The largest database of trusted experimental protocols

40 protocols using ics 2100

1

Fluoride Quantification in Supernatant

Check if the same lab product or an alternative is used in the 5 most similar protocols
The concentrations of released fluoride during growth with 4-fluorobenzoate were determined in cell-free supernatant after centrifugation of cell suspensions at 10,000 × g (4°C, 10 min). Measurements were performed by using a Dionex ICS-2100 ion chromatography system with Dionex IonPac AS11-HC column (analytical column; 2 mm by 250 mm). The samples (10 µl) were analyzed isocratically by using 5 mM KOH at a flow rate of 0.38 ml min−1. Fluoride was detected by suppressed conductivity with a retention time of 2.7 min. NaF dissolved in water was used as a calibration standard at different concentrations.
+ Open protocol
+ Expand
2

Geochemical Characterization of Lake Waters

Check if the same lab product or an alternative is used in the 5 most similar protocols
Lake water pH was measured in triplicate in situ by pH probe V3.0 (Atlas Scientific) at the time of sampling. Concentrations of major cations and anions ( F, Cl, NO2, SO42+, Br-, NO3, Li+, Na+, and K+) were measured using ion chromatography (Dionex, ICS-2100). Total organic carbon (TOC) and total nitrogen (TN) were measured by CN Analyzer (Vario Max CN, Elementar, Germany). GPS coordinates were recorded at each sampling point, ranged from 28° 58′ 29.53″ N to 31° 49′ 19.45″ N and 85° 44′ 57.78″ N to 91° 06′ 49.54″ N (Additional file 1: Table S1). The mean of lake water temperatures for the month (August 2015) prior to sampling dates was extracted from public data [23 (link)].
+ Open protocol
+ Expand
3

Ion Chromatography for Electrolyte Analysis

Check if the same lab product or an alternative is used in the 5 most similar protocols
The
HF concentration in the electrolytes was measured using an ion chromatography
(IC) system (Dionex, ICS-2100). The electrolytes were diluted 100
times for the purpose of accurate determination.
+ Open protocol
+ Expand
4

Water Sampling and Geochemical Measurements Protocol

Check if the same lab product or an alternative is used in the 5 most similar protocols
Water
sampling and geochemical
measurements were performed before initiating the experiment and subsequently
at time intervals of 30 min to 30 h for up to 200 h. Measurements
of pH and redox potential (Eh) were performed in conjunction with
water sampling. The pH electrode (VWR Symphony) was calibrated to
NIST-traceable 4, 7, and 10 buffer solutions. The electrode performance
was regularly checked, and recalibration was performed as necessary.
The performance of the redox electrode (Accumet) was verified using
ORP calibration solution (Orion). The water samples were collected
in PE syringes, passed through 0.2 μm polyethersulfone (PES)
syringe filter membranes, stored in high-density PE (HDPE) bottles,
and refrigerated until analysis. Inorganic anions (NO3, NO2, and SO42–) were quantified by ion chromatography (ICS2100;
Dionex Corporation) on nonacidified samples. Total Se was quantified
by inductively coupled plasma–optical emission spectroscopy
(SPECTROBLUE SOP, SPECTRO Analytical Instruments GmbH, Germany) on
samples acidified to pH < 2 with trace-metal-grade nitric acid
(Thermo Fisher Scientific). Spectrophotometric (DR2800; HACH Chemical
Co.) determination of total ammonium (NH4+)
concentrations (by the Nessler method) was performed for samples from
B3 and B4, which initially contained NO3.
+ Open protocol
+ Expand
5

Characterization of Cathode Electrolyte Loss

Check if the same lab product or an alternative is used in the 5 most similar protocols
Liquid chromatography for the organic component, and ion chromatography for the ions including inorganic materials. After the above measurement, the loss amount of the liquid electrolytes in cathode were extracted with water and analyzed by the Acquity H‐class Ultra High‐Pressure Liquid Chromatography (UPLC, Xevo G2‐S QTof, Waters) system coupled with a mass spectrometer and Ion chromatography (ICS‐2100, Dionex). The carbon paper, cathode electrode, and cathode separator were immersed in ultrapure water and sonicated for 10 min, then filtrated. The obtained cathode samples were diluted properly for each measurement. The liquid TEGDME volume was measured by LC‐MS by extracting the electrolyte from the positive electrode in the cell after the measurements.
+ Open protocol
+ Expand
6

Ion Chromatography Analysis of DL-Alanine Powder

Check if the same lab product or an alternative is used in the 5 most similar protocols

Example 1

The Musashino DL-alanine powder was prepared for Ion Chromatography (IC) analysis.

The powder was analyzed for fluoride, chloride, nitrite, sulfate, bromide, nitrate, phosphate, and ammonium concentration by IC (Dionex ICS-2100), and results were given in Table 1.

