV at m/z 300 to 65 eV at m/z >1300, automatic internal calibration with ESI-L low concentration tuning mix (Agilent Technologies, USA). Spectra were processed with BioPharma Compass 3.1.1 (Bruker Daltonik, Germany).
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!