The three-pulse ESEEM sequence (π/2)-τ-(π/2)-t-(π/2)-τ-echo was employed, with excitation at the maximum of the echo-detected EPR spectrum. The pulse lengths were 16 ns and the time delay τ was 204 ns, and the time delay t was scanned from 300 ns to 10 μs, with a 12 or 16 ns time step. The resonator was cooled with a stream of cold nitrogen gas. The temperature was controlled with a nitrogen flow stabilized by a Bruker ER4131VT temperature controller. The sample temperature was kept near 80 K.
Er 4118 x ms 3 resonator
The ER 4118 X-MS-3 resonator is a laboratory instrument designed for electron paramagnetic resonance (EPR) spectroscopy. It provides a controlled environment for analyzing and characterizing the magnetic properties of materials. The resonator is a key component in the EPR spectrometer, responsible for generating and detecting the electromagnetic signals used in the spectroscopic analysis.
2 protocols using er 4118 x ms 3 resonator
Pulsed EPR Spectroscopy at Cryogenic Temperatures
The three-pulse ESEEM sequence (π/2)-τ-(π/2)-t-(π/2)-τ-echo was employed, with excitation at the maximum of the echo-detected EPR spectrum. The pulse lengths were 16 ns and the time delay τ was 204 ns, and the time delay t was scanned from 300 ns to 10 μs, with a 12 or 16 ns time step. The resonator was cooled with a stream of cold nitrogen gas. The temperature was controlled with a nitrogen flow stabilized by a Bruker ER4131VT temperature controller. The sample temperature was kept near 80 K.
EPR Spectroscopy Protocol for Pulsed Experiments
In pulsed EPR studies, an X-band Bruker ELEXSYS E580 EPR spectrometer was used equipped with a split-ring Bruker ER 4118 X-MS-3 resonator and an Oxford Instruments CF-935 cryostat. A three-pulse ESEEM sequence (π/2)-τ-(π/2)-t-(π/2)-τ-echo was employed, with excitation at the maximum of the echo-detected EPR spectrum. The pulse lengths were 16 ns, time delay τ was 204 ns, the time delay t was scanned from 300 ns to 10 µs, with a 12 or 16 ns time step. The resonator was cooled with a stream of cold nitrogen gas. The temperature was controlled by a nitrogen flow stabilized by a Bruker ER4131VT temperature controller, the sample temperature was kept near 80 K.
About PubCompare
Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.
We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.
However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.
Ready to get started?
Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required
Revolutionizing how scientists
search and build protocols!