TABLE 1
IC Results of DL-alanine Powder
FluorideChlorideNitriteSulfateBromideNitratePhosphateAmmonium
(ppm)(ppm)(ppm)(ppm)(ppm)(ppm)(ppm)(ppm)
ND4.1NDBDNDNDNDBD

ND indicated that there was no peak detected; and BD indicated that the detected peak intensity was not high enough to quantify as the data, i.e., below detection limit (likely <1 ppm). Both ND and BD thus indicated that the phosphate was at least <1 ppm.

The data in Table 1 has shown that DL-alanine did not have any phosphate detected.

+ Open protocol
+ Expand
7

Aerosol Anion Analysis by IC

Check if the same lab product or an alternative is used in the 5 most similar protocols
Aerosols collected on teflon filters were extracted with 8 mL of deionized water under ultrasonic agitation for 20 min and extracts were analysed by ion chromatography (ICS-2100, Dionex, Sunnyvale, CA, USA) for 6 anions (i.e., fluoride, chloride, nitrite, nitrate, sulphate, and phosphate). Aerosols collected on nitrocellulose filters underwent gravimetric analysis to determine the mass concentrations of aerosol particles.
+ Open protocol
+ Expand
8

Sulfide and Metabolite Analysis Protocol

Check if the same lab product or an alternative is used in the 5 most similar protocols
Samples for sulfide
analysis were diluted in a solution containing 1 mM NaOH and 0.1 mM
zinc acetate for sulfide fixation. Dissolved sulfide was analyzed
using a Hach Lange kit LCK-653 (Hach, Germany) which is based on the
reaction of sulfide with dimethyl-p-phenylenediamine
and ferric chloride. The absorption was measured in a spectrophotometer
DR3900 (Hach, Germany). Sulfate was measured by ion chromatography
on a Dionex ICS 2100 equipped with an IonPac AS19 column (4 ×
2550 mm) (Dionex, USA) eluted at 30 °C with potassium hydroxide
(5 mM, 1 mL·min–1).
The headspace composition
of the batch bottles was analyzed with gas chromatography (Shimadzu
GC-2010 series) equipped with two columns in parallel, a Porabond
Q (50 m × 0.53 mm) and Molsieve 5A (25 m × 0.53 mm) for
nitrogen, oxygen, and methane. Hydrogen analysis was carried out with
a Hewlett Packard GC-5890 series equipped with a CP-Molsieve 5A column
(30 m × 0.53 mm). Acetate, propionate, butyrate, methanol, and
ethanol were measured with a GC system Agilent 7890B equipped with
an HP-FFAP column (25 m × 0.32 mm). Formate was analyzed with
an HPLC system Ultimate 3000 (Thermo Fisher, USA) equipped with an
Alltech OA-1000 column (300 mm × 6.5 mm) (Hichrom, UK) eluted
at 60 °C with sulfuric acid (1.25 mM, 0.6 mL·min–1).
+ Open protocol
+ Expand
9

Elemental Composition of Toothpaste Extracts

Check if the same lab product or an alternative is used in the 5 most similar protocols
Briefly, 2 g of toothpastes were mixed with 1 mL of DMEM culture medium and centrifuged at 4200 rpm. Afterwards, the supernatants were collected and filtered, and the proportion of calcium, potassium, magnesium, sodium, phosphorus, and silica released from each toothpaste was determined using ICP-MS (Agilent 7900 ICP-MS, Agilent, Santa Clara, CA, USA).
Fluoride concentration was analyzed using a Dionex ICS-2100 ion chromatograph (IC) with an AS19 column, using potassium hydroxide as the eluent.
+ Open protocol
+ Expand
10

Measuring Flue Gas CO2 and Sulfate

Check if the same lab product or an alternative is used in the 5 most similar protocols
The concentration of CO2 sparged into the reactor was
confirmed using GasLab software and a cozIR wide-range 0–20%
CO2 sensor (CM-0123, Gas Sensing Solutions Ltd., Glasgow,
UK).
Sulfate concentrations accumulated, from the sparged simulated
flue gases, during the 50 h bioreactor experiment were quantified
using SulfaVer 4 Method 8051 (HACH, Loveland, CO, USA) with a HACH
DR6000 UV–Vis spectrophotometer (see Figures S3 & S4 in
the Supporting Information).
During
the cultured simulated flue gas trials, sulfate, phosphate,
and nitrate concentrations were measured in daily samples of the culture
medium (0.2 μm filtered) using an ion chromatograph (Thermo
Fisher ICS-2100) equipped with a Dionex IonPac AS18 column. Combined
Seven Anion Standard II (Dionex, Sunnyvale, CA) was used to calibrate
the instrument.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!