The largest database of trusted experimental protocols

1260 infinity 2 high

Manufactured by Agilent Technologies
Sourced in United States

The Agilent 1260 Infinity II is a high-performance liquid chromatography (HPLC) system designed for efficient and reliable analytical separations. The core function of this system is to provide precise and reproducible chromatographic analysis across a wide range of applications.

Automatically generated - may contain errors

5 protocols using 1260 infinity 2 high

1

HPLC-based Peptide Purification and Analysis

Check if the same lab product or an alternative is used in the 5 most similar protocols
Agilent 1260 infinity II high performance liquid chromatography (HPLC) system was applied to remove interfering substances. Buffer A: 10 mM HCOONH4, 5% ACN, pH 10.0; buffer B: 85% ACN, 10 mM HCOONH4, pH 10.0. The peptides were eluted at a flow rate of 1 ml/min with a gradient of buffer B: 0% for 25 min, 0–7% for 5 min, 7–40% for 35 min, 40–100% for 5 min, and a final hold in 100% for 15 min. Elution was supervised by examining the absorbance at 214 nm, and the components were obtained per min (about 36 components). The samples were stored at − 80 °C [12 (link), 13 (link)].
+ Open protocol
+ Expand
2

Quantification of 11 PFAS compounds by HPLC-MS/MS

Check if the same lab product or an alternative is used in the 5 most similar protocols
Eleven PFAS (Table S1) were quantified in each extract using an Agilent 1260
Infinity II high-performance liquid chromatograph (HPLC) instrument coupled to
an Agilent 6460A triple quadrupole mass spectrometer. The mass spectrometer was
operated in negative electrospray ionization mode (HPLC-ESI-MS/MS). Separation
of analytes by LC was performed using a 4.6 mm (I.D.) × 50 mm Agilent
ZORBAX Eclipse XDB-C18 reversed-phase HPLC column (1.8 μm particle size)
preceded by a 4.6 mm × 5 mm XDB-C18 guard cartridge.
Mobile phases were 2 mM ammonium acetate in water (mobile phase A) and 2
mM ammonium acetate in methanol (mobile phase B) using a flow rate of 0.4
mL/min. Gradient conditions for chromatographic separation were as follows:
initial condition (30% B) was increased to 60% B over 1.5 minutes; then
increased to 95% B over 2 minutes and held for 5.5 minutes. The gradient was
then increased to 100% B over 3 minutes, before finally returning to initial
conditions (30% B) over 0.5 minutes, and holding for 5.5 minutes. The column
temperature was set at 45°C and the injection volume was 20 μL.
Data were acquired under multiple reaction monitoring (MRM) transitions using
optimized parameters. Additional methods information, including transitions, is
included in Tables S2,
S3, and S4.
+ Open protocol
+ Expand
3

Lutein Extraction and Quantification from Microalgae

Check if the same lab product or an alternative is used in the 5 most similar protocols
Lutein was extracted using a method modified from that of Xie et al. [50 (link)]. The lutein content of the extract was analyzed using 1260 Infinity II high-performance liquid chromatography (Agilent, Santa Clara, CA, USA), as proposed by Taucher et al. [51 (link)]. A reversed-phase YMC carotenoid C30 column (150 × 4.6 mm, 3 μm) operated at 22 °C with mobile phase 1 mL/min. The mobile phase consists of two solvent mixtures: (A solvent) methanol, methyl tert-butyl ether (MTBE), and water: 81:15:4; and (B solvent) methanol, MTBE, and water: 8:89:3 (v:v:v). The mobile phase is a gradient condition, first flowing at 98% (A solvent) for 11 min, then decreasing to 60% (A solvent) over the next 7 min. This is maintained for 6.5 min, then decreases to 0% (A solvent) in 2.5 min and held for 3 min; it then increases to 98% (A solvent) over the next 3 min to re-equilibrate the mobile phase and is held for 7 min. Lutein was detected by measuring the absorbance at the wavelength 445 nm. The lutein standard was purchased from Sigma Chemical Co. (St. Louis, MO, USA). The lutein content (mg/g) and lutein productivity (mg/L/d) of microalgae were calculated as shown in Equations (3) and (4).


+ Open protocol
+ Expand
4

HPLC Analysis of Caraway Extract

Check if the same lab product or an alternative is used in the 5 most similar protocols
For the C. carvi extract analysis, An 1260 infinity II high-performance liquid chromatography (HPLC) system (Agilent Technologies, Santa Clara, CA, USA) was used. Fitted with a quaternary pump, the apparatus was coupled with a UV detector. The mobile phase was composed of two solvents, A: 0.1 percent acidified water and B: acetonitrile. A C18 Zorbax eclipse plus C18 column (5 μm, 4.6 × 150 mm) was used for the separation. The temperature of the column furnace was set at 35 °C. The sample injection volume and flow rates were set at 10 μL (100 μg/mL of sample concentration) and 1 mL/min. The concentration was calculated according to the following Equation (2), depending on the retention time (RT) and spectral correspondence of each compound:
+ Open protocol
+ Expand
5

Synaptosome Peptide Fractionation by bRP

Check if the same lab product or an alternative is used in the 5 most similar protocols
Four mg of S-trap digested synaptosome peptides were fractionated by basic reversed-phase (bRP) chromatography. bRP solvent A was made in an aqueous solution of 5mM of Ammonium Formate and 2% ACN (pH 10). bRP Solvent B was made in a solution of 5 mM Ammonium Formate and 90% ACN (pH 10). A 4.6 mm inner diameter 250 mm long Agilent Zorbax 300 Extend-C18 (Agilent) was used as the bRP column.
Chromatography occurred over a 96-minute gradient with fractions collected in a Whatman 2ml 96 well plate (GE Healthcare). A 1260 Infinity II high-performance liquid chromatography (HPLC) instrument was used (Agilent). The chromatography gradient proceeded with a solvent B composition scheme of: 0% to 5 minutes, 5% at 7.66 minutes, 28.5% at 64.98 minutes, 34% at 70.48 minutes, 60% at 83.48 minutes, and 60% solvent B continuing for the remainder of the gradient. Peptides were concatenated into a 12-fraction scheme as previously described (34) . Peptides were then frozen and dried by vacuum centrifugation in preparation for O-GlcNAc enrichment.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!